Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K252888

    Validate with FDA (Live)

    Device Name
    NeoSculpt Plus
    Date Cleared
    2026-03-09

    (180 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeoSculpt Plus is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions, such as relief of pain and muscle spasms and increase in local circulation.

    Device Description

    The NeoSculpt Plus is a medical device designed for use by medical professionals in hospital settings. It delivers Radiofrequency (RF) energy to a patient for therapeutic treatment. The system is composed of a main console, one or more RF handpieces that connect to the console via umbilical cables, and neutral electrodes that attach to the patient's skin.

    The NeoSculpt Plus provides topical heating to elevate tissue temperature, which can be used to treat various medical conditions, including the relief of pain and muscle spasms and to increase local circulation. The system operates by generating a high-frequency current that flows from the handpiece, through the patient's body (which acts as a resistance), and back to the console via the neutral electrode. This intensive current flow generates a heating profile that produces a moderate temperature rise in the subcutaneous tissue.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K260292

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-27

    (29 days)

    Product Code
    Regulation Number
    870.5175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K081645, K183174, K170997, K151649, K822572, K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOTWIRE™ is indicated for creation of an atrial septal defect in the heart.

    Device Description

    HOTWIRE™ Kit (901XXX):

    The HOTWIRE™ is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and third-party RF electrosurgical generator(s), such as the Valleylab FT10 Electrosurgical Platform (K151649) which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).

    The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included Adapter Pin that connects to hand pieces used with compatible third-party RF electrosurgical generators.

    HOTWIRE™ + Handpiece Kit (902XXX):

    The HOTWIRE™ + Handpiece Kit is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ + Handpiece Kit is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and the HOTWIRE™ System RF Generator, which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).

    The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included handpiece that connects with the HOTWIRE™ System RF Generator.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233123

    Validate with FDA (Live)

    Device Name
    SILKRO
    Manufacturer
    Date Cleared
    2024-06-24

    (271 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SILKRO with RM. RN. RC. RV. PS handpieces is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis.

    The SILKRO with RN handpieces is intended for the removal and destruction of skin lesions and coagulation of tissue.

    Device Description

    This device is an instrument used for tissue coagulation by means of high-frequency current. It consists of the main unit, five handpieces, GP Cable, foot switch, LCD touch screen, ground pad and power cable.

    The SILKRO has two operating modes: monopolar mode and bipolar mode. In the monopolar mode. RF energy flows from the main unit and a patient loop is formed by pairing the active electrode tip with the neutral electrode pad. Heat is not generated in the neutral electrode pad due to its low contact resistance, but heat is generated in the active electrode tip which has higher contact resistance. The higher contact resistance heats up the tissue resulting in coagulation. In the bipolar mode, RF energy is delivered between adjacent needles in the electrode tip without use of the neutral electrode pad. The user can select the mode and adjust parameters through the touch screen user interface of electrosurgical device.

    RF(HF) enerqy is delivered to the target tissue using a handpiece (RM, RN, RC, RV, PS) and electrode tip, the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the RF(HF) energy passes through the skin, it generates an electro RF reaction which is capable of coagulating the tissue.

    • Electrosurgical Unit Main body
    • Five different handpieces
    • Neutral electrode pad and neutral electrode pad cable, cleared under K092761
    • Handpiece stands
    • Foot switch
    • Power cord
    AI/ML Overview

    I am sorry, but the provided text does not contain any information about acceptance criteria or a study that proves the device meets specific performance criteria. The document is a 510(k) summary for a medical device (SILKRO), primarily focusing on demonstrating substantial equivalence to predicate devices based on technological characteristics and adherence to various safety and performance standards. It mentions non-clinical ex vivo animal testing for histological data but does not present this data as acceptance criteria or a study proving device performance against such criteria. The document explicitly states, "No clinical test was performed."

