Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K182956

    Validate with FDA (Live)

    Date Cleared
    2019-01-18

    (86 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K883038, K925982, K093913

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

    Device Description

    The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The Puffin Lite Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, and manual ventilation as established by resuscitation guidelines to neonatal infants weighing less than 10 kg (22 lb). The resuscitation system includes two medical gas flowmeters, an integrated oxygen blender, airway pressure manometer, peak inspiratory pressure (PIP) control, positive end expiratory pressure (PEEP) control, a gas supply pressure alarm and T-piece resuscitator.

    AI/ML Overview

    The Puffin Lite Infant Resuscitation System underwent bench testing to demonstrate conformance to performance specifications.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Valve Function after VomitusThe proper function of the circuit shall be verified within 20 seconds of becoming disabled by vomitus. Function is verified by verifying flow valve accuracy.Passed
    Inspiratory ResistancePressure generated at the patient connection port during expiration should not exceed -5 cmH2O with inspiratory airflow set to 6 L/min.Passed
    Expiratory ResistancePressure generated at the patient connection port during expiration should not exceed 5 cmH2O with expiratory airflow set to 6 L/min.Passed
    Dead SpaceThe deadspace volume of the T-Piece circuit should be less than 7 mL.Passed
    FIO2 accuracyThe proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%.Passed
    Primary and Secondary Flow Valve - Peak FlowThe proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow.Passed
    Airway Manometer AccuracyThe proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure.Passed
    VOC TestingThe device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient.Passed
    Particulate AnalysisThe output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure.Passed

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each individual bench test. The testing was described as "bench testing," which typically involves laboratory-controlled environments and simulated conditions rather than real patient data. The data provenance is from this bench testing, not from human subjects or real-world clinical data. Therefore, there is no information on country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The study involved bench testing of a medical device, where acceptance criteria were based on engineering and performance specifications (e.g., pressure, flow, FIO2 accuracy) rather than expert interpretation of medical images or data. Ground truth was established by measuring the physical performance parameters of the device against predefined technical requirements.

    4. Adjudication method for the test set:

    Not applicable. As the study was bench testing against predefined technical specifications, there was no need for an adjudication method by human experts for the "ground truth." The "passing" or "failing" of a test was determined by whether the device's measured performance fell within the specified limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a 510(k) submission for a physical medical device (infant resuscitator) and not an AI or imaging diagnostic device. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is not an algorithm or an AI system. It is a physical medical device (infant resuscitator) that is intended for use by trained medical professionals.

    7. The type of ground truth used:

    The ground truth used for this study was the engineering and performance specifications outlined in the acceptance criteria. The device's performance was measured against these objective, predefined technical requirements. There was no reliance on expert consensus, pathology, or outcomes data in the context of this bench testing.

    8. The sample size for the training set:

    Not applicable. This device is not an AI or machine learning system that requires a "training set" of data.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142096

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (77 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Mercury NeoTee™ (K093913)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).

    Device Description

    The Mercury T-Piece Resuscitator is manually operated, gas powered resuscitator for use with patients greater than 10 kg (>22 lb).

    It is a simple T-piece, with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve for excessive pressure.

    The T-Piece Resuscitator can be connected to the patient via a face mask, Supraglottic airway or endotracheal tube.

    AI/ML Overview

    The provided document describes the Mercury T-Piece Resuscitator and its substantial equivalence to predicate devices, focusing on non-clinical testing for performance. This is a 510(k) premarket notification summary, which means the device is being compared to an already legally marketed device (predicate device) and does not involve AI or complex analytical studies in the way modern AI/ML medical devices would.

    Therefore, many of the requested categories in the prompt are not applicable to this document. I will focus on the information that can be extracted directly from the provided text.

    Here's the breakdown based on the provided document:

    Acceptance Criteria and Device Performance

    The acceptance criteria are generally aligned with the performance requirements outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators. The "Reported Device Performance" column reflects the specifications of the proposed Mercury Medical T-Piece Resuscitator as presented in comparison tables with its predicate devices.

    Acceptance Criteria (Derived from Predicate & ISO 10651-5)Reported Device Performance (Mercury Medical T-Piece Resuscitator)
    Indications for Use: Provide emergency respiratory support by face mask or tube; patients > 10kg (>22lbs).The T-Piece Resuscitator is a gas-powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).
    Environment of Use: Hospital, sub-acute facilities, and pre-hospital (EMS).Hospital, sub-acute facilities, and pre-hospital (EMS).
    Patient Population: Patients greater than 10 Kg (>22 lbs.).Patients greater than 10 Kg (>22 lbs.)
    Ventilation Frequency: Manually delivered by the user up to 60 BPM.Manually delivered by the user up to 60 BPM.
    Maximum Pressure Relief: Factory set at 40 cm H2O (predicate range 5-80 cm H2O).60 cm H2O (Factory set at 40 cm H2O).
    Delivered Pressure: Up to 60 cm H2O (predicate range 5-80 cm H2O).Up to 60 cm H2O.
    Delivered Volume: (Predicate range 190 - 675 ml with flow rates between 4 - 36 lpm).60 - 700 ml with flow rates between 5 - 35 lpm.
    Inspiratory Resistance: (Predicate: Less than - 5 cm H2O).2 cm H2O at minimum PEEP setting @ 60 lpm.
    Expiratory Resistance: (Predicate: Less than + 5 cm H2O).2.4 cm H2O at minimum PEEP setting @ 60 lpm.
    Oxygen Concentration with optional blender: (Predicate range: 21-100%).21 - 98% based upon blender setting.
    Manometer Range: (Predicate range: -20 to 80 cm H2O).Up to 60 cmH2O (Cleared under K954486).
    Manometer Accuracy: (Predicate: +/- 2 cm H2O).+/- 3 cm H2O up to 15 cm H2O; +/- 5 cm H2O > 15 cm H2O.
    Peak Inspiratory Pressure (PIP): (Predicate range: 5-80 cm H2O).0-60 cm H2O.
    Positive End-Expiratory Pressure (PEEP): (Predicate uses CPAP instead; reference NeoTee has PEEP).0 to 60 cm H2O.
    Operational Gas Flow Rate: (Predicate range: 0-60 lpm).0 - 35 lpm.
    Operational time with 400 L cylinder: (Predicate @ 36 lpm - 10 minutes).@ 35 lpm - 11 minutes.
    Dead space of circuit: (Predicate: < 15 ml).< 7.5 ml.
    Operating Temperature Range: (Predicate range: 10 to 40°C).-18 to 60°C.
    Storage Temperature: (Predicate range: -20 to 60°C).-40 to 50°C.
    Immersion Resistance: Temporary submersion of circuit does not affect functionality.Temporary submersion of circuit does not affect functionality.
    Vomitus Resistance: Meets ISO 10651-5.Tested per ISO 10651-5.
    Water Immersion: Meets ISO 10651-5.Tested per ISO 10651-5.
    Oxygen Concentration: Meets ISO 10651-5.Tested per ISO 10651-5.
    Pressure Limitation: Meets ISO 10651-5.Tested per ISO 10651-5.
    Storage/Operating Conditions: Meets ISO 10651-5.Tested per ISO 10651-5.
    Drop: Meets ISO 10651-5.Tested per ISO 10651-5.

