    K Number
    K133046

    Validate with FDA (Live)

    Date Cleared
    2014-05-14

    (229 days)

    Product Code
    Regulation Number
    878.3500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K123908, K083621

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OmniPore® Customized Surgical Implants are intended for non-weight bearing applications of craniofacial reconstruction/ cosmetic surgery and repair of craniofacial trauma. OmniPore Customized Surgical Implants are also intended for the augmentation or restoration of contour in the craniomaxillofacial skeleton.

    Device Description

    The OmniPore® Customized Surgical Implants are marketed as single patient use sterile implants that physicians request as customized surgical implants which use identical materials and manufacturing to the OmniPore® Surgical Implants but are made to a predetermined patient's measurements and size requirements. The applications include non-load bearing augmentation and/or reconstruction of the craniofacial skeleton. The OmniPore Customized Surgical Implants are created from the patient's CT imaging data provided from the physician. The OmniPore Customized Surgical Implants are manufactured from the same material, manufactured under the same processes, and packaged the same as the OmniPore Surgical Implants.

    The raw material used for the OmniPore Customized Surgical Implants is high-density polyethylene resin which has a long history of use in surgical implantable products. The interconnecting open pore structure of the OmniPore Customized Surgical Implants allow for tissue in-growth. Additionally, animal histology has shown fibrovascular in-growth occurs into the open pore structure of OmniPore Customized Surgical Implants.

    The implants are single use and provided sterile by ethylene oxide (EO) terminal sterilization.

    AI/ML Overview

    This document describes the OmniPore® Customized Surgical Implants, a device intended for non-weight-bearing applications in craniofacial reconstruction/cosmetic surgery, repair of craniofacial trauma, and augmentation/restoration of contour in the craniomaxillofacial skeleton.

    Here's an analysis of the provided information regarding acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state specific quantifiable acceptance criteria for the device's performance. Instead, it relies on demonstrating substantial equivalence to predicate devices. The "reported device performance" is implicitly that the device performs equivalently to its predicates across various aspects.

    Acceptance Criteria CategoryAcceptance Criteria (Explicitly Stated in Document)Reported Device Performance
    Material EquivalenceNot explicitly stated as a quantitative criterion, but implicit expectation is the materials are identical or perform equivalently to predicate devices."The OmniPore® Customized Surgical Implant materials are equivalent materials as the previously cleared OmniPore Surgical Implant devices..."
    BiocompatibilityNot explicitly stated as a quantitative criterion, but implicit expectation is the device is biocompatible as demonstrated by predicate testing."...so the biocompatibility and sterilization validation to validate that they are sterile devices for implantation was justified from the predicate testing." (Performance is inferred to be adequate based on predicate.)
    Sterilization EfficacyNot explicitly stated as a quantitative criterion, but implicit expectation is terminal sterilization is effective (e.g., Sterility Assurance Level of 10^-6)."...so the biocompatibility and sterilization validation to validate that they are sterile devices for implantation was justified from the predicate testing." (Performance is inferred to be adequate based on predicate.) The implants are "provided sterile by ethylene oxide (EO) terminal sterilization."
    Mechanical Properties (Impact, Purity, Porosity)Not explicitly stated as a quantitative criterion, but implicit expectation is that these properties are equivalent to predicate devices."The OmniPore Customized Surgical Implants are equivalent materials as the previously cleared OmniPore Surgical Implant devices so the mechanical testing specific to impact testing, purity testing per USP, and porosity testing was justified from the OmniPore Surgical Implant device testing." (Performance is inferred to be adequate based on predicate.)
    Software Verification (for Customization Process)Not explicitly stated, but implicit expectation is that the software used for customized design is validated and functions equivalently to predicate."The software verification testing documentation for the OmniPore Customized Surgical Implants was provided to justify equivalence against the MEDPOR® Customized Surgical Implant." (Performance is inferred to be adequate based on predicate, indicating the software part of the device's manufacturing process has been verified.)
    Clinical Performance / Safety and Effectiveness for Intended Use (Overall Substantial Equivalence)No specific quantitative clinical endpoints or safety profiles are provided as acceptance criteria in this document."The OmniPore® Customized Surgical Implants have the same intended use and indications for use, and the same technological characteristics and principles of operation as the predicate devices. The minor differences do not raise any issues of safety or effectiveness. Testing results support the determination of substantial equivalence with the results demonstrating that the OmniPore Customized Surgical Implants have equivalent results as the predicate devices." (Summary conclusion of equivalence.)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "test set" in the traditional sense of a clinical or retrospective study with patient data for evaluating the device's performance. The evaluation for this 510(k) submission is based on demonstrating equivalence to predicate devices, primarily through material and manufacturing equivalency, and justification from existing predicate testing data.

    Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or provided in this type of submission which focuses on substantial equivalence rather than a de novo clinical study. The "testing" mentioned refers to justification based on previously conducted tests for predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as there is no mention of a clinical "test set" requiring expert ground truth establishment for patient outcomes or diagnoses. The submission relies on scientific and engineering evaluations for material equivalence and manufacturing processes.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a "test set" requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The device described is a physical surgical implant, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. The device is a physical surgical implant, not an algorithm.

    7. The Type of Ground Truth Used

    Given the nature of the submission (demonstrating substantial equivalence for a physical implant), the "ground truth" is largely:

    • Material properties and manufacturing process specifications: Confirming the raw materials, manufacturing methods, and sterilization processes are the same or equivalent to the predicate devices.
    • Existing predicate device performance data: Relying on the established safety and effectiveness of the legally marketed predicate devices through their prior biocompatibility, sterilization validation, and mechanical testing results.
    • Software verification documentation: For the customization process, ensuring the software performs its intended function for creating the customized implant based on CT data, likely verified against functional requirements and possibly accuracy checks against reference CAD models or similar.

    There is no mention of pathology, expert consensus on patient outcomes from a new study, or specific outcomes data for the OmniPore® Customized Surgical Implants itself beyond the predicate justification.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is not an AI/ML algorithm that requires a training set. The customization is based on patient-specific CT imaging data for manufacturing, not on a machine learning model trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101835

    Validate with FDA (Live)

    Date Cleared
    2010-12-03

    (155 days)

    Product Code
    Regulation Number
    882.5360
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K832283, K922489, K952677, K040364, K083621

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDPOR Fixation System - Cranial is intended for use in cranial surgery to fixate MEDPOR cranial implants to the surrounding cranial skeleton.

    Device Description

    The MEDPOR Fixation System - Cranial is a plate and screw system to be used in Cranial Surgery. The components that comprise the system are composed of titanium plates. titanium alloy bone screws, titanium alloy implant screws, and surgical stainless steel instruments for the installation of the plates and screws. The plates are manufactured of grade 4 titanium and adhere to the American Society of Testing Materials (A.S.T.M.) F67 Standard. The bone and implant screws are manufactured of 6-4 ELI titanium that meets the ASTM F136 standard. The implant screws are designed to be used with MEDPOR polyethylene implants manufactured by Porex Surgical under K832283. K922489. K952677, K040364 and K083621. The screws and plates have been color anodized for ease of use by a surgeon. The blue anodized side of the plate corresponds to the blue anodized bone screw and the magenta anodized side of the plate corresponds to the magenta anodized implant screw. The MEDPOR Fixation System -- Cranial is sterilized by ethylene oxide sterilization. The instruments and unused components can be resterilized via steam sterilization.

    AI/ML Overview

    This document is a 510(k) premarket notification for the MEDPOR® Fixation System - Cranial. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting detailed acceptance criteria and a study to prove device performance against those criteria. In the context of a 510(k) submission for a device like plates and screws, the "acceptance criteria" are generally related to demonstrating that the new device is as safe and effective as existing legally marketed devices (predicate devices) and that any differences do not raise new questions of safety or effectiveness.

    Therefore, the requested information elements (1-9) which typically relate to performance studies for novel or significantly modified devices, are not explicitly provided in this 510(k) summary. A 510(k) does not usually include a formal "study" with stated acceptance criteria and device performance results in the same way a PMA or de novo submission might. Instead, it relies on comparisons to predicate devices and adherence to established material standards and manufacturing practices.

    Here's a breakdown of what can be extracted and what is not applicable or stated:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as pass/fail criteria from a specific performance study in this 510(k) summary. The acceptance criteria for a 510(k) is generally the demonstration of "substantial equivalence" to predicate devices, meaning the new device has "similar technological characteristics" and "same intended use."
    • Reported Device Performance: No specific performance metrics (e.g., tensile strength, fatigue life, etc.) are reported with numerical values in this summary. The submission implies that the material and design choices (titanium, specific alloys) meet established standards (ASTM F67, ASTM F136), which are themselves forms of performance criteria.

    2. Sample size used for the test set and the data provenance

    • Not applicable as no specific device performance testing study with a "test set" and "data provenance" is detailed in this 510(k) summary. The comparison is primarily against the characteristics of predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable as there is no "test set" ground truth established by experts in the context of this 510(k) summary. Substantial equivalence is assessed by regulatory reviewers based on design, materials, and intended use.

