Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY All-In-One STEAM Reusable Test Pack, when used with a VERIFY Bowie Dick Indicator Strip, is designed for testing the air removal efficiency of pre-vacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°C/-0°C) temperature difference between the center of the ANSI/AAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete color change in the indicator window). The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time.

    VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization a biological indicator, integrating indicator or both a biological and integrating indicator from the following list.

    • · 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company
    • · 3M ATTEST Super Rapid Read Biological Indicators catalog number 1492V manufactured by 3M Company
    • Smart-Read EZTest Self Contained Biological Indicators for steam manufactured by MesaLabs, Inc.
    • · Celerity 20 Steam Biological Indicator
    • · VERIFY STEAM Integrating Indicator
    • · VERIFY SixCess MF4 STEAM Emulating Indicator

    VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge the steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization cycles with the VERIFY STEAM Integrating Indicator.

    The VERIFY All-In-One STEAM Resuable Test Pack employing the All-in-One PV10 Assembly (a disposable polypropylene sleeve assembly) with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 can be used for routine monitoring of 10-Minute 270°P/132°C pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Resuable Test Pack.

    The VERIFY All-In-One STEAM Reusable Test Pack employing the Celerity Xtend 10 Assembly with a Celerity 20 STEAM Biological Indicator and a Celerity 10 STEAM Chemical Indicator can be used for routine monitoring of 10minute 270°F/132°C dynamic air removal steam sterilization cycles. The sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.

    Device Description

    The VERIFY All-in-One STEAM Reusable Test Pack consists of a reusable aluminum tube with a tortuous path challenge for air removal and steam penetration. The VERIFY All-in-One STEAM Reusable Test Pack will be used with the Celerity Xtend 10 Assembly containing the Celerity 20 STEAM Biological Indicator and the Celerity 10 STEAM Chemical Indicator.

    The Celerity Xtend 10 Assembly consists of a tube and cap design, comprised of one (1) polypropylene tube and one (1) silicone rubber cap containing one (1) Celerity 20 STEAM Biological Indicator and one (1) Celerity 10 STEAM Chemical Indicator and can only be used with the VERIFY A11-in-One STEAM Reusable Test Pack. The Celerity Xtend 10 Assembly has a tamper evident seal that will be broken once the assembly has been opened after use. A broken tamper evident seal will indicate to the user that it cannot be reused. The Celerity Xtend 10 Assembly is a single use device used for routine monitoring of 10-minute 270°F pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.

    The Celerity 20 STEAM Biological Indicator monitors the critical parameters of steam sterilization cycles by producing an optical change (signal) that is detected by the STERIS proprietary reader, Celerity STEAM Incubator, in 20 minutes to confirm viability of the biological indicator at the end of a steam sterilization process. Both the Celerity 20 STEAM Biological Indicator has been cleared under K173634 and K181429 and the Incubator has been cleared under K173670.

    The Celerity 10 STEAM Chemical Indicator is a single-use device used to monitor a 10-minute 270°F/132°C dynamic air removal steam sterilization cycle and is subject of a concurrent premarket notification.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the VERIFY All-In-One STEAM Reusable Test Pack, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test NameAcceptance CriteriaReported Device Performance
    Simulated UsePerformance of the Biological Indicator (BI) in the Reusable Test Pack is equivalent to the performance of the predicate.PASS
    Celerity Xtend 10 Assembly in pack vs Celerity Xtend 10 Assembly outside of packReusable Test Pack containing the Celerity Xtend 10 Assembly provides a greater challenge to the process than the Celerity Xtend 10 Assembly itself.PASS

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the "Simulated Use" and "Celerity Xtend 10 Assembly in pack vs Celerity Xtend 10 Assembly outside of pack" tests. The data provenance (country of origin, retrospective/prospective) is also not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    The document does not provide information on the number of experts or their qualifications used to establish ground truth for the test set. This type of information is usually relevant for studies involving subjective human interpretation (e.g., image analysis), which is not the primary focus of this device (a sterilization process indicator).

    4. Adjudication Method for the Test Set:

    The document does not mention an adjudication method for the test set. Given the nature of the device (a pass/fail indicator for sterilization), an adjudication method relying on multiple human readers is typically not applicable. The “ground truth” for effectiveness of sterilization indicators is typically objective, based on microbiological culture or physical/chemical changes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement with AI vs. Without AI Assistance:

    No, an MRMC comparative effectiveness study was not done. This device is a sterilization process indicator and does not involve AI assistance or human reader interpretation in the context of diagnostic decision-making.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

    The performance studies described ("Simulated Use" and "Celerity Xtend 10 Assembly in pack vs Celerity Xtend 10 Assembly outside of pack") appear to be evaluating the standalone performance of the device (the VERIFY All-In-One STEAM Reusable Test Pack in conjunction with its indicators). The "pass" conclusions suggest acceptable standalone performance.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

    For sterilization indicators like this device, the ground truth is typically based on:

    • Microbiological Culture: For biological indicators (BI), "pass" means the BI did not grow microorganisms after sterilization, indicating effective sterilization. "Fail" means growth, indicating ineffective sterilization.
    • Physical/Chemical Changes: For chemical indicators (CI) like the Celerity 10 STEAM Chemical Indicator, the ground truth is based on a specific, observable color change (e.g., yellow to blue/purple) which correlates with exposure to critical sterilization parameters.
    • Reference Standards: Comparison against established and validated sterilization process challenges (e.g., ANSI/AAMI/ISO standards for temperature differences or air removal efficiency).

    The document mentions "conditions are met and proper vacuum is achieved" and "2°C (+1°C/-0°C) temperature difference between the center of the ANSI/AAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature," suggesting a ground truth based on physical parameters validated against international standards for sterilization and microbiological efficacy.

    8. The Sample Size for the Training Set:

    The document does not provide information about a "training set" in the context of machine learning or AI. This device is a physical indicator system, not an AI/algorithm-based device that would typically use a training set. The development of such devices relies on engineering, material science, and microbiology, with validation studies rather than training sets.

    9. How the Ground Truth for the Training Set Was Established:

    This question is not applicable as there is no mention of a training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1