    Therefore, I cannot extract the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230586

    Validate with FDA (Live)

    Date Cleared
    2023-06-09

    (99 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K221830, K092761, K170188

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Renuvion® Micro Handpiece is intended for the delivery of radiofrequency energy and/or helium plasma where coagulation/contraction of soft tissue is needed. Soft tissue includes subcutaneous tissue.

    The Renuvion® Micro Handpiece is compatible with the Apyx One Console Generator, owned by Apyx Medical.

    Device Description

    The Renuvion® Micro System with all components and accessories is show in the picture below. The system is comprised of a compatible electrosurgical generator (Apyx One Console), a sterile, single use handpiece, and a supply of helium gas. Additional accessories include a digital gas regulator, grounding pad and optional footswitch.

    The Renuvion® Micro handpiece is a sterile, single use electrosurgical (monopolar) device intended to be used in conjunction with the Apyx One Console Electrosurgical Generator for the percutaneous delivery of radiofrequency and helium plasma energy to soft tissue. When connected to the Apyx One Generator, the device operates at an adjustable power of up to 12 W (expressed as 0 - 30% where 30% is 12 W) and an adjustable helium gas flow of 1 -1.5 L/Min. Radiofrequency energy is delivered to the handpiece by the generator and used to energize the electrode. When helium gas is passed over the energized electrode, a helium plasma is generated for soft tissue coagulation/contraction.

    The handpiece has a non-extendable electrode to generate helium plasma. The handpiece features a low-profile shaft that is 10cm in length. The distal end of the shaft and tip has a 2mm outer diameter that tapers down to 1.5mm for the remaining length of the shaft.

    AI/ML Overview

    The Renuvion® Micro Handpiece is cleared based on demonstrating substantial equivalence to a predicate device, the Renuvion® APR Handpiece (K223262 and K230272). This type of submission (510(k)) focuses on comparing a new device to an already legally marketed device, rather than proving absolute safety and effectiveness through extensive clinical trials. Therefore, the "acceptance criteria" are primarily based on demonstrating that the new device is as safe and effective as the predicate device by showing similar technological characteristics and performance.

    Here's a breakdown of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" in a separate section with specific numerical targets. Instead, it presents various tests performed and states whether the device "met requirements" or demonstrated equivalent/lesser effects compared to the predicate. The overall acceptance criterion is demonstrating that the Renuvion® Micro Handpiece is as safe and effective as the predicate device.

