Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K210699, K201693, K183377, K202785, K103583

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Consona N6, Consona N6 Pro, Consona N6 Super, Consona N6S, Consona AE, Consona AT, Consona N6 Exp, Consona No Elite, Consona N6T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates.

    It is intended for use in fetal, abdominal, Intra-operative, small organ(breast, thyroid, testes), neonatal cephalic,adult cephalic,trans-rectal, trans-vaginal, musculo-skeletal(conventional), musculo- skeletal(superficial), Thoracic/Pleural, cardiac adult, cardiac pediatric, Trans-esoph., peripheral vessel and urology exams.

    Modes of operation include: B, M, PWD, CWD, Color Doppler, Combined mode(B+M, PW+B, Color+B, PW+Color+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart 3D, iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast imaging (Contrast imaging (Contrast agent for Liver).

    This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healtheare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

    Device Description

    Consona N6, Consona N6 Pro, Consona N6 Super, Consona AR, Consona N6S, Consona AE, Consona AT, Consona N6 Exp, Consona N6 Elite, Consona N6T Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include : B. M. PWD. CWD. Color Doppler. Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, Amplitude PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart 3D, 4D(Real-time 3D), iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

    Subject device can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.

    AI/ML Overview

    This document is a 510(k) summary for a diagnostic ultrasound system (Consona N6 series). It appears the device does not rely on AI features requiring specific performance metrics or a human-in-the-loop study for its predicate equivalence. The summary explicitly states "Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence."

    Therefore, the requested information about acceptance criteria, specific study designs (MRMC, standalone), ground truth establishment for a test set and training set, and expert qualifications for AI/ML performance evaluation cannot be extracted from this document because such studies were not deemed necessary for this device's regulatory review.

    The document describes a general diagnostic ultrasound system and focuses on demonstrating substantial equivalence to predicate devices based on:

    • Intended Use: Similar applications (fetal, abdominal, cardiac, etc.)
    • Technological Characteristics: Similar imaging modes (B, M, Doppler), transducers, and features.
    • Non-clinical Tests: Compliance with acoustic output, biocompatibility, electrical safety, and software standards.

    The only "new option" mentioned that might be interpreted as having an "intelligent" component is "Smart Calc," described as: "Semi-automatically recognizes the contour of the target area and measures the diameter, area and circumference of the target area. The user needs to identify the target area first. When the measures results are not satisfactory, the user can use the trackball to manually edit the contour." However, this is presented as an enhancement based on cleared functions, not a new intended use or a feature requiring a specific AI/ML performance study as typically understood in the context of FDA submissions for AI-enabled devices.

    Summary of what cannot be provided given the document:

    • Table of acceptance criteria and reported device performance for AI features: Not provided as specific AI performance studies were not cited.
    • Sample size and data provenance for a test set: Not applicable; no such test set was described for AI performance.
    • Number of experts and qualifications for ground truth: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • MRMC comparative effectiveness study, effect size: Not applicable.
    • Standalone (algorithm only) performance: Not applicable.
    • Type of ground truth used: Not applicable for AI performance studies.
    • Sample size for the training set: Not applicable.
    • How ground truth for the training set was established: Not applicable.

    Information that can be extracted related to the device as a whole (not AI-specific):

    • Non-clinical Tests: The device underwent testing for acoustic output (NEMA UD 2-2004), electrical safety (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37), usability (IEC 60601-1-6, IEC 62366-1), risk management (ISO 14971), biocompatibility (ISO 10993-1), and software lifecycle processes (IEC 62304). This demonstrates the device meets general safety and performance criteria through adherence to recognized standards.
    • Clinical Studies: Explicitly stated as "Not applicable."
    • Device Performance: The conclusion states that the device "was found to have a safety and effectiveness profile that is similar to the predicate device" based on "performance data as documented in the study" (referring to the non-clinical tests described).

