510(k) Data Aggregation

    K Number
    K190176

    Validate with FDA (Live)

    Device Name
    MINAMO
    Date Cleared
    2019-08-01

    (181 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K063695, K021228

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA).

    The PCI Guide Wires are not intended for use in the neurovasculature.

    Device Description

    The MINAMO consists of a hybrid NiTi and stainless-steel core wire and a coil assembly on the distal end of the device. The coil assembly is soldered to the NiTi portion of the core. In addition, coatings are applied on the surface of the MINAMO. The proximal portion is coated with PTFE. The distal section is coated with hydrophilic coating, with the distal tip coated with silicone. The MINAMO is available in both 190cm and 300cm lengths and with different tip shapes.

    AI/ML Overview

    The MINAMO is a coronary guide wire. The 510(k) summary provides details about non-clinical testing performed to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "The in vitro bench tests demonstrated that the MINAMO met all acceptance criteria and performed similarly to the predicate devices." Specific numerical acceptance criteria values and reported device performance values are generally not detailed in 510(k) summaries for non-clinical tests. This is a common practice, as the focus is on demonstrating that the device meets established performance thresholds and is comparable to previously cleared devices.

    Test / AssessmentAcceptance Criteria (General Description)Reported Device Performance (General Description)
    Dimensional verification / AppearanceMet specified dimensions and visual criteria.Met all acceptance criteria.
    Tensile StrengthMet specified strength requirements, comparable to predicate.Met all acceptance criteria.
    Torque StrengthMet specified torque performance requirements, comparable to predicate.Met all acceptance criteria.
    TorqueabilityDemonstrated appropriate torque transmission, comparable to predicate.Met all acceptance criteria.
    Tip FlexibilityMet specified flexibility characteristics, comparable to predicate.Met all acceptance criteria.
    Coating AdhesionMet specified adhesion strength requirements, comparable to predicate.Met all acceptance criteria.
    Catheter CompatibilityDemonstrated compatibility with standard catheters.Met all acceptance criteria.
    LubricityMet specified lubricity requirements, comparable to predicate.Met all acceptance criteria.
    Corrosion ResistanceDemonstrated resistance to corrosion under specified conditions.Met all acceptance criteria.
    Kink ResistanceMet specified kink resistance requirements, comparable to predicate.Met all acceptance criteria.
    RadiodetectabilityDemonstrated adequate visibility under fluoroscopy.Met all acceptance criteria.
    Coating Integrity / Acute Particulate CharacterizationMaintained coating integrity and minimal particulate release.Met all acceptance criteria.
    Packaging / Transportation assessmentMaintained device integrity through packaging and simulated transport.Met all acceptance criteria.
    Biocompatibility Tests (Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic Toxicity, USP Rabbit Pyrogen, Material Mediated, Hemolysis, Partial Thromboplastin Time, In Vivo Thromboresistance, Sc5b-9 Complement Activation)Met international standards for biocompatibility (e.g., ISO 10993).Met all acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for each individual non-clinical test. It generally refers to "non-clinical laboratory testing." For medical devices, these tests are typically performed in a laboratory setting, and the provenance is the company's internal R&D or a contract testing lab. These are all retrospective tests conducted on manufactured device samples. There is no information regarding country of origin for the data for non-clinical bench testing, but the applicant's headquarters are in Japan.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not applicable to the non-clinical bench testing described. These tests do not involve human interpretation or the establishment of ground truth by clinical experts in the same way an AI diagnostic device would. Performance is measured against physical and chemical specifications.

    4. Adjudication Method:

    This information is not applicable. Adjudication methods are typically used in clinical studies or studies involving human readers, which is not the nature of the reported performance data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC study was not done. The device is a physical medical instrument (guide wire), not an AI diagnostic tool.

    6. Standalone Performance Study:

    A standalone performance study was not done in the context of an AI algorithm. The reported tests are for the physical properties and performance of the guide wire itself. The document describes "non-clinical laboratory testing" which inherently is a standalone evaluation of the device against predefined specifications.

    7. Type of Ground Truth Used:

    The "ground truth" for the non-clinical tests is established by:

    • Engineering Specifications: Predetermined physical and chemical properties that an equivalent guide wire must possess (e.g., specific tensile strength, flexibility parameters, corrosion resistance).
    • International Standards: Adherence to relevant ISO standards (e.g., for biocompatibility, sterility) or equivalent national standards.
    • Predicate Device Performance: Demonstrating comparable performance to legally marketed predicate devices through comparative bench testing.

