510(k) Data Aggregation

    K Number
    K223080

    Validate with FDA (Live)

    Date Cleared
    2022-11-22

    (53 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K081137,K123463,K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided text describes a 510(k) submission for a labeling modification to the da Vinci Xi and X Surgical Systems, not a study to prove a device meets acceptance criteria regarding active performance metrics. The submission asserts substantial equivalence based on the current device's performance being equivalent to predicate devices for new labeled uses.

    Here's an analysis of the provided information, focusing on how the request's points relate to the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in the typical sense of numerical thresholds for device performance. Instead, it aims to demonstrate substantial equivalence for new labeled uses by comparing the real-world performance of the da Vinci systems (robotic-assisted, "R") with traditional laparoscopic procedures ("L") for duodenal switch bariatric surgeries. This comparison functions as the "performance" data.

    Table of Device Performance (from the document, Table 1):

    OutcomesL (Unmatched)R (Unmatched)p-value (Unmatched)L (PS Matched)R (PS Matched)p-value (PS Matched)
    1. Operative Time (min), median (IQR)180 (150, 240)284 (240, 345)<.001210 (165, 270)270 (240, 330)<.001¶
    2. Length of Hospital Stay (days), median (IQR)2 (1, 3)2 (1, 3)0.5182 (1, 3)2 (1, 3)0.372§§
    3. Conversion to Open Surgery, n (%)5 (0.3)0 (0.0)0.5901 (0.3)0 (0.0)1.000§§
    4. Transfusion Rate, n (%)9 (0.6)4 (1.1)0.3083 (1.0)3 (1.0)1.000
    5. Intraoperative Complication Rate, n (%)2 (0.1)1 (0.3)0.4901 (0.3)1 (0.3)1.000
    6. 30 Day Post-Operative Complication Rate, n (%)309 (20.6)86 (22.8)0.34385 (27.1)67 (21.3)0.094
    7. 30 Day Readmission84 (5.6)14 (3.7)0.14228 (8.9)10 (3.2)0.003
    8. 30 Day Reoperation30 (2.0)7 (1.9)0.85914 (4.5)5 (1.6)0.036
    9. 30 Day Mortality2 (0.1)0 (0.0)1.0001 (0.3)0 (0.0)1.000

    Acceptance Criteria Interpretation: The "acceptance criteria" here are implied by the substantial equivalence argument: the da Vinci system performance for these new indications should be comparable to or better than traditional laparoscopic methods, without raising new questions of safety or effectiveness. The p-values are used to assess statistical differences between the robotic and laparoscopic cohorts. While the robotic operative time is significantly longer (p < 0.001), some key safety outcomes (conversion to open, transfusion, intraoperative complications, and mortality) show no significant difference or favorable trends for robotic assistance (e.g., lower reoperation rate and readmission rate in the PS matched data). The submission concludes that the RWE demonstrates the safety and effectiveness of robotic-assisted duodenal switch bariatric surgical procedures and that da Vinci-assisted procedures are substantially equivalent to laparoscopic procedures based on these outcomes.

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set):
      • Unmatched: 1,878 patients in total (377 robotic-assisted, 1501 laparoscopic).
      • Propensity Score Matched: 628 patients (314 robotic-assisted, 314 laparoscopic).
    • Data Provenance:
      • Country of Origin: USA.
      • Nature of Data: Retrospective Real World Evidence (RWE) from the Premier Health Database (PHD). The data covers the period from 2016 to 2020.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not describe a process of individual expert ground truth establishment for each case. Instead, the "ground truth" is derived from real-world clinical outcomes recorded in the Premier Health Database using ICD-10 and CPT codes. This implicitly relies on the medical professionals (e.g., surgeons, diagnosticians, hospital staff) who documented these codes and outcomes in clinical practice, rather than a separate panel of experts being convened to review the data for the purpose of this study.

    4. Adjudication method for the test set

    There is no mention of an adjudication method in the context of expert review for ground truth, as the data provenance is real-world clinical records rather than expert interpretation of medical images or other subjective assessments. The "adjudication" in this context is the inherent process of medical documentation and coding within the healthcare system.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This study compares the outcomes of robotic-assisted surgery (where the surgeon uses the da Vinci system) versus traditional laparoscopic surgery. It is not comparing human readers' performance with and without AI assistance in reading or diagnosis. The da Vinci system is a surgical tool, not an AI diagnostic assistant. Therefore, there's no "effect size of how much human readers improve with AI vs without AI assistance" to report here.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance was not done. The da Vinci system is an "Endoscopic Instrument Control System" that "is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments." It is fundamentally a human-in-the-loop system, where the surgeon controls the robotic arms. The study evaluates the outcomes of surgery performed by surgeons with the assistance of the da Vinci system, not an autonomous algorithm.

    7. The type of ground truth used

    The ground truth is based on outcomes data and clinical diagnoses/procedures (documented via ICD-10 and CPT codes) from real-world patient records in the Premier Health Database. These outcomes include operative times, length of hospital stays, conversion rates, various complication rates, readmission rates, reoperation rates, and mortality rates.

    8. The sample size for the training set

    The document does not mention a separate "training set" in the context of an AI/machine learning model being developed or trained. This submission is for a labeling modification of an existing surgical system, not the initial clearance of a new AI algorithm. The real-world evidence presented is used to support the safety and effectiveness of the existing device for new labeled uses, but not to train a model in the typical sense.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" for an AI model, this question is not applicable to the provided document. The real-world data itself serves as the basis for demonstrating equivalence for the new indications, not as data for training an algorithm where ground truth needs to be established by experts.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211784

    Validate with FDA (Live)

    Date Cleared
    2021-08-06

    (58 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedure of "Hepatectomy/Liver Resection" under the cleared "general laparoscopic surgical procedures" Indications for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294). There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided text describes a 510(k) submission for a labeling modification to the da Vinci Xi Surgical System (Model IS4000) and the da Vinci X Surgical System (Model IS4200) to include "Hepatectomy/Liver Resection" as a representative specific procedure under the existing Indications for Use. The submission argues for substantial equivalence based on clinical data from published literature.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The submission does not explicitly define "acceptance criteria" in a quantitative manner (e.g., "mortality rate must be less than X%"). Instead, it presents a comparative effectiveness study showing that the da Vinci-assisted procedures for Hepatectomy/Liver Resection are "substantially equivalent" to minimally invasive/laparoscopic procedures based on several clinical endpoints. The acceptance criterion is implicitly that the robotic-assisted procedure's performance on these endpoints isnot worse than the comparator procedures.

    The reported device performance is presented in Tables 1A and 1B: da Vinci vs. Minimally Invasive/Laparoscopic Hepatectomy/Liver Resection Procedures. These tables detail various clinical outcomes from 32 retrospective studies, 5 database studies, and 1 prospective study comparing robotic-assisted and laparoscopic/minimally invasive approaches for hepatectomy/liver resection.