    In conclusion, this document does not contain the information requested about AI acceptance criteria and study designs because the device, as described for this 510(k) submission, does not appear to be an AI-enabled device requiring such specific performance evaluations for its regulatory clearance, or at least, those details are not provided in this public summary.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K210699, K201693, K183377, K202785, K103583

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Consona N9, Consona N9 Pro, Consona N9 Super, Consona Nova, Consona N9S, Consona NI, Consona NT, Nuewa N9, Recho N9, Consona N9 Exp, Consona N9 Elite, Resona N9, Consona N9T, Consona N8, Consona N8 Pro, Consona N8 Super, Consona OR, Consona N8S, Consona OI, Consona OT, Consona N8 Exp, Consona N8 Elite, Consona N8T, Consona N7, Consona N7 Pro, Consona N7 Super, Consona VR, Consona N7S, Consona VI, Consona VT, Consona N7 Exp, Consona N7 Elite, Consona N7T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates.

    It is intended for use in fetal, abdominal, Intra-operative, small organ(breast, thyroid, testes), neonatal cephalic.adult cephalic.trans-rectal, trans-vaginal, musculo-skeletal(conventional), musculo-skeletal(superficial), Thoracic Pleural, cardiac adult, cardiac pediatric, Trans-esoph., peripheral vessel and urology exams.

    Modes of operation include: B, M, PWD, Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart 3D, 4D(Real-time 3D), iScape View, TDI, Color M, Strain Elastography, STE, STQ, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

    This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging,measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

    Device Description

    Consona N9, Consona N9 Pro, Consona N9 Super, Consona Nova, Consona N9S, Consona NI, Consona NT, Nuewa N9, Recho N9, Consona N9 Exp, Consona N9 Elite, Resona N9, Consona N9T, Consona N8, Consona N8 Pro, Consona N8 Super, Consona OR, Consona N8S, Consona OI, Consona OT, Consona N8 Exp, Consona N8 Elite, Consona N8T, Consona N7, Consona N7 Pro, Consona N7 Super, Consona VR, Consona N7S, Consona VI, Consona VT, Consona N7 Exp, Consona N7 Elite, Consona N7T Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include : B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart 3D, 4D(Real-time 3D), iScape View, TDI, Color M, Strain Elastography, STE, STQ, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver). Subject device can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state quantitative acceptance criteria or reported device performance metrics for the Consona N9/N8/N7 series Diagnostic Ultrasound System. The submission is a 510(k) premarket notification for substantial equivalence, not a performance study against specific acceptance criteria.

    The document primarily focuses on demonstrating that the subject device is substantially equivalent to legally marketed predicate devices (primarily the Mindray MX7, K212900) in terms of intended use, imaging modes, features, functions, and technological characteristics.

    Instead of a table of specific performance metrics, the document highlights:

    • Indications for Use: The subject device's indications are comparable to the predicate.
    • Patient Contact Materials: Same as predicate or tested under ISO 10993-1.
    • Acoustic Power Levels: Below FDA limits and same as the predicate (MX7).
    • Electrical and Physical Safety Standards: Complies with FDA recognized standards, same as the predicate (MX7).
    • Features and Functions: Same as predicate devices, with new options being either improvements or enhancements of cleared functions, with no new intended uses or clinical risks introduced.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence."

    Therefore, there is no test set, no sample size, and no data provenance (e.g., country of origin, retrospective/prospective) related to clinical performance for this submission. The evaluation is based on non-clinical tests and comparison to predicate devices.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    As there were no clinical studies conducted and no test set, this information is not applicable. Ground truth for clinical performance was not established by experts for this particular submission.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used for performance evaluation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done, as the document states that clinical studies were not required and were not performed. Therefore, there's no reported effect size of human readers improving with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone performance study of an algorithm was conducted or reported, as this is a general-purpose diagnostic ultrasound system, not an AI-driven diagnostic algorithm. The new features like "Smart Calc" are described as semi-automatic tools that assist the user, implying a human-in-the-loop design, but no performance study specific to this feature is detailed.

    7. Type of Ground Truth Used

    Given the absence of clinical studies, there is no mention of ground truth established through expert consensus, pathology, or outcomes data. The basis for substantial equivalence relies on technical specifications, compliance with safety standards, and functional comparison to predicate devices, rather than clinical performance data against a defined ground truth.

