Found 2 results

510(k) Data Aggregation

    K Number
    K183116

    Validate with FDA (Live)

    Device Name
    Unity Network ID
    Manufacturer
    Date Cleared
    2018-12-06

    (27 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Edwards HemoSphere (K163381), GE Aisys CS2 (K170872), GE Giraffe Incubator (K101788), GE Giraffe Omnibed
    Incubator (K101788), GE Giraffe Warmer (K070377), GE Panda Warmer (K070377), Maquet Flow-i (K160665)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (information systems) it is connected to.

    Device Description

    The Unity Network ID system communicates patient data from sources other than GE Medical Systems Information Technologies, Inc. equipment to a clinical information system, central station, and/or GE Medical Systems Information Technologies Inc. patient monitors.

    The Unity Network ID acquires digital data from eight serial ports, converts the data to Unity Network protocols, and transmits the data over the monitoring network to a Unity Network device such as a patient monitor, clinical information system or central station.

    AI/ML Overview

    This document primarily describes a 510(k) premarket notification for the GE Healthcare Unity Network ID, focusing on its substantial equivalence to a predicate device, Unity Network ID V8 (K170199). It does not contain information about acceptance criteria for device performance with specific metrics or detailed study results where a device's performance is measured against those criteria.

    The information provided describes the device's function (data collection and clinical information management), its intended use, and the changes made from the predicate device (primarily software updates to support new third-party devices).

    However, it explicitly states:
    "The Unity Network ID V9 was tested to assure that the device meets its design specifications. Testing included all new or modified features."
    and
    "The subject of this premarket submission, Unity Network ID V9, did not require clinical studies to support substantial equivalence."

    Therefore, based on the provided text, I cannot describe the acceptance criteria and study as requested, because specific performance acceptance criteria and a study demonstrating the device meets those criteria are not detailed.

    The document only states that non-clinical tests were performed to ensure compliance with voluntary standards and design specifications. It lists general quality assurance measures applied during development and testing but does not provide specific performance metrics, sample sizes, ground truth establishment, or expert involvement as typically found in a clinical performance study for AI/machine learning devices.

    Here's a breakdown of the specific points you requested, noting what is and isn't available in the provided text:

    1. A table of acceptance criteria and the reported device performance

      • Not Available: The document does not provide a table of acceptance criteria nor reported device performance metrics against such criteria. It states the device "meets its design specifications" and "comply with, applicable voluntary standards," but no specifics are given.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not Available: No test set sample sizes or data provenance are mentioned as no clinical studies were performed. The testing described is non-clinical verification and validation of design specifications.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable/Not Available: Since no clinical studies were required and no test sets with ground truth are described, there is no information about experts establishing ground truth.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable/Not Available: No clinical test set or adjudication method is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not Applicable/Not Available: This device is a data collection and management system, not an AI-assisted diagnostic tool. No MRMC study was performed or is relevant for this type of device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not Applicable/Not Available: This device is not an algorithm for diagnostic or prognostic purposes, but rather an interface for data transmission. Standalone performance in the context of an algorithm is not relevant here.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not Applicable/Not Available: No ground truth in the context of a clinical performance study is mentioned.
    8. The sample size for the training set

      • Not Applicable/Not Available: This device is not an AI/machine learning model that requires a training set in the conventional sense. Its "training" would involve configuring it to correctly interpret and transmit data from specific third-party devices, which is part of its design and verification process, not a machine learning training process.
    9. How the ground truth for the training set was established

      • Not Applicable/Not Available: See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152814

    Validate with FDA (Live)

    Date Cleared
    2016-03-17

    (171 days)

    Product Code
    Regulation Number
    880.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K101788, K071175, K020543, K993407

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

    Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.

    This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

    Device Description

    The Giraffe OmniBed Carestation is an updated version of the cleared predicate Giraffe OmniBed. The Giraffe OmniBed Carestation is a combination device that can function as an incubator (with the canopy closed) or as an infant radiant warmer (with the canopy open) based on the user's selection.

    Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. With the canopy closed, the bed functions as an incubator, maintaining the infant's temperature by circulating heated air within the enclosed bed compartment. The operator may select either the air or skin temperature control method. Depending on the control method selected, heat is regulated based on either the air temperature or the infant's skin temperature compared to the operated selected control temperature. Physical access to the patient is obtained through the side portholes or by opening one of the side doors.

    With the canopy opened, this bed operates like a conventional open, radiantly heated infant bed. Radiant heat from an infrared heat source is focused onto the bed to warm the patient.

    The Giraffe OmniBed Carestation incorporates an optional weighing scale, Uninterruptible Power Supply (UPS) & Shuttle, Mounting Accessories Rail and Shelves and Storage drawers.

