510(k) Data Aggregation

    K Number
    K201347

    Validate with FDA (Live)

    Date Cleared
    2020-06-18

    (28 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K111433, K103756, K140881

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Attune Total Knee System is intended for cemented use as a total knee replacement system.

    Candidates for total knee replacement include patients with and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.

    Device Description

    The subject device is a polyethylene tibial inserts and patella component made from AOX Polyethylene that mates with existing cemented and noncemented femoral and tibial base components

    AI/ML Overview

    The provided text does not describe an AI medical device or a study involving AI. It is a 510(k) submission for a Total Knee System and specifically concerns an extension of the product's shelf life (from 5 years to 8 years) for certain components made from AOX Polyethylene that are gamma irradiated.

    Therefore, I cannot extract information related to:

    • Acceptance criteria for an AI device.
    • Sample size for a test set or data provenance for an AI study.
    • Number of experts for ground truth establishment for an AI study.
    • Adjudication method for an AI study.
    • Multi-reader multi-case (MRMC) comparative effectiveness study for AI.
    • Standalone performance for an AI algorithm.
    • Type of ground truth for an AI study.
    • Sample size for a training set for AI.
    • How ground truth for a training set was established for AI.

    The document discusses non-clinical testing to support the shelf-life extension, specifically:

    • Material Oxidation Index/Resistance Testing (ASTM F2102-17, ASTM F2003)
    • Wear Testing with Particle Analysis (ISO 14243-2, ISO 14243-3, ASTM F1877-05)

    It explicitly states: "Clinical testing was not necessary to determine substantial equivalence between the ATTUNE Polyethylene Fixed Bearing Tibia Inserts and Patella components and the predicate devices."

    In summary, the provided text does not contain the information requested as it pertains to a mechanical medical device (knee implant) and a shelf-life extension, not an AI device or an AI study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180446

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-12

    (234 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    Zimmer NexGen LPS, K960279, Zimmer NexGen Fluted Stemmed Tibial Component, K963148, Depuy Attune PS, K111433

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The b-ONE Total Knee System is intended for total knee arthroplasty due to the following conditions:

    · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post -traumatic arthritis,

    · Post-traumatic loss of knee joint configuration and function.

    · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

    · Revision of previous unsuccessful knee replacement or other procedure.

    • · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture- management techniques.
      Additional indications for Posterior Stabilized (PS) and Posterior Stabilized Plus (PS+) components:

    • · Ligamentous instability requiring implant bearing surface geometries with increased constraint.

    • · Absent or nonfunctioning posterior cruciate ligament.

    • · Severe anteroposterior instability of the knee joint.

    The b-ONE Total Knee System is intended for implantation with bone cement only. b-ONE Total Knee System components are not intended for use with other knee systems.

    Device Description

    The b-ONE Total Knee System is a modular artificial knee replacement system comprised of symmetric cemented femoral components with optional femoral distal pegs, symmetric cemented tibial baseplate, symmetric tibial inserts with locking wires, symmetric patellar resurfacing button, and reusable surgical instruments. The therapeutic effect is replacement of the diseased joint with artificial components to restore joint function. The system is a posterior stabilized device type and includes two options for posterior stabilization, a PS and a PS+. The PS+ is designed with additional constraint when more stability is required. Compatibility of the system components is only claimed with the b-ONE Total Knee System. There is no allowed interchangeability with systems manufactured by other companies.

    The System includes left and right femoral components manufactured from cast cobalt chrome. The system includes 30 sizes. Sizes 1-10 are provided in right and left versions. Sizes 3-7 are provided in both standard and narrow configurations, with the narrow sizes having a relatively smaller aspect ratio of the medial-lateral to anterior-posterior widths when compared to the standard sizes. Optional modular distal pegs are included in the system.

    The system includes a Patella Component offered in seven sizes of variable diameter and thicknesses. The bone fixation surface incorporates three fixation pegs. The Patella Components are made from Conventional UHMWPE.

    The system includes Tibial Inserts which are offered in 40 sizes, size A/1-4 through HJ/7-10 with 8 thicknesses and 48 sizes for PS+ inserts, size A1-2+ through HJ/7-10+ with 8 thicknesses. The Tibial Inserts are made from Conventional UHMWPE. The Tibial Inserts are pre-assembled with a locking wire which is manufactured from Cobalt Chrome alloy.

    The system includes Tibial Baseplates offered in 9 sizes ranging from A-J. Tibial Baseplates are manufactured from forged titanium alloy.

    All system components are supplied sterile and are single use devices.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification from the FDA for a medical device called the b-ONE Total Knee System. This type of document is used to demonstrate that a new device is "substantially equivalent" to an existing, legally marketed device (predicate device) and therefore does not require a full Pre-Market Approval (PMA).

    It is important to understand that a 510(k) summary for a prosthetic joint like a knee replacement does NOT involve clinical studies or "acceptance criteria" related to diagnostic performance as would be found in AI/ML medical devices.

    Instead, the "acceptance criteria" for a knee replacement relate to the performance of the physical components (e.g., strength, fatigue life) and their biocompatibility, demonstrating that the new device functions as safely and effectively as the predicate devices. The "study that proves the device meets the acceptance criteria" refers to non-clinical (bench) tests of the device's materials, design, and mechanical properties.

    Therefore, the requested information elements related to AI/ML device evaluations (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of device and submission.

