Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K172998

    Validate with FDA (Live)

    Device Name
    uWS-MI
    Date Cleared
    2018-04-05

    (190 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K101749, K063324, K130451, K130902

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    uWS-MI is a software solution intended to be used for viewing, manipulation, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:

    The PET/CT Oncology application is intended to provide tools to display and analyze the follow-up PET/CT data, with which users can do image registration, lesion segmentation, and statistical analysis.

    The PET/CT Dynamic Analysis application is intended to display the dynamic PET image data and its associated timeactivity curve.

    The PET/CT Brain Analysis (NeuroQ™) application is intended to analyze the brain PET scan, give quantitative results of the relative activity of 240 different brain regions, and make comparison of activity of normal brain regions in AC database.

    The PET/CT Cardiac Analysis (ECTbTM) application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis and cardiac function analysis of the cardiac short axis.

    Device Description

    uWS-MI is a comprehensive software solution designed to process, review and analyze PET, CT and PET/CT patient studies. It can transfer images in DICOM 3.0 format over a medical imaging network or import images from external storage devices such as CD/DVDs or flash drives. These images can be functional data, such as PET as well as anatomical datasets, such as CT. It can be at one or more time-points or include one or more time-frames. Multiple display formats including MIP and volume rendering and multiple statistical analysis including mean. maximum and minimum over a user-defined region is supported. A trained, licensed physician can interpret these displayed images as well as the statistics as per standard practice.

    AI/ML Overview

    The acceptance criteria and study proving device performance are not explicitly detailed in the provided text in the format requested. However, based on the scattered information, I can infer some points and explicitly state what is missing.

    The device is uWS-MI, a software solution for viewing, manipulation, and storage of medical images, with specific applications for PET/CT Oncology, PET/CT Dynamic Analysis, PET/CT Brain Analysis (NeuroQ™), and PET/CT Cardiac Analysis (ECTb™).

    The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed performance study with specific acceptance criteria and detailed quantitative results.

    Here's an attempt to extract and synthesize the information based on your requested format:


    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/ApplicationAcceptance Criteria (Implied)Reported Device Performance
    General Image ProcessingEquivalent functionality to predicate device (K123920 Syngo.via)All listed general functions (2D/3D review, filming, fusion, ROI/VOI, MIP display, compare) are "Same" as predicate.
    PET/CT Dynamic AnalysisEquivalent functionality to reference device (K101749 Syngo™TrueD Software)All listed functions (ROI Analysis, Pseudo color, Automatic cine, Curve Analysis, Table Statistics, Save, Filming) are "Same" as reference device. Reframe/Rebin functionality is present but differs from reference device, with a remark that it "will not impact the safety and effectiveness of the device."
    PET/CT OncologyEquivalent functionality to reference device (K063324 GE PET VCAR)All listed functions (Compare display, Auto registration, Manual registration, Fix Segmentation, Adaptive Segmentation, Spread, Statistical Analysis, Save) are "Same" as reference device.
    PET/CT Brain Analysis (NeuroQ™)Equivalent functionality to reference device (K130451 NeuroQ™3.6)All listed functions (Reformat, Quality Control, Slice Display, Compare, PET/CT Fusion, EQuAL analysis, AmyQ, Save results, Capture region/display, Exit) are "Same" as reference device.
    PET/CT Cardiac Analysis (ECTb™)Equivalent functionality to reference device (K130902 Emory Cardiac Toolbox™3.2)All listed functions (Reconstruction, SSS, Polor Maps, Perfusion Analysis, Viability Analysis, Functional Analysis, Save results, Capture region/display, Exit) are "Same" as reference device.
    Safety and EfficacyOverall safety and effectiveness profile similar to predicate device."uWS-MI was found to have a safety and effectiveness profile that is similar to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). However, it mentions "Performance evaluation Report for PET/CT Oncology and Image Fusion" as part of Performance Verification. Details about this report, including the data used, are not provided. Given the nature of a 510(k) summary, the data often comes from internal testing designed to show equivalence, rather than large-scale clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Ground truth establishment details are absent.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not specifically mentioned or detailed. The document focuses on demonstrating substantial equivalence based on features and software verification, and states "No clinical study was required." for general device clearance. The performance of specific applications (NeuroQ™ and ECTb™) leverages their prior clearances, implying that their functionalities, rather than a new comparative effectiveness study of the full uWS-MI system, were reviewed.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    The device is described as "stand-alone software," and its primary verification listed is "Software Verification and Validation." This suggests that the performance verification would have evaluated the algorithm's output independently, as a standalone system. However, the exact metrics and results of this standalone performance are not explicitly provided beyond the statement of functional equivalence to predicate/reference devices.

