510(k) Data Aggregation

    K Number
    K162775

    Validate with FDA (Live)

    Date Cleared
    2017-01-13

    (102 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080625, K103256, K123012, K951987

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Legion Cones are intended to be used with the Legion Revision and Legion Hinge knee systems. The Legion Cones are indicated for the following:

    1. Rheumatoid arthritis.

    2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.

    3. Failed osteotomies, unicompartmental replacement, or total knee replacement.

    4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

    5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.

    6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are absent of incompetent.

    These are single use implants and are intended for use with and without bone cement.

    Device Description

    Subject of this Traditional premarket notification is the Smith & Nephew Legion™ Cone System. The Smith & Nephew Legion Cone System consists of porous coated tibial and femoral augments to address cavitary defects in the proximal tibia or distal femur. The system will also include the necessary instrumentation for the surgeon to prepare the tibia and femur for implantation of the augments.

    The components of the Smith & Nephew Legion Cone System are as follows:

    • Titanium alloy (Ti-6Al-4V) tibial cones of 2 heights and 7 sizes .
    • · Titanium alloy (Ti-6Al-4V) hand-specific femoral cones of 1 height and 7 sizes

    The Leqion Tibial and Femoral Cones are a one piece, conically shaped device with cannulation all the way through the cone which allows fixation to the Legion Revision and Legion Hinge tibial baseplates, femoral components, offset couplers, augments/wedges, and stem constructs.

    The Legion Tibial and Femoral Cones are similar to the primary predicate Zimmer Trabecular™ Metal Tibial and Femoral Cone Augments cleared via premarket notification K053340. The Legion Tibial Cones will be available in heights of 25mm (short) and 40mm (long) and inner diameters (ID) of 18mm-30mm, in 2mm increments. The Legion Femoral Cones will be available in a height of approximately 35mm and inner diameters of 18mm-30mm, in 2mm increments. The Legion Femoral Cones will be provided in left and right orientations. The subject tibial and femoral cones are manufactured from forged titanium alloy (Ti-6Al-4V) that conforms to ASTM F 1472. The subject tibial and femoral cones also have an external asymmetric (Stiktite™) porous coating specification identical to the porous coating of the reference predicate Smith & Nephew SMF™ Hip Stems cleared via premarket notifications K080625. K103256, and K123012. The interior side of the subject cones has a grit-blasted surface specification for cement adhesion identical to the gritblasted surface of reference predicate Smith & Nephew Genesis II Total Knee System cement-on-augment devices cleared via premarket notification K951987.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "Legion Cone System." This type of document is a submission to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. It typically does not contain information about acceptance criteria for a study proving device performance in the way that a clinical trial or a formal performance study for an AI/CADe device would.

    Instead, for this type of device (knee prosthesis components), the "acceptance criteria" are generally met through mechanical testing and a comparison to predicate devices, demonstrating substantial equivalence in design, materials, and intended use. The "study" proving it meets acceptance criteria is primarily a series of mechanical tests and a predicate comparison.

    Therefore, I will interpret your request in the context of this document type, focusing on the mechanical testing performed and the comparison to predicate devices, rather than a clinical study evaluating AI performance.

    Here's the breakdown of the information you requested based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list "acceptance criteria" in a quantitative table with specific pass/fail values for each test. Instead, it states that a "mechanical evaluation" was conducted and that this evaluation "demonstrated that there are no new issues related to the safety or effectiveness of the subject devices." This implies that the results of these tests were considered acceptable when compared to established standards for such devices or to the performance of predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Meet safety and effectiveness standards comparable to predicate devices for mechanical properties relevant to an orthopedic implant.Mechanical evaluation demonstrated "no new issues related to the safety or effectiveness." The subject devices were found to be substantially equivalent to predicate devices based on these tests.

    Specific Mechanical Tests Conducted:

    • Fatigue Strength
    • Push-Out Strength
    • Torsional Shear
    • Tensile Attachment Strength

    The document states, "A review of testing has demonstrated that there are no new issues related to the safety or effectiveness of the subject devices." This indicates the device passed these implied criteria by performing adequately in these mechanical tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The "tests" mentioned are mechanical tests on device components, not typically "data sets" in the context of AI/software or clinical trials. Therefore, concepts like country of origin or retrospective/prospective do not apply. The sample sizes for the mechanical tests (e.g., how many cones were tested for fatigue) are not disclosed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. The "ground truth" for a mechanical test is typically determined by the physics and engineering principles governing the test and measurement. There isn't an "expert" establishing ground truth in the way a clinical expert would for a diagnostic image. The "experts" involved would be the engineers and scientists conducting and analyzing the mechanical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods like 2+1 are used in clinical studies, especially for ground truth establishment in ambiguous cases. This is not relevant for mechanical testing of an orthopedic implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study or AI assistance is not relevant to the Legion Cone System, which is a physical implant component, not an AI/software device. The document explicitly states: "Clinical data was not needed to support the safety and effectiveness of the subject devices."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This refers to AI algorithm performance studies, which are not relevant to this physical orthopedic implant.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the mechanical tests, the "ground truth" is established by the physical and engineering measurements against accepted standards and the performance of predicate devices. There is no biological "ground truth" like pathology or outcomes data from patients in this premarket notification for a mechanical device.

    8. The sample size for the training set

    This information is not applicable/not provided. There is no "training set" in the context of mechanical testing for a physical implant. This concept applies to machine learning models.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As there is no training set mentioned, the establishment of its ground truth is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081566

    Validate with FDA (Live)

    Date Cleared
    2008-08-21

    (78 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew Oxinium DH femoral heads are for single use only.