    Table of Performance Comparison (Derived from Tables 1A and 1B and supporting text):

    Clinical EndpointAcceptance Criteria (Implicit: Not worse than minimally invasive/laparoscopic results)Reported Device Performance (da Vinci-assisted vs. Minimally Invasive/Laparoscopic)
    Mortality RatesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B generally show comparable (often 0%) mortality rates for both robotic and laparoscopic groups, or small differences that are not highlighted as significant.
    Estimated Blood Loss (EBL) VolumesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B show similar EBL ranges and means for both groups, or sometimes lower for robotic, with variations across studies.
    Transfusion RatesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B show comparable or sometimes lower transfusion rates for robotic, with variations across studies.
    Lengths of Hospital Stay (LOS)No statistically significant increase compared to comparator.Data in Tables 1A and 1B show comparable or sometimes slightly varied LOS for both groups, with no consistent trend indicating worse outcomes for robotic.
    Post/Peri-Operative Complication RatesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B show comparable complication rates for both groups, with no consistent trend indicating worse outcomes for robotic.
    Conversion RatesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B show comparable or sometimes lower conversion rates for robotic, with variations across studies.
    Readmission RatesNo statistically significant increase compared to comparator.Data in Tables 1A and 1B show comparable or sometimes lower readmission rates for robotic, with variations across studies.
    Perforation RatesNo statistically significant increase compared to comparator.Not explicitly detailed in the tables, but stated as an endpoint demonstrating substantial equivalence.
    Operative TimesAcceptable, considering potential benefits or known learning curve for robotic.Data in Tables 1A and 1B show variable operative times; in some studies robotic times are longer, in others comparable or shorter. This is often accepted for robotic procedures due to precision and benefits.
    Biliary/Bile Leak RatesNo statistically significant increase compared to comparator.Data in Table 1B shows comparable or low bile leak rates for both robotic and laparoscopic groups.
    Liver Failure RatesNo statistically significant increase compared to comparator.Data in Table 1B shows comparable or low liver failure rates for both robotic and laparoscopic groups.
    R0 Resection Rate (Negative Margins)No statistically significant decrease compared to comparator.Data in Table 1B shows comparable R0 resection rates for both groups, indicating similar oncologic efficacy where applicable.
    PSM/R1 Resection Rate (Positive Margins)No statistically significant increase compared to comparator.Data in Table 1B shows comparable PSM/R1 resection rates for both groups.

    2. Sample size used for the test set and the data provenance

    The "test set" in this context refers to the published clinical studies analyzed.

    • Sample size: The study identified 38 publications. Within these publications, individual study sizes (N) for both Robotic and Laparoscopic cohorts vary widely, as shown in Tables 1A and 1B. For example, some studies have N as low as 9 (Berber 2010 Robotic) or 10, while others are significantly larger, such as Lap 6186 (Stewart 2019) or Robotic 354 (Stewart 2019). The total number of patients across all 38 studies is not aggregated but is certainly substantial.
    • Data provenance: The data is retrospective and prospective clinical data from published literature.
      • One (1) prospective study (LOE 2b)
      • Five (5) database studies (LOE 2c)
      • Thirty-two (32) retrospective studies (LOE 3b)
        The country of origin is not specified in the provided text, but it's international clinical literature.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the outcomes reported in the individual studies (e.g., mortality, complications, EBL) would have been established by the clinical teams involved in those primary studies. This submission is a literature review, not a primary clinical study establishing new ground truth. Therefore, the "ground truth" here is derived from the aggregated findings of numerous published studies conducted by various clinical experts worldwide. The submission itself doesn't mention specific experts establishing a ground truth for its own analysis, but rather synthesizes existing expert-derived clinical data.

    4. Adjudication method for the test set

    Not applicable in the conventional sense. This is a review of published literature. The outcomes of the individual studies would have been adjudicated according to the methods of those respective studies. The selection of the studies for this submission was based on specific search criteria and filters as outlined in Figure A, implying a systematic review process rather than expert adjudication of individual cases for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI/imaging device. It's a surgical system where human surgeons directly control the robotic instruments. The comparative effectiveness study referenced is between human surgeons using the da Vinci system versus human surgeons using traditional minimally invasive/laparoscopic techniques. There is no mention of "human readers" or "AI assistance" in the context of improving diagnostic/interpretive tasks.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. The da Vinci system is a robotically-assisted surgical tool that requires a human surgeon in the loop. It is not an autonomous algorithm.

    7. The type of ground truth used

    The "ground truth" in this context is the clinical outcome data (e.g., mortality, complications, blood loss, length of stay, resection margins) reported in published peer-reviewed clinical studies. These outcomes are established through standard clinical practice, surgical records, patient follow-up, and, where applicable, pathology reports for resection margins.

    8. The sample size for the training set

    Not applicable. This device is a surgical system, not an AI/machine learning algorithm that requires a "training set" in the computational sense. The "training" for this device is the training of surgeons to use the system, which is mentioned in the Indications for Use: "It is intended to be used by trained physicians."

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no "training set" in the context of an AI algorithm. The training of surgeons involves established surgical curricula, proctoring, and credentialing processes within medical institutions, based on clinical experience and outcomes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182140

    Validate with FDA (Live)

    Date Cleared
    2018-10-24

    (78 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedure under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294) and the associated labeling claims: Gastrectomy. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided document describes a 510(k) submission for a labeling modification of the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) to include "Gastrectomy" as a recognized procedure under the existing "general laparoscopic surgical procedures" indication for use. This is explicitly stated as a labeling modification only, with no changes to the technological characteristics of the cleared devices. As such, the "acceptance criteria" and "device performance" in a traditional sense (e.g., diagnostic accuracy, sensitivity, specificity) for a novel device or AI algorithm are not applicable here.

    Instead, the study aims to demonstrate that adding "Gastrectomy" to the labeling does not introduce new safety or effectiveness concerns compared to the already cleared general laparoscopic indications, and that the da Vinci systems are substantially equivalent for this new labeled use.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a labeling modification and not a new device or AI algorithm with specific performance metrics like accuracy, sensitivity, or specificity, there are no predefined numerical acceptance criteria or typical "device performance" values.

    Instead, the "acceptance criteria" can be inferred as demonstration of comparable safety and effectiveness outcomes for Gastrectomy procedures performed with the da Vinci system when compared to laparoscopic and/or open procedures, based on literature review.