    The study that proves the device meets the acceptance criteria is a series of "Bench testing" performed according to ISO 10651-5 for gas powered resuscitators. The document explicitly states: "We have demonstrated that the T-Piece Resuscitator meets the performance and design specifications requirements as outlined in requirements and specifications as outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators."


    Additional Information on the Study:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • The document describes "bench testing" which implies tests on physical devices rather than patient data. Therefore, details like "country of origin of the data" or "retrospective/prospective" are not applicable. The sample size for the bench tests is not specified in the provided summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This is not applicable as the study involves bench testing of a physical medical device, not a diagnostic or AI algorithm relying on expert interpretation of data. The ground truth is established by the performance standards outlined in ISO 10651-5.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable to a physical device performance test assessed against engineering standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This is a physical medical device (resuscitator), not an AI/ML diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not applicable. This is a physical medical device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the device's performance is based on engineering and safety standards, specifically ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators.
    7. The sample size for the training set:

      • This is not applicable. The device is a physical resuscitator, not an AI/ML algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable, as there is no training set for a physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140707

    Validate with FDA (Live)

    Date Cleared
    2014-05-28

    (69 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K883038, K925982, K093913

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LifeBorne Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

    Device Description

    The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The LifeBorne Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, manual ventilation, and emergency airway clearance as established by resuscitation guidelines to neonates and infants weighing less than 10 kg (22 lb).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the LifeBorne Infant Resuscitator, based on the provided text:

    Acceptance Criteria and Device Performance

    TESTTEST REQUIREMENTS (Acceptance Criteria)REPORTED DEVICE PERFORMANCE
    Valve Function after VomitusThe proper function of the circuit shall be verified within 20 seconds of becoming disabled by vomitus. Function is verified by verifying flow valve accuracy.Passed.
    Inspiratory ResistancePressure generated at the patient connection port during expiration should not exceed -5 cmH2O with inspiratory airflow set to 6 L/min.Passed.
    Expiratory ResistancePressure generated at the patient connection port during expiration should not exceed 5 cmH2O with expiratory airflow set to 6 L/min.Passed.
    Dead SpaceThe deadspace volume of the T-Piece circuit should be less than 7 mL.Passed.
    FIO2 accuracyThe proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%.Passed.
    Primary and Secondary Flow Valve - Peak FlowThe proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow.Passed.
    Airway Manometer AccuracyThe proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure.Passed.
    VOC TestingThe device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient.Passed.
    Particulate AnalysisThe output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure.Passed.

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a separate "test set" sample size. The document refers to "bench testing" as the method for performance evaluation. The tests themselves are functional evaluations of the device's components and overall system.
    • Data Provenance: The testing was "bench testing," meaning it was conducted in a laboratory or simulated environment. The country of origin of the data is not specified, but the submission is from International Biomedical, located in Austin, TX, U.S.A. It is retrospective in the sense that the testing was completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device's performance was evaluated through bench testing against predefined technical specifications, not by expert interpretation of data. Therefore, no experts were used to establish ground truth in this context.

    4. Adjudication method for the test set:

    • Not applicable. As described above, the evaluation was based on meeting objective technical requirements through bench testing, not on human interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (resuscitator) that does not involve "human readers" interpreting medical cases or utilizing AI. Its effectiveness is determined by its mechanical and functional performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, in a way. The device itself operates "standalone" in performing its functions (e.g., delivering precise FIO2, controlling PIP/PEEP). The "bench testing" represents the performance of the algorithm/device without human intervention during the test itself (though humans operate the test setup). There is no "algorithm" in the sense of AI or image analysis involved.

    7. The type of ground truth used:

    • The ground truth for the device's performance was based on predefined technical specifications and engineering standards. Each test had a specific, measurable requirement (e.g., FIO2 within 5%, dead space less than 7 mL) which served as the objective ground truth against which the device's performance was measured.

    8. The sample size for the training set:

    • Not applicable. This device is a mechanical/electronic system, not one that uses machine learning or AI that would require a "training set" of data.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1