    8. Sample Size for the Training Set:

    This information is not applicable. The MINAMO is a physical medical device, not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set for a physical guide wire.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161702

    Validate with FDA (Live)

    Device Name
    Endura guidewire
    Date Cleared
    2016-12-06

    (169 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K063695, K052339

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endura guidewire is intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and/or peripheral vasculature.

    Device Description

    The Endura guidewire is a straight tip guidewire with a maximum outer diameter of 0.014" available in 190 cm and 300 cm lengths. The Endura guidewire consists of a bi-metal stainless steel and nitinol core wire covered on the distal end by a radiopaque coil at the distal end of the Endura guidewire has a hydrophilic coating (HPC) and the proximal end is coated with polytetrafluoroethylene (PTFE). The 190 cm length model is compatible with a guidewire extension.

    AI/ML Overview

    The Endura Guidewire is a medical device intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and/or peripheral vasculature. The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Tensile StrengthNot explicitly stated, but implicitly passes performance testingPassed
    Torque StrengthNot explicitly stated, but implicitly passes performance testingPassed
    Torque Response (Torqueability)Not explicitly stated, but implicitly passes performance testingPassed
    Guidewire Support ProfileNot explicitly stated, but implicitly passes performance testingPassed
    Tip Load (Tip Flexibility)Not explicitly stated, but implicitly passes performance testingPassed
    Fracture ResistanceNot explicitly stated, but implicitly passes performance testingPassed
    Flex ResistanceNot explicitly stated, but implicitly passes performance testingPassed
    Hydrophilic Coating Particulate (Coating Adherence/Integrity)Not explicitly stated, but implicitly passes performance testingPassed
    PTFE Coating Adhesion (Coating Adherence/Integrity)Not explicitly stated, but implicitly passes performance testingPassed
    Corrosion ResistanceNot explicitly stated, but implicitly passes performance testingPassed
    Track ForceNot explicitly stated, but implicitly passes performance testingPassed
    Friction Force (HPC and PTFE)Not explicitly stated, but implicitly passes performance testingPassed
    Fluoroscopy VisualizationNot explicitly stated, but implicitly passes performance testingPassed
    Length ExtensionNot explicitly stated, but implicitly passes performance testingPassed
    Cytotoxicity (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    Sensitization (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    Irritation/Intracutaneous Reactivity (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    Systemic Toxicity (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    Pyrogenicity (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    Hemocompatibility (Biocompatibility)In accordance with ISO 10993-1, implicitly passes acceptable bio-responsePassed
    SterilitySterility Assurance Level (SAL) 10-6 via Ethylene Oxide (EO) sterilizationAchieved SAL 10-6

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes used for each individual performance test or biocompatibility test. It mentions that "device samples" passed the biocompatibility tests and that the design was "verified through the following tests." The data provenance can be inferred as originating from Vascular Solutions, Inc., USA, the manufacturer of the device, through prospective testing conducted to demonstrate the device's performance and safety.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This section is not applicable as the Endura Guidewire is a physical medical device. The "ground truth" for its performance and safety is established through laboratory and bench testing, as well as biocompatibility assessments, rather than expert interpretation of data or images.

    4. Adjudication Method for the Test Set:

    This section is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation or subjective assessments, which is not the case for the pre-market testing of this device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This section is not applicable. An MRMC study is relevant for AI algorithms or diagnostic tools where human reader performance is being evaluated with and without AI assistance. The Endura Guidewire is a physical interventional device, not a diagnostic or AI-powered system.

    6. Standalone (Algorithm Only) Performance:

    This section is not applicable. The Endura Guidewire is a physical medical device, not an algorithm, so the concept of "standalone algorithm performance" does not apply.

    7. Type of Ground Truth Used:

    The ground truth for the Endura Guidewire's performance is based on:

    • Engineering specifications and test methodologies: Each performance test (e.g., tensile strength, torque response, flexibility, coating adherence) has predefined engineering benchmarks and standards that the device must meet.
    • Biocompatibility standards: Compliance with ISO 10993-1 guidelines for biological evaluation of medical devices.
    • Sterilization standards: Achieving a Sterility Assurance Level (SAL) of 10-6.

    8. Sample Size for the Training Set:

    This section is not applicable. The Endura Guidewire is a physical medical device and does not involve AI or machine learning, therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This section is not applicable as there is no training set for this device.