Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110906

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-07-26

    (117 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APEX Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic. Iumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).

    The APEX Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.

    The APEX Spine System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with deqeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

    When used in a percutaneous, posterior approach with AIM Spine MIS instrumentation, the APEX Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), turnor, pseudoarthrosis, and failed prvious fusion in skeletally mature patients . Levels of fixation are for the thoracic. lumbar and sacral spine.

    Device Description

    The purpose of this submission is to add cannulated screws to the APEX Spine System and add indications for use of the APEX Spine System in a posterior percutaneous approach and product code NKB. The APEX Spine System Cannulated Screws have exactly the same geometry of the APEX screws with the addition of a cannulation. The screws are available in a variety of diameters and lengths and can be used with the components of the previously cleared APEX Spine System.

    Materials: Titanium alloy per ASTM F136

    Function: The APEX Spine System is indicated for 1)non-cervical spinal fixation devices intended for posterior, non-pedicle fixation, 2)for non-cervical pedicle screw fixation and 3)hook fixation systems of the non-cervical spine, 4)for degenerative disc disease.

    AI/ML Overview

    The APEX Spine System is a pedicle screw spinal system. The submission (K110906) addresses the addition of cannulated screws to the system and new indications for use in a posterior percutaneous approach.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Performance: - Static Compression Bending - Static Torsion - Dynamic Compression BendingMet: Results were "equal or higher than the predicate systems" for all tests (ASTM F1717-09).
    Material Composition: - Titanium alloy per ASTM F136Met: The device utilizes Titanium alloy per ASTM F136.
    Biocompatibility: (Implicitly covered by substantial equivalence to predicates)Met: Substantially equivalent to predicate devices with established biocompatibility.
    Sterility: - Non-sterile, steam sterilized at hospitalMet: Device is non-sterile and intended for steam sterilization at the hospital.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for specific tests. The document refers to "testing" and "results," implying a sufficient number of samples were tested to generate comparative data.
    • Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted by the manufacturer or a contracted testing facility, following ASTM standards. This is a retrospective analysis in the context of device design and comparison to predicates.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable. This submission focuses on the mechanical and material equivalence of a spinal implant system to predicate devices, not on diagnostic accuracy requiring expert interpretation or ground truth establishment in a clinical sense.

    4. Adjudication Method for the Test Set

    • Not applicable. As noted above, this submission does not involve human readers or clinical interpretation that would require an adjudication method. The "adjudication" here is the comparison of mechanical test results against established predicate device performance and ASTM standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. This type of study is not applicable to a spinal implant system focused on mechanical performance and material composition.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This device is a physical spinal implant, not an AI algorithm.

    7. The Type of Ground Truth Used

    • Engineering/Performance Standards: The "ground truth" used for testing was primarily based on:
      • ASTM F1717-09: "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
      • Predicate Device Performance: The performance of previously cleared spinal systems (e.g., APEX Spine System (K062513), Viper Spine System (K090648), Polaris Spinal System (K103393), Synthes Pangea System (K052123)).
      • Material Specifications: ASTM F136 for Titanium alloy.

    8. The Sample Size for the Training Set

    • Not applicable. There is no "training set" in the context of this kinesiologic and material equivalence submission for a physical medical device. The device's design is based on established engineering principles and prior device knowledge, not a data-driven training process.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set, this question is not relevant. The "ground truth" in development of such a device is rooted in biomechanical principles, material science, and the established safety and effectiveness of predicate devices.

    Summary of the Study Proving Acceptance Criteria:

    The study summarized in K110906 is a non-clinical bench testing study designed to demonstrate the substantial equivalence of the modified APEX Spine System (with added cannulated screws and percutaneous approach indication) to its predicate devices.

    • Study Name: Non-clinical Test Summary
    • Study Type: Bench Testing, Mechanical Performance Comparison.
    • Methodology:
      • Tests were conducted according to ASTM F1717-09, "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
      • Specific tests included: Static Compression Bending, Static Torsion, and Dynamic Compression Bending.
      • The "acceptance criteria" here were that the results of these tests for the new device configuration should be "equal or higher" than those of the predicate devices.
    • Results: The new device configuration for the APEX Spine System with cannulated pedicle screws met these performance criteria, with results reported as "equal or higher than the predicate systems."
    • Conclusion: Based on these non-clinical tests, the manufacturer concluded that the APEX Spine System with the added cannulated pedicle screws is substantially equivalent to the predicate devices in terms of intended use, design, materials, performance, and function. No clinical studies were performed for this submission, as substantial equivalence was demonstrated through non-clinical testing and comparison to legally marketed predicate devices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1