    Device Name
    pixium 4343RCE
    Date Cleared
    2017-02-09

    (28 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K141736, K153318, K063781

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As a part of a radiographic system, the Pixium 4343RCE is intended to acquire digital radiographic images. The Pixium 4343RCE is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    The pixium 4343RCE is a Stationary x-ray system that converts x-ray patterns into electrical signals. The signals are converted into visible images for use in medical diagnosis. In the device, a cesium iodide scintillator absorbs the input x-ray photons. The scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

    The detector is permanently installed and intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct x-ray imaging. It is electrically powered by and connected with the x-ray system. The device is connected to the Philips Eleva Workspot to create a complete x-ray imaging chain, and it is intended to be used in Philips x-ray systems such as the DigitalDiagnost.

    AI/ML Overview

    The provided text describes the Philips pixium 4343RCE, a stationary x-ray system. The document is a 510(k) summary for premarket notification to the FDA, asserting substantial equivalence to a predicate device (Philips pixium 4343RC, K131483).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present specific numerical acceptance criteria (e.g., minimum DQE, MTF values that must be met) for the pixium 4343RCE, nor does it provide a direct side-by-side comparison of claimed performance data against such defined criteria. Instead, the acceptance criteria are implicitly stated as demonstrating similar performance to the predicate device and compliance with various standards. The "Discussion" column in the comparison table serves as the "reported device performance" against the implicitly accepted criterion of not negatively impacting safety or effectiveness compared to the predicate.

    Here's a table summarizing relevant performance metrics and the discussion provided, which indicates how the device meets (or is considered to meet) implied acceptance criteria of non-inferiority or slight improvement:

    FeaturePredicate Device pixium 4343RC (K131483)Proposed Device pixium 4343RCE (K170113)Discussion / Performance vs. Implied Acceptance Criteria
    Design Features
    X-Ray AbsorberCsI ScintillatorCsI ScintillatorIdentical; no impact to safety and effectiveness.
    Installation typeStationary, permanently installedStationary, permanently installedIdentical; no impact to safety and effectiveness.
    Readout MechanismThin Film TransistorThin Film TransistorIdentical; no impact to safety and effectiveness.
    Detector Size500 x 490 x 45.5 mm³500 x 490 x 45 mm³Similar, slightly smaller; no impact to safety or effectiveness.
    Detector Weight<14 kg11.7kg ± 0.85kgSimilar, slightly lighter; no impact to safety or effectiveness.
    Image Size (Pixel)2840 x 28742840 x 2874Identical; no impact to safety and effectiveness.
    Image Size (X-ray field)420 x 425 mm²420 x 425 mm²Identical; no impact to safety and effectiveness.
    Distance Image to Rim34.9 mm40mmSimilar, slightly greater; no impact to safety or effectiveness.
    Pixel Size148µm148µmIdentical; no impact to safety and effectiveness.
    Nyquist Frequency3.37 lp/mm3.37 lp/mmIdentical; no impact to safety and effectiveness.
    ADC Digitization16 bit16 bitIdentical; no impact to safety and effectiveness.
    SENR (@ 1 µGy)42.3 dB42.3 dBIdentical; no impact to safety and effectiveness.
    Max X-ray Dose for Linear Response50 µGy50 µGyIdentical; no impact to safety and effectiveness.
    Image Readout Duration1.6 s1.36 sSlightly faster; no impact to safety and effectiveness.
    Number of Modes23Additional default mode included; no impact safety and effectiveness.
    Exposure Window Durations1-8192 ms0-8191 msMinimal decrease; insignificant, no impact to safety or effectiveness.
    Sequence Time Detector6 s6 sIdentical; no impact to safety and effectiveness.
    Image Data16.3 MBytes16.3 MBytesIdentical; no impact to safety and effectiveness.
    Use w and w/o Radiographic Grid?YesYesIdentical; no impact to safety and effectiveness.
    Maximum Lifetime Dose100 Gy100 GyIdentical; no impact to safety and effectiveness.
    Warm-up Duration before Calibration4 h2 hDecreased; no impact to safety or effectiveness.
    Data Interface to Workstation100 Mbit/s Ethernet1 Gbit/s EthernetIncreased; no impact to safety or effectiveness.
    Power Consumption20.4WAvg: 8W. Max: 15WDecreased; no impact to safety or effectiveness.
    Cover Factor (Optical Fill Factor)60.3%63%Essentially the same; no impact to safety or effectiveness.
    Image Quality
    Modulation Transfer Function (MTF)1 lp/mm 64%, 2 lp/mm 32%, 3 lp/mm 17%, Nyquist 13% (3.37 lp/mm)1 lp/mm 62%, 2 lp/mm 35%, 3 lp/mm 19%, Nyquist 15% (3.37 lp/mm)Similar, with some improvements and one minimal decrease; no impact to safety and effectiveness.
    Detective Quantum Efficiency (DQE) at 2 µGy0.05 lp/mm 65%, 1 lp/mm 52%, 2 lp/mm 42%, 3 lp/mm 25%, 3.37 lp/mm 18%0.05 lp/mm 67%, 1 lp/mm 51%, 2 lp/mm 42%, 3 lp/mm 27%, 3.37 lp/mm 18%Similar, with one improvement and one minimal decrease; no impact to safety and effectiveness.

