Found 4 results

510(k) Data Aggregation

    K Number
    K210770

    Validate with FDA (Live)

    Date Cleared
    2021-07-29

    (136 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K083246, K102939, K120672

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

    Device Description

    Fully Automatic Electronic Blood Pressure Monitor (KD-5810, KD5810B, KD-5811, KD-5920TL and KD-552) is designed and manufactured according to IEC 80601-2-30.

    The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Andon Health Co., Ltd.'s Fully Automatic Electronic Blood Pressure Monitor. The document indicates that the device's accuracy was verified by meeting criteria 1 and 2 of ISO 81060-2. However, the text does not explicitly provide a table of acceptance criteria and reported device performance for the subject device itself. Instead, it refers to prior clinical test reports for predicate devices (KD-5961, BP3, and BP5) as references.

    Therefore, not all requested information can be extracted directly for the subject device from the provided text. I will answer based on the information available and will highlight where specific details are not provided for the subject device but are referenced through predicate devices.

    1. A table of acceptance criteria and the reported device performance

    The provided text states that the accuracy of the blood pressure monitors in the clinical test reports was verified by meeting Criteria 1 and Criteria 2 of ISO 81060-2. However, it does not provide a specific table with numerical acceptance criteria and the reported performance for the subject device models (KD-5810, KD-5810B, KD-5811, KD-5920TL, and KD-552). It refers to the clinical test reports of predicate devices.

    Based on general knowledge of ISO 81060-2, the criteria typically involve:

    ISO 81060-2 Criteria (General Reference, not explicitly stated in document for subject device):

    Acceptance Criteria (ISO 81060-2:2009)Reported Device Performance (Reference to predicate studies)
    Criterion 1: Mean difference and standard deviationNot explicitly stated for the subject device. The document states that the clinical test reports of the predicate devices met these criteria.
    The mean difference between the device measurements and reference measurements should be within ±5 mmHg.
    The standard deviation of the differences should be ≤ 8 mmHg.
    Criterion 2 (for device approval for general population): Distribution of individual differencesNot explicitly stated for the subject device. The document states that the clinical test reports of the predicate devices met these criteria.
    For SYS/DIA, 65% of the differences should be within ±5 mmHg.
    85% of the differences should be within ±10 mmHg.
    95% of the differences should be within ±15 mmHg.

    2. Sample size used for the test set and the data provenance

    For the subject devices (KD-5810, KD-5810B, KD-5811, KD-5920TL, and KD-552), the text does not explicitly state the sample size used for the test set or the data provenance directly for these devices.
    It states:

    • "The proposed devices KD-5810, KD-5810B, KD-5811 and KD-5920TL has the same algorithm and design principle with cleared device KD-5961 (K083246), so the clinical test report of KD-5961can be used as a reference when considering the clinical effect of KD-5810, KD-5810B, KD-5811 and KD-5920TL."
    • "The proposed devices KD-552 has the same algorithm and design principle with cleared device BP3 (K102939) and BP5 (K120672), so the clinical test report of BP3 (K102939) and BP5 (K120672) can be used as a reference when considering the clinical effect of KD-552."

    To determine the sample sizes and data provenance, one would need to access the clinical test reports for K083246 (KD-5961), K102939 (BP3), and K120672 (BP5). The provided document does not contain this information.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The text does not provide details on the number or qualifications of experts used to establish the ground truth for the clinical studies mentioned (those for KD-5961, BP3, and BP5). For blood pressure measurement validation studies, ground truth is typically established by trained observers (e.g., physicians or nurses) using a sphygmomanometer following a standardized protocol (such as auscultation with a mercury sphygmomanometer or an equivalent reference device).

    4. Adjudication method for the test set

    The text does not specify an adjudication method for the test set. In blood pressure validation studies, ground truth is usually established by highly trained technicians or physicians taking multiple reference readings, and often, paired readings are averaged or discrepancies between two observers are reconciled, but these specific methods are not described in this document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device described is a Noninvasive Blood Pressure Measurement System. This type of device is not typically evaluated using MRMC studies as it is a standalone measurement device, not an interpretive imaging or diagnostic system where human readers interpret results. Therefore, this section is not applicable. The document does not describe any MRMC studies or human reader improvement with AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the clinical validation referenced through the predicate devices (KD-5961, BP3, BP5) would assess the standalone performance of the algorithm in measuring blood pressure against a reference standard. The text states: "Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2." This implies a standalone performance assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For blood pressure monitors validated against ISO 81060-2, the "ground truth" (or reference measurement) is typically established by simultaneous or sequential measurements taken by trained observers using a calibrated reference sphygmomanometer (e.g., mercury or an equivalent validated device) using the auscultatory method. This is a direct expert measurement, not pathology or outcomes data.