    Here's a breakdown of the applicable information based on the provided text:


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (Type of Non-Clinical Test)Reported Device Performance (Implied by submission)
    Tibial Baseplate FatigueDemonstrated to be equivalent to predicate devices in fatigue resistance.
    Locking Mechanism StrengthDemonstrated to be equivalent to predicate devices in locking mechanism strength.
    Tibial Post FatigueDemonstrated to be equivalent to predicate devices in tibial post fatigue resistance.
    ConstraintDemonstrated to have appropriate constraint comparable to predicate devices.
    Contact Area and Contact StressDemonstrated contact area and stress characteristics within acceptable limits and comparable to predicate devices.
    Tibiofemoral Range of MotionDemonstrated range of motion comparable to predicate devices.
    Characterization of UHMWPE Insert MaterialMaterial properties of the Ultra-High Molecular Weight Polyethylene (UHMWPE) inserts were characterized and found to be acceptable and comparable to predicate devices.
    Bacterial Endotoxin TestingMet established safety criteria for bacterial endotoxin levels.
    Shelf Life StudiesDemonstrated appropriate shelf-life for sterility and material integrity.
    BiocompatibilityDemonstrated to be biocompatible, indicating no harmful biological reactions.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This is a physical medical device (knee implant), not a diagnostic algorithm. The "test sets" refer to individual device components or representative samples subjected to specific engineering tests, not patient data sets. The data provenance would be laboratory testing results.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. "Ground truth" in this context refers to established engineering standards, material specifications, and mechanical performance criteria, not expert consensus on medical images or diagnoses. These tests are conducted by engineers and technicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for subjective interpretations of data, typically in clinical evaluations or AI/ML ground truth establishment. These are objective engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a physical knee implant, not an AI/ML diagnostic or assistive device. MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical knee implant, not an AI/ML algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for the non-clinical tests is based on established engineering standards, material specifications, mechanical performance criteria, and comparison to the predicate devices' known performance and specifications.

    8. The sample size for the training set

    • Not Applicable. This is a physical medical device. There is no "training set" in the context of machine learning. The design and manufacturing processes are developed through engineering and materials science principles.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant. The "ground truth" for the device's design and manufacturing is established through adherence to recognized industry standards, regulatory requirements, and engineering principles.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140881

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-13

    (312 days)

    Product Code
    Regulation Number
    888.3565
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K101433, K111433, K103756, K062654

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATTUNE Cementless CR and PS Femoral Components are intended for cementless use within the ATTUNE total knee replacement system.

    Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant (provided that adequate bone is present).

    Device Description

    The ATTUNE Cementless cruciate retaining (CR) and posterior stabilized (PS) femoral components have an asymmetric trochlear groove and are available in sizes 1-10 in right and left options. Sizes 3-6 are available in standard and narrow options. The fixation surface is porous coated with Porocoat. The porous coated fixation surface comprises: the bone cut surfaces of the anterior flange, anterior chamfer, distal surface, posterior chamfer, posterior condylar resection, as well as the lugs for additional stability. The fixation of the femoral component to the femoral bone is achieved by biologic fixation via ingrowth into the Porocoat porous coating. The ATTUNE femoral components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The ATTUNE Cementless CR Femoral Components are compatible with the ATTUNE CR Fixed Bearing inserts (K101433), fixed bearing tibial bases (K101433), and patellae (K103756). The ATTUNE Cementless PS Femoral Components are compatible with the ATTUNE PS Fixed Bearing inserts (K111433), fixed bearing tibial bases (K101433), and patellae (K103756). The ATTUNE CR Total Knee is designed to accommodate knee flexion to 150 degrees in those patients able to attain a high degree of knee flexion. The ATTUNE PS Total Knee is designed to accommodate knee flexion to 145 degrees in those patients able to attain a high degree of knee flexion.

    AI/ML Overview

    This document pertains to the 510(k) premarket notification for the ATTUNE® Knee System-Cementless CR and PS Femoral Components. It describes the device and its substantial equivalence to predicate devices, focusing on the lack of need for new studies rather than describing the acceptance criteria or studies of the device itself.

    Based on the provided text, there is no information available regarding acceptance criteria or a study that proves the device meets those acceptance criteria for the ATTUNE® Knee System-Cementless CR and PS Femoral Components.

    The document explicitly states:

    • "Non-Clinical Testing: None provided as it was not necessary to determine substantial equivalence between the ATTUNE Cementless CR and PS Femoral Components and the predicate devices. As the coating for the subject device is identical to that in K062654, the characterization in K062654 also applies to this 510(k)." (Page 6)
    • "Clinical Testing: None provided as it was not necessary to determine substantial equivalence between the ATTUNE Cementless CR and PS Femoral Components and the predicate devices." (Page 6)

    The submission argues for substantial equivalence based on existing predicate devices:

    • The ATTUNE Cementless CR and PS Femoral Components are "identical in design to the ATTUNE CR and PS Femoral Components cleared for cemented use only in K101433 and K111433."
    • "The Porocoat porous coating of the ATTUNE Cementless CR and PS Femoral Components is identical to the Porocoat porous coating used on the SIGMA CR Porocoat Femoral Components cleared for Cementless use in K062654."

    Therefore, I cannot provide the requested information from the given text as it explicitly states that new testing was not performed for this 510(k) submission.