Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K180015

    Validate with FDA (Live)

    Date Cleared
    2018-03-20

    (77 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K091195, K162025, K090462

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Incisive CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Incisive CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

    These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

    *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    Device Description

    The proposed Philips Incisive CT is a whole-body computed tomography (CT) X-Ray System featuring a continuously rotating x-ray tube, detectors, and gantry with multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. This system also includes signal analysis and display equipment, patient and equipment supports, components, and accessories. The Philips Incisive CT has a 72cm bore and includes a detector array that provides 50cm scan field of view (FOV). The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the proposed Philips Incisive CT have the same fundamental design characteristics and are based on comparable technologies as the current market predicate Philips Ingenuity CT (K160743, 08/08/2016). The main system modules and functionalities are: 1. Gantry. The Gantry consists of 4 main internal units: a. Stator - a fixed mechanical frame that carries HW and SW b. Rotor - A rotating circular stiff frame that is mounted in and supported by the stator. c. X-Ray Tube (XRT) and Generator,- fixed to the Rotor frame d. Data Measurement System (DMS) - a detectors array, fixed to the Rotor frame 2. Patient Support (Couch) - carries the patient in and out through the Gantry bore synchronized with the scan 3. Console - Containing a Host computer and display that is the primary user interface. In addition to the above components and the software operating them, each system includes hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.

    AI/ML Overview

    The Philips Incisive CT scanner is compared to a predicate device, the Philips Ingenuity CT (K160743), for substantial equivalence. The provided document focuses on technical comparisons and non-clinical performance data rather than a clinical study with specific acceptance criteria that would typically be seen for a new AI-powered diagnostic device.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a quantitative sense with pass/fail thresholds for clinical performance. Instead, it demonstrates substantial equivalence to a predicate device by comparing technical specifications and imaging features. The core "acceptance criterion" is proving substantial equivalence to the predicate device, Philips Ingenuity CT (K160743).