    4. Adjudication method for the test set

    • Not applicable for the same reasons as #3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a medical device for surgical fixation, not an AI or imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical device for surgical fixation, not an algorithm.

    7. The type of ground truth used

    • The "ground truth" for a 510(k) submission for this type of device is the evidence that the device's technological characteristics (materials, design, sterilization) and intended use are substantially equivalent to legally marketed predicate devices, which are themselves considered safe and effective. The reference to conformity with ASTM standards (F67 and F136) serves as a form of "ground truth" for material properties.

    8. The sample size for the training set

    • Not applicable as there is no "training set" in the context of this 510(k) submission. Design and manufacturing are based on established engineering principles and material standards.

    9. How the ground truth for the training set was established

    • Not applicable for the same reasons as #8.

    Summary based on the provided document:

    The provided document is a 510(k) Summary. For devices like the MEDPOR® Fixation System - Cranial, the "acceptance criteria" are intrinsically linked to demonstrating substantial equivalence to predicate devices already on the market. The "study" proving this largely involves a comparison of technological characteristics (materials, design, indications for use, sterilization) and adherence to recognized standards.

    Table of Acceptance Criteria (Implied by Substantial Equivalence) and Reported Features:

    Criteria Category (Implied)Acceptance Criteria (How substantial equivalence is demonstrated)Reported Device Features / Performance
    Intended UseMust be substantially equivalent to predicate devices."Intended for use in cranial surgery to fixate MEDPOR cranial implants to the surrounding cranial skeleton." (Compared to predicate intended uses which include cranial, craniofacial, and reconstructive procedures of the craniofacial skeleton).
    Device ClassificationMust align with predicate devices or existing classifications.Class II, Code HBW, Reg. No. 882.5360 - Cranioplasty plate fastener (Matches or is consistent with classifications of predicate devices).
    MaterialsMust be substantially equivalent to predicate devices, preferably with a history of safe use, and adhere to recognized standards.Plates: CP Grade 4 Titanium (Adheres to ASTM F67 Standard). Screws: 6-4 ELI Titanium Alloy (Meets ASTM F136 standard). (Comparable to titanium/titanium alloy used in multiple predicate devices, though one predicate uses bioabsorbable material).
    DesignMust be substantially equivalent in form and function (e.g., plates and screws for fixation).Plate and screw system. (Exhibited by all predicate devices). Includes color anodization for ease of use.
    SterilizationMust use appropriate and validated sterilization methods.Ethylene Oxide & Steam sterilization. (Comparable to predicate devices which use steam, or are supplied sterile/non-sterile with instructions for steam sterilization).
    InstrumentationPresence of dedicated instrumentation.Yes. (Comparable to most predicate devices).

    Explanation of Study (Substantial Equivalence Demonstration):

    The "study" in this context is the detailed comparison provided in the "Substantial Equivalence" section and the comparative table on page 2 of the document. This comparison directly addresses how the MEDPOR Fixation System - Cranial aligns with the identified predicate devices:

    • Comparison of Intended Use: The document argues that the intended use of the MEDPOR system (fixating cranial implants to the cranial skeleton) is comparable to the broad craniofacial, cranial, and reconstructive procedures of the predicate devices.
    • Comparison of Materials: The use of specific titanium grades (CP 4 Titanium for plates, 6-4 ELI Titanium Alloy for screws) is stated to meet ASTM standards F67 and F136, respectively. The document highlights that titanium has a "long history of use in surgical implantable products" and that these materials are consistent with most predicate devices (Predicate Device #1 is an exception, being bioabsorbable).
    • Comparison of Technological Characteristics: The document states that the "technological characteristics of the design are substantially equivalent to the predicate devices and any slight differences do not raise new issues of safety and effectiveness." This includes design as a plate and screw system, dedicated instrumentation, and sterilization methods.

    Missing/Not Applicable Information:

    • Sample Size for Test Set/Training Set: Not applicable for a 510(k) based on substantial equivalence comparison of device characteristics.
    • Data Provenance: Not applicable.
    • Experts/Ground Truth for Test/Training set: Not applicable in the sense of clinical study experts. The "ground truth" hinges on the established safety and effectiveness of the predicate devices and conformity to recognized standards.
    • Adjudication Method: Not applicable.
    • MRMC or Standalone AI Study: Not applicable as this is not an AI/diagnostic device.