Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K161506

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-26

    (117 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Aspiration Catheters and Separators: As part of the Penumbra Aspiration System, the Penumbra Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.
    Penumbra Aspiration Tubing: As part of the Penumbra Aspiration System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Aspiration Catheters to the Penumbra Aspiration Pump.
    Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the coronary and peripheral vasculature. The Aspiration Catheter and Separator are available in multiple configurations. The devices are provided sterile, non-pyrogenic, and intended for single use only. Intended users for this device are physicians who have received appropriate training in interventional radiology. The Penumbra Aspiration System is designed to remove thrombus from the coronary and peripheral vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054 in or larger. The Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The Aspiration Catheter is used with the Penumbra Pump MAX to aspirate thrombus from an occluded vessel. As needed, the Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Penumbra Pump MAX, which is connected using the Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Separator is provided with an introducer and torque device. The Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheter and Separator are visible under fluoroscopy. The Penumbra Pump MAX is the aspiration source for the Penumbra Aspiration System. The Penumbra Pump MAX operates using AC power and is designed to be portable if needed. The Penumbra Pump MAX provides vacuum of up to 29 inHg. The pump is available in both 110Vac and 230Vac versions. The front face of the pump has a display panel with a vacuum gauge, vacuum regulator dial, and power switch. The pump connects to the canister reservoir with a tubing assembly (Penumbra Pump/Canister Tubing), which is provided as an accessory. The Penumbra Pump/Canister Tubing consists of a short tubing segment with an inline filter and connectors on each end to facilitate attachment to the pump's vacuum port. The tubing is provided pre-attached to the canister reservoir lid. The Penumbra Pump/Canister Tubing is provided non-sterile and is used outside the sterile field.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Penumbra Aspiration System. It details the device's indications for use, its classification, and a comparison to predicate and reference devices. However, the document primarily focuses on demonstrating substantial equivalence through non-clinical (benchtop) and animal study data, largely leveraging existing data from previously cleared similar devices.

    Crucially, this document does NOT contain information about specific "acceptance criteria" or "device performance" in terms of clinical outcomes, nor does it describe a study involving human subjects or experts for establishing ground truth for AI/algorithm performance.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, along with the specified sub-points (table of acceptance criteria/performance, sample size, experts, adjudication, MRMC study, standalone performance, type of ground truth, training set details) because this information is not present in the provided text.

    The document discusses:

    • Leveraged Non-Clinical Data (Section 1.10): This section lists benchtop tests (Friction, Visual & Dimensional, Pouch Seal Strength, Flow Rate, Tensile, Elongation, Bond Strength, Hub Air Aspiration, Torsion, Burst, Simulated Use, Particulate, Flexibility, Packaging, Corrosion). It states that "all established requirements and acceptance criteria were met," but it does not specify what those acceptance criteria were in measurable terms (e.g., "flow rate X L/min," "tensile strength Y N"). It also does not define "device performance" in the way requested (e.g., diagnostic accuracy, sensitivity, specificity).
    • Leveraged Animal Studies (Section 1.11): This section summarizes findings from design validation in GLP animal testing, noting "No vessel injury," "No abnormal gross or histology findings," and "no significant vascular response." This is performance data in an animal model, but not clinical performance in humans, nor does it relate to acceptance criteria for an AI/algorithm-driven device.

    In summary, the provided text describes the regulatory clearance process for a medical device (an aspiration system) based on substantial equivalence to existing devices, supported by benchtop testing and animal studies. It does not contain the kind of information typically found in a study proving an AI/algorithm-driven device meets specific clinical acceptance criteria, which would involve human data, expert review, and metrics like accuracy, sensitivity, or specificity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160449

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-05-25

    (97 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators: As part of the Penumbra System, the Reperfusion Catheters and Separators are in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
    Penumbra Aspiration Tubing: As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Pump MAX.
    Penumbra Pump MAX: The Penumbra Pump MAX is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System is designed to remove thrombus from the neurovasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an I.D. of 0.054in [1.37mm] or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Pump MAX to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Penumbra Pump MAX, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy. The Penumbra Reperfusion Catheter, Separator, and Aspiration Tubing are provided sterile, non-pyrogenic, and intended for single use only. Additionally, a pre-packaged configuration (KIT packaging) for all Penumbra System Reperfusion Catheters with Aspiration Tubing is available.

