Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K102688

    Validate with FDA (Live)

    Date Cleared
    2010-12-03

    (77 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K081513, K090063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Merete Locking Bone Plate System can be used for adult and pediatric patients. Indications for use include fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers, and toes.

    Device Description

    The Merete Locking Bone Plate System II consists of contoured, anatomically shaped locking bone plates in various sizes and different curvatures for Merete 3.0mm and 3.5mm locking screws (K081513) and cannulated 3.0mm compression screws (K090063). The system is made of titanium alloy Ti-6Al-4V and unalloyed titanium.

    AI/ML Overview

    This document is a 510(k) summary for the Merete Locking Bone Plate System II. It describes a medical device, specifically a bone plate system, and its intended use. There is no information about acceptance criteria or a study proving the device meets acceptance criteria in the provided text.

    The document focuses on:

    • Identifying the device and its manufacturer.
    • Listing predicate devices to establish substantial equivalence.
    • Describing the device's materials and technological characteristics (stating they are similar to predicate devices).
    • Stating its intended use (fixation of fractures, revision procedures, etc., in small bones).
    • Providing FDA correspondence regarding its 510(k) clearance.

    Therefore, I cannot provide the requested table and study details as they are not present in the given input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090063

    Validate with FDA (Live)

    Date Cleared
    2009-04-22

    (103 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K067808, K063487, K081513

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Merete Locking Bone Plate System can be used for adult and pediatric patients. Indications for use include fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers, and toes.

    Device Description

    The Merete Locking Bone Plate System consists of contoured, anatomically shaped locking bone plates in various sizes and different curvatures, Merete 3.0mm and 3.5mm locking screws and cannulated 3.0mm compression screws. The system is made of titanium alloy Ti-6Al-4V.

    AI/ML Overview

    The provided text describes a 510(k) submission for the "Merete Locking Bone Plate System" and focuses on its substantial equivalence to predicate devices, rather than a study demonstrating acceptance criteria met by the device itself through performance evaluation.

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance: The document does not specify quantitative acceptance criteria or report performance metrics for the Merete Locking Bone Plate System. The submission is based on demonstrating substantial equivalence to existing devices.
    • Sample size used for the test set and the data provenance: No test set or associated data are mentioned for evaluating device performance.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: There is no mention of a test set requiring ground truth establishment by experts.
    • Adjudication method for the test set: No test set is described.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: This type of study is not mentioned. The device is a bone plate, not an imaging or diagnostic AI tool.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is irrelevant for a physical bone plate device.
    • The type of ground truth used: Not applicable as no performance study is detailed.
    • The sample size for the training set: Not applicable as this is a physical medical device, not an AI/algorithm-based one.
    • How the ground truth for the training set was established: Not applicable.

    Summary from the provided document:

    The 510(k) submission focuses on demonstrating substantial equivalence for the Merete Locking Bone Plate System to legally marketed predicate devices, rather than presenting a study with acceptance criteria and performance data. The FDA's letter states that the device is substantially equivalent "for the indications for use stated in the enclosure" to predicate devices, meaning it has similar technological characteristics, materials, and intended use as devices already on the market. This pathways allows medical devices to be marketed without requiring new clinical trials, if they can show they are as safe and effective as a legally marketed predecessor.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1