    7. Type of Ground Truth Used

    The type of ground truth used for specific performance evaluations is not explicitly stated. However, considering the device's function as an image post-processing and analysis tool, the ground truth would typically involve:

    • Expert consensus or expert readings: For evaluating the accuracy of image analysis, measurements, or lesion segmentation outputs.
    • Pathology or clinical outcomes data: For validating the diagnostic accuracy of features, particularly for oncology applications, though "no clinical study was required" suggests this was not the primary method for this 510(k).
    • Phantom or synthetic data: For testing algorithm robustness and accuracy in controlled environments.

    Since the PET/CT Brain Analysis (NeuroQ™) and PET/CT Cardiac Analysis (ECTb™) applications were cleared previously, their ground truth would have been established during their initial clearances, likely involving expert-derived truth based on clinical images or phantoms.

    8. Sample Size for the Training Set

    The document does not provide information regarding the sample size of any training set. As a 510(k) for a software device, the focus is often on functional equivalence and verification/validation, rather than detailing the machine learning model training process if such models are embedded.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162337

    Validate with FDA (Live)

    Device Name
    Symbia 6.5
    Date Cleared
    2016-09-08

    (17 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K151752, K101749

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

    SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; Planar imaging, and tomographic imaging for isotopes with energies up to 588 keV.

    CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles.

    SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.

    Software: the syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT and other imaging modalities.

    Device Description

    The Siemens Symbia systems consist of Single Photon Emission Computed Tomography (SPECT) scanners and integrated hybrid X-Ray Computed Tomography (CT) and SPECT scanners. The SPECT subsystem images and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and integrates CT's anatomical detail for precise reference of the location of the metabolic activity. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The system can be used as an integrated SPECT and CT modality while also enabling independent functionality of SPECT and CT as standalone diagnostic imaging devices.

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification from Siemens Medical Solutions, USA, Inc. to the FDA for their Symbia 6.5 device. This document primarily focuses on establishing substantial equivalence to a predicate device (Symbia 6.0) rather than presenting a detailed clinical study for a novel AI/software component, which is what the prompt's questions imply by asking for "human readers improve with AI vs without AI assistance" or "stand alone (i.e. algorithm only without human-in-the-loop performance)."

    The document discusses updates to the Symbia device, including upgraded software and the integration of commercially marketed CT software and a viewing application. The "syngo MI Applications software" is described as a display and analysis package intended to aid clinicians. However, the performance testing section does not describe a clinical study of an AI algorithm in the way the prompt specifies. Instead, it focuses on physical performance characteristics of the SPECT/CT system, specifically quantitative accuracy using phantoms, and refers to electrical, mechanical, and radiation safety standards.

    Therefore, many of the requested details about acceptance criteria and study design for an AI-powered diagnostic device are not present in this document. The document describes performance testing for a medical imaging system, not a specific AI-driven diagnostic or assistive algorithm for image interpretation that would involve human readers or ground truth established by experts.

    Given the information provided in the document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria and performance reported relate to the quantitative accuracy of the SPECT system, not a specific AI diagnostic algorithm.

    Acceptance CriteriaReported Device Performance
    Absolute quantification accuracy of the system shall be within 10% in phantoms for objects larger than three times the system resolution when acquired for count rates up to 160 kcps.Pass (Deviation from true <=10%) for both 3/8" and 5/8" phantoms

    Study Information (Based on the document's content)

    The document describes phantom testing for system performance, not a clinical study involving human readers or AI algorithm performance on patient data. Therefore, many of the prompt's questions are not applicable or cannot be answered from the provided text.

