Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K090490

    Validate with FDA (Live)

    Device Name
    MIRAGE ECHO
    Manufacturer
    Date Cleared
    2009-05-06

    (70 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071808, K072940, K081321, K050359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Echo channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Echo is: to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.

    Device Description

    The Mirage Echo provides an interface such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face. Mirage Echo is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product. Mirage Echo is a prescription device supplied non-sterile.

    AI/ML Overview

    Here's an analysis of the Mirage Echo Traditional 510k, based on the provided text, focusing on acceptance criteria and supporting studies.

    Based on the provided information, the 510(k) submission for the Mirage™ Echo nasal mask describes a comparative study against predicate devices (Mirage Micro and Ultra Mirage II Mask) to demonstrate substantial equivalence, rather than a study with explicit, quantitative acceptance criteria for device performance as one might see for an AI algorithm.

    The core of the "study" is a set of comparisons and assertions of substantial equivalence, relying on bench testing and the existing clinical acceptance of similar technology.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a traditional 510(k) for a medical device (nasal mask), the "acceptance criteria" are primarily established implicitly by demonstrating substantial equivalence to predicate devices, focusing on safety, performance characteristics, and intended use. The performance is reported as "substantially equivalent" to the predicates.

    Acceptance Criteria (Implied from Substantial Equivalence to Predicates)Reported Device Performance (Mirage Echo)
    Safety: Materials deemed safe (ref: ISO 10993-1).All components are fabricated using materials deemed safe (ref: ISO 10993-1).
    Intended Use: Channels airflow noninvasively from a PAP device to adult patients (> 66 lb/30 kg) for single-patient home reuse and multipatient hospital/institutional reuse.Same intended use as predicates. Channels airflow noninvasively to a patient from a PAP device (CPAP/bilevel system), for adult patients (> 66 lb/30 kg), for single-patient home reuse and multipatient hospital/institutional reuse.
    Technological Characteristics:
    - Provides a seal via silicone interface.Provides a seal via silicone interface.
    - Offered in various sizes for adequate fit.Offered in various sizes to ensure adequate fit.
    - Incorporates vent holes for continuous air leak to flush CO2 and minimize rebreathing without interfering with performance.Incorporates vent holes to provide continuous air leak to flush out and minimize the amount of CO2 rebreathed. The design ensures vents do not interfere with intended performance.
    - Connects to conventional air delivery hose via standard conical connectors (ISO 5356-1:2004).Connects to a conventional air delivery hose via standard conical connectors (ref: ISO 5356-1:2004).
    - Constructed using molded plastic and silicone components.Constructed using molded plastic and silicone components.
    - Operates on the same ResMed flow generator settings.Designed to operate on the same ResMed flow generator settings.
    - Substantially equivalent pressure-flow characteristics and flow impedance.Pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent.
    - Can be reused in home and hospital/institution environment.Can be reused in the home and hospital/institution environment.
    - CO2 performance is substantially equivalent.The CO2 performance of the new device and the predicate device are substantially equivalent.
    Effectiveness: At least as safe and effective as the predicate devices and does not raise new questions of safety and effectiveness.Asserted to be at least as safe and effective and does not raise new questions of safety and effectiveness. "Use of vented nasal masks with CPAP or Bilevel therapy is Clinical Data proven technology and is well accepted by the medical community." "Bench testing is sufficient to demonstrate safety and efficacy of the Mirage ECHO, as was the case with the predicate device."

    2. Sample Size for the Test Set and Data Provenance

    The provided text does not mention a "test set" in the context of clinical data or patient-specific evaluation. The performance data refers to bench testing. Therefore, information regarding:

    • Sample size for the test set
    • Data provenance (e.g., country of origin, retrospective/prospective)
      is not applicable as no clinical test set data is described. The submission relies on "bench testing" and comparison of technical specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. Since no clinical test set data is described, there's no mention of experts establishing a ground truth for such a set. The "ground truth" for this type of submission is typically derived from established engineering principles, international standards (e.g., ISO), and the performance of legally marketed predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable. No clinical test set data is described that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, potentially with AI assistance. The Mirage Echo is a medical device (nasal mask) for therapy delivery, not a diagnostic tool requiring interpretation.

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study was not done, nor would it be relevant for this type of device. The Mirage Echo is a physical medical device, not an AI algorithm.

    7. Type of Ground Truth Used

    The "ground truth" in this context is established through bench testing results, adherence to international standards (e.g., ISO 10993-1 for biocompatibility, ISO 5356-1:2004 for connectors), engineering specifications, and the established performance and safety profile of legally marketed predicate devices. There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for this submission, as it focuses on demonstrating substantial equivalence in physical and functional characteristics.

    8. Sample Size for the Training Set

    This information is not applicable. The Mirage Echo is a physical medical device, not an AI algorithm that undergoes a "training set" process.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As it's not an AI algorithm, there is no "training set" or ground truth for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072940

    Validate with FDA (Live)

    Device Name
    MIRAGE MICRO
    Manufacturer
    Date Cleared
    2007-12-13

    (57 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K071808, K063122

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

    The Mirage Micro is:

    • to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed.
    • intended for single patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.
    Device Description

    The Mirage Micro provides seal such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.