    The Penumbra Pump MAX is designed to provide aspiration for the Penumbra System. The Penumbra Pump MAX operates using AC power. The Penumbra Pump MAX provides vacuum of up to 29 inHg and is available in both 110Vac and 230Vac versions. The Penumbra Pump MAX and Pump/Canister Tubing are provided non-sterile and is used outside the sterile field.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focused on the KIT configuration, as that's where new testing was performed:

    1. Table of Acceptance Criteria and Reported Device Performance (KIT Configuration)

    AttributeSpecificationAcceptance CriteriaReported Performance
    Visual InspectionThese evaluations confirm that the units used in this Design Verification testing meet all inspection criteria for release of finished goods (clinically acceptable) product.Products meet all inspection criteria.Pass
    Simulated Use (Intracranial Access, Vessel Access Entry Performance & Clot Removal)Simulated use testing of the Catheter and Separator was performed with accessory devices in an anatomical model which simulated the tortuosity of the neurovasculature. Devices were delivered through the tortuous anatomical model to evaluate the effectiveness of the devices to remove clots and that the Reperfusion Catheter does not collapse under vacuum.Devices effectively remove clots and catheters do not collapse under vacuum.100% Pass
    Catheter CoatingCoating has not delaminated, peeled, or flaked after simulated use.Coating remains intact (not delaminated, peeled, or flaked).100% Pass
    Gross LeakPouch seals, pouch front, and pouch back material do not leak.No leaks in pouch seals, front, or back material.100% Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the "Design Verification testing" for the KIT configuration. It only mentions "units used in this Design Verification testing" for Visual Inspection, and implicitly for Simulated Use, Catheter Coating, and Gross Leak tests as they all report 100% Pass.

    • Sample Size: Not explicitly stated as a number, but all tested units passed.
    • Data Provenance: The tests are described as "additional testing performed for the KIT configuration packaging," suggesting these are prospective, benchtop tests conducted by Penumbra, Inc. The country of origin for the data generation would logically be the USA, given the company's location.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts or clinicians being used to establish a ground truth for these non-clinical tests. The tests are focused on physical and mechanical properties and simulated use in anatomical models.

    4. Adjudication Method for the Test Set

    Not applicable, as these are non-clinical, objective tests and not based on expert review or consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This document describes a medical device (Penumbra System and Penumbra Pump MAX) for treating acute ischemic stroke, not an AI-based diagnostic or treatment guidance system that would involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for these non-clinical tests is based on:
    * Defined specifications and performance expectations: For Visual Inspection, Catheter Coating, and Gross Leak, the "ground truth" is adherence to predefined quality and integrity standards.
    * Simulated physiological conditions and objective performance: For Simulated Use, the "ground truth" is the effective removal of clots and prevention of catheter collapse in an anatomical model that mimics neurovasculature.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device, not a machine learning model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142458

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-05-22

    (262 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K072718, K090752, K100769, K113163, K133317

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. The Reperfusion Catheters ACE 64 and ACE 68 are intended for use in revascularization within the Internal Carotid Artery (ICA) within 8 hours of symptom onset.