    The study that proves the device meets the acceptance criteria is detailed as "Non-clinical verification and validation tests." These tests cover the intended use, technical claims, requirement specifications, and risk management results. The document explicitly states: "Non-clinical verification and validation test results demonstrate that the pixium 4343RCE... Meets the acceptance criteria and is adequate for its intended use." The basis for this is "compliance with the aforementioned international and FDA-recognized consensus standards and device specific guidance document" and showing "identical or equivalent technological characteristics," "equivalent diagnostic capability," and no "new potential hazards or safety risks" compared to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a separate "test set" in the context of data used for a clinical performance evaluation, as it explicitly states that no clinical study was required. The non-clinical verification and validation tests are performed on the device itself (hardware and software) and against established technical specifications and standards (e.g., ISO, IEC, NEMA, FDA guidance). Therefore, there is no mention of a "sample size" of patient data or "data provenance" (country of origin, retrospective/prospective) for a clinical test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical study was conducted, there was no "test set" of patient data requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable as no clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a digital x-ray detector, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device itself (the x-ray detector) as a standalone component without human interpretation being part of the primary evaluation. The "Summary of Non-clinical Performance Data" and the "Comparison" tables effectively serve as the standalone performance assessment. The device's performance metrics (MTF, DQE, SENR, image readout speed, etc.) are evaluated directly for the detector. The conclusion is that "the device is as safe, as effective, and performs as well or better than the predicate device" in terms of its technological characteristics and ability to produce images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance evaluation, the "ground truth" is based on technical specifications, international and FDA-recognized consensus standards, and device-specific guidance documents. For instance, the MTF and DQE values are compared against those of the predicate device, with the implicit ground truth being that comparable or improved values demonstrate equivalent performance.

    8. The sample size for the training set

    Not applicable. The document describes an x-ray detector, not an AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable as there is no training set mentioned for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131483

    Validate with FDA (Live)

    Date Cleared
    2013-10-07

    (137 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K123005, K06371, K931071, K063781

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As a part of a radiographic system, the Pixium 4343RC is intended to acquire digital radiographic images. The Pixium 4343RC is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    The Pixium 4343RC is a Solid State X-ray Imaging Device that converts x-ray patterns into electrical signals are converted into visible images for use in medical diagnosis. In the device, a cesium iodide scintillator absorbs the input x-ray photons. The scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