    Metric (Comparison Group)Acceptance Criteria (Inferred)Reported Device Performance Statement for da Vinci (vs. Laparoscopic)Reported Device Performance Statement for da Vinci (vs. Open)
    MortalityComparable mortality ratesComparable mortality ratesComparable mortality rates
    Estimated Blood Loss (EBL)Comparable or lower EBL volumesComparable or lower EBL volumesLower EBL volumes
    Blood Transfusion RatesComparable blood transfusion ratesComparable blood transfusion ratesComparable blood transfusion rates
    Lengths of Hospital Stay (LOS)Comparable or shorter lengths of hospital stayComparable or shorter lengths of hospital stayComparable or shorter lengths of hospital stay
    Intraoperative Complication RatesComparable intraoperative complication ratesComparable intraoperative complication ratesComparable intraoperative complication rates
    Postoperative Complication RatesComparable postoperative complication ratesComparable postoperative complication ratesComparable postoperative complication rates
    Conversion RatesComparable conversion ratesComparable conversion ratesNot explicitly stated for open, but relevant for comparing da Vinci to other minimally invasive approaches.
    Reoperation RatesComparable reoperation ratesComparable reoperation ratesComparable reoperation rates
    Readmission RatesComparable readmission ratesComparable readmission ratesComparable readmission rates
    Anastomotic Leak RatesComparable anastomotic leak ratesComparable anastomotic leak ratesComparable anastomotic leak rates
    Anastomotic Stenosis RatesComparable anastomotic stenosis ratesComparable anastomotic stenosis ratesComparable anastomotic stenosis rates
    Operative TimeComparable or longer operative times, without an increase in mortality or complication ratesComparable or longer operative times; this increase was not associated with an increase in mortality or complication rates.Longer operative times; this increase was not associated with an increase in mortality or complication rates.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for the Test Set:
      The study is a literature review, not a primary clinical trial with a "test set" in the traditional sense of a prospectively collected cohort. The sample sizes are derived from the N values reported in the individual studies included in the meta-analysis/systematic review. For example, some studies reviewed had sample sizes for Robotic Gastrectomy ranging from 53 to 1830 patients, for Laparoscopic Gastrectomy from 61 to 4123 patients, and for Open Gastrectomy from 145 to 8585 patients.
    • Data Provenance: The data provenance is stated as "Published clinical data from literature." The document extensively cites 16 publications. No specific countries of origin are mentioned, but based on typical medical literature, these would likely be international. The studies cited include:
      • Two (2) randomized controlled trials (level 1b evidence)
      • Ten (10) systematic reviews/meta-analysis (level 2a evidence)
      • Four (4) database studies (level 2c evidence)
        The nature of these studies (e.g., retrospective database analyses, prospective RCTs within the included literature) would vary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of study does not involve "experts establishing ground truth" for individual cases in a test set. The "ground truth" here is the aggregated clinical outcomes reported in published studies, which are considered evidence-based medical conclusions. The experts involved would be the authors of the 16 published clinical studies (surgeons, statisticians, researchers, etc.) who conducted the original research and established their findings. Their specific qualifications are not detailed in this 510(k) summary, but are implicitly assumed based on the peer-reviewed nature of the published literature.


    4. Adjudication Method for the Test Set

    No adjudication method is applicable here as this is a review of published literature, not a study involving individual case assessments by reviewers. The "adjudication" is inherent in the peer-review process of the published articles and the methodology of systematic reviews and meta-analyses.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This submission is for a surgical system, not an AI diagnostic/interpretive tool requiring human reader performance evaluation.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No standalone algorithm performance study was done. The da Vinci Xi and X Surgical Systems are electro-mechanical systems intended to assist surgeons (human-in-the-loop), not standalone algorithms. The submission explicitly states "The Intuitive Surgical Endoscopic Instrument Control System...is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments...It is intended to be used by trained physicians in an operating room environment..."


    7. The Type of Ground Truth Used

    The ground truth used is clinical outcomes data (e.g., mortality, EBL, LOS, complication rates, reoperation rates, etc.) collected and analyzed in previously published clinical studies (Randomized Controlled Trials, Systematic Reviews/Meta-analyses, and Database Studies).


    8. The Sample Size for the Training Set

    This submission does not involve a "training set" in the context of an AI/ML algorithm. The "training data" for making the determination about substantial equivalence for the new indication are the 16 published clinical studies. The sum of patients across all studies' da Vinci cohorts, laparoscopic cohorts, and open cohorts constitutes the entirety of the evidence base reviewed. These numbers are substantial, with individual studies reporting hundreds to thousands of patients (e.g., Robotic 1830, Lap 4123, Open 8585 in some studies).


    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the AI/ML sense, the question refers to how the evidence for the labeling modification was established. The "ground truth" (i.e., the clinical outcomes and conclusions) in the selected 16 publications was established through:

    • Clinical Research Methodologies: Randomized controlled trials, systematic reviews, meta-analyses, and database studies.
    • Data Collection: Gathering patient data on various clinical outcomes (e.g., EBL, LOS, complications, mortality) from surgical procedures.
    • Statistical Analysis: Rigorous statistical methods applied by the authors of the original publications to analyze and interpret the collected data, leading to the reported conclusions regarding comparability or differences in outcomes between da Vinci, laparoscopic, and open surgical approaches.
    • Peer Review: The scientific rigor of these publications being accepted into peer-reviewed journals.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173585

    Validate with FDA (Live)

    Date Cleared
    2018-07-19

    (241 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided document, K173585, describes the a 510(k) premarket notification for a labeling modification to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System (Model IS4000) and the da Vinci X Surgical System (Model IS4200). The document does not describe the acceptance criteria and study proving a device's performance against acceptance criteria in the manner typically seen for AI/ML-enabled devices. Instead, it focuses on demonstrating the substantial equivalence of adding VHR procedures to the existing robotic surgical systems by comparing the outcomes of robotic-assisted VHR to traditional open and laparoscopic VHR.

    Therefore, many of the specific questions about acceptance criteria for an AI/ML device's performance, sample sizes for test sets, establishment of ground truth by experts, MRMC studies, and standalone algorithm performance, as requested in the prompt, are not applicable or directly addressed by this document. This document is about a surgical system, not an AI/ML diagnostic or prognostic tool.

    However, I can extract information related to the clinical study data used to support the labeling modification, which serves a similar purpose to a performance study for a new indication.

    Here's the information as it relates to the provided document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't define explicit numerical "acceptance criteria" in the way one would for an AI/ML device (e.g., "sensitivity must be >X%"). Instead, it demonstrates comparability and, in some cases, superiority of the robotic-assisted VHR to existing surgical methods across several clinical outcomes and measures. The "acceptance" for this 510(k) is based on demonstrating substantial equivalence and safety and effectiveness for the new indication compared to existing methods.

    The tables provided (Tables 1, 2, 3, and 4a) present the "reported device performance" in terms of clinical outcomes for robotic-assisted VHR procedures compared to open and laparoscopic VHR. Below is a summary of the key findings, which serve as the "performance" data used to support the new indication.

    Summary of Device Performance (Robotic-Assisted VHR vs. Comparators)

    Outcome/MeasureRobotic-Assisted vs. Open (Non-Complex VHR)Robotic-Assisted vs. Laparoscopic (Non-Complex VHR)Robotic-Assisted vs. Open (Complex VHR)Robotic-Assisted vs. Laparoscopic (Complex VHR) - Unmatched AHSQC Data & Published Literature
    Length of StayComparable (2 ± 7 days vs. 2 ± 14 days)Shorter (2 ± 7 days vs. 4 ± 13 days)Shorter (2 ± 3 days vs. 5 ± 7 days)Shorter in AHSQC data (2 ± 3 days vs. 4 ± 3 days). Published literature varies (1-9.2 days).
    Intraoperative ComplicationsComparableComparableComparableComparable in AHSQC data (2% vs. 11%). Published literature (0-1.5%).
    Transfusions (Intra/Post-op)ComparableComparableComparableComparable (0% across all AHSQC cohorts). Published literature (0 or 51-91.7ml EBL).
    Postoperative Complications (through 30 days)Comparable (9% vs. 11%)Comparable (10% vs. 11%)Comparable (24% vs. 20%)Mixed: Robotic-assisted (24%) vs. AHSQC Laparoscopic (11%). Published literature (0-60%).
    Readmission Rates (through 30 days)Comparable (2% vs. 3%)Comparable (2% vs. 3%)Comparable (7% vs. 4%)Comparable in AHSQC data (7% vs. 7%). Published literature (0-13%).
    Re-encounter Rates (through 30 days)Lower clinic re-encounter for robotic-assisted (29% vs. 41%). Comparable ER.ComparableLower clinic re-encounter for robotic-assisted (16% vs. 27%). Comparable ER.Mixed: Robotic-assisted (17% Clinic, 5% ER) vs. AHSQC Laparoscopic (22% Clinic, 15% ER). Published literature (0).
    Reoperation Rates (through 30 days)Comparable (<1% vs. 1%)Comparable (<1% vs. 1%)Comparable (3% vs. 1%)Mixed: Robotic-assisted (3%) vs. AHSQC Laparoscopic (0%). Published literature (0-20%).
    Recurrence Rates (through 30 days)Comparable (1% vs. 0%)Comparable (1% vs. 1%)Comparable (1% vs. 1%)Comparable in AHSQC data (1% vs. 0%). Published literature "Not Reported" or low.
    Mortality (through 30 days)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable in AHSQC data (<1% vs. 0%). Published literature (0-1.9%).
    Operative TimeLonger for robotic-assisted (e.g., 45% between 60-119 min vs. 35%).Longer for robotic-assisted (e.g., 43% between 60-119 min vs. 50%).Longer for robotic-assisted (e.g., 45% 240+ min vs. 17%).Longer for robotic-assisted (46% 240+ min vs. 19%). Published literature (125-372 min).
    Conversion Raten/an/an/aRobotic-assisted (4%) vs. AHSQC Laparoscopic (70%). Published literature (0%).