    1. Sample size used for the test set and the data provenance: Not applicable in the context of an AI study. The "test set" here refers to phantoms (3/8" and 5/8" NEMA IEC phantoms). Data provenance is from testing within the Siemens system (Symbia Intevo, SW version VB20A).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for phantom testing is the known physical properties of the phantom and the expected system response, not expert consensus on patient images.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Phantom testing does not involve adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No. The document does not describe an MRMC study comparing human readers with or without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No. The performance testing is for the overall imaging system (SPECT/CT), not a standalone AI algorithm interpreting images. The software "syngo MI Applications" is described as a "display and analysis package intended to aid the clinician," suggesting human-in-the-loop, but no performance study of this aid is detailed.
    6. The type of ground truth used: For the performance testing described, the ground truth is the known properties of the NEMA IEC phantoms (e.g., true activity concentration). It is not expert consensus, pathology, or outcomes data from patients.
    7. The sample size for the training set: Not applicable. The document does not describe the development or training of an AI algorithm from a dataset.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document is a regulatory submission for a medical imaging device (SPECT/CT system) demonstrating substantial equivalence, focusing on hardware and core software updates, and basic quantitative performance using phantoms. It does not present a clinical study or performance evaluation of an AI-driven diagnostic component in the manner implied by the prompt's detailed questions about AI acceptance criteria and study design.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123920

    Validate with FDA (Live)

    Device Name
    SYNGO.VIA
    Date Cleared
    2013-01-18

    (29 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K123375, K120579, K112020, K121434, K101749

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    syngo.via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo.via supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.

    Device Description

    syngo.via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo.via supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S. The system is a software only medical device. It defines minimum requirements to the hardware it runs on. The hardware itself is not seen as a medical device and not in the scope of this 510(k) submission. It supports the physician in diagnosis and treatment planning. syngo.via also supports storage of Structured DICOM Reports. In a comprehensive imaging suite syngo.via integrates Radiology Information Systems (RIS) to enable customer specific workflows. The predicate device, syngo.via allows for the use of a variety of advanced applications (clinical applications) These applications are medical devices on their own rights and filed separately. They are not part of this 510(k) submission and not part of the syngo.via medical device. syngo.via has a universal component called generic reader application which is part of this medical device and it allows no newly introduced imaging and post processing algorithms compared to the above mentioned predicate devices. syngo.via is based on Windows. Due to special customer requirements and the clinical focus syngo.via can be configured in the same way as the predicate device with different combinations of syngo- or Windows based software options and clinical applications which are intended to assist the physician in diagnosis and/or treatment planning. This includes commercially available post-processing software packages.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding acceptance criteria and supporting studies:

    This 510(k) pertains to "syngo.via," a PACS system. Based on the document, this 510(k) is for enhanced functionalities of syngo.via, making it an update or extension of a previously cleared device (K123375). The key here is that it's not a new AI algorithm designed for a
    specific diagnostic task with associated performance metrics. Instead, it seems to be primarily a software platform update that integrates functionalities already present in other cleared Siemens products.

    Therefore, the typical structure for reporting AI/CADe/CADx device performance (sensitivity, specificity, AUROC, etc.) involving a test set, ground truth, and expert readers is not applicable in this submission. The "acceptance criteria" discussed are likely related to software verification and validation, adherence to standards, and demonstrating substantial equivalence to existing devices with similar functionalities.


    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned above, this 510(k) is for an enhanced PACS system and does not present specific diagnostic performance metrics. The "performance" is primarily demonstrated through compliance with standards and equivalence to predicate devices. There are no explicit quantitative acceptance criteria for diagnostic performance in terms of sensitivity, specificity, etc., as it's not a new diagnostic algorithm.