    8. The sample size for the training set

    The provided text does not provide any information regarding a "training set" or its sample size. Blood pressure monitors like this typically use established algorithms (oscillometric method) and are validated against clinical standards rather than being "trained" in the machine learning sense with a distinct training set. If the algorithm involved machine learning, this information would be critical, but the description ("operational principle is based on oscillometric and silicon integrates pressure sensor technology") suggests a traditional signal processing approach.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" in the context of machine learning, the question of how its ground truth was established is not applicable based on the provided document. The ground truth for the validation (test) set would be established as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162668

    Validate with FDA (Live)

    Date Cleared
    2017-02-14

    (141 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102939, K120672

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KD-926 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

    Device Description

    KD-926 Fully Automatic Electronic Blood Pressure Monitor is designed and manufactured according to IEC 80601-2-30.

    The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the KD-926 Fully Automatic Electronic Blood Pressure Monitor:

    No specific acceptance criteria or performance numbers are explicitly stated in the provided document. The document primarily focuses on the substantial equivalence argument, referencing standards and prior clearances.

    However, based on the mentioned standards (IEC 80601-2-30), we can infer the implied performance criteria that the device must meet to conform to these standards.

    Inferred Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria:

    Given the nature of the submission (510(k) for a blood pressure monitor) and the referenced standards, the primary acceptance criteria would revolve around the accuracy of blood pressure and pulse rate measurements.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from IEC 80601-2-30)Reported Device Performance (Implied)
    Accuracy of Systolic Blood Pressure: Mean difference and standard deviation between device and reference standard within specified limits (e.g., typically ≤ ±5 mmHg mean difference and ≤ ±8 mmHg standard deviation for individual measurements).Not explicitly stated with numerical values in the provided text for KD-926. The document states "KD-926 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards: ... IEC 80601-2-30:2009 & A1:2013." Conformance to this standard implies meeting its accuracy requirements.
    Accuracy of Diastolic Blood Pressure: Mean difference and standard deviation between device and reference standard within specified limits (e.g., typically ≤ ±5 mmHg mean difference and ≤ ±8 mmHg standard deviation for individual measurements).Not explicitly stated with numerical values in the provided text for KD-926. Conformance to IEC 80601-2-30 implies meeting its accuracy requirements.
    Accuracy of Pulse Rate: Mean difference between device and reference standard within specified limits.Not explicitly stated with numerical values in the provided text for KD-926. Conformance to IEC 80601-2-30 implies meeting its accuracy requirements.
    Safety: Compliance with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC)."Electrical safety according test to IEC 60601-1;" "Electromagnetic compatibility test according to IEC 60601-1-2;" "None of the test demonstrates that KD-926 Fully Automatic Electronic Blood Pressure Monitor bring new questions of safety and effectiveness." (Implies compliance).
    Performance Characteristics (General): Compliance with IEC 80601-2-30."Safety and performance characteristics of the test according to IEC 80601-2-30." (Implies compliance).

    2. Sample size used for the test set and the data provenance

    The document states: "The cuff and algorithm of KD-926 is the same as K102939 and K120672, so we use the clinical data of K102939 and K120672 as the clinical proof of the new device."

    • Sample Size: The sample size for the clinical data used for K102939 and K120672 is not specified in the provided document.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only indicates that "clinical data of K102939 and K120672" were used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. For blood pressure clinical trials, ground truth is typically established by trained observers using mercury sphygmomanometers as per recognized protocols (e.g., ANSI/AAMI/ISO 81060-2). The number and qualifications of these observers are not detailed here.

    4. Adjudication method for the test set

    This information is not provided in the document. For blood pressure clinical trials, multiple observers may be used, and their readings are often averaged or assessed for agreement, but the specific adjudication method is missing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for interpretation by human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone (algorithm only) performance assessment was done. The clinical proof relies on the accuracy of the device itself (the "cuff and algorithm") against a reference standard, without human interpretation in the loop beyond initial application and reading of the reference. The device directly outputs blood pressure and pulse rate measurements.

    7. The type of ground truth used

    The type of ground truth used for validating blood pressure monitors is typically simultaneous measurements by trained observers using a reference standard device, such as a mercury sphygmomanometer or another validated oscillometric device, following an established protocol. While not explicitly stated, this is standard practice for devices conforming to IEC 80601-2-30.

    8. The sample size for the training set

    • Not applicable / Information not provided. As a standalone medical device measuring physiological parameters, it's unlikely to have a "training set" in the machine learning sense for the device's core function. The device's algorithm development involves calibration and validation against established physiological principles and clinical data, but not typically a "training set" for an AI model interpreting images or signals. The clinical data referenced (K102939 and K120672) serves as the validation/test set for the algorithm's accuracy.