    Test TypeObjectiveReported Device Performance (Result)
    Bench Testing
    Mechanical Verification & FunctionalityVerify Mechanical functionality of the Renuvion® Micro Handpiece.Mechanical functionality requirements met.
    System Compatibility TestingVerify the device, when connected to the generators (with all the components and accessories working together as a system), are working as intended and are compatible with all system components.System compatibility requirements met.
    Inspectional VerificationDocument parameters that can be verified through inspection of the Micro Handpiece Labeling and IFU.Inspectional requirements met.
    Packaging ValidationEvaluate simulated (and/or real-time) aging and associated package integrity and shelf-life claims. Assess the potential for damage to the device (e.g., drop tests of the instrument in its packaging).Packaging requirements met.
    Electrical Safety & EMCCompliance with ANSI/AAMI/IEC 60601-1:2015/(R)2012 and A1:2012, AAMI/ANSI/IEC 60601-2-2:2017, and AAMI/ANSI/IEC 60601-1-2:2014 (4th Edition) standards.The Renuvion® Micro System was determined to be in conformance with these standards.
    Biocompatibility TestingCompliance with FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and AAMI/ANSI/ISO 10993-1:2018 standard.
    CytotoxicityEvaluate cytotoxicty.Non-cytotoxic.
    SensitizationEvaluate sensitization reaction.No sensitization reaction.
    IrritationEvaluate irritation.Non-irritant.
    Acute Systemic ToxicityEvaluate acute systemic toxicity.Non-toxic.
    Material Mediated Pyrogen/PyrogenicityEvaluate pyrogenicity.Non-pyrogenic.
    Pre-Clinical & Ex-Vivo Studies
    Thermal Tissue EffectEvaluate the thermal effect of the Renuvion Micro Handpiece against its predicate device on different tissue types at different Renuvion settings.The test device (APYX-10-MICRO) demonstrated an equal or less tissue thermal effect when compared to the predicate device (APYX-15-TP) on five different tissue types at different Renuvion settings.
    Tissue Temperature Over 85 Degrees Celsius ComparisonEvaluate the tissue time over 85 degrees Celsius between the Renuvion Micro Handpiece and its predicate device when used at recommended settings for coagulation/contraction of soft tissue.The test device (APYX-10-MICRO) met the acceptance criteria demonstrating that on average the device is at or above 85 degrees for greater than 45 msec. The predicate device showed similar results as a control.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify numerical sample sizes for the bench, electrical safety, biocompatibility, or ex-vivo tests. It states that tests were conducted according to protocols.
    • Data Provenance:
      • All summarized tests appear to be pre-clinical/bench-top studies and ex-vivo (tissue) tests conducted by the manufacturer, Apyx Medical Corporation.
      • The "Ex-vivo tissue testing" section implies that animal or cadaveric tissue was used, but the specific origin (e.g., species, country) is not mentioned.
      • There is no mention of human subjects or clinical data in this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Since these are primarily bench and ex-vivo tests for a 510(k) submission, the "ground truth" is established by engineering and scientific measurements/observations against predefined specifications and industry standards.
    • The document does not mention the involvement of external medical experts (e.g., radiologists) for establishing ground truth in these specific tests. The experts involved would likely be engineers, scientists, and quality assurance personnel from Apyx Medical performing the tests and comparing results to established criteria. There is no mention of "ground truth" in the context of clinical interpretation, as clinical studies were not performed.

    4. Adjudication Method for the Test Set:

    • Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving interpretation by multiple readers.
    • Given that this submission relies on bench and ex-vivo testing, such adjudication methods are not applicable and are not mentioned in the document. Performance is measured against physical or biological criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." (Section 6: Clinical Studies).
    • Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance. The device is not an AI-based diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • This question is not applicable as the Renuvion® Micro Handpiece is an electrosurgical device, not an algorithm or AI-based system. It does not perform diagnostic tasks independently.

    7. The Type of Ground Truth Used:

    • For the bench and ex-vivo tests, the "ground truth" is based on:
      • Defined engineering specifications (e.g., mechanical functionality, system compatibility, electrical safety, packaging integrity).
      • Biological responses to known stimuli (e.g., non-cytotoxic, no sensitization reaction for biocompatibility tests).
      • Physical measurements and comparisons (e.g., thermal effect and tissue temperature measurements in ex-vivo tissue).
    • The ground truth is established by the test protocols against established standards and internal design requirements, often compared to the performance of the predicate device.
    • No pathology or outcomes data is mentioned as forming the ground truth for this set of studies.

    8. The Sample Size for the Training Set:

    • This question is not applicable for this device. Training sets are relevant for machine learning or AI-based devices. The Renuvion® Micro Handpiece is a physical surgical tool, and its development involves engineering design and testing, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established:

    • This question is not applicable as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222709

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-21

    (194 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    Retraction, CWM-910T and APOLEX Tite are RF electrosurgical system. These are identical except for model designation. The RF electrosurgical system consists of an electrosurgical device (ESU), 3 types of detachable handpieces (mono-polar active electrode), a grounding plate (neutral electrode) and a foot switch. The hand piece consists of a hand grip and an electrode. There are 3types of mono-polar electrode depend on diameter and length. The electrosurgical device generates 480KHz RF current. RF output energy (Max. 50W) is delivered through a mono-polar active electrode and returned through the grounding plate (neutral electrode).