    Device Description

    The Penumbra System ACE components are additional components to the currently available Penumbra System / Penumbra System MAX. The Penumbra System ACE components provide a larger lumen to assist in the removal of thrombus from the neurovasculature. The devices are provided sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Penumbra System ACE 64 and ACE 68 Reperfusion Catheters:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device PerformancePass / Fail
    Biocompatibility Testing
    In Vitro CytotoxicitySample extracts must yield cell lysis grade 2 or lowerGrade 1: SlightPass
    SensitizationTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control Grade < 1)Grade 0: No visible changePass
    Acute Intracutaneous ReactivityThe difference in the mean test article and mean control score must be grade 1.0 or lowerGrade ≤ 1.0 difference between mean test article and mean control scorePass
    Acute Systemic ToxicitySample extracts must not cause: > 10% weight loss in 3 or more test animals, Mortality of 2 or more test animals, Abnormal behavior in 2 or more test animalsNo evidence of systemic toxicity from sample extracts: No weight loss (all gained weight), No death, All test animals appeared normalPass
    Rabbit Pyrogen StudySample Extracts must not cause a total rise in body temperature of ≥ 0.5°CNon-pyrogenic: No evidence of material-mediated pyrogenicity; no single animal had a total body temperature rise of ≥ 0.5°CPass
    Hemocompatibility In Vitro HemolysisSample extracts must be non-hemolytic (≤ 2% hemolytic index)Non-hemolytic: Hemolytic Index = 0.70%, Corrected Hemolytic index = 0.00%Pass
    Complement ActivationThe concentrations of C3a and SC5b-9 in the test samples are statistically similar to the predicate (Exposure Control & Ref Material) control and statistically lower than the positive control for all exposure timesThe test sample concentrations of C3a and SC5b-9 were statistically similar or lower than the predicate control sample concentrations, and statistically lower than the positive control sample concentrations at all three exposure timesPass
    Dog ThrombogenicityThe device must be non-thrombogenic after 4 hours in vivo when compared to a control device (Boston Scientific Excelsior SL-10 microcatheter)No significant thrombosis with a Grade of 0 was observed in 2 out 2 test site and 2 out of 2 control sites. Based on the evaluation criteria, the amount of thrombosis was not considered significantPass
    Bench-top Testing
    Dimensional/Visual InspectionUnits meet all inspection criteria for release of finished goods (clinically acceptable) product.PassPass
    Simulated Use (Intracranial Access, Vessel Access Entry Performance & Clot Removal)Effectiveness of devices to remove clots and that Reperfusion Catheter does not collapse under vacuum.100% PassPass
    Coating IntegrityCoating has not delaminated, peeled, or flaked after simulated use.100% PassPass
    Particulate Testing (Hydrophilic Coating)Max particles: ≥ 10 um ≤ 6000 particles; ≥ 25 um ≤ 600 particles.10um 100% Pass; 25 um 100% PassPass
    Particulate Testing (Catheter/Separator)Max particles: ≥ 10 um ≤ 6000 particles; ≥ 25 um ≤ 600 particles.10μm 100% Pass; 25 um 100% PassPass
    Coating Integrity (after particulate testing)Coating is not grossly damaged after undergoing particulate testing.100% PassPass
    Hub/Catheter Air AspirationWhen negative pressure is pulled, no air may leak into hub.100% PassPass
    Pressure Test45 psi for 30 sec MIN100% PassPass
    Reperfusion Catheter/Sheath or 8F Guide & 0.014" Guidewire compatibility (Friction Force)Maximum value per specification.100% PassPass
    Markerband Section Bond StrengthMinimum value per specification.100% PassPass
    Joint Sections Bond StrengthMinimum value per specification.100% PassPass
    Hub to Shaft & Hub to Hypotube Bond StrengthMinimum value per specification.100% PassPass
    Steam-Shaped Distal Tip TensileMinimum value per specification.100% PassPass
    Elongation to Failure% Elongation ≥ 5%100% PassPass
    Kink ResistanceNo kinking when formed in a defined radius.100% PassPass
    CorrosionNo visible corrosion on Reperfusion Catheter immediately after Corrosion Testing procedure.100% PassPass
    Animal Study
    Vessel InjuryNo vessel injury on final angiograms.No vessel injury was noted.Pass
    Gross/Histology FindingsNo abnormal gross or histology findings in test vessel segments.No abnormal gross or histology findings were noted.Pass
    Vascular ResponseNo significant vascular response.No significant vascular response.Pass

    2. Sample Size and Data Provenance (for test sets, where applicable)

    • Biocompatibility Testing:

      • In Vitro Cytotoxicity: Not specified in terms of sample size for the test itself, but implies multiple samples for extract testing.
      • Sensitization: "Test Group" - size not specified.
      • Acute Intracutaneous Reactivity: "mean test article and mean control score" - implies multiple samples, size not specified.
      • Acute Systemic Toxicity: "3 or more test animals" (for weight loss criteria), "2 or more test animals" (for morality/behavior) - implies at least 3 animals for each extract/test (likely mice or rats, as common for this test).
      • Rabbit Pyrogen Study: "No single animal" suggests multiple rabbits, specific number not stated.
      • Hemocompatibility In Vitro Hemolysis: Sample extracts, size not specified.
      • Complement Activation: "test samples" compared to "predicate (Exposure Control & Ref Material) control" and "positive control" at "all three exposure times." Specific number of samples not stated.
      • Dog Thrombogenicity: "2 out of 2 test site and 2 out of 2 control sites" - suggests at least two test animals (dogs) for the in-vivo evaluation.
      • Data Provenance: Retrospective, conducted by the manufacturer, or by external labs following GLP. Country of origin not specified, but following EN ISO 10993 guidelines.
    • Bench-top Testing:

