Found 2 results

510(k) Data Aggregation

    K Number
    K123660

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-17

    (231 days)

    Product Code
    Regulation Number
    862.1825
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102432, K112725, K110619, K091849, K110586, K071480

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EUROIMMUN's 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Li-heparin). Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.

    Device Description

    EUROIMMUN 25-OH Vitamin D ELISA consists of a microwell ELISA plate coated with (sheep) monoclonal anti-25-OH vitamin D antibodies, 6 calibrators, 2 controls, Biotin, sample buffer, conjugate. wash buffer concentrate. enzyme TMB chromogen/substrate solution and stop solution.

    This ELISA test kit is designed for the in vitro determination of 25-OH vitamin D in human serum or plasma samples. In the first analysis step, the calibrators and patient samples are diluted with biotin-labelled 25-OH vitamin D and added to microplate wells coated with (sheep) monoclonal anti-25-OH vitamin D antibodies. During the incubation an unknown amount of 25-OH vitamin D in the patient sample and a known amount of biotinlabelled 25-OH vitamin D compete for the antibody binding sites in the microplate wells plate. Unbound 25-OH vitamin D is removed by washing. For the detection of bound biotin-labelled 25-OH vitamin D, a second incubation is performed using peroxidase-labelled streptavidin. In a third incubation using the peroxidase substrate tetramethylbenzidine (TMB) the bound peroxidase promotes a colour reaction. The colour intensity is inversely proportional to the 25-OH vitamin D concentration in the sample. Results for the samples can be calculated directly using a standard curve.

    Antibodies: sheep monoclonal antibodies which identify specifically 25-OH vitamin D3 and 25-OH vitamin D2.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    ReproducibilityCV's < 12% for samples > 10 ng/ml; CV's < 20% for concentrations < 10 ng/mlIntra-Assay CVs: (Data from table is unreadable; however, the text states "All results passed the acceptance criteria.") Inter-Assay CVs: (Data from table is unreadable; however, the text states "All results passed the acceptance criteria.") Lot-to-Lot CVs:
    Limit of Blank (LoB)N/A (determined for the device)0.4 ng/ml
    Limit of Detection (LoD)N/A (determined for the device)1.8 ng/ml
    Functional Sensitivity (FS)Lowest concentration at which the potential regression line crosses the 20% CV line4.0 ng/ml (Results below 4 ng/mL are reported as "< 4 ng/mL")
    Linearity/Reportable RangeAmount of nonlinearity above functional sensitivity of 4.0 ng/ml acceptable below 15%. Sufficiency linear from 4.0 to 120 ng/mlAlthough polynomial regression p-values were significant (< 0.05) for two of three sample sets, the amount of nonlinearity above 4.0 ng/ml was found acceptable below 15%. Conclusion: Assay is sufficiently linear from 4.0 to 120 ng/ml.
    InterferenceIndividual recovery within 90 – 110% when spiked with potential interfering substances.Hemoglobin (up to 750 mg/dl): 90 - 110% recovery. Triglycerides (up to 2000 mg/dl): 98 - 110% recovery. Bilirubin (up to 40 mg/dl): 90 - 109% recovery. Cholesterol (up to 400 mg/dl): 94 – 109% recovery. Biotin (up to 1000 mg/dl): 91 - 107% recovery. Ascorbic acid (up to 10.0 mg/ml): 94 - 110% recovery. Note: Significant interference (> 10% deviation) seen with 1000 mg/dl of hemoglobin.
    Method Comparison (vs. Predicate)High correlation to predicate devicen: 240 samples. Concentration Range (Predicate): 4.1 - 110.4 ng/ml. Concentration Range (Candidate): 4.1 - 119.1 ng/ml. Regression Equation (y = Candidate, x = Predicate): y = 0.78 + 1.08 x. 95% C.I. of Intercept: -0.06 - 1.63. 95% C.I. of Slope: 1.06 - 1.11. Correlation Coefficient R: 0.9858. 95% C.I. of R: 0.9817 - 0.9890.
    Matrix Comparison (Plasma vs. Serum)Regression equation near ideal correlation (intercept 0; slope 1.0); Coefficients of determination > 0.99; %recovery compared to serum in range of 87 to 127 %EDTA plasma (n=38): Y = 0.29 + 0.99x (95% CI Intercept: -0.37 - 1.18, 95% CI Slope: 0.93 - 1.02); R² = 0.996; Mean %recovery = 100% (Range: 87 - 116%). Li-heparin plasma (n=38): Y = 0.55 + 0.97x (95% CI Intercept: -0.65 - 1.33, 95% CI Slope: 0.93 - 1.04); R² = 0.993; Mean %recovery = 101% (Range: 87 - 127%).
    StabilityPredicted shelf-life of at least 12 months at 2-8°C for control materials and calibrators; Open-vial stability of 3 months at 2-8°C for kit.Predicted shelf-life of at least 12 months at 2-8°C for control materials and calibrators (based on accelerated testing at 37°C, real-time ongoing). Open-vial stability of 3 months when stored at 2-8°C. Biotin stable for at least 2 weeks when diluted to working strength.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Reproducibility:
      • Intra- and Inter-Assay: Minimum of 8 natural serum samples (2 spiked) for CV determination; 40 determinations for intra-assay CVs; 40 determinations performed in 10 different runs on 5 different days (with 4 replicates per run) for inter-assay CVs.
      • Lot-to-Lot: Minimum of 8 natural serum samples (2 spiked); 32 determinations performed in 8 different runs on 4 different lots (with 2 runs per lot and 4 replicates per run).
      • Data Provenance: "natural serum samples collected freshly in-house from obvious healthy blood donors".
    • LoD: 200 determinations from 5 samples in the low range (2-10 ng/ml), measured in 5 independent runs with 8 replicates per run.
    • Linearity/Assay Reportable Range: Sets of 11 sample preparations (natural negative sample + high positive samples), tested in double determinations.
    • Cross-Reactivity: 25-OH Vitamin D free sample aliquoted and spiked with 7 potential cross-reacting Vitamin D metabolites.
    • Interference: Sera at different 25-OH Vitamin D concentrations spiked with 6 potential interfering substances.
    • Method Comparison w/Predicate Device:
      • Sample Size: 240 samples in total.
      • Data Provenance:
        • 141 prospective samples for 25-OH-Vitamin D testing from a clinical laboratory.
        • 28 samples from 25-OH-Vitamin D quality assessment programs (harvested from blood donated by venesection undergoing therapeutic for patients with hemochromatosis or polycythemia).
        • 5 samples from a 25-OH-Vitamin D quality control panel (untreated routine patient material).
        • 30 samples from normal blood donors.
        • 36 samples (15%) were spiked with 25-OH Vitamin D3 stock solution to ensure concentration distribution.
        • All patients gave informed consent. Samples are natural and not treated.
    • Matrix Comparison: 38 sample pairs of serum and corresponding plasma (EDTA, Li-heparin) from donors. 3 sample pairs were spiked.
    • Expected Values/Reference Range: 206 samples from healthy subjects (70 men, 136 women; average age 64 years; age range: 22–99 years) from a US commercial source (normal US blood donors from the US Midwest region, drawn early Oct 2010).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their qualifications for establishing ground truth for the test set in terms of clinical interpretation. However, the performance studies (e.g., reproducibility, linearity) rely on in-house personnel and laboratory procedures, with calibrator and control value assignments confirmed by "2 different technicians" and "a different person for final release." Method comparison was against an "FDA-cleared reference assay," implying the predicate device serves as a reference for comparison.