    The detector is permanently installed and intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct x-ray imaging. It is electrically powered and connected with the x-ray system. The device is connected to the Philips XD-S Eleva Radiography Workstation to create a complete x-ray imaging chain, and it is intended to be used exclusively in the Philips Digital Diagnost x-ray system, one of the predicate devices.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Philips Pixium 4343RC, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission for the Philips Pixium 4343RC primarily focuses on demonstrating substantial equivalence to predicate devices through technical characteristic comparisons and a clinical concurrence study. It does not explicitly list "acceptance criteria" in a quantitative performance metric sense (e.g., target accuracy, sensitivity/specificity thresholds for diagnostic tasks). Instead, the acceptance criteria are implicitly that the new device performs at least as well as, or is equivalent to, the predicate devices across various technical and clinical aspects.

    The table below summarizes the key technical characteristics and performances compared between the new device and its predicates, which serve as the implicit "acceptance criteria" for demonstrating equivalence.

    Acceptance Criteria (Predicate Device Performance)Reported Device Performance (Philips Pixium 4343RC)Discussion/Outcome (as per document)
    Indications for Use:
    Acquire digital radiographic images for routine radiography, including specialist areas (intensive care, trauma, pediatric work), excluding fluoroscopy, angiography, and mammography. (Aligned with FD-W17)Acquire digital radiographic images for routine radiography, including specialist areas (intensive care, trauma, pediatric work), excluding fluoroscopy, angiography, and mammography.Identical / Equivalent
    X-Ray Absorber: CsI Scintillator (Both Predicates)CsI ScintillatorIdentical
    Installation type: Stationary, permanently installed (Pixium 4600)Stationary, permanently installedIdentical
    Readout Mechanism: Switching Diode (Pixium 4600) / Thin Film Transistor (FD-W17)Thin Film TransistorEquivalent
    MTF (Modulation Transfer Function):
    1 lp/mm: 62% (Pixium 4600), 60% (FD-W17)1 lp/mm: 64%Equivalent
    2 lp/mm: 33% (Pixium 4600), 30% (FD-W17)2 lp/mm: 32%Equivalent
    3 lp/mm: 17% (Pixium 4600), 15% (FD-W17)3 lp/mm: 17%Equivalent
    3.4 lp/mm: 13% (Both Predicates)3.4 lp/mm: 13%Equivalent
    3.5 lp/mm (Nyquist): 12% (Both Predicates)3.5 lp/mm (Nyquist): 12%Equivalent
    DQE (Detective Quantum Efficiency) at 1 μGy:
    1 lp/mm: 50% (Pixium 4600), 51% (FD-W17 at 2.5 µGy)1 lp/mm: 51% (at 1 µGy)Equivalent
    2 lp/mm: 40% (Pixium 4600), 39% (FD-W17 at 2.5 µGy)2 lp/mm: 42% (at 1 µGy)Equivalent
    3 lp/mm: 23% (Pixium 4600), 22% (FD-W17 at 2.5 µGy)3 lp/mm: 25% (at 1 µGy)Equivalent
    3.4 lp/mm: 17% (Pixium 4600), 15% (FD-W17 at 2.5 µGy)3.4 lp/mm: 18% (at 1 µGy)Equivalent
    3.5 lp/mm (Nyquist): 15% (Pixium 4600), 13% (FD-W17 at 2.5 µGy)3.5 lp/mm: Not explicitly stated, but implies high similarity to 3.4 lp/mm.Equivalent
    Image Processing: Philips Thoravision (Pixium 4600) / XD-S Eleva Workstation (FD-W17)XD-S Eleva WorkstationIdentical

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions a "concurrence study" was conducted but does not specify the sample size (number of images or cases) used for the test set.

    The data provenance is not explicitly stated as country of origin, nor is it explicitly labeled as retrospective or prospective. Given it's a "concurrence study" for regulatory submission, it is typically a controlled study, which could be either.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth in the concurrence study. It only states that the study confirmed the device "provides images of equivalent diagnostic capability to the predicate radiographic system." This implies expert assessment, but details are missing.

