Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142023

    Validate with FDA (Live)

    Date Cleared
    2014-11-20

    (118 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Luna 360 Interbody Fusion System consists of the Luna 360 Implant and associated accessories. This system is indicated for spinal fusion procedures in skeletally mature patients with symptomatic degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). The Luna 360 System is to be used with autogenous bone graft. Patients receiving the device should have had at least six months of nonoperative treatment prior to receiving the Luna 360 Implant. The Luna 360 System is to be used with supplemental fixation.

    Device Description

    The Benvenue Luna 360 System consists of the Luna 360 Implant and associated accessories set of disposable accessories for use in lumbar fusion procedures to treat degenerative disc disease. The proposed indications for use for the Luna 360 System are identical to predicate interbody lumbar cages. The Luna 360 Implant is provided pre-loaded and sterile within a single-use Insertion Tool.

    The Luna 360 Implant is available in heights ranging from 8mm to 13mm in 1mm increments. A series of vertically oriented slots allows the device to flex and enables it to be inserted from a straight cannula and then attain a closed, fixed, and circular shape upon being placed into the disc space with a bone graft pocket. Teeth engage the implant into the adjacent endplates.

    The Luna 360 Implant is manufactured from polyetheretherketone (PEEK Optima LT-1), stainless steel, tantalum, and silicone lubricant (NuSil MED-360).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Luna 360 Interbody Fusion System." This document is primarily concerned with establishing substantial equivalence to legally marketed predicate devices, rather than presenting a clinical study with detailed acceptance criteria and performance metrics typically found in AI/ML medical device submissions.

    Therefore, the requested information regarding acceptance criteria, reported performance, sample sizes for test and training sets, expert qualifications, and adjudication methods is not available in the provided text, as this type of information is generally not required for this specific device type and approval pathway.

    The device is an intervertebral body fusion device, not an AI/ML algorithm, and thus the concepts of "acceptance criteria" for algorithm performance, "test set," "training set," "ground truth," "MRMC studies," or "standalone performance" as they relate to AI/ML devices do not apply here.

    However, the document does contain information about performance testing relevant to a medical device's physical and mechanical properties.

    Here's a breakdown of the relevant information provided:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Implicit): The implicit acceptance criterion is that the device demonstrates mechanical properties that are "substantially equivalent" to legally marketed predicate devices, particularly regarding static/dynamic compression, shear-compression, torsion, subsidence, and expulsion. The specific numerical thresholds for "acceptance" are not explicitly stated as they would be in an AI/ML performance summary, but rather implied by meeting the standards of similar devices.
    • Reported Device Performance:
      • Mechanical Testing: The tests demonstrated that the Luna Interbody System is "substantially equivalent" to legally marketed predicate devices in static/dynamic compression, static/dynamic shear-compression, static/dynamic torsion testing (per ASTM F2077), static subsidence (ASTM F2267), and expulsion testing (ASTM Draft Standard F-04.25.02.02).
      • Cadaveric Testing: Performed to validate the surgical technique.
      • MR Compatibility Testing: Performed to demonstrate MR safety.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Not applicable for this type of device and testing. The "test set" in this context refers to the physical devices undergoing mechanical and material testing, not a dataset of patient images or clinical outcomes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. "Ground truth" in this context would likely refer to engineering specifications or validated test methods, rather than expert diagnostic opinions.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the mechanical testing, the "ground truth" would be the established ASTM standards (F2077, F2267, and ASTM Draft Standard F-04.25.02.02) and engineering principles for mechanical performance. For MR compatibility, the "ground truth" would be established MR safety protocols.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this device.

    9. How the ground truth for the training set was established

    • Not applicable.

    In summary, the provided document details the regulatory clearance of a physical medical device (intervertebral body fusion system) based on substantial equivalence to existing devices, supported by a series of mechanical and material performance tests. It does not involve a clinical study or AI/ML algorithm, and therefore the specific questions related to AI/ML validation are not addressed by the text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1