    4. Adjudication Method for the Test Set

    No specific adjudication method (e.g., 2+1, 3+1) is mentioned or implied for the evaluation of the test set results for clinical interpretation. The performance studies describe analytical methods and statistical comparisons.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) assay, not an imaging device typically evaluated with MRMC studies. The "Method Comparison" section compares the device's analytical results against a predicate device, not human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are for the standalone performance of the EUROIMMUN 25-OH Vitamin D ELISA assay, which is an algorithm-only (test kit) device. Its performance characteristics like linearity, reproducibility, limits of detection, and interference are evaluated intrinsically, and its results are then compared to a predicate device. There is no human-in-the-loop component for the device's operation, only human involvement in performing the laboratory tests (technicians) and interpreting the output for clinical use.

    7. The Type of Ground Truth Used

    The ground truth used varies based on the type of study:

    • Reproducibility, LoB, LoD, Functional Sensitivity, Linearity, Cross-Reactivity, Interference: The ground truth is effectively the known concentrations of the analytes in the prepared samples (e.g., spiked samples, calibrators, controls, or precisely characterized samples). These are established through gravimetric methods, UV spectrophotometric analysis, and comparison with NIST and DEQAS standards, as well as predicate device measurements and HPLC.
    • Method Comparison w/Predicate Device: The ground truth for comparison is the results obtained from the FDA-cleared predicate device, which is considered the reference method for this type of comparison.
    • Matrix Comparison: The ground truth is the serum sample concentrations when comparing to plasma.
    • Expected Values/Reference Range: The ground truth for establishing population reference intervals is the 25-OH Vitamin D levels measured in healthy subjects using the new device itself.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning, as this is an in-vitro diagnostic ELISA kit. The term "training set" is not applicable here. The assay is based on chemical reactions and optical density measurements, not a learned algorithm in the typical sense.

