Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K183045

    Validate with FDA (Live)

    Device Name
    Eclipse 2L
    Manufacturer
    Date Cleared
    2019-04-30

    (179 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K101570

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECLIPSE 2L is indicated for use in the blood vessels of the peripheral and neurovasculature where temporary occlusion is desired. These catheters offer a vessel selective technique of temporary vascular occlusion which is useful in selectively stopping or controlling blood flow and for balloon assisted embolization of intracranial aneurysms.

    ECLIPSE 2L is indicated for use in the peripheral vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic devices, such as embolization coils.

    ECLIPSE 2L is also indicated for neurovascular use for the infusion of diagnostic agents, such as contrast media, and therapeutic devices, such as DMSO based embolization materials, that have been approved for use in the neurovasculature and are compatible with the diameter of the ECLIPSE 2L.

    Device Description

    ECLIPSE 2L balloon catheters are catheters with dual lumen arranged in a parallel configuration. The guidewire lumen or the working lumen is DMSO compatible and can be used for infusion/delivery of diagnostic and therapeutic devices. The inflation lumen is used for the inflation and deflation of the balloon. The inflation of the balloon can be achieved independently of the presence of the guidewire. ECLIPSE 2L balloon catheter is equipped with a non-detachable low inflation pressure super compliant balloon welded to the distal end of the catheter.

    The ECLIPSE 2L balloon catheter incorporates 3-4 radiopaque markers. The radiopaque marker bands are located at the distal end of the working lumen and at either end of the balloon to facilitate fluoroscopic visualization during procedure. The short nose (SN) variant has an additional proximal marker located 3 cm from the distal tip of the catheter to aid in positioning detachable coils. The ECLIPSE 2L also has distal purge air from the balloon prior to use. The catheter and the outer surface of the balloon is coated with hydrophilic coating to reduce friction and increase lubricity. An introducer sheath is provided with the device as an accessory.

