Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101545

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-10-28

    (147 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K051614, K053088, K061176, K062432, K050705, K071370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good primary stability is established and appropriate occlusal loading is applied.

    Device Description

    The Genesis Implant System includes implants, and associated surgical, restorative and dental laboratory components. Genesis implants are surgically inserted into the upper and/or lower jawbone and serve as a replacement tooth root, which provides a stable foundation for restorations. Genesis implants are manufactured from Grade 4 titanium, have a tapered or straight cylindrical design with an internal indexing connection, and are available in various platform diameters and lengths. The implants have a macro-, micro- and nano-topography and are treated with the BioSpark process, which results in a hydrophilic surface enriched with calcium and phosphorous ions. The implant collar is micro-roughened and treated with the AnaTite" process, which results in a pink color for enhanced esthetics. The majority of Genesis abutments are manufactured from Grade 5 titanium and are treated with the AnaTite process. Other Genesis abutments are made from Grade 5 titanium/plastic or gold alloy/plastic. Abutments intended for fixed restorations utilize a Grade 5 titanium screw for attachment to the implant. Genesis abutments and associated restorative components are manufactured in a variety of sizes and configurations to be compatible with the implant platforms.

    AI/ML Overview

    The provided text is a 510(k) summary for the Genesis Implant System, a dental device. It focuses on demonstrating substantial equivalence to predicate devices, primarily through non-clinical testing. This type of submission generally doesn't include the same kind of detailed acceptance criteria and clinical study results that an AI/software as a medical device (SaMD) would. Therefore, much of the requested information regarding AI performance metrics, sample sizes for test/training sets, expert adjudication, MRMC studies, or standalone algorithm performance is not applicable or cannot be extracted from this document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in a quantitative table as would be expected for a SaMD that relies on specific performance metrics like sensitivity or specificity. Instead, the "acceptance criteria" are implicitly met by demonstrating that the device is substantially equivalent to legally marketed predicate devices through a combination of material properties, design characteristics, and non-clinical testing confirming those characteristics are maintained or improved.

    The device performance is reported in qualitative terms, demonstrating properties consistent with or superior to predicate devices.

    Feature / StandardAcceptance Criteria (Implicit)Reported Device Performance
    MaterialsBiocompatible materials used in predicate devices (Grade 4 & 5 titanium, gold alloy, plastic)Genesis implants are manufactured from Grade 4 titanium. Genesis abutments from Grade 5 titanium, Grade 5 titanium/plastic, or gold alloy/plastic. All are biocompatible materials used in predicate devices.
    DesignSimilar design (threaded, rootform implants, internal indexing connection, various sizes/configurations)Genesis implants have a tapered or straight cylindrical design with an internal indexing connection, available in various platform diameters and lengths. Genesis abutments are in a variety of sizes and configurations compatible with implant platforms. Explicitly states "Genesis implants are similar in design and materials to the predicates in that all are threaded, rootform implants." and "The Genesis system has the same implant/abutment TiLobe™ internal connection as the PrimaConnex system."
    Mechanical Properties (Fatigue Testing)Consistent with ISO 14801 and FDA Special Controls Guidance for root-form endosseous dental implants and abutments.Genesis implants with angled abutments were fatigue tested in accordance with ISO 14801 and FDA's Special Controls Guidance. (No specific numerical results are provided in this summary, but the implicit acceptance is that they passed).
    Insertion Torque / Primary StabilityOptimized insertion torque to limit pressure on bone while ensuring good primary stability.In a bench study using simulated bone material, moderate insertion torque was shown when Genesis implants were placed utilizing appropriate surgical instrumentation.
    Surface Treatment (AnaTite - Coloration)Produces a pink color resembling natural gingival tissue for enhanced esthetics.An in vivo study has shown that pink coloration (from AnaTite treatment) most closely resembles natural gingival tissue. The color is produced by anodization, a common process. This aligns with or improves upon aesthetic characteristics.
    Surface Treatment (BioSpark - Osseointegration)Promote osseointegration, hydrophilic surface, enriched with calcium and phosphorus ions, nano-rough topography.Genesis implants have a macro-, micro-, and nano-topography and are treated with the BioSpark process, resulting in a hydrophilic surface enriched with calcium and phosphorous ions. Contact angle testing confirmed hydrophilicity. Energy dispersive spectroscopy and sputter mass spectroscopy confirmed presence of calcium and phosphorus ions. Scanning electron microscopy demonstrated surface roughness at macro-, micro-, and nano-levels. Mineralization potential reported in published literature.
    BioSpark Biological Efficacy (Osseointegration Speed)Facilitate faster osseointegration.Demonstrated by significant bone-to-implant-contact (BIC) in the cortical bone of sheep at 4 weeks compared to machined or blasted titanium. In vitro studies showed: significant increase in cellular metabolic activity (1-7 days), significant increase in osteoblast proliferation (1-3 days), and early fibronectin absorption and enhanced adhesion, proliferation, and differentiation of osteoblasts (24, 48, 72 hours) on BioSpark treated cp-Ti disks compared to controls. Note: Results in animal studies are not necessarily predictive of human clinical results.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes non-clinical testing (bench studies, in vitro studies, animal studies). There is no "test set" in the context of human data or AI model evaluation.

    • Fatigue Testing: Conducted in accordance with ISO 14801 and FDA Special Controls Guidance. The number of implants/abutments tested is not specified.
    • Insertion Torque Bench Study: Used "simulated bone material." The sample size is not specified.
    • AnaTite In Vivo Study (Coloration): "An in vivo study has shown..." No details on sample size or species.
    • BioSpark Animal Study (Osseointegration): "significant bone-to-implant-contact (BIC) in the cortical bone of sheep at 4 weeks." Sample size not specified.
    • BioSpark In Vitro Studies: These involved "cp-Ti disks" (commercially pure titanium disks). Sample sizes for these experiments are not specified.

    Data Provenance:

    • Bench studies (fatigue, insertion torque)
    • In vitro studies (cellular metabolic activity, osteoblast proliferation, fibronectin absorption, adhesion, proliferation, differentiation)
    • Animal studies (sheep for BIC, unspecified animal for AnaTite coloration)
    • Published literature (for mineralization potential)

    All these are non-clinical, controlled experiments. No human data (retrospective or prospective) is mentioned for the performance evaluation of the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. This is not an AI/SaMD submission evaluating human data against expert ground truth. The "ground truth" for the non-clinical tests is based on objective measurements and established scientific methods (e.g., ISO standards, microscopy, spectroscopy, biological assays, bone-to-implant contact measurements).

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable, as there is no human-annotated test set requiring adjudication in this context.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/SaMD. There are no "human readers" or "AI assistance" relevant to an MRMC study described in this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical dental implant system, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance evaluations relies on:

    • Physical measurements and material science: e.g., chemical composition (energy dispersive spectroscopy, sputter mass spectroscopy), surface topography (scanning electron microscopy), wettability (contact angle testing), mechanical strength (fatigue testing per ISO 14801).
    • Biological assays in vitro: measuring cellular metabolic activity, osteoblast proliferation, fibronectin absorption, adhesion, and differentiation.
    • Histomorphometry in animal models: bone-to-implant contact (BIC) in sheep.
    • Visual assessment (in vivo): comparing pink coloration to natural gingival tissue.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/SaMD. There is no concept of a "training set" for physical device testing. The "learning" for the device design would come from long-standing scientific and engineering principles in dental implantology, feedback from predicate devices, and internal R&D.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set in the AI sense. The design and material choices are informed by existing scientific knowledge, performance of predicate devices, and internal research and development validated by the non-clinical tests described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1