    Note: The document explicitly states that longer operative times for robotic-assisted cases were "not associated with increases in the complication, readmission, reoperation or mortality rates." This addresses a potential negative finding.

    2. Sample sizes used for the test set and the data provenance:

    • Test Set (Clinical Data): Real-world evidence (RWE) from the AHSQC registry database.
      • Non-Complex VHR Procedures:
        • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=871 for each group.
        • Robotic-Assisted vs. Laparoscopic: Propensity score matched cohorts of N=615 for each group.
        • Data Provenance: The registry data for these procedures occurred between July 7, 2013, and January 1, 2017. The document does not specify the country of origin, but AHSQC (America's Hernia Society Quality Collaborative) suggests it is primarily from the United States. The data is retrospective as it's from a registry of past procedures.
      • Complex VHR Procedures:
        • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=297 for each group.
        • Robotic-Assisted vs. Laparoscopic: Unmatched data from the AHSQC Registry Database (Robotic-Assisted N=305, Laparoscopic N=27) and a summary of published literature of laparoscopic VHR (N=3-53 per study).
        • Data Provenance: Similar to non-complex, July 7, 2013, and January 1, 2017, primarily United States, retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable in the context of this document. The "ground truth" for the performance data is derived directly from clinical outcomes recorded in a surgical registry (AHSQC) and published clinical literature. These are factual clinical occurrences (e.g., length of stay, complications, readmissions, mortality) rather than subjective interpretations by experts, as would be the case for image-based diagnostic AI/ML.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As noted above, the data are clinical outcomes from a registry, not expert-labeled data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human "readers." The "device" (da Vinci Xi/X Surgical System) is a surgical tool. The study compares surgical approaches (robotic, open, laparoscopic), not the performance of human readers with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The da Vinci system is a human-controlled robotic surgical system; it does not operate "stand-alone" as an algorithm would. Performance is always human-in-the-loop (i.e., the surgeon uses the system).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth for the clinical performance data (test set) is Real-World Evidence (RWE) from a clinical registry (AHSQC). This constitutes outcomes data (e.g., length of stay, complication rates, readmission rates, recurrence rates, mortality rates) directly recorded from patient care. For certain comparisons (Complex VHR Robotic-Assisted vs. Laparoscopic), it also incorporates published literature on laparoscopic VHR outcomes.

    8. The sample size for the training set:

    The document does not describe a "training set" in the context of machine learning. The da Vinci Xi and X Surgical Systems are existing electromechanical systems; this submission is for an expanded indication. The "training" for the system itself has already occurred and is inherent in its design and previous clearances. The data used in this submission serves as clinical validation data for the expanded Indication for Use.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no "training set" in the AI/ML sense described.

    In summary, this 510(k) submission demonstrates the substantial equivalence and safety/effectiveness of adding a new surgical indication to an already cleared robotic surgical system, not the performance of a new AI/ML device against specific quantitative acceptance criteria. The "study" here is a retrospective analysis of clinical outcomes from a large surgical registry and published literature.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173842

    Validate with FDA (Live)

    Date Cleared
    2018-04-23

    (126 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294): Gastric Bypass (Roux-en-Y) and Gastric Sleeve. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    This document is a 510(k) summary for a labeling modification of the Intuitive Surgical da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200). It focuses on adding specific gastric procedures (Gastric Bypass and Gastric Sleeve) to the "general laparoscopic surgical procedures" indication for use. Therefore, the acceptance criteria and supporting studies are related to the safety and effectiveness of the device for these additional procedures, primarily against existing surgical methods (open and laparoscopic).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state formal "acceptance criteria" with specific numerical thresholds for performance metrics. Instead, it demonstrates substantial equivalence to predicate devices and established surgical practices (open and laparoscopic) by showing comparable or improved outcomes in various metrics. The acceptance is implicitly based on these comparisons showing non-inferiority or superiority in critical safety and effectiveness parameters.

    Metric (Acceptance Criteria implicitly: comparable or better than predicate/laparoscopic/open surgery)Reported Device Performance (da Vinci-assisted vs. Open)Reported Device Performance (da Vinci-assisted vs. Laparoscopic)
    Mortality Rate (in-hospital or 30 days)ComparableComparable
    Lengths of Hospital StayShorterComparable or Shorter
    Intraoperative Complication RatesComparableComparable
    Postoperative Complication RatesComparable or LowerComparable or Lower
    Anastomotic Leak RateComparableComparable or Lower (depending on anastomosis technique)
    Anastomotic Stricture RateComparableComparable or Lower (depending on anastomosis technique)
    Reoperation RateComparable or LowerComparable or Lower
    Readmission RateNot reported for vs. OpenComparable
    Operative TimeIncreased (but not associated with increased mortality/complication)Increased (in N=7 publications) or Comparable/Shorter (in N=3 publications), not associated with increased mortality/complication
    Estimated Blood Loss (EBL)Not reported for vs. OpenComparable
    Conversion RatesNot reported for vs. OpenComparable or Lower

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" in this context refers to the published clinical data analyzed for Gastric Bypass (Roux-en-Y), which served as the "umbrella procedure" to support both Gastric Bypass and Gastric Sleeve.
    • Sample Size (for Gastric Bypass umbrella procedure): The clinical data summarized in Table 1A involved 10 publications. The total number of patients included across these studies is:
      • da Vinci (Robotic) Cohort: 172 + 61 + 100 + 100 + 100 + 125 + 100 + 65 + 388 + 143 = 1,354 patients
      • Laparoscopic (Lap) Cohort: 173 + 46 + 100 + 100 + 100 + 121 + 100 + 54 + 389 + 323 = 1,506 patients
      • Open Cohort: 95 + 524 = 619 patients
    • Data Provenance: The data are from published clinical studies. The specific countries of origin are not mentioned, but clinical literature is generally international. The studies include both non-randomized, controlled, prospective studies (2 studies) and retrospective cohort studies (8 studies).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not provided in the given text. The "ground truth" for these clinical outcome studies is inherent in the reported patient outcomes and surgical metrics gathered and published by the original study authors. There is no mention of independent experts establishing a ground truth for a test set for the purpose of this 510(k) submission. The FDA's review process inherently involves medical experts, but they are assessing the submitted evidence, not establishing a new ground truth.