    8. Sample Size for the Training Set

    Not applicable. This submission concerns a diagnostic ultrasound system and does not involve an AI model with a distinct "training set" in the conventional sense of machine learning algorithms requiring large datasets for training. The "new options" like Smart Calc are described as improvements or enhancements but not as standalone AI algorithms whose performance requires training and validation datasets.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183377

    Validate with FDA (Live)

    Date Cleared
    2019-04-04

    (119 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K171579, K103583, K171233, K163690, K122010

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DC-40/DC-35/DC-40S/DC-40 Pro Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic.adult.cephalic.trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo- skeletal (superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

    Device Description

    The DC-40/DC-35/DC-45/DC-40S/DC-40 Pro Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW-mode, Color-Mode, Power/Dirpower mode, THI Mode, 3D/4D Mode, iScape mode, TDI mode, Color M mode, Biospy Guidance, Elastography, Contrast imaging (Liver) or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include linear array, Phased array and convex array.

    AI/ML Overview

    This document is an FDA 510(k) summary for a Diagnostic Ultrasound System (DC-40/DC-35/DC-45/DC-40S/DC-40 Pro). The information provided focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than on proving performance against specific acceptance criteria through a clinical study involving AI or human readers.

    Therefore, many of the requested categories in your prompt are Not Applicable (N/A) based on the provided text, as this submission is for an ultrasound system, not an AI/CADe device requiring performance metrics.

    Here's the breakdown of the available information:

    1. Table of acceptance criteria and the reported device performance:

    The document does not specify quantitative acceptance criteria for the device performance in terms of diagnostic effectiveness for specific clinical tasks, as it's a general ultrasound system and not an AI or CADe device. Instead, it demonstrates compliance with regulatory standards for safety and fundamental performance.

    Acceptance Criteria CategoryDevice Performance (Based on provided text)
    Acoustic OutputBelow the limits of FDA
    BiocompatibilityTransducers and needle-guided brackets tested under ISO 10993-1
    Cleaning and DisinfectionEffectiveness evaluated
    Electrical SafetyComplies with ANSI/AAMI ES60601-1 and related standards
    Mechanical SafetyComplies with ANSI/AAMI ES60601-1 and related standards
    Software Life CycleComplies with IEC 62304
    Usability (Human Factors)Complies with IEC 62366-1 and IEC 60601-1-6
    Risk ManagementComplies with ISO 14971

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. The document refers to non-clinical tests (e.g., electrical safety, acoustic output), not performance data from a patient test set for diagnostic accuracy.
    • Data Provenance: Not applicable. The tests were compliance-based rather than data-driven from patient cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable. This information is relevant for studies validating diagnostic algorithms against ground truth established by medical experts (e.g., radiologists, pathologists). The provided document describes the clearance of an ultrasound system, not an AI/CADe product.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as no human reader study or test set requiring adjudication of ground truth is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted device. The submission is for a diagnostic ultrasound system.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm but a diagnostic ultrasound system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. The "ground truth" in this context refers to compliance with established engineering and safety standards, rather than medical diagnostic ground truth.

    8. The sample size for the training set:

    Not applicable. This device is not an AI/Machine Learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI/ML model described.

    In summary: The provided document is a 510(k) summary for a general-purpose diagnostic ultrasound system. It details the device's intended use, modifications from a predicate device, and compliance with various recognized safety and performance standards. It explicitly states, "Clinical Tests: Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence." Therefore, the type of performance data and studies requested in your prompt (which are typical for AI/CADe devices) are not included in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123185

    Validate with FDA (Live)

    Date Cleared
    2012-11-02

    (23 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

    Device Description

    DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode, 3D/4D mode, Elastography or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 3 MHz to 10.0 MHz.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System. This type of submission is a statement to the FDA that the device is substantially equivalent to another legally marketed device, not a claim of superior performance or a detailed study of acceptance criteria against novel performance metrics.

    Therefore, the provided text does not contain the acceptance criteria or a study proving the device meets those criteria in the context of specific performance metrics for diagnosis or improved human reader performance with AI assistance. Instead, it focuses on demonstrating substantial equivalence to predicate devices and adherence to general safety and performance standards.