    The "performance" described is in terms of:

    • Software Functionality: Viewing, manipulation, communication, and storage of medical images.
    • Integration: HL7-/DICOM-compatible RIS workflow.
    • Technological Characteristics: Runs on Windows OS, supports DICOM images, image data compression (lossless and lossy).
    • Imaging Algorithms (inherited/similar to predicates): MPR, MIP, MinIP, VRT, SSD, Digitally Reconstructed Radiograph, Editor functionality, Registration, Region Growing, Quantitative measurements.
    • Automatic Spine Labeling (inherited/similar to predicates): Anatomy Labeling of Vertebra bodies, automatically suggested labels with manual override.
    Acceptance Criteria CategoryReported Device Performance/Characteristics
    Intended Use Fulfillmentsyngo.via is intended for viewing, manipulation, communication, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions.
    Technological CharacteristicsSoftware-only system (runs on specified IT hardware). Backend: Windows 2008. Client: Windows XP, Vista, 7. Supports DICOM formatted images and objects. Image data compression: Lossless (factor 2-3), lossy (higher rate). Receives/decompresses JPEG2000. Incorporates imaging algorithms like MPR, MIP, MinIP, VRT, SSD, DRR, Editing, Registration, Region Growing, Quantitative measurements (distance, angle). Supports Automatic Spine Labeling: Anatomy Labeling of Vertebra bodies, with automatically suggested labels and manual override. Supports multi-time point registration and user verification.
    IntegrationWorkflow Management with HL7-/DICOM-compatible RIS (IHE Year 5).
    Safety and Effectiveness ControlsSoftware verification and validation (Unit, Integration, System Test Levels) performed according to: DICOM Standard [2011], ISO/IEC 15444-1:2005+TC 1:2007, ISO/IEC 10918-1:1994 + TC 1:2005, HL7 [2006], IEC 62304:2006, IEC 62366:2007, ISO 14971:2007, IEC 60601-1-4:2000. Risk analysis performed to identify and control potential hazards. Device labeling contains instructions, cautions, and warnings. Adheres to recognized industry practices and standards. Supports quality assurance methods (e.g., SMPTE, HIPAA). Major software self-tests/checks are performed. Device is a post-processing software with no capability to control connected modalities.
    Substantial EquivalenceDemonstrated substantial equivalence to several Siemens predicate devices (syngo.via K123375, SOMATOM Definition Edge CT System K120579, syngo.CT Vascular Analysis K112020, Software syngo MR D13A K121434, syngo TrueD K101749) by incorporating similar functionalities without introducing new significant safety risks.

    2. Sample size used for the test set and the data provenance

    The document does not describe a "test set" in the context of diagnostic performance evaluation (e.g., a set of medical images used to evaluate an algorithm's diagnostic accuracy). The testing performed was software verification and validation testing at Unit, Integration, and System levels, as per IEC 62304. This type of testing uses various software inputs and configurations to ensure functional correctness, rather than a diagnostic image dataset. No specific sample size of images or data provenance (country, retrospective/prospective) is provided because it's not relevant for this type of submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. There was no diagnostic "test set" requiring expert ground truth for diagnostic accuracy evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no diagnostic "test set" requiring expert ground truth or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This 510(k) does not present an MRMC study comparing human reader performance with and without AI assistance, as it is a PACS system enhancement, not a new AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not a standalone diagnostic algorithm. syngo.via is a platform for viewing, manipulation, communication, and storage of medical images, intended to "support the physician in diagnosis and treatment planning." The functionalities described (like automatic spine labeling) are features within this broader platform, and their performance is indicated as being similar to those from previously cleared predicate devices.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. As there was no diagnostic test set in the traditional sense, there was no ground truth for diagnostic accuracy established through expert consensus, pathology, or outcomes data. The "ground truth" for the software's functional performance would be defined by the software requirements and design specifications, verified through testing procedures.

    8. The sample size for the training set

    Not applicable. This 510(k) does not describe a new AI algorithm that requires a training set. The enhanced functionalities are stated to have "similar technological characteristics as the predicate device" and incorporate "imaging and post processing algorithms compared to the above mentioned predicate devices." This implies that any underlying algorithms for features like "Automatic Spine Labeling" are either existing, well-established, or derived from components previously cleared, rather than newly developed and trained models.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for a new algorithm in this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1