    4. Adjudication Method for the Test Set

    • This information is not provided. The text describes a review of published clinical data. Adjudication methods (like 2+1 or 3+1 consensus) are typically associated with image-based diagnostic studies or clinical trials where expert review of individual cases is part of the study design. For a meta-analysis or review of published literature on surgical outcomes, the "adjudication" is primarily how each original study defined and reported its outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. MRMC studies are typically used to evaluate the diagnostic performance of a device (often imaging-based AI) by comparing how multiple readers perform with and without the device. This submission is for a surgical system and its labeled indications, and the performance evaluation relies on real-world surgical outcomes and peer-reviewed literature, not diagnostic reader performance.
    • Effect size of human readers improving with AI vs. without AI assistance: Not applicable, as it's not an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm-only) performance study was not done. The da Vinci system is a surgical assistance system, inherently involving a human surgeon (human-in-the-loop). Its "performance" is evaluated by the outcomes of surgeries performed with the system assisting the surgeon, compared to traditional methods.

    7. The Type of Ground Truth Used

    • The "ground truth" for the clinical performance evaluation is based on patient outcomes and surgical metrics reported in peer-reviewed clinical literature (e.g., mortality, complication rates, length of stay, operative time, reoperation rate, etc.). This is essentially outcomes data collected and documented in clinical studies.

    8. The Sample Size for the Training Set

    • This information is not applicable in the conventional sense for this type of submission. The da Vinci system is an electro-mechanical surgical system, not an AI/ML algorithm that undergoes a distinct "training" phase on a dataset of cases to learn to perform a task. The "experience" or "learning" of the system happens through engineering design, software development, preclinical testing, and extensive clinical validation over many years.
    • However, if one were to consider the continuous development and refinement of the system's design and software, it would be based on an enormous "dataset" of engineering principles, surgical knowledge, user feedback, and prior clinical experience.

    9. How the Ground Truth for the Training Set Was Established

    • As explained above, a "training set" with established ground truth in the context of machine learning is not directly applicable here. The development of the surgical system relies on established surgical practices, engineering standards, and medical knowledge. Pre-clinical animal studies (mentioned as 6 evaluations in 24 animals) contribute to the developmental understanding and validation of the system's design, but this isn't a "training set" in the AI sense.
    • The "ground truth" for surgical device development is rooted in the safety and effectiveness principles derived from surgical anatomy, physiology, surgical techniques, and clinical outcomes, primarily established by expert surgeons and medical researchers over decades.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172643

    Validate with FDA (Live)

    Date Cleared
    2018-01-31

    (152 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System, Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model 184200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294): Pancreatectomy (partial, distal, segmental), Pancreaticoduodenectomy (Whipple Procedure), Splenectomy and Cholecystectomy. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a labeling modification of the da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200. This modification adds specific representative procedures (Pancreatectomy, Pancreaticoduodenectomy, Splenectomy, and Cholecystectomy) under the existing "general laparoscopic surgical procedures" indication for use.

    Crucially, the submission states that there are no changes to the technological characteristics of the device. Therefore, the "device" in question (the da Vinci surgical system itself) is already cleared, and the performance data presented is to establish substantial equivalence for the expanded indications for use, not to prove the fundamental performance of the surgical robot. The acceptance criteria and "device performance" in this context refer to the clinical outcomes of procedures performed using the da Vinci system compared to traditional open or laparoscopic methods, demonstrating that the new indications are safe and effective.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the document. Instead, the study aims to demonstrate that da Vinci Xi/X-assisted procedures for the newly added indications are comparable or superior to existing open or laparoscopic surgical procedures across a range of clinical outcomes.

    The "reported device performance" is derived from the clinical literature review for the "umbrella procedures" (Pancreatectomy and Pancreaticoduodenectomy), which are considered representative of the more complex/higher-risk operations. The less complex/lower-risk "covered procedures" (Splenectomy and Cholecystectomy) were supported by inference from the umbrella procedures.

    Below is a summary of the findings, which implicitly serve as the "acceptance criteria" and "reported performance" through comparative equivalence:

    Performance MetricDa Vinci-assisted vs. Open/Laparoscopic Pancreatectomy (Umbrella Procedure #1)Da Vinci-assisted vs. Open/Laparoscopic Pancreaticoduodenectomy (Umbrella Procedure #2)
    MortalityComparable mortality rates.Comparable mortality rates.
    Estimated Blood Loss (EBL) / Transfusion RatesComparable or lower blood transfusion rates and/or EBL volumes.Comparable EBL volumes.
    Length of Hospital StayComparable or shorter length of hospital stay as compared to laparoscopy and open surgery.Comparable length of hospital stay as compared to laparoscopy and comparable or shorter length of hospital stay as compared to open surgery.
    Complication RatesComparable or lower complication rates.Comparable or lower complication rates.
    Spleen Preservation RatesHigher or comparable spleen preservation rates. (Specific to pancreatectomy)Not applicable / Not reported in the context of this procedure.
    Gastric Emptying RatesLower or comparable delayed gastric emptying rates.Lower or comparable delayed gastric emptying rates.
    Reoperation RatesComparable reoperation rates.Comparable reoperation rates.
    Readmission RatesComparable or higher readmission rates.Not explicitly summarized as "comparable or higher" but individual studies show variations (e.g., Orti-Rodriguez not reported, Zhang not reported). Needs closer interpretation of Tables 2A/2B.
    Pancreatic Fistula RatesComparable pancreatic fistula rates as compared to both open and laparoscopic surgical procedures.Comparable pancreatic fistula rates as compared to both open and laparoscopic surgical procedures.
    Operative TimeIncreased operative times were associated with da Vinci procedures as compared to open and laparoscopic procedures. (Not associated with increased mortality/complications).Increased operative times were associated with da Vinci procedures as compared to open and laparoscopic procedures. (Not associated with increased mortality/complications).
    Conversion RatesComparable or lower conversion rates reported for da Vinci-assisted procedures as compared to laparoscopic procedures.Comparable or lower conversion rates for da Vinci-assisted procedures as compared to laparoscopic procedures.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Studies (Test Set):

      • Pancreatectomy (Umbrella Procedure #1): Eleven (11) publications were identified. The sample sizes within these publications varied significantly. For example, in Table 1A, individual study cohorts ranged from 22 to 3743 patients for da Vinci, laparoscopic, and open procedures. The largest single cohort for da Vinci was 258 (Zureikat, 2017) and for open was 3743 (Zureikat, 2017).
      • Pancreaticoduodenectomy (Umbrella Procedure #2): Eight (8) publications were identified. Individual study cohorts ranged from 102 to 3743 patients. The largest single cohort for da Vinci was 476 (Pedziwiatr, 2017) and for open was 3743 (Zureikat, 2017).
      • Covered Procedures (Splenectomy and Cholecystectomy): No specific clinical studies were provided for these procedures; their clearance was supported by the data from the umbrella procedures.
    • Data Provenance: The data are derived from published clinical literature, including:

      • Prospective randomized controlled trials
      • Meta-analyses
      • Systematic reviews
      • Large database comparative studies

      The specific countries of origin are not specified, but peer-reviewed medical literature typically encompasses international data. The studies are retrospective and prospective, depending on the nature of the individual publications included in the literature review.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable in the traditional sense for this submission. The "ground truth" for the clinical effectiveness and safety of the da Vinci system for these procedures is established by the peer-reviewed published clinical literature, which is generated by numerous clinical researchers and practicing surgeons. The FDA review process itself involves experts (e.g., "Director, Division of Surgical Devices, Office of Device Evaluation") who evaluate the scientific rigor and conclusions of the submitted literature review.