      • Sample sizes are not explicitly stated for all individual tests, but most indicate "100% Pass," which suggests the testing was performed on a sample of devices and all met the criteria. For particulate testing, the maximum number suggests a specific measurement from a sample.
      • Data Provenance: Retrospective, conducted by the manufacturer.
    • Animal Study:

      • Sample Size: A "swine model" was used. The number of individual animals (swine) is not explicitly stated, but the conclusions "No vessel injury was noted on the final angiograms following the vessel response procedure," "No abnormal gross or histology findings were noted in test vessel segments," and "The use of the Penumbra System ACE devices resulted in no significant vascular response in these experimental conditions," suggest sufficient animal subjects were used to support the claim.
      • Data Provenance: Prospective, animal study (GLP Animal Testing).

    3. Number of Experts Used and Qualifications (for ground truth establishment)

    • Biocompatibility Testing: Experts in toxicology, microbiology, and animal studies would have been involved in the design and interpretation of these studies. Their specific number and qualifications are not detailed in this summary.
    • Bench-top Testing: Engineers and material scientists within the manufacturer's R&D and Quality departments would have developed the specifications and assessed the results.
    • Animal Study: Veterinarians, interventionalists (to perform the procedures), pathologists (for gross and histology findings), and researchers expert in animal models for vascular devices.
    • Note: This document does not pertain to AI/ML or image data, so the concept of experts establishing ground truth for a test set (e.g., radiologist for image interpretation) as typically understood in AI studies is not directly applicable here. The "ground truth" here is physical/biological measurements and observations.

    4. Adjudication Method (for the test set)

    • Not applicable in the context of this device's non-clinical testing. Adjudication methods (like 2+1 or 3+1) are typically used for medical image interpretation where there is subjective assessment by multiple human readers (e.g., radiologists) that needs to be reconciled to establish a ground truth. The tests described are objective physical, chemical, or biological measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. This document describes the pre-market non-clinical testing of a physical medical device (catheter), not an AI/ML algorithm for diagnostic imaging or similar applications where human reader performance is augmented.

    6. Standalone (Algorithm Only) Performance Study

    • No, this is not an AI/ML device. Therefore, a standalone algorithm performance study was not performed.

    7. Type of Ground Truth Used

    • Biocompatibility Testing: The ground truth is objective biological and chemical reactions/measurements (e.g., cell lysis grade, inflammation scores, weight changes, temperature changes, hemolytic index, C3a/SC5b-9 concentrations, histopathology for thrombogenicity).
    • Bench-top Testing: The ground truth is objective physical and mechanical measurements against technical specifications (e.g., dimensions, force measurements, flow rates, visual integrity, particulate counts).
    • Animal Study: The ground truth is direct in-vivo observation and pathological assessment (e.g., angiographic evidence of injury, gross pathology findings, histological examination of vessel segments).

    8. Sample Size for the Training Set

    • Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" for such devices involves engineering design, material selection, and iterative prototyping based on established engineering principles and prior knowledge.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" in the AI/ML sense for this device. The development process relies on engineering specifications, material science data, and established test methods, rather than a labeled dataset for algorithm training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113163

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-23

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K072718,K090752,K100769

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.

    Device Description

    The Penumbra System MAX components are additional components of the currently available Penumbra System. The Penumbra System MAX components provide a larger lumen to assist in the efficient removal of thrombus from the brain. The devices are provided sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The Penumbra System® MAX components are additional components of the currently available Penumbra System. The summary of non-clinical data details the testing that substantiates the safe and effective performance of the Neuron MAX System and its substantial equivalence to predicate devices. This includes Biocompatibility testing, Design Verification (Bench-Top Testing), and an Animal study.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria for each test were "Pass," meaning the device met the established requirements for each attribute.