    9. How the ground truth for the training set was established

    • Not applicable / Information not provided. See the explanation for point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122013

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-13

    (307 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K060686, K102939

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model Cardio+ ( BP4GAPO-2M ) is a device intended to measure the systolic and diastolic blood pressure, pulse rate of an adult individual by using a non-invasive oscillometric technique in one inflatable cuff is wrapped around the single upper arm.

    The Cardio+ (BP4GAPO-2M) achieves its function by integrating the device with an iPhone, iPad, or iPod Touch.

    The device can be used in connection with your personal computer (PC) running the software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC.

    Device Description

    Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model Cardio+ (BP4GAPO-2M) is designed to measure systolic and diastolic blood pressure, pulse rate of an individual by using a non-invasive technique in which one inflatable cuff is wrapped around the single upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but use two resistive pressure sensors rather than a stethoscope and mercury manometer. The sensors convert tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

    The Cardio+ (BP4GAPO-2M) achieves its function by integrating the device with an iPhone, iPad, or iPod Touch. Because the device does not contain an LCD or other display components, it is necessary for the device to connect with an iPhone, iPad, or iPod Touch containing supporting software to constitute a complete blood pressure measure system.

    In addition, the device can be used in connection with your personal computer (PC) running the software. The memory data can be transferred to the PC by connecting the monitor with the PC via cable.

    AI/ML Overview

    Due to the nature of the provided text, which primarily describes a 510(k) premarket notification for a blood pressure monitor, it does not contain the detailed information typically found in a study report regarding acceptance criteria and performance of the device against those criteria.

    Specifically, the document focuses on demonstrating substantial equivalence to predicate devices based on non-clinical tests (reliability, EMC, safety) and stating that clinical testing for blood pressure accuracy is "not required" because the core measurement technology is identical to a previously cleared device.

    Therefore, for the requested information, the following points can be extracted or noted as explicitly not provided in the given text:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicit in text for blood pressure accuracy. The document states "repeated clinical testing in accordance with ANSI/AAMI SP10 is therefore not required" because the blood pressure measurement technology is considered identical to a predicate device that would have met such standards.Not explicitly reported in this document. Performance is inferred to meet the standards of the predicate device.
    Reliability Test (Storage, Operating, Vibration, Drop, Life)Device "met all relevant requirements" for these tests.
    EMC TestDevice "met all relevant requirements" for this test.
    IEC 60601-1 Safety TestDevice "met all relevant requirements" for this test.

    Regarding the study that proves the device meets the acceptance criteria:

    The document states:

    • "The subject modified device Model Cardio+ (BP4GAPO-2M) is, from the technical point of view, identical to the blood pressure monitor Model BP3AC1-1 PC. The differences between these two devices do not relate to blood pressure measurement technology. Therefore, the clinical accuracy in terms of blood pressure detection will not be affected. Based on this, repeated clinical testing in accordance with ANSI/AAMI SP10 is therefore not required."

    This indicates that no new clinical study was performed or presented in this 510(k) submission to directly prove the device's accuracy against specific acceptance criteria. Instead, the argument for substantial equivalence relies on the technical identity of the blood pressure measurement technology with a predicate device (Model BP3AC1-1 PC) that would have previously met such clinical accuracy standards (e.g., ANSI/AAMI SP10).


    Additional Information (Extracted or Not Provided):

    • 2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not provided. No new clinical test set was described for blood pressure accuracy.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not provided. No new clinical test set was performed.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not provided. No new clinical test set was performed.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable/Not provided. This device is a blood pressure monitor, not an AI-assisted diagnostic tool.
    • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable/Not provided in the context of diagnostic performance. The device is an automated blood pressure monitor. Its "algorithm" (oscillometric method) is inherent to its function, and its performance is assumed to be equivalent to the predicate.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not explicitly stated for this submission. For blood pressure monitors, ground truth for clinical accuracy studies typically involves concurrent measurements by trained observers using a mercury sphygmomanometer (auscultatory method), often following standards like ANSI/AAMI SP10. The submission states that the predicate device would have met such standards.
    • 8. The sample size for the training set:

      • Not applicable/Not provided. This is not an AI/machine learning device that involves a training set in the typical sense. The "training" would refer to the development and calibration of the oscillometric algorithm in the predicate device, for which details are not in this document.
    • 9. How the ground truth for the training set was established:

      • Not applicable/Not provided. (See point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K122443

    Validate with FDA (Live)

    Device Name
    UFIT TEN-10
    Date Cleared
    2012-11-20

    (102 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K102939, K103046

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UFIT® TEN-10 Automated Wrist Blood Pressure Monitor with Computer Controls and Web Services is a non-invasive blood pressure measurement system intended to measure the systolic and diastolic blood pressure and pulse rate. It is intended for patient use at home and by healthcare professionals in their respective practice.