    4. Adjudication Method for the Test Set

    Not directly applicable in the terms of human reader adjudication of images. The "adjudication" in this context is the consensus reached through the scientific and statistical analysis presented in the peer-reviewed meta-analyses, systematic reviews, and comparative studies. The FDA then adjudicates the strength of this evidence for the purpose of granting substantial equivalence.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The submission relies on a literature review comparing clinical outcomes of different surgical approaches (da Vinci, laparoscopic, open).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This question is not applicable. The da Vinci Surgical System is a human-in-the-loop device; it is controlled by a surgeon. The "performance" being evaluated is the outcome of surgical procedures when performed using the da Vinci system (with a human surgeon operating it), compared to other surgical methods. This is not an AI algorithm performing a task autonomously.

    7. The Type of Ground Truth Used

    The "ground truth" for the clinical performance claims is based on patient outcomes data reported in published clinical trials and studies. These outcomes include:

    • Mortality rates
    • Estimated blood loss
    • Length of hospital stay
    • Postoperative complication rates
    • Spleen preservation rates (for pancreatectomy)
    • Gastric emptying rates
    • Reoperation rates
    • Readmission rates
    • Pancreatic fistula rates
    • Operative time
    • Conversion rates
    • R0/R1 resection rates
    • Lymph node yield

    8. The Sample Size for the Training Set

    No explicit "training set" for an AI algorithm is mentioned as this device is not presented as an AI-driven autonomous system in this submission. The performance data is for the surgical system itself. The publications reviewed in the submission can be considered the evidence base that informs the regulatory decision, rather than a training set for an AI model.

    9. How the Ground Truth for the Training Set Was Established

    As no training set for an AI algorithm is mentioned, this question is not applicable. The clinical "ground truth" for the effectiveness of the surgical system is established through the standard methodologies of clinical research (e.g., prospective trials, retrospective analyses, meta-analyses) that generated the published literature.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171699

    Validate with FDA (Live)

    Date Cleared
    2017-07-28

    (50 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K090993, K131861, K171294

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System, Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System Model IS3000) is intended to r no multive accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp assist in the dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation approximation against, orceral procedures, general laparoscopic surgical procedures, gynecologic probos and acoussent anning actrage tal otolaryngology surgical procedures restricted to benign and malignant itumors classified as T1 and T2, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy rumors clasmed as 1 x and 12, gateda adjunctive mediastinotomy to perform coronary anastomosis during procedurer The system is indicated for adult and pediatric use. It is intended to be used by trained earchies in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model 184200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The da Vinci Surgical Systems (Models IS3000, IS4000 and IS4200) are software-controlled, electromechanical systems designed for surgeons to perform minimally invasive surgery. The da Vinci Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided text is a 510(k) summary for the da Vinci Surgical System, focusing on a labeling change rather than a study proving performance against acceptance criteria for a new device or significant modification.

    Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment is not available in this document. This 510(k) is specifically for a minor administrative change (a labeling update to contraindications), not a clinical performance study.

    Here's an explanation of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Not applicable / Not provided. The document describes a change to labeling (contraindications) for existing devices already on the market, not a new device or a performance enhancement that would require a study with acceptance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided. No test set or clinical study data is presented as this submission is for a labeling change.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. No ground truth establishment is described as there is no clinical study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. No test set or adjudication process is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable / Not provided. This device is a surgical system, not an AI-assisted diagnostic tool or an "AI" in the typical sense that would involve human "readers" or an MRMC study comparing AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable / Not provided. The da Vinci system is a human-controlled surgical system. There is no standalone "algorithm-only" performance to evaluate in this context.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable / Not provided. No ground truth is established as there is no performance study.

    8. The sample size for the training set

    • Not applicable / Not provided. No new algorithm or trainable component is described in this 510(k) submission that would require a training set.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. No training set is involved.

    Summary of the document:

    This 510(k) submission (K171699) by Intuitive Surgical, Inc. is a Special 510(k) for their da Vinci Surgical Systems (Models IS3000, IS4000, IS4200). A Special 510(k) is typically used for well-defined modifications to a manufacturer's own legally marketed device that do not affect the indications for use or alter the fundamental scientific technology of the device, and for which the risk of the modification can be adequately assessed by the manufacturer's design controls.

    In this specific case, the modification is a labeling change to the "General Warnings and Cautions" section in the User Manuals, specifically concerning the endoscopic contraindication statement. The submitter argues that this change:

    • Does not change device usage or the patient population.
    • Does not change the risk profile of the devices.
    • Adequately instructs the user.

    Therefore, the submission concludes that the modified devices (with the updated labeling) are substantially equivalent to the predicate devices because the labeling modification has no impact on the safety or effectiveness of the da Vinci Surgical Systems themselves. This type of submission does not involve clinical studies or performance data against specific acceptance criteria for a new or significantly modified device function.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170713

    Validate with FDA (Live)

    Date Cleared
    2017-06-13

    (96 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endloscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparossopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatic use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The da Vinci Xi Surgical System, Model IS4000, is a softwarecontrolled, electro-mechanical system designed for surgeons to perform minimally invasive surgery. The Model IS4000 Surgical System consists of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and is used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    It appears that the provided text is a 510(k) summary for a medical device (da Vinci Xi Surgical System) and focuses on a labeling modification rather than presenting a performance study of the device itself against specific acceptance criteria for a new algorithmic feature. The submission is to include "Inguinal Hernia Repair" procedures under an already cleared "general laparoscopic surgical procedures" indication.

    Therefore, the information I can extract to directly answer your request about device acceptance criteria and a study proving it meets them for a specific performance metric (like algorithm accuracy) is limited, as this is primarily a procedural and comparative effectiveness submission for a surgical procedure, not an AI algorithm.

    However, I can reframe the "acceptance criteria" based on the comparative effectiveness study presented for the procedure performed with the device. The study aims to demonstrate that robotic-assisted inguinal hernia repair is "safe and effective" and compares several clinical outcomes to open surgery and laparoscopic procedures.

    Here's the closest I can get to your request based on the provided document:


    Acceptance Criteria and Study for da Vinci Xi Surgical System (for Inguinal Hernia Repair Procedure Labeling)