    AI/ML Overview

    The provided document, an FDA 510(k) K222709 submission summary, describes the design, testing, and comparison of the Retraction, CWM-910T, APOLEX Tite RF electrosurgical system to a predicate device. However, it does not contain the specific information required to complete a table of acceptance criteria and reported device performance related to an AI/ML-driven medical device study.

    The document focuses on demonstrating substantial equivalence for an electrosurgical device, primarily through bench testing, electrical safety, biocompatibility, and reprocessing validation. It's a traditional medical device submission, not one for an AI-enabled diagnostic or therapeutic device that would involve performance metrics like sensitivity, specificity, or reader improvement.

    Therefore, I cannot create the requested table or answer the specific questions related to AI/ML study components (e.g., sample size for test set, ground truth experts, MRMC study, standalone performance, training set details) because this information is not present in the provided text.

    The document does mention:

    • Device Type: RF Electrosurgical System (not AI/ML)
    • Purpose: Electrocoagulation and hemostasis in dermatological and general surgical procedures.
    • Bench Testing: Conducted for RF output power and a comparison test with the predicate device, including graphical display of output waveform and power output over a range of loads.
    • Animal Study: A "mini pig" thermal effect study was conducted for safety assessment, comparing the subject device and predicate device. Evaluated thermal imaging area analysis, maximum temperature, and time to reach basal temperature.
    • Software: Firmware verification and validation according to FDA guidance and IEC 62304. This is firmware, not an AI algorithm.
    • Electrical Safety & EMC: Tested against IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2.
    • Biocompatibility: Tested against ISO 10993 series.
    • Reprocessing: Validated according to FDA reprocessing guidance.

    Without the context of an AI-driven device, the questions regarding acceptance criteria for AI performance, test set and training set details, ground truth establishment by experts, and MRMC studies are not applicable to the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201685

    Validate with FDA (Live)

    Device Name
    Potenza
    Date Cleared
    2021-01-19

    (211 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POTENZA is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The POTENZA is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis. The POTENZA consists of the following components: Electrosurgical Unit - Main body, Two different handpieces ( motor and AC), Six different electrode tips for the motor handpiece and three needle tips for the AC handpiece; provided as single use, sterile products, Neutral electrode pad and neutral electrode pad cable, cleared under K092761, Handpiece stand, Foot switch, Power cord. The POTENZA generates radiofrequency (RF) energy by means of high RF at 1MHz or 2MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. The POTENZA has two operating modes: monopolar mode and bipolar mode.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the POTENZA device, specifically for the addition of four new electrode tips. The document details the performance data collected to demonstrate that the expanded device meets acceptance criteria and is substantially equivalent to its predicate.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides tables of acceptance criteria and results for biocompatibility and sterilization residual testing. Other tests mention conformance to specified standards as their acceptance criteria.

