Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192135

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-24

    (139 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K101542, K110813, K110741, K121785

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIA 21, 27, 33 - The VIA® Microcatheter is intended for the introduction of interventional devices (such as the WEB device/stents/flow diverters) and infusion of diagnostic agents (such as contrast media) into the neuro, peripheral, and coronary vasculature.

    VIA 17, 17 Preshaped 45°, 17 Preshaped 90° - The VIA® Microcatheter is intended for the introduction of interventional devices (such as coils/stents) and infusion of diagnostic agents (such as contrast media) into the neuro, peripheral, and coronary vasculature.

    Device Description

    The VIA® Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature. The physician inserts the catheter into the vein or artery through the skin (percutaneous) using a sheath or guidewire. The device can then be navigated to the treatment site. Navigation is aided by the coated surface of the catheter which assists with manipulation while in the vasculature. Throughout the procedure the physician can obtain the position of the catheter by the radiopaque marker bands using fluoroscopic techniques (VIA17 Microcatheter has 2 radiopaque marker bands. VIA 21, 27 and 33 Microcatheters have one radiopaque tip marker band). Diagnostic and interventional devices can be delivered through the lumen of the catheter to the treatment site. The proximal end of the catheter incorporates a standard luer adapter to facilitate attachment of accessories.

    AI/ML Overview

    The provided FDA 510(k) summary for the VIA® Microcatheter does not contain the specific information required to answer all parts of your request, as it pertains to a medical device's 510(k) clearance rather than a diagnostic AI algorithm study. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing (biocompatibility and bench testing), not on a clinical effectiveness study with human readers and AI assistance.

    Therefore, for aspects related to AI performance, human expert adjudication, effect size with AI assistance, and large-scale training/test sets for an AI algorithm (questions 2-9), the information is not applicable or not provided in this document.

    However, I can extract the acceptance criteria and reported performance for the device itself (the VIA® Microcatheter) based on the provided bench testing data.


    Acceptance Criteria and Reported Device Performance (Based on Bench Testing)

    Acceptance Criteria / TestReported Device Performance
    Visual and Dimensional InspectionPass
    Hub ISO 80369-7Pass
    WEB RetractionPass
    Kink ResistancePass
    TensilePass
    Tip BucklingPass
    Steam Shaping / Shape RetentionPass
    Catheter Leakage and Static BurstPass
    Coating Friction & DurabilityPass
    Coating ParticulatePass
    Tracking ForcePass

    Note: Some tests (Coating Integrity, Catheter Torque Strength, Flow Rate, Dead Space) were reported "For characterization only," implying they were not subject to specific pass/fail acceptance criteria for this submission but were measured to understand device properties.


    Study Details (for the VIA® Microcatheter, not an AI study)

    1. A table of acceptance criteria and the reported device performance:

      • See table above.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: The document does not specify exact sample sizes for each bench test. For biocompatibility tests, specific numbers of animals (e.g., 17 male Hartley Guinea Pigs, 3 female New Zealand White Rabbits, 20 female Albino Swiss Mice, 2 Canine) are mentioned, but these are for biological testing, not for evaluating an AI algorithm on a medical dataset.
      • Data Provenance: Not applicable for device performance bench testing in the context of AI. The tests are laboratory-based.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. This is not an AI study requiring expert ground truth for medical images.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is not an AI study requiring expert adjudication.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This document outlines the premarket notification for a physical medical catheter, not an AI-powered diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For biocompatibility testing, the "ground truth" is defined by biological responses in test organisms and in vitro systems, evaluated against established ISO standards and criteria for non-toxicity, non-sensitization, non-irritation, non-hemolysis, etc.
      • For bench testing, the "ground truth" is adherence to predefined engineering and functional specifications or standards (e.g., ISO 10555-1:2014, ISO80369-7:2016).
    8. The sample size for the training set:

      • Not applicable. There is no training set for an AI algorithm in this document specific to the device's performance evaluation.
    9. How the ground truth for the training set was established:

      • Not applicable. There is no AI training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1