    The acceptance criteria are not explicitly stated in terms of a quantifiable threshold for a device's performance (e.g., AUC > 0.90 for an AI algorithm). Instead, the study aims to demonstrate that performing Inguinal Hernia Repair with the da Vinci Xi Surgical System is safe and effective and comparable or superior in key clinical outcomes to existing standard-of-care procedures (open and traditional laparoscopic surgery).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Implicit for Procedure)Reported Device Performance (Robotic-Assisted vs. Open/Laparoscopic)
    Safety:
    Intraoperative Complications (comparable to open)Comparable (0.3% robotic-assisted vs. 0% open; within 0-8% for laparoscopic)
    Transfusions (comparable to open)Comparable (0.3% perioperative robotic-assisted vs. 0.3% perioperative open; intraoperative 0% for both)
    Postoperative Complications (Discharge through 30 days) (comparable/better)Lower rate (4.3% robotic-assisted vs. 7.3% open); propensity-matched analysis also showed robotic-assisted associated with lower likelihood of complications.
    Readmission Rates (comparable to open)Comparable (3.5% robotic-assisted vs. 4.3% open)
    Reoperation Rates (Postoperative to Discharge) (comparable to open)Comparable (0.5% robotic-assisted vs. 0.3% open; within 0-2.5% for laparoscopic)
    Reoperation Rates (Discharge through 30 days) (comparable/better)Lower rate (0% robotic-assisted vs. 1.3% open); however, propensity-matched analysis showed comparable rates.
    Mortality (comparable to open)Comparable (0% robotic-assisted vs. 1 open)
    All Complications (comparable to open/laparoscopic)Comparable (6.9% robotic-assisted vs. 8.5% open; within 7.9-8.7% for laparoscopic)
    Effectiveness:
    Length of Stay (comparable to open/laparoscopic)Comparable outpatient; for inpatient, comparable, but propensity-matched analysis showed statistically lower inpatient length of stay for robotic-assisted.
    Operative Time (explicitly acknowledged as longer but without adverse impact)Longer (79.7 ± 31.7 min robotic-assisted vs. 45.0 ± 21.7 min open); however, this longer time was "not associated with increases in the complication, readmission, reoperation or mortality rates." (This implies an "acceptable" longer operative time if other outcomes are preserved or improved).
    Conversion Rate (for robotic-assisted to open)0.9% (reported for robotic-assisted, no comparable data for other cohorts directly in table format)
    Absence of adverse events (death or serious injury)44 adverse events reported for robotic-assisted, 49 for open, with a higher concentration at one site for both. The study defines this as serious injury (Clavien Grade IIIa or higher).

    2. Sample Size and Data Provenance

    • Test Set (Clinical Study Data):

      • Robotic-assisted cohort: 652 subjects
      • Open cohort: 602 subjects
      • Laparoscopic cohort: Variable N from 20 to 3,457 (derived from 11 published studies for comparison)
      • Data Provenance: Retrospective, multi-center, non-randomized controlled clinical study conducted at six (6) investigational sites in the United States. The laparoscopic comparison data is from published literature (Bittner, et al., 2015).
    • Training Set: Not applicable as this is not a machine learning model being trained and tested. This is a post-market clinical study for a labeling modification of an existing surgical system.

    3. Number of Experts and Qualifications for Ground Truth for Test Set

    • The "ground truth" for the clinical study is the actual clinical outcomes observed in patients undergoing the procedures (e.g., actual complications, readmissions, reoperations, length of stay, mortality). These are objective clinical endpoints, not subjective interpretations by experts. Therefore, the concept of "experts establishing ground truth" as it applies to an image interpretation AI model does not fit here. The study involved practicing surgeons and medical staff at the investigational sites who collected the data.

    4. Adjudication Method for the Test Set

    • The document does not describe a formal "adjudication method" for the clinical endpoints in the same way one would for an AI model's output. Clinical data collection would typically follow standard protocols at each investigational site, potentially with source data verification, but not an "adjudication" of the data itself by independent experts in the context of resolving disagreements on results. The data are quantitative clinical outcomes.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC study was not done. This study is a clinical comparative effectiveness study comparing surgical approaches (robotic-assisted vs. open vs. laparoscopic) for a specific procedure. It does not involve multiple human readers interpreting images with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone study was not done. The da Vinci Xi Surgical System is a surgical robotic system with human-in-the-loop (the surgeon). The study evaluates the performance of the procedure when performed with this system, not an isolated algorithm. The submission is for a labeling modification to include a new procedure, not a new algorithm or AI functionality.

    7. Type of Ground Truth Used

    • The ground truth used for the clinical study was clinical outcomes data collected from patient records (e.g., hospital discharge records, patient follow-ups). These include objective metrics like length of stay, occurrence of complications, reoperations, readmissions, and mortality.

    8. Sample Size for the Training Set

    • Not applicable; see point 2.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable; see point 2.

    Summary for this specific 510(k) context:

    This 510(k) is about expanding the indications for use of an existing surgical robotic system to include Inguinal Hernia Repair. The "study" here is a clinical comparative effectiveness study demonstrating that the procedure performed with the device is safe and effective and compares favorably or acceptably to existing surgical methods. It is not an AI/algorithm performance study on a test set, so many of the requested items (like expert ground truth for interpretation, MRMC, standalone algorithm performance, or specific AI training/test set details) are not directly addressed in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153276

    Validate with FDA (Live)

    Date Cleared
    2016-08-07

    (269 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    0.08 - 2
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K081137, K123463, K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedures under the cleared "thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures" general indication for the da Vinci Xi Surgical System (K131861): Lobectomy, Mediastinal Mass Resection, Thymectomy, Segmentectomy, Wedge Resection and Lymphadenectomy. There are no changes to the technological characteristics of the cleared da Vinci Xi Surgical System proposed in this submission. The da Vinci Xi Surgical System, Model IS4000 is a software-controlled, electro-mechanical system designed for surgeons to perform minimally invasive surgery. The Model IS4000 Surgical System consists of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and is used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Da Vinci Xi Surgical System, focusing on a labeling modification to include additional specific procedures under existing general indications for use. This involves demonstrating substantial equivalence to predicate devices, rather than establishing entirely new performance criteria for a novel device. Therefore, the information typically requested regarding acceptance criteria and a study proving a new device meets those criteria is not directly applicable in this context as new performance metrics are not being established.

    However, I can extract and structure the available information about the demonstration of suitability for the expanded indications, which functions as the system's "acceptance" for these new procedures.

    Here's an interpretation based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are implicitly demonstrating that the da Vinci Xi Surgical System, when used for the expanded "umbrella" and "covered" procedures, shows comparable or improved patient outcomes (mortality, complications, blood loss, length of stay, operative time) compared to established surgical methods (Open and VATS procedures).

    Acceptance Criteria (Comparable to Predicate or Established Methods)Reported Device Performance (da Vinci-assisted procedures)
    Lobectomy- Mortality Rates: Comparable or lower. - Complications: Comparable. - Blood Transfusion Rates: Comparable or lower. - Length of Hospital Stay: Comparable or shorter. - Operative Times: Comparable or shorter (5 studies), or increased (13 studies) (but no correlation with increased mortality). - Conversion Rates: Comparable or lower compared to VATS.
    Mediastinal Mass Resection- Mortality Rates: Comparable. - Operative Times: Comparable. - Complication Rates: Comparable or lower. - Length of Hospital Stay: Comparable or shorter compared to open surgical procedures. - Single-arm studies: No reports of mortality, low complication rates (intraoperative <1%, postoperative <=14%), low conversion rates (0.65-4.3%), low transfusion rates (0-1.4%).
    Thymectomy- Mortality Rates: Comparable. - Complication Rates: Comparable or lower. - Length of Hospital Stay: Comparable or shorter. - Transfusion Rates/EBL: Comparable or lower compared to open and VATS. - Operative Times: Comparable or longer (but no correlation with higher mortality rates). - Single-arm studies: No reports of mortality, low complication rates (intraoperative 1.3-15.4%, postoperative 3.6-15.4%), conversion rates 0-15.4%.
    Segmentectomy, Wedge Resection, Lymphadenectomy- The published data on the "umbrella" procedures (Lobectomy, Mediastinal Mass Resection, Thymectomy) were deemed sufficient to support the clearance of these "covered" procedures, implying similar expected performance based on the general surgical context and the device's technological characteristics remaining unchanged.