    CategoryAcceptance Criteria (Implied: Substantially Equivalent to Predicate)Reported Device Performance (Philips Incisive CT)Conclusion
    Scan Characteristics
    Number of Slices64/12864/128Identical. Substantially equivalent.
    Scan ModesSurview, Axial Scan, Helical ScanSurview, Axial Scan, Helical ScanIdentical. Substantially equivalent.
    Minimum Scan Time0.42 sec for 360° rotation0.35 sec for 360° rotationFaster rotation speed to meet wider heart rate application. Safety and effectiveness are not affected. Substantially equivalent.
    Image (Spatial) ResolutionHigh resolution: 16 lp/cm, Standard resolution: 13 lp/cmHigh resolution: 16 lp/cm, Standard resolution: 13 lp/cmIdentical. Substantially equivalent.
    Image Noise0.27% at 120 kV, 250 mAs, 10 mm slice thickness0.27% at 120 kV, 230 mAs, 10 mm slice thicknessIdentical (despite slightly different mAs, the noise level is the same). Substantially equivalent.
    Slice ThicknessesHelical: 0.67mm-5mm, Axial: 0.625mm-12.5mmHelical: 0.67mm-5mm, Axial: 0.625mm-10.0mmEssentially the same, does not affect safety and effectiveness. Substantially equivalent.
    Scan Field of ViewUp to 500 mmUp to 500 mmIdentical. Substantially equivalent.
    Image MatrixUp to 1024 * 1024Up to 1024 * 1024Identical. Substantially equivalent.
    Imaging Features(Function/User Interface/Workflow similar to Predicate)
    2D ViewerYesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    MPRYesYesUser interface, algorithm principle, function, and workflow are similar/same. Substantially equivalent.
    3D (volume mode)YesYesVolume rendering protocol, function, and workflow are similar/same. Substantially equivalent.
    Virtual Endoscopy (Endo)YesYesVE rendering protocol, function, and workflow are similar/same. Substantially equivalent.
    FilmingYesYesBasic function (display, layout, editing, print management) similar/same. Substantially equivalent.
    Image matrix1024 * 10241024 * 1024Both are 1024 * 1024. Substantially equivalent.
    O-MARYesYesAlgorithm Principle and workflow are same. Substantially equivalent.
    Dose ModulationYesYesFunction and workflow are same. Substantially equivalent.
    iPlanningManualiPlanning (automated adjustment)Workflow improvement for user assistance. Safety and effectiveness are not affected. Substantially equivalent.
    On line MPRYes (with tilt and trim)Yes (without trim and tilt)Can generate sagittal and coronal results. Other functions and workflow are same. Substantially equivalent.
    iBatchManual BatchiBatch (automated identification)Workflow feature to improve productivity. Safety and effectiveness are not affected. Substantially equivalent.
    Bolus TrackingYesYes (Post Threshold Delay longer)Function and workflow are same (despite longer Post Threshold Delay). Substantially equivalent.
    SAS (Spiral Auto Start)YesYes (manual trigger only)Other functions and workflow are same. Substantially equivalent.
    WorklistYesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    MPPSYesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    ReportingYes (including PDF)Yes (no PDF support)Format of exported report similar. Other functions and workflow are same. Substantially equivalent.
    CCT (Continuous CT)Yes (with Volume display)Yes (no Volume display support)Other functions and workflow are same. Substantially equivalent.
    Brain PerfusionYesYesUser interface, principle, mechanism, and analysis parameters are similar/same. Substantially equivalent.
    Dental (Dental planning)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    iDose4YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    Helical Retrospective TaggingYesYesECG viewer user interface, function, and workflow are similar/same. Substantially equivalent.
    Axial Prospective Gating calcium scoringYesYesECG viewer user interface, function, and workflow are similar/same. Substantially equivalent.
    Step & ShootYes (with arrhythmia handling)Yes (no arrhythmia handling)Other functions and workflow are same. Substantially equivalent. The lack of arrhythmia handling is noted but not deemed to affect substantial equivalence.
    CCS (Cardiac calcium scoring)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    Supplementary Imaging Features (Compared to Philips MX 16 SLICE K091195)
    CTC (CT Colonoscopy)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    VA (Vessel Analysis)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    LNA (Lung Nodule Analysis)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    Supplementary Imaging Features (Compared to IntelliSpace Portal Platform K162025)
    CAA (Cardiac Artery Analysis)YesYesUser interface, analysis of cardiac coronary artery, and workflow are similar/same. Substantially equivalent.
    CFA (Cardiac Function Analysis)YesYesUser interface, function, and workflow are similar/same. Substantially equivalent.
    Supplementary Imaging Features (Compared to BRILLIANCE DUAL ENERGY OPTION K090462)
    DE (Dual Energy)YesYesUser interface, function, Algorithm Principle, and workflow are similar/same. Substantially equivalent.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "The proposed Philips Incisive CT did not require clinical study since substantial equivalence to the legally marketed predicate device was proven with the verification/validation testing."

    Therefore, there is no mention of a "test set" in the context of patient data, nor any information about data provenance (country of origin, retrospective/prospective). The evaluation was based on non-clinical performance data, primarily engineering verification and validation testing, as well as comparisons to the predicate device's specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Given that no clinical study was performed and no patient-based "test set" was described, there were no experts used to establish ground truth in the traditional sense of clinical outcome assessment for the Incisive CT device. The ground truth for the technical comparisons was the established performance and specifications of the predicate device.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set requiring adjudication of findings was described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a CT scanner, not an AI-powered diagnostic software that assists human readers. The context is about the substantial equivalence of the imaging hardware and associated software functionalities.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This refers to the performance of the CT scanner itself, a hardware device with integrated software, not a separate standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the "study" (which was non-clinical verification/validation and substantial equivalence comparison) was the established technical specifications and performance characteristics of the predicate device (Philips Ingenuity CT, K160743) and compliance with various international and FDA-recognized consensus standards.