    Mirage Micro is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product.

    Mirage Micro is a prescription device supplied nonsterile.

    AI/ML Overview

    The provided document is a 510(k) summary for the ResMed Mirage Micro Traditional, a vented nasal mask, seeking substantial equivalence to predicate devices. It is not a study that proves the device meets acceptance criteria in the way typically seen for AI/ML-based medical devices or diagnostics. Instead, it demonstrates substantial equivalence through comparisons of technological characteristics and performance data via bench testing.

    Therefore, many of the requested fields (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable to this type of submission.

    Here's a breakdown of the available information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state formal "acceptance criteria" with numerical thresholds in the sense of a diagnostic device. Instead, it relies on demonstrating substantial equivalence to predicate devices through comparative performance and technological characteristics. The "reported device performance" is essentially a statement of equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    CO2 performance substantially equivalent to predicate device."The CO2 performance of the new device and the predicate device are substantially equivalent."
    Pressure-flow characteristics and flow impedance substantially equivalent to predicate device."The pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent."
    Operation on the same standard flow generator setting as predicate device."Both the new device and the predicate are designed to operate on the same standard flow generator setting."
    Provision of seal via dual wall silicone interface."The new device and the predicate mask, provide seal via dual wall silicone interface."
    Incorporation of vent holes for continuous air leak."Both the masks incorporate vent holes to provide continuous air leak to flush out the dead space within the mask and minimize the amount of CO2 rebreathed by the patient."
    Connection to conventional air delivery hose via standard conical connectors."Both the masks connect to conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)."
    Provisions for connecting oxygen and pressure sensing tubing."Both the masks have provisions for connecting oxygen and pressure sensing tubing via luer ports."
    Construction using molded plastic components and fabric headgear with safe materials."Both the masks are constructed using molded plastic components and fabric headgear. All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1)."
    Intended use matching predicate devices."it has the same intended use;"
    Technological characteristics similar to predicate devices."it has similar technological characteristics to both predicates:"
    Does not raise new questions of safety and effectiveness."it does not raise new questions of safety and effectiveness:"
    At least as safe and effective as predicate devices."it is at least as safe and effective as the predicate devices Mirage Micro and Mirage Quattro."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for the test set: Not applicable. The submission refers to "bench testing" and "comparison with predicate" rather than a clinical trial or a test set of data points in the context of an AI/ML device.
    • Data provenance: Not applicable. Bench testing is typically performed in a laboratory setting. No country of origin for clinical data is mentioned as none was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not applicable. Ground truth establishment by experts is not described as part of this bench testing for a mechanical device.
    • Qualifications of experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication method: Not applicable. This concept is relevant for expert disagreement resolution in diagnostic studies, not for mechanical device bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: No. This is a medical device, not an AI/ML diagnostic.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: Not applicable. This is a medical device, not an AI/ML diagnostic.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of ground truth: Not explicitly stated as "ground truth" in the AI/ML sense. The performance of the new device was compared against the performance of the predicate devices (Mirage Micro K071808 and Mirage Quattro K063122) through bench testing parameters like CO2 performance, pressure-flow characteristics, and flow impedance. The "truth" is based on the established safety and efficacy of the predicate devices.

    8. The sample size for the training set

    • Sample size for training set: Not applicable. This is a medical device, not an AI/ML system.

    9. How the ground truth for the training set was established

    • Ground truth for training set establishment: Not applicable. This is a medical device, not an AI/ML system.

    Summary of the Study (as described in the 510(k) Summary):

    The submission for the Mirage Micro Traditional nasal mask relies on a bench testing study to demonstrate substantial equivalence to its predicate devices (Mirage Micro Mask K071808 and Mirage Quattro Mask K063122).

    The core of the "study" involved:

    • Comparison of Technological Characteristics: Detail how the new device's design, materials, and features are similar to the predicates. This includes aspects like the dual-wall silicone interface, vent hole design, standard connectors, luer ports, and biocompatible materials.
    • Performance Data Comparison (Bench Testing): Direct comparison of the new device's performance against the predicate Mirage Micro regarding:
      • CO2 performance: Stated as "substantially equivalent."
      • Pressure-flow characteristics: Stated as "substantially equivalent."
      • Flow impedance: Stated as "substantially equivalent."
      • Operation on standard flow generator settings: Stated as designed to operate on the same settings.

    The conclusion drawn from this comparison and bench testing is that the new Mirage Micro is substantially equivalent because it shares the same intended use, similar technological characteristics, does not raise new questions of safety and effectiveness, and is at least as safe and effective as the predicate devices. Clinical data was deemed unnecessary as "Use of vented nasal masks with CPAP or Bilevel therapy is proven technology and is well accepted by the medical community. Bench testing is sufficient to demonstrate safety and efficacy..."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1