    Test Type / MethodAcceptance CriteriaReported Device Performance
    Biocompatibility Testing (for new electrode tips)
    Extractables (ISO 10993-12:2012):Property - per standardPass
        pHpH ≤ 1.5Pass
        Potassium permanganate reducing substances≤ 2.0 mLPass
        Residue after evaporation≤ 1.0 mgPass
        UV spectrum (250 nm - 350 nm)≤ 0.1Pass
        Heavy Metalsper standardPass
    Cytotoxicity (ISO 10993-05:2009, USP 41:2018 <87>)Meet requirements of standardPass
    Sensitization (Guinea Pig Maximization Test (GPMT): ISO 10993-10:2010)Meet requirements of standardPass
    Irritation or Intracutaneous Reactivity (Animal Intracutaneous (Intradermal) Reactivity Test: ISO 10993-10:2010)Meet requirements of standardPass
    Acute Systemic Toxicity (ISO10993-11:2017)Meet requirements of standardPass
    Hemolysis (ISO 10993-4, ASTM F756)Meet requirements of standardPass
    USP Rabbit Pyrogen (ISO 10993-11, USP, General Chapter <151>)Meet requirements of standard / USPPass
    Electrical Safety and Electromagnetic Compatibility (EMC)Conformance to IEC 60601-1, IEC 60101-2-2, IEC 60601-1-2, IEC 60601-1-6Confirmed
    Software Verification and Validation TestingIn accordance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005, commensurate with a moderate level of concernConducted
    Bench Testing (for new electrode tips):Operate safely and within predefined design specificationsAssured
        Output accuracy (Monopolar at 1MHz and 2 MHz)(Implicitly safe and within spec)Tested
        Output accuracy (Bipolar at 1MHz and 2MHz)(Implicitly safe and within spec)Tested
        Frequency: manual and standard(Implicitly safe and within spec)Tested
        Power fluctuation characteristics(Implicitly safe and within spec)Tested
        Negative output protection(Implicitly safe and within spec)Tested
        Impedance measurement accuracy and range(Implicitly safe and within spec)Tested
        HP count accuracy(Implicitly safe and within spec)Tested
        Safety test of various warnings / failsafe mechanisms(Implicitly safe and within spec)Tested
        Needle depthWithin range of predicate device (0.5~2.5 mm)Tested
        Motor speed level(Implicitly safe and within spec)Tested
        Thermal testingIn accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014Tested
    Sterilization (Ethylene Oxide (EO) residuals)According to ISO 10993-7:2008Pass
    Sterility Assurance Level (SAL)10^-6 (per ISO 11135:2014)Assured
    Shelf-life (Accelerated Aging)2-year shelf-life based on 65-days accelerated aging per ASTM F1980-16, USP <71>, ASTM F1929-15Established

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state sample sizes for specific tests with numerical values like "N=X". For biocompatibility testing, it mentions "the electrode tips were tested as shown in Table 1," implying a sufficient sample was used for each test type according to the standard. Similarly, for bench testing, it states "Jeisys conducted bench testing to assure that the four new electrode pins operate safely and within the predefined design specifications," suggesting testing was performed on the new electrode tips.

    The data provenance is from Jeisys Medical, Inc. (South Korea), as the submitter is based in Seoul, Korea. The studies are retrospective in the sense that they are conducted by the manufacturer for the specific purpose of demonstrating substantial equivalence to a predicate device, rather than actively collecting new clinical data on human patients in a prospective manner.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The studies described are primarily in vitro (biocompatibility, sterilization), in silico (software), and benchtop engineering tests, not studies that require expert-established ground truth from clinical images or patient data.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies or studies involving expert review of data (e.g., medical images). The tests described here are technical performance tests, not clinical evaluations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This information is not provided and not applicable. The device is an electrosurgical cutting and coagulation device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This information is not applicable. The POTENZA is a hardware device with associated software, not a standalone algorithm/AI for diagnostic interpretation. Its performance is evaluated through its physical and electrical characteristics as well as its software's functionality. The bench testing and electrical safety tests demonstrate the device's standalone performance.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the various tests is based on recognized consensus standards and predefined design specifications. For example:

    • Biocompatibility: Standards like ISO 10993 and USP provide the "ground truth" for acceptable biological response.
    • Electrical Safety & EMC: IEC 60601 series standards define the ground truth for safe electrical operation and electromagnetic compatibility.
    • Software Verification: FDA guidance provides the framework for acceptable software validation.
    • Bench Testing: The device's own predefined design specifications, safety mechanisms, and performance ranges (e.g., needle depth range of the predicate device) serve as the ground truth.
    • Sterilization & Shelf-life: ISO standards (e.g., 10993-7, 11135), ASTM standards (e.g., F1980-16, F1929-15), and USP standards provide the ground truth for acceptable sterilization residuals, sterility assurance level, and packaging integrity.

    8. The Sample Size for the Training Set

    This information is not provided and not applicable. The described tests are for device validation and verification, not for training an AI or machine learning model.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided and not applicable. No training set for an AI/ML model is mentioned or relevant to this device's submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1