    AI/ML Overview

    The document provides acceptance criteria and study results for the ECLIPSE 2L device. Here is the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Performance Bench Testing
    Corrosion ResistanceAll test samples meet acceptance criteria.All test samples met acceptance criteria.
    Particulate Matter CharacterizationAll test samples meet acceptance criteria.All test samples met acceptance criteria. Particulate matter characterization of the Subject Device was comparable to Predicate Device.
    Coating Integrity and AdherenceAll test samples meet acceptance criteria.All test samples met acceptance criteria.
    Compatibility and Coil RemodelingNo coating particulate shall be observed during and post simulated use.All test samples met acceptance criteria.
    Simulated Use/Coating Lubricity and DurabilityNo coating particulate shall be observed during and post simulated use. Overall performance comparable or better than competitor devices.All test samples met the acceptance criteria. The overall performance of the Subject Device is equivalent or better than the Predicate Device.
    Flow Rate(For reference only)Testing was conducted for reference only.
    Dynamic BurstCatheter hub and shaft junctions shall not burst.All test samples met the acceptance criteria.
    Liquid LeakNo liquid shall be leaking from hub and catheter shaft under pressure.All test samples met the acceptance criteria.
    Air LeakAir bubbles shall not leak into the hub assembly during aspiration via syringe.All test samples met the acceptance criteria.
    Static BurstGuidewire lumen shall not burst per specification.All test samples met the acceptance criteria.
    Torque StrengthComparison of torque strength to Predicate Device.All test samples met the acceptance criteria. The torque strength of the Subject Device is equivalent or better to the Predicate Device.
    Physical AttributesDimensions inspect and measure per product specification.All test samples met the acceptance criteria.
    Dead Space Volume(For reference only)Testing was conducted for reference only.
    Balloon Deflation TimeDetermine time required to deflate balloon from rated burst volume.All test samples met the acceptance criteria.
    Balloon ComplianceMeasure balloon OD while inflating incrementally to max rated inflation volume.All test samples met the acceptance criteria.
    Balloon FatigueNo leakage or rupture after 20 inflation/deflation cycles to rated burst volume.All test samples met the acceptance criteria.
    Balloon Burst VolumeDetermine burst volume by incrementally inflating until burst failure.All test samples met the acceptance criteria.
    Kink ResistanceNo OD change or kinking when wrapped with appropriate pin gauge (180° turn), comparable to Predicate Device.All test samples met the acceptance criteria. The kink resistance performance of the Subject Device is equivalent or better to the Predicate Device.
    Surface ContaminationEntire catheter surface free of particulate under 2.5x magnification.All test samples met the acceptance criteria.
    Tensile ForcePeak tensile strength of distal tip ≥ 3N (for OD 0.55mm to 0.75mm); Peak tensile strength of catheter bonds ≥ 5N (for OD 0.75mm to 1.15mm).All test samples met the acceptance criteria.
    Gauge TestReference conical gauge and port opening align in limit planes; no rocking.All test samples met the acceptance criteria.
    Ease of AssemblySatisfactory fit with inflation and guidewire ports and reference fittings.All test samples met the acceptance criteria.
    Unscrewing TorquePort remains attached to reference fitting after specified unscrewing torque.All test samples met the acceptance criteria.
    Resistance to OverridingReference fitting does not override threads of port after specified torque.All test samples met the acceptance criteria.
    Separation ForcePort remains attached to reference fitting after specified max force.All test samples met the acceptance criteria.
    Stress CrackingNo evidence of stress cracks on the hub.All test samples met the acceptance criteria.
    Catheter Flexural FatigueAcceptable if Simulated Use, Dynamic Burst, Liquid Leakage, Air Leakage, Static Burst tests passed.All test samples met the acceptance criteria.
    DMSO CompatibilityAcceptable if Simulated Use, Dynamic Burst, Liquid Leakage, Air Leakage, Static Burst, Balloon Deflation, Balloon Compliance, Balloon Fatigue, Balloon Burst, Tensile Strength tests passed after DMSO Conditioning.All test samples met the acceptance criteria.
    Packaging Integrity Test
    Visual InspectionPackaging free of damage, seals free of defects, labels legible and intact.All test samples met the acceptance criteria.
    Bubble Leak TestAll samples must exhibit no leaks.All test samples met the acceptance criteria.
    Seal StrengthPerform seal strength test per ASTM F88/F88M-15.All test samples met the acceptance criteria.
    Performance Animal Testing
    Animal Testing (GLP)Evaluate in vivo performance, trackability, handling, radiopacity, and balloon occlusion, comparable to predicate devices.All test samples met the acceptance criteria. Trackability and handling of the catheter, radiopacity, and the balloon test occlusion were comparable to predicate device.
    Biocompatibility Testing
    Cytotoxicity - ExtractionAverage grade of 0 (no reactivity).Non-cytotoxic
    Sensitization - ISO Guinea Pig Maximization Sensitization TestSensitization response of “0”. No delayed sensitization.Did not elicit sensitization response.
    Irritation or Intracutaneous - ISO Intracutaneous Study in RabbitsDifferences in mean test and control scores of dermal observations ≤ 1.0.Non-irritant
    Acute Systemic Toxicity - ISO Acute Systemic Toxicity Study in MiceNo signs of toxicity, body weight changes within acceptable parameters.Non-toxic
    Systemic Toxicity - ISO Materials Mediated Rabbit PyrogenNo temperature rise ≥ 0.5 °C.Non-pyrogenic
    Hemocompatibility-Hemolysis (Direct)Test article in non-hemolytic range.Non-Hemolytic
    Hemocompatibility - ASTM Hemolysis (Extract Method)Test article in non-hemolytic range.Non-Hemolytic
    Hemocompatibility – Complement Activation – SC5b-9Statistically less activation than predicate device.Passed
    Hemocompatibility – Partial Thromboplastin Time (PTT)Not statistically different from predicate device; clotting times similar. All acceptance criteria met.Passed
    Hemocompatibility – Platelet & Leukocyte Counts (PLC)Not statistically significant (p>0.05) compared to comparison article.Passed
    Sterilization
    Original Sterilization Validation and adoptionValidation demonstrated reliable and consistent sterilization to SAL of 10-6. Fractional, half, and full cycles met acceptance criteria. Subject device successfully adopted.The validation study demonstrated that the sterilization process and equipment are capable of reliably and consistently sterilizing the devices to a minimum SAL of 10-6. Fractional cycles, half cycles, and full cycles met the acceptance criteria. The subject device was successfully adopted to the original sterilization validation by performing one (1) full cycle.
    EO and ECH ResidualsBelow limits specified in ISO 10993-7.The residual traces of EO and ECH for the subject device are below the limits specified in ISO 10993-7.
    Bacterial Endotoxin Levels< 2.15 EU/device< 2.15 EU/device
    Shelf LifeAll acceptance criteria met after accelerated aging.All acceptance criteria were met after 1 year of accelerated aging.
    PackagingAll acceptance criteria met after accelerated aging.All acceptance criteria were met after 1 year of accelerated aging.

    2. Sample sizes used for the test set and the data provenance
    The document does not explicitly state specific sample sizes for each test in the acceptance criteria, nor does it specify data provenance (country of origin, retrospective/prospective). It generally states "All test samples" met the criteria. For animal testing, it mentions evaluation in an "acute porcine model."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not provided in the document. The tests seem to rely on objective measurements and established standards (e.g., ISO, ASTM) rather than expert consensus on a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    This information is not applicable and not provided. The testing described involves bench and animal studies with objective measurements against predefined criteria, not diagnostic interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This information is not applicable and not provided. The device (ECLIPSE 2L) is a physical medical device (occlusion balloon catheter), not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This information is not applicable and not provided, as the device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    For the performance and biocompatibility studies, the "ground truth" is based on:

    • Established industry standards: ISO, ASTM specifications (e.g., ISO 10555-1 for static burst, ISO 594 for gauge testing, ISO 10993 for biocompatibility, ISO 11135 for sterilization).
    • Predefined acceptance criteria: Specific quantitative and qualitative thresholds for various physical, chemical, and biological properties.
    • Comparison to predicate devices: For several performance aspects (e.g., particulate matter, simulated use, torque strength, kink resistance, animal testing trackability).
    • Direct measurement: For various physical attributes like dimensions, flow rates, burst volumes, and deflation times.
    • In vivo observation: For animal testing.

    8. The sample size for the training set
    This information is not applicable and not provided. The device is a physical product and does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established
    This information is not applicable and not provided, as there is no "training set" for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1