    Here's an analysis of the provided information relative to your requested points:


    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices and compliance with recognized safety and performance standards, rather than defining specific outcome-based acceptance criteria and reporting performance against them.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided as there is no clinical performance study detailed in this 510(k) summary. The document states "DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety." These are non-clinical hardware tests, not clinical performance studies involving a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable/not provided. As no clinical performance study is detailed, there's no mention of a test set, ground truth establishment, or experts for that purpose.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. No clinical performance study or test set requiring adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done, and there is no mention of AI assistance in this document. The device is a general-purpose diagnostic ultrasound system, not explicitly an AI-powered diagnostic tool. The purpose of this 510(k) is to establish substantial equivalence, not to demonstrate improved human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. The device is a diagnostic ultrasound system, which inherently involves human operation and interpretation. There is no mention of a standalone algorithm or AI performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable/not provided. Since no clinical performance study with a test set is detailed, there's no mention of the type of ground truth used.

    8. The sample size for the training set

    This information is not applicable/not provided. The device is an ultrasound system. There is no mention of a "training set" in the context of an algorithm or AI.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As there's no training set mentioned in the context of an algorithm or AI, there's no corresponding information on how ground truth would be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121010

    Validate with FDA (Live)

    Date Cleared
    2012-04-19

    (16 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103677, K103583, K113647, K03218, K093563

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M7/M7T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in gynecology, obstetric, abdominal, pediatric, small parts (breast, thyroid), neonatal cephalic, transcranial, cardiac, transvaginal, transrectal, peripheral vascular, urology, orthopedic, and musculoskeletal (conventional and superficial), intraoperative and transesophageal(cardiac) exams.

    Device Description

    M7/M7T Diagnostic Ultrasound System is a general purpose, portable/mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 2.5 MHz to 10.0 MHz.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and study information for the M7/M7T Diagnostic Ultrasound System:

    Summary of Device Performance Study Information

    The provided 510(k) summary for the M7/M7T Diagnostic Ultrasound System is for a general-purpose ultrasound device. As such, it primarily focuses on establishing substantial equivalence to predicate devices and adherence to safety and performance standards rather than presenting a clinical study with specific acceptance criteria related to diagnostic accuracy.

    The document indicates that the device has been evaluated for:

    • Acoustic output
    • Biocompatibility
    • Cleaning and disinfection effectiveness
    • Thermal, electrical, and mechanical safety

    It states that the device conforms to applicable medical safety standards, including UD 2, UD 3, IEC 60601-1 series, UL 60601-1, ISO14971, ISO 10993-1, IEC62304, and IEC60601-2-18.

    Acceptance Criteria and Reported Device Performance

    Since this is a filing for a general-purpose ultrasound system based on substantial equivalence, there are no specific diagnostic accuracy acceptance criteria or performance metrics (like sensitivity, specificity, AUC) reported in this summary. The "performance" being demonstrated here is primarily the device's adherence to relevant safety and performance standards, and its technological characteristics being comparable to predicate devices.

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Safety StandardsConforms to: UD 2, UD 3, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, UL 60601-1, ISO14971, ISO 10993-1, IEC62304, IEC60601-2-18.
    FunctionalityAcquires and displays ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode or combined modes.
    Technological EquivalenceSame technological characteristics and comparable key safety/effectiveness features as predicate devices.
    Intended Use EquivalenceSame intended uses and basic operating modes as predicate devices.
    Quality SystemsConformance with 21 CFR 820, ISO 9001, and ISO 13485 quality systems.
    Acoustic OutputEvaluated and found to conform to limits (implied by conformance to standards).
    BiocompatibilityEvaluated and found to conform (implied by conformance to standards).
    Cleaning/DisinfectionEvaluated for effectiveness (implied by conformance to standards).
    Thermal/Electrical/Mechanical SafetyEvaluated and found to conform (implied by conformance to standards).

    Specific Study Information Not Provided in this Document:

    Based on the provided text, the following information is not available:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): No clinical test set or data from a diagnostic performance study is mentioned. The evaluation is focused on engineering and safety standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no clinical ground truth study is described.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable as no clinical test set is described.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance: Not applicable. This document describes a general diagnostic ultrasound system, not an AI-powered diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable as no clinical ground truth study is described.
    7. The sample size for the training set: Not applicable as this is not an AI/machine learning device.
    8. How the ground truth for the training set was established: Not applicable as this is not an AI/machine learning device.

    Conclusion from the Document:

    The conclusion states that the "Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the M7/M7T Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market."