    The device is intended for use by adults 18 years or older. The cuff is to be used for wrist sizes from 13.5 cm to 23 cm (5.3 in to 9.1 in).

    Device Description

    UFIT® TEN-10 "Automated Wrist Blood Pressure Monitor with Computer Controls and Web Services" is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-23cm.

    UFIT® TEN-10 "Automated Wrist Blood Pressure Monitor with Computer Controls and Web Services" is designed and manufactured according to ANSI/AAMI SP10-manual, electronic or automated sphygmanometers.

    The operational principle is based on oscillometric method and pressure sensor technology. It calculates the systolic and diastolic blood pressure. UFIT® TEN-10 "Automated Wrist Blood Pressure Monitor with Computer Controls and Web Services" achieves its function by connecting the device to a Personal Computer via USB connection. Furthermore the device does not contain an LCD or other display components, so it is necessary for the device to connect to a Personal Computer containing a support software to constitute a complete blood pressure measurement system.

    When a reading is taken by the device through the software, it is sent through the Internet to servers that perform the analysis of the reading and the result is returned through the Internet to the end user - all by means of Cloud Computing. This provides secure accounts for the users to submit their data so that they can monitor their results and progress.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (from ANSI/AAMI SP10:2002)Reported Device Performance (UFIT® TEN-10)
    All applicable safety and performance requirements for an automated sphygmomanometer (as defined by ANSI/AAMI SP10:2002, and its amendments A1:2003 and A2:2006)The UFIT® TEN-10 complies with and meets all applicable requirements of the standards.
    Compliance to IEC 60601-1-2 (Electromagnetic Compatibility)Complied with EN 60601-1-2:2001, ICES-001 Issue 4 June 2006, and FCC 47 CFR Part 15, Subpart B - Verification.
    Compliance to IEC 60601-1 (General Requirements for Safety)Successfully subjected to the test requirements of IEC 60601-1:1988, CAN/CSA C22.2 No.601.1-M90 and UL 60601-1:2003.

    2. Sample Size for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the number of subjects (sample size) used for the clinical tests that demonstrated compliance with ANSI/AAMI SP10. It only mentions that "UFIT® TEN-10 Automated Wrist Blood Pressure Monitor with Computer Controls and Web Services tests in accordance with ANSI/AAMI SP10, and the device meets all applicable requirements of the standard."
    • Data Provenance: Not explicitly stated. The document doesn't provide information on the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth:

    • The document does not specify the number of experts used to establish ground truth or their qualifications. The clinical performance data references compliance with the ANSI/AAMI SP10 standard, which implies a standardized method for validation, usually involving a comparison to a reference standard (like a mercury sphygmomanometer) performed by trained observers. However, the details of these observers or their specific roles in establishing "ground truth" are not provided.

    4. Adjudication Method:

    • Not explicitly stated. The document refers to compliance with ANSI/AAMI SP10, which outlines specific methodologies for validating blood pressure devices. These methodologies often involve simultaneous measurements by multiple observers (adjudication). However, the specific adjudication method (e.g., 2+1, 3+1) is not detailed in this summary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. The document does not describe an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.

    6. Standalone Performance (Algorithm only without Human-in-the-loop performance):

    • Yes. The entire testing described in the "Non-Clinical Performance Data" and "Clinical Performance Data" sections, specifically compliance with ANSI/AAMI SP10, refers to the standalone performance of the UFIT® TEN-10 device in measuring blood pressure and pulse rate. The software merely processes and displays the data, and the web services provide storage and analysis, but the core measurement is performed by the device itself.

    7. Type of Ground Truth Used:

    • Compliance with a recognized standard: The ground truth for validating the device's accuracy would have been established through a comparison to a referencestandard according to ANSI/AAMI SP10 protocols. This typically involves simultaneous measurements against a highly accurate, calibrated reference device (e.g., mercurysphygmomanometer) performed by trained observers.

    8. Sample Size for the Training Set:

    • The document does not provide a sample size for a training set. This is because the device is a medical measurement device, not an AI/machine learning model that typically requires a separate training set. The performance is validated through clinical testing against a standard, not by training an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable, as there is no mention of a training set for an AI/machine learning algorithm. The "ground truth" for the device's performance evaluation was established by adherence to the methodologies outlined in the ANSI/AAMI SP10:2002 standard and its amendments, as described in point 7.