    9. How the Ground Truth for the Training Set was Established

    As noted in #8, there isn't a "training set" in the machine learning context. The calibration and standardization of the assay (which could be considered analogous to "establishing ground truth" for the assay's internal function) involves:

    • Gravimetric calibration using UV-Vis verified stock standards.
    • Comparison with NIST standards and DEQAS quality assessment data.
    • In-house quality control sera.
    • Initial values for quality control sera assigned using the predicate device in conjunction with liquid chromatography (HPLC).
    • Final calibrator values verified and assigned by adjusting initial values to meet specified ranges when tested against predicate assays.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130321

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-04-09

    (60 days)

    Product Code
    Regulation Number
    862.1045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    LIAISON® 25 OH Vitamin D Control K071480, LIAISON® 25 OH Vitamin D Calibration Verifier K090104

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Aldosterone assay uses chemiluminescent immunoassay (CLIA) technology and is intended for the quantitative determination of Aldosterone in human serum. EDTA plasma and treated urine samples. Aldosterone measurements are intended for use in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states and other conditions of electrolyte imbalance. The test has to be performed on the LIAISON® Analyzer.

    The DiaSorin LIAISON® Aldosterone Control Set is intended for use as assayed quality control samples to monitor the accuracy of the DiaSorin LIAISON® Aldosterone assay on the LIAISON® Analyzer.

    The DiaSorin LIAISON® Aldosterone Calibration Verifiers are assayed quality control materials intended for the quantitative verification of calibration and reportable range of the LIAISON® Aldosterone assay when performed on the LIAISON® Analyzer.

    Device Description

    The LIAISON® Aldosterone assay is a competitive modified 2 step chemiluminescent assay that uses sheep monoclonal antibody for capture of the Aldosterone molecule. Results are determined by a 2 point calibration conversion of the master curve to a working curve. The light signal is measured by a photomultiplier as relative light units (RLU) and is inversely proportional to the concentration of aldosterone present in the calibrators, controls or samples.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the LIAISON® Aldosterone assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific pass/fail thresholds for the LIAISON® Aldosterone assay. However, it provides performance data that would implicitly be compared against desired clinical or analytical requirements. Below is a table summarizing the reported performance, which serves as evidence of the device meeting its intended analytical characteristics.

    Performance MetricReported Device Performance (LIAISON® Aldosterone)Clinical Context / Implied Acceptance
    Method Comparison (vs. Predicate RIA)High correlation and agreement with established method
    - Serum Samples (n=144)Slope: 0.98 (95% CI: 0.94 to 1.02)Slope near 1.0, indicating proportional agreement.
    Intercept: 1.10 ng/dL (95% CI: 0.43 - 1.49)Intercept near 0, indicating minimal constant bias.
    R: 0.988High correlation (close to 1).
    - Urine Samples (n=104)Slope: 0.98 (95% CI: 0.91 to 1.05)Slope near 1.0, indicating proportional agreement.
    Intercept: 34 ng/dL (95% CI: 11.43 to 56.7)Intercept near 0 (though 34 ng/dL for urine is noted).
    R: 0.948High correlation (close to 1).
    Limit of Blank (LoB)Serum: 0.97 ng/dLAbility to distinguish analyte absence from presence.
    Urine: 1.26 ng/dL
    Limit of Detection (LoD)Serum: 1.45 ng/dLLowest concentration detectable with reasonable certainty.
    Urine: 2.00 ng/dL
    Limit of Quantitation (LoQ)Serum: 3.0 ng/dLLowest concentration quantifiable with acceptable precision and accuracy.
    Urine: 2.80 ng/dL
    Precision (20-Day Reproducibility)Consistent and reproducible results across runs, days, sites, and reagent lots.
    - Serum (Aldo-S1 to Aldo-S6, n=480 each)Total %CV: 5.6% - 10.5%Low Coefficient of Variation (CV) indicating high precision.
    - Urine (Aldo-U1 to Aldo-U3, n=480 each)Total %CV: 8.6% - 9.8%Low Coefficient of Variation (CV) indicating high precision.
    Dilution LinearitySerum: Observed = 0.994(Expected) + 0.71, R = 1.000Linear response across dilution range. (Slope near 1, intercept near 0, R near 1)
    EDTA Plasma: Observed = 1.01(Expected) + 1.43, R = 0.998
    Urine: Observed = 0.996(Expected) + 0.69, R = 0.999
    Interfering SubstancesNo interference observed at specified concentrations for various substances.Robustness against common physiological and therapeutic interferences.
    Cross-Reactivity< 0.02% cross-reactivity for all tested substances.High specificity, minimizing false positives from similar compounds.