    Test / Test SubjectAttributeReported Device Performance (Result)
    Biocompatibility Testing
    In Vitro CytotoxicityNo evidence of cell lysis or toxicityNo evidence of cell lysis or toxicity
    Acute Intracutaneous Reactivity (Irritation)No evidence of irritationNo evidence of irritation
    Acute Systemic ToxicityNo evidence of systemic toxicityNo evidence of systemic toxicity
    Rabbit Pyrogen StudyNo evidence of material-mediated pyrogenicityNo evidence of material-mediated pyrogenicity
    SensitizationNon-SensitizingNon-Sensitizing
    Hemo-compatibility - In Vitro HemolysisNon-hemolyticNon-hemolytic
    Hemo-compatibility - In Vitro Coagulation (PT, PTT)Coagulation times are within the normal range; Non-ThrombogenicWithin normal range; Non-Thrombogenic
    Hemo-compatibility - Complement ActivationNo greater biological response than corresponding controlNo greater biological response than corresponding control
    Bench-top Testing (Reperfusion Catheters)
    Pouch SealPouch Seal StrengthPass
    Dimensional / Visual InspectionUnits meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Simulated Use [Intracranial Access & Thrombus Removal]Units meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXHub /Shaft & Hub / HypotubeTensile StrengthPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXJoint Tensile StrengthPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXHub Air AspirationPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXBurst TestPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXParticulate Testing (Hydrophilic Coating)Pass
    Reperfusion Catheter4MAX / Sheath or 8F Guide compatibility (Friction Force)Pass
    Reperfusion Catheter 4MAX / 0.014" Guidewire compatibility (Friction Force)Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 054 (Friction Force)Friction ForcePass
    Reperfusion Catheter 3MAX / 0.014" Guide wire compatibility (Friction Force)Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXFlow RatePass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXElongation to FailurePass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXCorrosionPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXTorsionPass
    Bench-top Testing (Separator 3MAX)
    Dimensional / Visual Inspection (all sizes)Units meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Simulated Use: Hub TransitionPass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceTrackingPass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX & Reperfusion Catheter CompatiblePass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX / Reperfusion Catheter / Pump and accessories / Aspiration Tubing (Aspiration Remove Clot)Pass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX and Reperfusion Catheter Compatibility (Separator Advance / Retract)Pass
    Separator Bond Joint TestSeparator 054 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 041 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 032 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 026 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 3MAX and Wire Joint Break ForcePass
    Animal Study
    Safe use of Penumbra System MAX in swine modelNo vessel injury noted on final angiograms following the vessel response procedure. No abnormal gross or histology findings in test vessel segments. No significant vascular response.No vessel injury; No abnormal findings; No significant vascular response

    2. Sample Size Used for the Test Set and the Data Provenance

    • Bench-top Testing: Sample sizes for individual tests ranged from 3 to 30 units (e.g., N=30 for Pouch Seal Strength, N=10 for Particulate Testing, N=3 for Flow Rate). The data provenance is internal laboratory testing ("All studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. All testing was performed using units which were 2x sterilized and met finished goods release requirements."). This is considered retrospective data from a manufacturing/design verification process.
    • Biocompatibility Testing: The document does not specify exact sample sizes for each biocompatibility test but states that "All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices," which implies standard laboratory animal testing where applicable (e.g., Rabbit Pyrogen Study, Acute Intracutaneous Reactivity, Acute Systemic Toxicity, Sensitization). This data is from controlled laboratory experiments.
    • Animal Study: The document mentions "An animal study was conducted to evaluate the safe use of the Penumbra System MAX in a swine model." The exact sample size for the animal study is not explicitly provided in the summary, but it refers to "experimental conditions" and "test vessel segments," suggesting a controlled study within a lab setting. This is prospective animal data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    Not applicable. The studies described are non-clinical (bench-top, biocompatibility, animal studies) and do not involve human diagnostic or treatment decision-making where expert consensus would establish ground truth for a test set. The "ground truth" here is determined by objective physical, mechanical, biological, or physiological measurements against predefined engineering and biological safety standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, these are non-clinical studies. The results are objective measurements/observations (e.g., "Pass," "No evidence of toxicity," "No vessel injury") rather than interpretations requiring adjudication among human experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes the safety and effectiveness testing of a physical medical device (catheter system) for mechanical thrombus removal, not an AI or imaging-based diagnostic device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document describes a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for these non-clinical studies was established by:

    • Predefined engineering specifications and performance parameters for mechanical and physical properties.
    • Established international and national standards for biocompatibility (e.g., ISO 10993, USP, ASTM, 21 CFR Part 58).
    • Physiological observations, angiographic evaluations, gross pathology, and histology in the animal model.

    8. The Sample Size for the Training Set

    Not applicable. This document refers to the testing of a physical medical device. There is no "training set" in the context of machine learning or AI. The design verification and biocompatibility testing involved various "test units" or "samples" for specific tests, as detailed in the table, to ensure the device met its design specifications.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, for the same reason as #8. Ground truth in this context refers to the defined specifications and standards the device was designed to meet and was tested against.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1