    The proposed modification of the Giraffe OmniBed is referred to as the Giraffe OmniBed Carestation.

    The Giraffe OmniBed Carestation updates the graphical monochrome display user interface (UI) on the predicate to a digital touchscreen UI, with the required software changes to support the new format and layout. The main system control software was not changed; however the software for the UI was updated because of the UI format change. The modified Giraffe OmniBed Carestation maintains the predicate Giraffe OmniBed functionality, performance, and clinical workflows. The changes also include a modified device visual indicator light and the capability for Hands Free Alarm Silencing (HFAS).

    Other modifications being made for the Giraffe OmniBed Carestation include upgrading the power supply from 75W to 120W to support the increased power requirements of the touchscreen and associated electronics.

    There is no change in the indications for use or intended use of the system. There are no changes to the patient contacting materials of the device, and they remain identical to the predicate. The changes made do not affect the function, performance, safety, or clinical use of the device.

    AI/ML Overview

    This document is a marketing submission (510(k) summary) for a medical device, specifically a neonatal incubator/warmer, and not a study proving a device meets acceptance criteria. Therefore, the detailed information requested regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment cannot be extracted from this document, as it does not contain the specifics of a performance study with defined acceptance criteria and results.

    The document states: "The subject of this premarket submission, Giraffe OmniBed Carestation, did not require clinical studies to support substantial equivalence." This means that the manufacturer is claiming substantial equivalence based on non-clinical tests (such as electrical safety, performance verification against standards, software testing) and the device's technological similarity to previously cleared predicate devices, rather than through a comparative clinical study.

    Here's what can be inferred and what cannot, based on the provided text:

    What Can Be Inferred/Extracted (to some extent):

    • Device Name: Giraffe OmniBed Carestation CS1 (K152814)
    • Device Type: Combination infant incubator and infant warmer.
    • Indication for Use: To provide heat in a controlled manner to neonates unable to thermoregulate. Can operate as an incubator (closed canopy) or radiant warmer (open canopy). May incorporate a Servo Controlled Oxygen Delivery System for stable oxygen concentration (21-65%).
    • Predicate Device(s): Giraffe OmniBed (K101788, K071175, K020543, K993407)
    • Key Modification: Updates the graphical monochrome display user interface (UI) to a digital touchscreen UI, with associated software changes. Also, an upgraded power supply (75W to 120W) and updated visual indicator light and Hands Free Alarm Silencing (HFAS) capability.
    • Determination of Substantial Equivalence: Based on non-clinical tests and technological similarity to predicates.
    • Non-Clinical Tests Mentioned (examples of types of tests, not specific results or acceptance criteria):
      • Risk Analysis
      • Design Reviews
      • Unit level testing (Module verification)
      • Integration testing (System verification)
      • Software testing (Verification and Validation, IEC 62304:2006)
      • Performance testing (Verification of performance specifications, including IEC 60601-2-21:2009 for warmers, IEC 60601-2-19:2009 for incubators)
      • Safety and EMC testing (Verification ES 60601-1:2005+A1 2012, IEC60601-1-2:2007)
      • Usability testing (Validation IEC 62366: 2014)

    What Cannot Be Extracted (because the document is not a study report):

    1. A table of acceptance criteria and the reported device performance: This document lists types of tests performed (e.g., performance testing against standards), but it does not provide the specific numerical acceptance criteria (e.g., "temperature stability must be within +/- 0.5°C") nor the measured device performance against those criteria.
    2. Sample sizes used for the test set and the data provenance: Not applicable as it's not a clinical study. The non-clinical tests would have involved specific test units, but the "sample size" of patients/data as in a clinical study is not relevant here.
    3. Number of experts used to establish the ground truth for the test set and qualifications: Not applicable as no clinical ground truth was established for "testing" in the sense of patient data. Usability testing might involve experts, but details are not provided.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and effect size: No, the document explicitly states: "The subject of this premarket submission... did not require clinical studies to support substantial equivalence." An MRMC study would be a clinical study.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is electro-mechanical, not AI-driven in the sense of an algorithm interpreting data to generate a finding.
    7. The type of ground truth used: Not applicable. Ground truth for clinical studies would be, for example, pathology results for a lesion, or confirmed diagnosis; here, "ground truth" would be the engineering specifications and performance standards.
    8. The sample size for the training set: Not applicable. This device is not an AI/ML device that requires a training set of data.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided text is a regulatory submission demonstrating substantial equivalence of a medical device based on non-clinical testing and comparison to predicate devices, not a report of a study designed to prove the device meets specific clinical acceptance criteria.