    2. Sample Sized Used for the Test Set and the Data Provenance

    • Method Comparison Test Set:

      • Serum: 155 samples (144 included in analysis as 11 were below measuring range).
      • Urine: 106 samples (104 included in analysis as 2 were above measuring range).
      • Data Provenance: Samples were "collected from apparently healthy individuals." A portion (approximately 15%) were spiked with aldosterone to span the measuring range. The study was conducted using the LIAISON® Aldosterone assay and the predicate RIA assay. Although not explicitly stated, the context of a 510(k) submission strongly implies these are likely prospective or retrospective clinical samples (for the healthy individuals) used in a controlled laboratory setting, potentially from the region where DiaSorin Inc. operates (Stillwater, MN, USA), though this is not explicitly stated.
    • LoB/LoD/LoQ Test Set: Not specified, but generally refers to a set of blank and low-concentration samples. The determination was performed according to CLSI EP17-A2.

    • Reference Range/Expected Values Test Set:

      • Serum and EDTA Plasma: 126 apparently healthy subjects.
      • Urine (24 hour): 91 individuals.
      • Data Provenance: Samples collected from apparently healthy subjects aged 21-65 years with normal blood pressure and normal fasting glucose levels. These are prospectively collected clinical samples following specific patient preparation protocols (fasting, upright/supine positions, 24-hour collection). No specific country of origin is mentioned.
    • Reproducibility/Precision Test Set:

      • Panel: 6 frozen serum samples and 3 frozen urine samples.
      • Controls: 2 levels of controls.
      • Data Provenance: A "coded panel" prepared by DiaSorin Inc. Tested at "DiaSorin Inc. and 2 external sites." This indicates laboratory-prepared samples tested in a multi-site, controlled study.
    • Dilution Linearity Test Set: Not specified, but samples of "each sample type" (serum, EDTA plasma, urine) were diluted. This implies laboratory-prepared or clinical samples serially diluted.

    • Interfering Substances & Cross-Reactivity Test Set: Not specified, but involved samples spiked with various substances at defined concentrations. These are laboratory-prepared samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of immunoassay (measuring a specific analyte concentration) does not typically involve human experts to establish "ground truth" in the same way an imaging or diagnostic AI does. The "ground truth" for this device's performance studies is established by:

    • Predicate Device: For method comparison, the Siemens Coat-A-Count® Aldosterone RIA assay (K831178) served as the reference standard. This is a previously cleared and established clinical assay, representing the accepted "truth" for aldosteronemeasurements at the time.
    • Known Concentrations: For studies like LoB/LoD/LoQ, Dilution Linearity, Interfering Substances, and Cross-Reactivity, the "ground truth" is based on known, prepared concentrations of the analyte and potential interfering/cross-reacting substances. These are analytically derived standards.
    • Statistical Analysis: CLSI guidelines (EP9-A2, EP17-A2, EP6-A) provide the statistical and methodological frameworks for establishing these analytical performance characteristics.

    Therefore, no information about human experts (e.g., radiologists) with specific qualifications or experience in establishing ground truth for this analytical device is provided or expected.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned or applicable. As a quantitative immunoassay, the output is a numerical concentration. The comparison is between the new device's numerical result and the predicate device's numerical result (or known spike concentrations), not subjective interpretations requiring adjudicated consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was performed or is applicable to this type of in vitro diagnostic device (immunoassay). This device directly measures an analyte concentration, and there is no "human reader" involved in interpreting its output in the context of an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented (Method Comparison, LoB/LoD/LoQ, Precision, Dilution Linearity, Interfering Substances, Cross-Reactivity) all represent standalone performance of the LIAISON® Aldosterone assay. The device provides a quantitative measurement directly, without human interpretation for the primary output. Its performance is evaluated purely on its analytical accuracy, precision, and specificity against established methods or known standards.

    7. The Type of Ground Truth Used

    The ground truth used in these studies consists of:

    • Predicate Device Measurements: For method comparison, the results from the Siemens Coat-A-Count® Aldosterone RIA assay served as the comparative ground truth.
    • Known/Expected Concentrations: For LoB/LoD/LoQ determinations, dilution linearity, interference, and cross-reactivity studies, the ground truth was based on analytically derived and known concentrations of aldosterone or other substances.
    • Biological Referencing: For the reference range studies, the "ground truth" was established by measuring aldosterone levels in clinically defined healthy populations under controlled conditions.

    8. The Sample Size for the Training Set

    No information about a "training set" is provided. This type of immunoassay (CLIA technology) is a chemical and biological measurement system, not a machine learning or AI algorithm that typically has a distinct training phase requiring a training dataset. The device's operational parameters (e.g., calibration, reagent concentrations, reaction times) are established through analytical development and optimization, rather than machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of a machine learning algorithm, this question is not applicable. The device's functionality is based on its underlying chemical and immunoassay principles, which are developed and validated through a series of analytical studies as described in the performance data section, rather than through a machine learning training process.