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method used in the concurrence study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention or describe an MRMC comparative effectiveness study evaluating how much human readers improve with AI vs. without AI assistance. The study described is a "concurrence study" focused on the diagnostic capability of the new device's images being equivalent to a predicate, not on human-AI collaboration.

    6. Standalone (Algorithm Only) Performance Study

    The device itself is a Solid State X-ray Imaging Device (a digital detector), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance study, often associated with AI software, is not applicable in this context. The entire device, including its hardware and integrated image processing software (XD-S Eleva Workstation), is being evaluated as a system that produces diagnostic images.

    7. Type of Ground Truth Used

    The document implies that the ground truth for the concurrence study was based on the diagnostic capability of images from the predicate device. The study's goal was to show that the new device "provides images of equivalent diagnostic capability to the predicate radiographic system." This means the predicate device's images (presumably deemed diagnostically acceptable by current clinical practice) served as the reference for equivalence. It does not explicitly mention pathology or outcomes data as direct ground truth for the study itself, but rather clinical equivalence.

    8. Sample Size for the Training Set

    The document does not mention a training set sample size. This is understandable as the Pixium 4343RC is a hardware device (X-ray imager) with associated image processing software. While the image processing software itself might have been "trained" or optimized during its development, this submission focuses on the performance of the integrated hardware and software system, comparing it to predicate devices, not detailing the training of specific algorithms within the image processor.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set and associated ground truth are explicitly mentioned for this device's submission, this information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112132

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-18

    (115 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K1100403, K063781

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist. As part of a radiographic system, the KrystalRad 660 is intended to acquire digital radiographic images. It is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    This device is simply the combination of two cleared devices, the same Wireless Portable Detector as used in the FD-W I (K90062) marketed by Philips Medical Systems and the image processing software cleared in our K080522 DDR MAK Series. Alternately the image processing software is the same as in Omil'ision, K. 100403 made by Modern Module Inc. The Wireless Portable Detector consists of three main parts: Portable in radiography detector (x-ray sensitive part); Docking station which is directly connected to the radiographic workshow and a backup cable which can connect the decector to the docking station if the wireless connection cable which can connect the detector to the docking if the wireless connection cannot be used. Detector size: 35 x 43 cm (14 x 17") Image matrix size: 3000 pixels x 2400 pixels. Pixel size 144 um Image matrix size: 3000 pixels x 2400 pixels. Pixel size: 144 µm, 1mage resolution up to 3.5 LP/mm. The device is intended as an upgrade to existing film x-ray systems. It should be installed by a qualifics trained by a qualifics trained

    AI/ML Overview

    The KrystalRad 660 Digital Radiographic Portable Retrofit System's acceptance criteria and study details are provided below based on the 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state acceptance criteria in quantitative terms (e.g., minimum sensitivity, fidelity metrics). Instead, it focuses on demonstrating substantial equivalence to predicate devices. The primary performance metric is the qualitative comparison of clinical images and integration testing.

    Acceptance Criteria (Implied)Reported Device Performance
    Clinical images are not significantly different from predicate.Clinical images acquired were compared to predicate images; no significant differences.
    Integration testing successfully completed.Integration testing was performed.
    Device is as safe and effective as predicate devices.The new device is as safe and effective as the predicate device.
    Conforms to US Performance Standards.The modified device conforms to US Performance Standards.
    Hardware is UL Listed to US Standards for medical device safety.The hardware is UL Listed to US Standards for safety for medical devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "Clinical images were acquired and compared to our predicate images." However, it does not specify the sample size for the test set (number of images or patients).

    The data provenance is not explicitly stated regarding its country of origin or whether it was retrospective or prospective. It only mentions "clinical images were acquired," which could imply prospective acquisition, but this is not confirmed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information on the number of experts used or their qualifications to establish ground truth for the clinical images. The comparison is described simply as "compared to our predicate images," suggesting an internal comparison rather than a formal expert review process for ground truth.

