Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K091980

    Validate with FDA (Live)

    Device Name
    ACL TOP 700 LAS
    Date Cleared
    2009-09-22

    (83 days)

    Product Code
    Regulation Number
    864.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K073377

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACL TOP 700 LAS is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.

    The system provides results for both direct hemostasis measurements and calculated parameters.

    Device Description

    The ACL TOP 700 LAS is being introduced as a new family member to the ACL TOP (K073377), with the added feature of an extra arm and hardware to interface with laboratory automation systems (LAS). A Point-in-Space design solution is utilized where the patient sample remains under the control of the laboratory automation system (i.e., the automation track) and the ACL TOP 700 LAS analyzer aspirates an aliquot for sample analysis without removing the primary container from the automation track. The instrument's system software was also modified to interface with an IM (interface module) computer, which controls the communications to the LAS system.

    AI/ML Overview

    The provided text describes the regulatory clearance for the ACL TOP 700 LAS coagulation instrument, asserting its substantial equivalence to a predicate device (ACL TOP K073377). However, the document (a 510(k) summary) focuses on the reason for submission and comparison to predicate device, stating that testing demonstrated substantial equivalence rather than providing specific acceptance criteria and detailed study results for the new device.

    Therefore, much of the requested information (acceptance criteria, specific performance metrics, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) is not present in the provided text, as this type of detailed study report is typically found in the full 510(k) submission, not the publicly available summary.

    Based on the provided text, here's what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not explicitly provided in the given 510(k) summary. The summary states: "Testing demonstrated that the performance of the ACL TOP 700 LAS is substantially equivalent to the performance of the current legally marketed ACL TOP family (K073377)." This implies that the acceptance criteria for the new device were implicitly met by demonstrating performance equivalent to the predicate, but the specific criteria (e.g., precision, accuracy ranges) and the numerical results are not detailed.

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not provided in the 510(k) summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not provided in the 510(k) summary. For a coagulation instrument, "ground truth" would typically relate to the accuracy and precision of its measurements compared to reference methods or established standards, rather than expert consensus on a diagnostic image.

    4. Adjudication Method for the Test Set:

    This information is not provided in the 510(k) summary.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    This information is not provided in the 510(k) summary. MRMC studies are typical for diagnostic imaging devices where human interpretation is involved. This device is a fully automated laboratory instrument; therefore, an MRMC study is unlikely to be relevant or performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The device is described as "fully automated," suggesting its performance would inherently be standalone (algorithm/instrument only). However, specific performance metrics of a standalone study are not detailed beyond the statement of substantial equivalence.

    7. The Type of Ground Truth Used:

    For a coagulation instrument, the "ground truth" would likely be established through reference methods, calibrated standards, and known-concentration samples. The 510(k) summary does not specify the exact type of ground truth used, but it implicitly relies on established laboratory practices for validating instrument performance.

    8. The Sample Size for the Training Set:

    This information is not provided in the 510(k) summary. For a coagulation instrument, "training set" might refer to data used for internal calibration or algorithm development, but it's not discussed in the context of clinical validation data in this summary.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided in the 510(k) summary.


    Summary of Available Information from the Provided Text:

    • Device Name: ACL TOP 700 LAS
    • Predicate Device: ACL TOP (K073377)
    • Nature of Study: A substantial equivalence demonstration (Special 510(k)) based on testing that confirmed the new device's performance is equivalent to the predicate.
    • Key Change: Addition of an extra arm and hardware for interface with Laboratory Automation Systems (LAS), allowing aspiration without removing the primary container from the automation track. System software was also modified for LAS interface.
    • Claim: The ACL TOP 700 LAS shares the same intended use/indications for use, analytical technology/operating principle, analytical specifications, labeled performance characteristics, analytical data reduction, software, test parameters, and uses the same consumables and racks as the predicate.

    To obtain the detailed acceptance criteria and study results, one would typically need to review the full 510(k) submission available through FDA, as the summary often omits these specifics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1