Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162471

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-30

    (205 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K974899, K082182, K093244, K100237, K121022

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hach CM130 is an instrument that is intended for use by hemodialysis professionals to automatically monitor low levels of total chlorine (i.e. total chloramines plus free chlorine) in feed water used to prepare dialysate in hemodialysis systems.

    The CM130 is a component of the complete water treatment system for hemodialysis and does not treat or alter the water used in dialysate. The CM130 instrument's automated monitoring records total chlorine values in feed water at intervals between 5 and 20 minutes.

    Device Description

    The Hach CM130 Chlorine Monitor is a microprocessor-controlled analyzer used to monitor the chlorine content of water which is used to prepare dialysate for hemodialysis. The monitor is mounted in the water room in a location that allows sampling of feed water between the primary (scrubbing) and secondary (polishing) carbon tanks. Water that flows into the CM130 exits the device to a drain. The water samples, once tested, are also discarded to a drain.

    The monitor measures and displays total chlorine (free chlorine and combined chloramines) as Cl2, ranging from 0.03 to 0.20 mg/L.

    The CM130 Chlorine Monitor employs a DPD Colorimetric Method for the detection of total chlorine. The analyzer introduces N,N-Diethyl-p-phenylenediamine (DPD) indicator with potassium iodide and a buffer to the water sample under test. The indicator and buffer are allowed to react with the chlorine and chloramines present in the water. The reaction results in a dye which forms a red color in the water sample. The intensity of the red color is proportional to the total chlorine concentration. It is measured photometrically and the results are automatically recorded and displayed.

    The Remote Indicator (RI) is a subsystem that is mounted in the patient treatment area, typically near the ceiling, so that it can be easily seen. Its primary function is to communicate the status of the chlorine concentration of the feed water and the status of the analyzer to the staff while they are working in the patient treatment room.

    The CM130 Chlorine Monitor and the Remote Indicator will each provide both an audible and a visual signal.

    AI/ML Overview

    The provided text describes the HACH CM130 Chlorine Monitor, a device intended to automatically monitor total chlorine levels in feed water for hemodialysis systems. The document is a 510(k) premarket notification summary submitted to the FDA.

    Based on the provided text, here's a breakdown of the acceptance criteria and the study proving the device meets them:

    Note: This document is a 510(k) summary, which focuses on demonstrating "substantial equivalence" to a predicate device rather than outright clinical efficacy studies. Therefore, many of the typical acceptance criteria for AI/ML medical devices (such as sensitivity, specificity, PPV, NPV) and the detailed study elements like MRMC studies, large-scale ground truth establishment by multiple experts, and specific training set details, are not directly applicable or explicitly stated in this type of submission for this particular device. The focus here is on analytical performance and safety.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily focused on the analytical and safety performance of the device, rather than diagnostic accuracy in a clinical context.

    Acceptance CriteriaReported Device Performance
    Bias (Accuracy)Demonstrates the ability of the CM130 to match a known reference reading. Performance is compared to the predicate device's ability to match a known reading. Bias testing covered the entire measurement range (0.03 to 0.20 mg/L).
    Repeatability (Precision - within instrument)Demonstrates within instrument variations.
    Reproducibility (Precision - between instruments)Demonstrates between instrument variations. Comparative reproducibility testing showed less variation with the CM130 compared to the predicate.
    Interference TestingDevice performs acceptably even when potentially "interfering" chemicals (e.g., iron cations, metal cations, additional inorganics) are present at limits allowed by water standards. Results are published in the instructions for use.
    Electrical SafetyEvaluated against IEC61326-1 (electromagnetic compatibility) and IEC61010-1 (electrical safety).
    Software Verification and ValidationDeveloped, verified, and validated in accordance with applicable FDA guidance documents. Results demonstrate software performs as intended.
    Alarm PerformanceTested for alarm performance following CLSI EP12-A2. Positive and negative alarm performance demonstrated the device is safe and effective.
    Reagent Shelf Life and Use LifeReagents met specified shelf life and use life when stored upright in the marketing container at specified temperatures.
    Human Factors UsabilityTested per ANSI/AAMI HE75:2009. Found to be safe and effective for intended users, uses, and use environments based on subjective and objective data from formative and summative (simulated and actual use) testing with representative users.

    Study Details Proving Device Meets Acceptance Criteria

    The studies conducted are primarily non-clinical performance, analytical, and safety validations.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: Not explicitly stated in terms of a specific number of samples for each test (e.g., how many water samples for bias, how many for repeatability). The document mentions "same solutions and test setups" were used for the CM130 and the predicate device.
      • Data Provenance: The studies were "Non-clinical testing" conducted by Hach Company. The location of the testing or the origin of the solutions is not specified, but it implies laboratory-based or simulated environment testing rather than clinical patient data. The nature is prospective in the sense that the device was actively tested under controlled conditions.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This is not applicable in the typical sense for this device. The "ground truth" for chemical measurements (total chlorine) was established by a reference method for independent verification of "true" measurement values, not by human expert interpretation like in image analysis. Humans were involved in establishing the "ground truth" for human factors testing (representative users), but their qualifications are not specified beyond being "representative users" in target groups (hemodialysis professionals).
    3. Adjudication Method for the Test Set:

      • Not applicable as the primary tests are quantitative measurements against a reference method or engineering standards, not subjective interpretations. Human factors testing involved observation and subjective/objective data collection, but no explicit adjudication method (e.g., 2+1 consensus) is described.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical images with and without AI assistance to measure improvements in diagnostic performance. The CM130 Chlorine Monitor is a chemical monitoring device, not an imaging diagnostic device.
    5. Standalone (Algorithm-Only) Performance:

      • Yes, the core of the performance testing (bias, repeatability, reproducibility, interference) assessed the standalone analytical performance of the CM130 Chlorine Monitor. The "algorithm" here refers to the instrument's measurement and display capabilities. The comparison was made against a "reference method" and the predicate device.
    6. Type of Ground Truth Used:

      • Reference Method: For accuracy (bias) testing, a "reference method used to independently verify the 'true' measurement" of total chlorine was utilized. This is equivalent to a highly accurate laboratory standard or gold standard measurement.
      • Engineering Standards: For electrical safety, software validation, and alarm testing, the ground truth was defined by compliance with specific, recognized engineering and performance standards (e.g., IEC61326-1, IEC61010-1, CLSI EP12-A2).
      • Usability Objectives: For human factors testing, the "ground truth" was the ability of representative users to safely and effectively use the device, as assessed against predefined usability objectives and observed performance.
    7. Sample Size for the Training Set:

      • Not Applicable / Not Explicitly Stated. The CM130 is not described as an AI/ML device that requires a "training set" in the sense of supervised learning from data. It's a deterministic chemical analyzer. Its "calibration" or internal parameters would be set through engineering and testing, not by training on a large dataset in the AI sense. Software verification and validation refer to traditional software engineering processes, not machine learning model training.
    8. How the Ground Truth for the Training Set Was Established:

      • Not Applicable. Since there isn't a "training set" for an AI/ML model as described in the typical context, this question does not apply. The device's operational parameters would be established through manufacturer specifications, calibration procedures, and adherence to design control processes.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1