    8. The sample size for the training set

    Not applicable. This device is a CT scanner, not a machine learning model that requires a training set of data in the AI sense. Its underlying technology and algorithms are based on established CT principles, and its performance is verified through engineering tests.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160743

    Validate with FDA (Live)

    Date Cleared
    2016-08-08

    (144 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

    These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

    *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information,

    Device Description

    The Philips Ingenuity CT consists of three system configurations, the Philips Ingenuity CT, the Philips Ingenuity Core and the Philips Ingenuity Core128. These systems are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of Jung nodules that may represent cancer*.

    The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the Philips Ingenuity CT have the same fundamental design characteristics and are based on comparable technologies as the predicate.

    The main system modules and functionalities are:

      1. Gantry. The Gantry consists of 4 main internal units:
      • a. Stator a fixed mechanical frame that carries HW and SW.
      • b. Rotor A rotating circular stiff frame that is mounted in and supported by the stator.
      • c. X-Ray Tube (XRT) and Generator fixed to the Rotor frame.
      • d. Data Measurement System (DMS) a detectors array, fixed to the Rotor frame.
      1. Patient Support (Couch) carries the patient in and out through the Gantry bore synchronized with the scan.
      1. Console A two part subsystem containing a Host computer and display that is the primary user interface and the Common Image Reconstruction System (CIRS) - a dedicated powerful image reconstruction computer.

    In addition to the above components and the software operating them, each system includes a workstation hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: The provided document is a 510(k) submission for a CT scanner (Philips Ingenuity CT), which focuses on demonstrating substantial equivalence to a predicate device (Philips Plus CT Scanner), rather than establishing new performance claims with specific acceptance criteria and clinical trial results typical for entirely novel AI/ML devices. Therefore, much of the requested information, particularly regarding AI-specific performance (like effect size of human reader improvement with AI, standalone AI performance, ground truth for training AI models) is not directly present. The clinical evaluation described is a comparative image quality assessment rather than a diagnostic accuracy clinical trial.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) submission, the "acceptance criteria" are primarily established against international and FDA-recognized consensus standards for medical electrical equipment and CT systems, and against the performance of the predicate device. The "reported device performance" refers to the successful verification against these standards and equivalence to the predicate.

    Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance
    Safety and Essential Performance (General)IEC 60601-1:2006 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance)All verification tests were executed and passed the specified requirements.
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2007 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests)All verification tests were executed and passed the specified requirements.
    Radiation ProtectionIEC 60601-1-3 Ed 2.0:2008 (Medical electrical equipment Part 1-3: General requirements for basic safety - Collateral standard: Radiation protection in diagnostic X-ray equipment)All verification tests were executed and passed the specified requirements, including radiation metrics.
    UsabilityIEC 60601-1-6:2010 (Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability)All verification tests were executed and passed the specified requirements.
    Safety of X-ray Equipment (Specific)IEC 60601-2-44:2009 (Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment)All verification tests were executed and passed the specified requirements.
    Software Life Cycle ProcessesIEC 62304:2006 (Medical device software Software life cycle processes)Software Documentation for a Moderate Level of Concern (per FDA guidance) was included. All verification tests were executed and passed the specified requirements.
    Risk ManagementISO 14971 (Medical devices Application of risk management to medical devices (Ed. 2.0, 2007))Traceability between requirements, hazard mitigations and test protocols described. Test results per requirement and per hazard mitigation show successful mitigation.
    Image Quality Metrics (Comparative to Predicate)CT number accuracy and uniformity, MTF, noise reduction performance (i.e., iDose4 vs. FBP), slice thickness, slice sensitivity profiles. Diagnostic image quality for brain, chest, abdomen, pelvis/orthopedic.Bench tests included patient support/gantry positioning repeatability and accuracy, laser alignment accuracy, CT image quality metrics testing. Sample phantom images provided. Clinical evaluation found no difference in image quality between iDose4 and FBP, with iDose4 scoring higher in most cases, maintaining diagnostic quality.
    Functional and Non-Functional Requirements (System Level)System Requirements Specification, Subsystem Requirement Specifications, User Interface VerificationFunctional and non-functional regression tests, as well as user interface verification, provided in the Traceability Matrix (successful).
    Clinical Validation (Workflow & Features)Covered requirements related to clinical workflows and features.Validation test plan executed as planned, acceptance criteria met for each requirement. All validation tests demonstrate safety and effectiveness.
    Serviceability ValidationCovered requirements related to upgrade, installation, servicing, and troubleshooting.Validation test plan executed as planned, acceptance criteria met for each requirement.
    Manufacturing ValidationCovered requirements related to operations and manufacturing.Validation test plan executed as planned, acceptance criteria met for each requirement.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" sample size in the sense of a number of clinical cases or patient images used for a diagnostic accuracy study. Instead, it refers to:

    • Bench tests: These involved phantom images and physical testing of the system (e.g., patient support/gantry positioning repeatability and accuracy, laser alignment accuracy, CT image quality metrics testing). No sample size for these is given.
    • Clinical Evaluation: An "image evaluation" was performed involving "images of the brain, chest, abdomen and pelvis/peripheral orthopedic body areas." The number of images or patient cases used for this evaluation is not specified.
    • Data Provenance: Not explicitly stated, but given it's a Philips product, it's likely internal development and validation data. There is no mention of external datasets or specific countries of origin. The evaluation compares FBP and iDose4 reconstructions of the same images. The clinical evaluation implicitly relates to retrospective data as it compares reconstructed images.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: "a qualified radiologist". So, one expert.
    • Qualifications of Experts: Described only as "a qualified radiologist." No specific experience (e.g., years of experience, subspecialty) is provided.

    4. Adjudication Method for the Test Set

    The evaluation was performed by a single radiologist using a 5-point Likert scale. Therefore, no adjudication method (like 2+1, 3+1 consensus) was used as there was only one reviewer.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document describes an image evaluation by a single radiologist, not multiple readers. It also describes a comparison of image quality between reconstruction techniques (FBP vs. iDose4), not a comparison of human reader diagnostic performance with vs. without AI assistance.

    • Effect size of human readers improving with AI vs without AI assistance: This information is not applicable as this type of study was not performed. The study evaluated if iDose4-reconstructed images (which is an iterative reconstruction technique for image quality improvement and dose reduction, not an AI for diagnosis) maintained diagnostic quality compared to standard FBP.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, in spirit, the primary evaluation is about the algorithm's output quality. The iDose4 iterative reconstruction algorithm directly produces images without human intervention, and these images were then evaluated by a radiologist. The core of this 510(k) is about the technical performance and safety of the CT scanner and its components, including its reconstruction algorithms. The evaluation described ("image evaluation...") is a standalone assessment of the image quality produced by the iDose4 algorithm compared to standard FBP. It is not an "AI diagnostic algorithm" standalone performance, but rather an "image reconstruction algorithm" standalone performance.


    7. The Type of Ground Truth Used

    For the clinical image evaluation, the "ground truth" was established by the evaluation of a qualified radiologist using a 5-point Likert scale to determine if images were of "diagnostic quality" and for comparing image quality between reconstruction methods. This could be considered a form of "expert consensus," albeit from a single expert in this case. There is no mention of pathology or outcomes data being used as ground truth for this specific image quality assessment.


    8. The Sample Size for the Training Set

    Not applicable in the context of this 510(k) as presented.

    The device (Philips Ingenuity CT) is a hardware CT scanner with associated software, including image reconstruction algorithms (like iDose4). While iterative reconstruction algorithms might involve some form of "training" or optimization during their development, the document does not speak to a "training set" in the sense of a dataset used to train a machine learning model for a specific diagnostic task that would typically be described in an AI/ML device submission. The description focuses on technical modifications and adherence to engineering and safety standards, and performance against a predicate device.


    9. How the Ground Truth for the Training Set was Established

    Not applicable for the reasons stated above (no "training set" for an AI/ML diagnostic model described).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1