    This heavily implies that due to the nature of the device (a general-purpose ultrasound system) and the regulatory pathway chosen (510(k) substantial equivalence), the "study" demonstrating its meeting of acceptance criteria consists of:

    • Verification and validation against engineering and safety standards (as listed above).
    • Bench testing and performance verification to ensure the device functions as intended (e.g., imaging modes, frequency ranges).
    • Comparison of its intended uses and technological characteristics to already cleared predicate devices.

    There is no mention of a clinical trial or a study assessing the diagnostic accuracy or efficacy of the M7/M7T Diagnostic Ultrasound System in this summary. Such studies are typically not required for general-purpose ultrasound systems cleared via the 510(k) pathway when substantial equivalence to existing devices is established based on technical and functional similarities and adherence to recognized standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113647

    Validate with FDA (Live)

    Date Cleared
    2011-12-16

    (4 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

    Device Description

    DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode, 4D mode or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 3 MHz to 10.0 MHz.

    AI/ML Overview

    The provided text is a 510(k) summary for the Shenzhen Mindray Bio-medical Electronics Co., LTD DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System. This document focuses on demonstrating substantial equivalence to predicate devices and adherence to medical safety standards rather than establishing new performance criteria through clinical studies.

    Therefore, the document does NOT contain information regarding:

    • Specific acceptance criteria framed as quantitative metrics (e.g., sensitivity, specificity, accuracy).
    • Details of a study proving the device meets acceptance criteria in terms of clinical performance.
    • Sample sizes used for test or training sets for an AI/algorithm.
    • Data provenance, number or qualifications of experts, or adjudication methods for establishing ground truth related to clinical performance.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes for human readers with/without AI assistance.
    • Standalone algorithm performance studies.
    • The type of ground truth used as it relates to clinical performance studies.
    • The sample size for the training set or how ground truth was established for a training set.

    The document primarily outlines the device's technical specifications, intended uses, and lists applicable safety and electrical standards it conforms to. It states that the device has been evaluated for "acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety" and found to conform with applicable medical safety standards.

    However, based on the provided text, I can infer the "acceptance criteria" and "study" in a regulatory context, which is demonstrating substantial equivalence to predicate devices and compliance with recognized standards.

    Here's a breakdown of what can be extracted based on that understanding:

    1. Table of "Acceptance Criteria" and "Reported Device Performance":

    In the context of a 510(k) submission for a diagnostic ultrasound system like this, the "acceptance criteria" are not reported as specific performance metrics (like sensitivity/specificity) but as conformance to established safety and performance standards, and substantial equivalence to legally marketed predicate devices. The "reported device performance" is the statement of conformity.

    Acceptance Criteria (Regulatory Context)Reported Device Performance (as stated in the 510(k) Summary)
    Substantial Equivalence to Predicate Devices (similar technological characteristics, comparable safety & effectiveness, same intended uses)"DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices: Mindray DC-7 (K103583), Mindray M7 (K103677), GE Voluson E8 (K101236), Mindray DC-T6 (K110199), Sonosite M-Turbo (K101757), Siemens ACUSON S2000 (K112596), Siemens ACUSON SEQUOIA 512 (K063085), GE LOGIQ e (K102256). They have the similar technological characteristics, are comparable in key safety and effectiveness features, and have the same intended uses and basic operating modes as the predicate devices."
    Conformance with Applicable Medical Safety Standards"DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical safety standards. This device has been designed to meet the following standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, IEC 62304, IEC 62366, UL 60601-1, ISO14971, UD 2, UD 3 and ISO 10993-1."
    Quality System Compliance"The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems."
    Intended Uses are Consistent with Traditional Clinical Practices and FDA Guidelines"Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination."

    2. Sample size used for the test set and the data provenance:

    The document does not detail specific clinical studies with test sets, sample sizes, or data provenance (country, retrospective/prospective) for proving diagnostic performance. The evaluations mentioned are primarily for safety, electrical, and mechanical conformity.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as no described clinical performance study with a test set requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The device described is a general diagnostic ultrasound system, not an AI-powered diagnostic algorithm aiding human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a diagnostic ultrasound system, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable in the context of clinical diagnostic performance. For technical evaluations, the "ground truth" would be the specifications and requirements defined by the applicable standards (e.g., standard measures for acoustic output, material properties for biocompatibility testing).

    8. The sample size for the training set:

    Not applicable, as no described AI/algorithm training process.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1