    4. Adjudication Method for the Test Set

    The document does not describe any formal adjudication method for the test set. The statement "no significant differences between them" suggests a qualitative assessment without detailing a specific process like 2+1 or 3+1 consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported in this submission. The study focuses on demonstrating equivalence to predicate devices, not on comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is a hardware combination (digital X-ray panel and image processing software). The "performance" being evaluated is the image quality and functionality of this combined system itself. Therefore, the "standalone" performance in this context refers to the system acquiring images, which was done by collecting and comparing clinical images. It's not an "algorithm-only" performance in the sense of an AI diagnostic aid.

    7. The Type of Ground Truth Used

    The ground truth for the comparison was based on "our predicate images." This implies comparing images from the KrystalRad 660 system to images produced by the previously cleared predicate devices (Philips Wireless Portable Detector FD-W17 K090625 and the image processing software from K080582 DDR MAK Series or K100403 OmniVision). The "ground truth" is therefore the established image quality and clinical utility of the predicate devices.

    8. The Sample Size for the Training Set

    The device is a combination of two already cleared predicate devices. The "study" described is a comparison and integration test of this combination, not the development or training of a new algorithm. Therefore, there is no specific "training set" mentioned for the KrystalRad 660 as it's not a de novo AI algorithm requiring training. The image processing software itself (cleared under K080582 or K110040) would have had its own validation/training, but details are not provided here for KrystalRad 660.

    9. How the Ground Truth for the Training Set was Established

    As stated in point 8, there isn't a "training set" for the KrystalRad 660 as a new device. The ground truth for the predicate image processing software or detector would have been established during their original clearance, likely involving various methods (e.g., expert review, physical phantoms for image quality metrics), but these details are not provided in this submission for K112132. The KrystalRad 660 leverages the already established "ground truth" (i.e., safety and effectiveness) of its predicate components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090625

    Validate with FDA (Live)

    Date Cleared
    2009-03-24

    (15 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K051523,K063781

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Detector is combined with a Philips XD-S workstation (K063781) which consists of a computer, keyboard, display, mouse. The complete X-ray system would further include other Philips subsystems and components, like patient table, X-ray control(s), X-ray high voltage generator, X-ray tube(s), collimator(s), accessories, etc. The XD-S workstation and the complete X-ray systems are not changed other than by replacing an Xray receptor with the Wireless Portable Detector FD-W17.

    AI/ML Overview

    The provided text describes a 510(k) submission for the "Wireless Portable Detector FD-W17." However, it does not contain any information about acceptance criteria, device performance metrics, sample sizes for test sets or training sets, expert qualifications, adjudication methods, multi-reader multi-case studies, or standalone performance studies.

    The submission focuses on:

    • Device Identification: Manufacturer, submitter, product names, classification.
    • Compliance with Standards: Federal X-Ray performance standards (CFR 1020.30, .31), electrical safety (UL 60601-1, IEC 60601-1), radiation protection (IEC 60601-1-3), Electromagnetic Compatibility (IEC-60601-1-2), and risk management (ISO 14971).
    • System Description: How the detector integrates into a larger X-ray system.
    • Intended Use: Acquiring digital radiographic images for routine, intensive care, trauma, and pediatric exams, excluding fluoroscopy, angiography, and mammography.
    • Equivalence Information: Demonstrating substantial equivalence to a predicate device (Pixium 4600) and other existing Philips components for the workstation and pre-processing.
    • Safety Information: Mature technology, compliance with standards, wireless transmission evaluation, and risk management.
    • Conclusion: Stating substantial equivalence based on similar indications for use, technological characteristics, and no new hazards.
    • FDA Clearance Letter: Confirming substantial equivalence based on the provided indications for use.

    Therefore, I cannot fulfill your request for information regarding acceptance criteria and performance studies based on the provided text. The document is a regulatory submission for substantial equivalence, not a detailed performance study report.