    2. Sample Size Used for the Test Set and Data Provenance

    • Pre-Clinical Animal Study Data:

      • Sample Size: A total of 20 animals were used in three evaluations.
      • Provenance: This was a preclinical study. Specific country of origin is not mentioned but it's an internal study supporting the submission.
    • Clinical Study Data (Human Data):

      • Sample Size: The clinical data relies on published retrospective, comparative, and single-arm studies.
        • Lobectomy: The provided table (Table 1) lists 18 publications with varying sample sizes per study group (ranging from N=22 to N=20,238 for comparison groups and N=22 to N=2,498 for da Vinci groups). The total number of patients across all studies is not explicitly summed, but it is substantial.
        • Mediastinal Mass Resection: Table 2 lists 7 publications with sample sizes ranging from N=14 to N=75 (da Vinci groups).
        • Thymectomy: Table 3 lists 15 publications with sample sizes ranging from N=6 to N=125 (da Vinci groups).
      • Provenance: These are published clinical data. The specific countries of origin are not detailed in this summary but are derived from a global body of medical literature. These studies are retrospective, comparative, and single-arm.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of independent experts being used to establish a "ground truth" for a test set in the traditional sense of an AI/ML device evaluating images or signals. The "ground truth" or clinical outcome data in this context comes from the published clinical studies themselves, where the outcomes (e.g., mortality, complications, operative time) were determined and reported by the treating physicians and study authors. The qualifications of these physicians are implicitly high, as they are authors of peer-reviewed clinical publications performing complex surgical procedures.

    4. Adjudication Method for the Test Set

    Not applicable. This submission relies on aggregated and summarized outcomes from published clinical literature, not on a new test set requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission is not for an AI-assisted diagnostic or interpretative device where human readers interact with AI. It's for a surgical system where the performance is measured by patient outcomes.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. The da Vinci Xi Surgical System is a tool used by human surgeons; it is not a standalone AI device performing tasks without human intervention. Its performance is measured in the context of human surgeons using the system.

    7. The Type of Ground Truth Used

    The "ground truth" used is outcomes data (e.g., mortality, complication rates, operative times, length of stay, blood loss) as reported in published clinical studies. This data represents real-world clinical results rather than an independent expert consensus on a test set or pathology reports for specific diagnoses.

    8. The Sample Size for the Training Set

    Not applicable. The da Vinci Xi Surgical System is an electro-mechanical system controlled by software, not a machine learning algorithm that undergoes a "training phase" to learn from a specific dataset in the way an AI diagnostic tool would. Its development and refinement are based on engineering principles and clinical experience, not a "training set" for an algorithm in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for an algorithm in the context of this device. The system's design and functionality are established through engineering, preclinical testing, and clinical experience, not algorithmic training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123463

    Validate with FDA (Live)

    Date Cleared
    2012-12-03

    (24 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K090993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, transoral otolaryngology surgical procedures restricted to benign and malignant tumors classified as T1 and T2, general thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use (except for transoral otolaryngology surgical procedures). It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The Intuitive Surgical da Vinci Si Surgical System is a computer-assisted device designed to facilitate complex surgery using a minimally invasive approach. The system consists of three main components: the Surgeon Console, the Patient Cart, and the Vision Cart. The Surgeon Console houses the hand (Master Tool Manipulators or MTMs) and foot controls for the system. These controls allow the surgeon to manipulate the endoscopic instruments and camera located on the Patient Cart. The Vision Cart houses the system's image processing unit as well as the Instrument Control Box and any electrosurgical generators used in conjunction with the endoscopic instruments. The proposed modification to the Intuitive Surgical da Vinci Si Surgical System user interface will result in a change from Functional Footpedal Mapping to Left/Right Associative Footpedal Mapping. Currently, under Functional Footpedal Mapping, instruments are assigned to footpedal banks according to their function (monopolar, bipolar, etc). The proposed change to Left/Right Associative Footpedal Mapping will provide a more intuitive user interface by associating each pedal bank with a hand controller or "master tool manipulator" (MTM). Therefore, instruments assigned to the left MTM will be controlled by the left pedal bank and the instruments assigned to the right MTM will be controlled by the right pedal bank.

    AI/ML Overview

    The provided text describes a Special 510(k) Pre-market Notification for a device modification, focusing on a change in footpedal mapping for the da Vinci Si Surgical System. The purpose of the submission is to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Human Factors/UsabilitySubstantially equivalent in terms of error rate.Human Factors testing was conducted to demonstrate that the subject device is substantially equivalent to the predicate device in terms of error rate.
    Human Factors/UsabilitySubstantially equivalent in terms of risk profile.Human Factors testing was conducted to demonstrate that the subject device is substantially equivalent to the predicate device in terms of... risk profile.
    Software VerificationDesign specifications meet design requirements.Software verification testing was performed to confirm that the design specifications meet the design requirements.
    Safety and EffectivenessNo new types of safety or effectiveness questions raised.The results of the testing did not raise any new types of safety or effectiveness questions.
    Technological CharacteristicsSubstantially equivalent in terms of intended use.The subject device... is substantially equivalent in technological characteristics in terms of the intended use... as compared to the da Vinci Si System, with Functional Footpedal Mapping (the predicate device).
    Technological CharacteristicsSubstantially equivalent in terms of indications for use.The subject device... is substantially equivalent in technological characteristics in terms of the... indications for use... as compared to the da Vinci Si System, with Functional Footpedal Mapping (the predicate device).
    Technological CharacteristicsSubstantially equivalent in terms of technological characteristics.The subject device (da Vinci Si System, with Left/Right Associative Footpedal Mapping) is substantially equivalent in technological characteristics... as compared to the da Vinci Si System, with Functional Footpedal Mapping (the predicate device).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "Human Factors testing" and "Software verification testing," but does not specify the sample size used for the test set for either.
    The data provenance is not explicitly mentioned, but given that Intuitive Surgical is a U.S.-based company and the submission is to the FDA, it is highly likely that testing was conducted in the United States. The text does not specify if the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the human factors testing. It only states that human factors testing was performed.

    4. Adjudication Method for the Test Set:

    The document does not specify any adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    This device is not an AI-based diagnostic tool for interpretation by human readers. It's a surgical system modification changing the input mechanism (footpedal mapping). Therefore, a MRMC comparative effectiveness study is not relevant or applicable in this context, nor is an "effect size of how much human readers improve with AI vs without AI assistance." The study focuses on usability and safety of a control interface change.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done:

    The document describes "Software verification testing" which confirms design specifications. This can be considered a standalone assessment of the software's functionality in fulfilling its requirements. However, the primary performance evaluation for the user interface change is the "Human Factors testing," which inherently involves humans (users of the system) in the loop to assess usability and error rates. So, while software verification might be considered standalone, the overall performance evaluation of the device modification is not.

    7. The Type of Ground Truth Used:

    For the "Human Factors testing," the ground truth implicitly refers to objective measures of errors and risk profiles in system operation, likely compared against a benchmark or expected safe performance, and against the predicate device's performance. For "Software verification testing," the ground truth is simply whether the software meets its predefined design requirements/specifications.

    8. The Sample Size for the Training Set:

    The concept of a "training set" is typically associated with machine learning or AI models. Since this device modification is a change in a user interface (footpedal mapping) for a surgical system, rather than an AI algorithm being trained, there is no mention or relevance of a training set in this document.

    9. How the Ground Truth for the Training Set was Established:

    As there is no training set for this type of device modification, this question is not applicable to the provided text.

    Ask a Question

    Ask a specific question about this device