    Device Description

    The Oxinium DH femoral heads are designed for use with existing Smith & Nephew hip stems featuring a 12/14 taper and will articulate against existing acetabular shell and liner constructs. A complete list of hip stems and acetabular components intended for use with the Oxinium DH heads is provided in Tables 1 and 2, respectively. The overall design of the Oxinium DH femoral heads is based upon the existing Total Hip 12/14 Taper Femoral Heads cleared via K021673 and the Oxinium femoral heads cleared as part of K022958 for Total Hip Femoral Heads & Liners.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Smith & Nephew Oxinium DH Femoral Heads:

    The provided document (K081566) is a 510(k) Summary of Safety and Effectiveness for a medical device. It's important to understand that 510(k) clearances are based on demonstrating "substantial equivalence" to a predicate device, rather than proving stand-alone clinical efficacy through randomized controlled trials. Therefore, the information you're requesting regarding "acceptance criteria" and a "study that proves the device meets the acceptance criteria" in the context of performance metrics like sensitivity, specificity, human reader improvement with AI, etc., will likely not be present in this type of regulatory submission.

    Instead, the "acceptance criteria" for a 510(k) device like this are typically related to:

    • Mechanical performance: Ensuring the device meets strength, durability, and biocompatibility standards.
    • Indications for Use: Confirming the device is intended for the same or similar purposes as the predicate device.
    • Design and Materials: Demonstrating similarity in overall design and materials to a legally marketed predicate.

    The "study" proving these criteria are met is often a combination of mechanical testing and a comparison to predicate devices.

    Given this context, here's the information extracted from the provided text, addressing your questions where possible:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: As explained above, "acceptance criteria" for a 510(k) submission are generally not expressed in clinical performance metrics (like sensitivity/specificity) but rather in engineering and design equivalence.

    Acceptance Criteria CategoryReported Device Performance (from text)
    Mechanical Performance"A review of the mechanical testing results indicated that the Smith & Nephew Oxinium DH femoral heads are equivalent to devices currently used clinically and are capable of withstanding expected in vivo loading without failure."
    Indications for UseThe stated Indications for Use are identical to the predicate devices, covering rehabilitation of hips damaged by trauma, inflammatory/degenerative joint disease, avascular necrosis, fractures, etc.
    Design & Materials"The Smith & Nephew Oxinium DH femoral heads are similar in overall design, indications, and materials to the Total Hip 12/14 Taper Femoral Heads cleared via K021673 and the Oxinium femoral heads cleared as part of K022958 for Total Hip Femoral Heads & Liners."
    Biocompatibility(Not explicitly detailed in the provided text, but implied through substantial equivalence to predicate devices likely already deemed biocompatible.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of clinical performance evaluation (like for diagnostic AI). The "test set" here refers to the samples used in mechanical testing. The document states "A review of the mechanical testing results," implying tests were conducted, but does not specify the sample size (e.g., number of heads tested for fatigue, static strength, etc.).
    • Data Provenance: Not applicable in the context of clinical data. The "data" primarily comes from mechanical laboratory testing performed on the device prototypes or production samples. The country of origin for such testing is not specified. It is a prospective test in the sense that the device was manufactured and then tested to confirm its properties.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. For mechanical testing, "ground truth" is established by engineering standards, specifications, and physical measurements, not by expert consensus in a clinical sense. The review of these results would be done by internal engineers, potentially external consultants, and ultimately FDA reviewers.
    • Qualifications of Experts: Not specified. If "experts" refers to those who designed or evaluated the mechanical tests, they would typically be biomechanical engineers, materials scientists, or other relevant technical professionals.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing clinical ground truth (e.g., diagnosis from imaging). For mechanical testing, the "adjudication" is typically a direct comparison of measured values against established engineering specifications or performance of predicate devices. There isn't a "consensus" process for adjudicating if a material broke at a certain load.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic tools or AI systems where human readers interpret results. This submission is for a physical implantable device (femoral head) and focuses on mechanical and material equivalence.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No. This question is irrelevant to the device described. This is a physical orthopedic implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • Ground Truth Type:
      • For mechanical performance: Engineering specifications, material properties, and performance data from predicate devices. The "ground truth" is that the device must withstand "expected in vivo loading without failure," which is assessed via standardized biomechanical testing (e.g., fatigue testing, static load testing to ISO or ASTM standards, though specific standards aren't cited in this summary).
      • For design, materials, and indications: The characteristics and performance of the legally marketed predicate devices (K021673 and K022958).

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This document is not describing an AI/machine learning model, so there is no "training set" in that context. If interpreted very broadly as the historical data used to design the device, it would be the cumulative knowledge and clinical experience with previous hip prostheses, but specific numbers are not given.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as there is no "training set" in the common understanding for an AI/ML device. The "ground truth" for the design of the device is based on established biomechanical principles, historical clinical performance of similar devices, and regulatory standards for orthopedic implants.

    In summary: The K081566 document is a regulatory submission for substantial equivalence for a physical medical device. It relies on a comparison to existing, legally marketed predicate devices and demonstrates equivalence through mechanical testing and design/material similarity, rather than clinical performance studies (like those used for diagnostic AI or new drug approvals). The terms "acceptance criteria" and "study" in this context are interpreted differently than they would be for software-as-a-medical-device (SaMD) or AI medical devices.