510(k) Data Aggregation

    K Number
    K072932

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-04-25

    (192 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052908,K980851,K050577,K051337,K042188

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LipiScan Coronary Imaging System is intended for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography. The System is intended for the detection of lipid-core-containing plaques of interest. The System is intended for the assessment of coronary artery lipid core burden.

    Device Description

    The LipiScan Coronary Imaging System is comprised of the catheter, catheter accessories, pull-back and rotation device and laser console with accessories.

    AI/ML Overview

    The provided text describes a 510(k) summary for the LipiScan Coronary Imaging System. However, it does not contain specific acceptance criteria or details of a study that directly proves the device meets such criteria. The document focuses on establishing substantial equivalence to a predicate device, highlighting similar intended use, operating principles, catheter design, shelf life, packaging, and compliance with general safety and performance standards.

    Here's an analysis based on the information provided and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not explicitly provided in the document. The 510(k) summary states, "Further preclinical testing has shown that the product can function as intended and meets all internal design specifications," but it does not list these specifications or acceptance criteria. There are no quantitative performance metrics (e.g., sensitivity, specificity, accuracy for detecting lipid-core-containing plaques) presented as acceptance criteria or reported device performance.

    2. Sample Size Used for the Test Set and Data Provenance

    Not explicitly provided for any specific performance study. The document mentions "Ex vivo and in vivo data is presented to support expanded indications for use," but no details about the sample size, type of test set, or data provenance (e.g., country of origin, retrospective/prospective) are given for these studies.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not explicitly provided. Since no specific performance study with a test set is detailed, information about experts, their number, or their qualifications for establishing ground truth is absent.

    4. Adjudication Method for the Test Set

    Not explicitly provided. As no specific performance study is detailed, there is no information about any adjudication methods (e.g., 2+1, 3+1).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not mentioned and likely not performed or reported in this summary. The document focuses on demonstrating substantial equivalence, not on comparative effectiveness with or without AI assistance. The device is for imaging, and there's no mention of AI assistance for human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable/not described. The LipiScan Coronary Imaging System is presented as an imaging system that provides an image of the artery wall. It's a diagnostic tool intended for use by clinicians, not an autonomous AI algorithm. Therefore, a "standalone algorithm only" performance study is not relevant in the context described.

    7. Type of Ground Truth Used

    Not explicitly stated for specific performance evaluations. The summary does not detail the nature of the ground truth for any studies supporting "expanded indications for use." For such devices, ground truth might involve histology, pathology, or correlative imaging modalities, but this is not specified here.

    8. Sample Size for the Training Set

    Not applicable/not provided. The device is an imaging system, not explicitly an AI/machine learning algorithm that requires a training set in the conventional sense. The "preclinical testing" and "ex vivo and in vivo data" likely refer to validation and verification testing for safety and performance, not training data for a learning model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable/not provided. As explained in point 8, there's no mention of a training set or an AI algorithm that would necessitate establishing ground truth for training.

    In summary, the provided 510(k) summary serves to demonstrate substantial equivalence to predicate devices and adherence to general safety and performance standards. It mentions "preclinical testing" and "ex vivo and in vivo data" to support its expanded indications, but it lacks the detailed information about specific acceptance criteria and the comprehensive study methodology asked for in the prompt. This level of detail is often found in the full 510(k) submission, not typically in the public summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071554

    Validate with FDA (Live)

    Date Cleared
    2008-01-04

    (211 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Volcano s5/s5i Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.

    ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.

    VH IVUS is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.

    The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.

    Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature. As an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. The pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.

    Device Description

    When operating in IVUS mode, the system console gathers and displays high-resolution intraluminal images that can be analyzed both qualitatively and quantitatively. In addition to supplying diagnostic information, the Volcano s5/s5i system is an adjunct to interventional therapies, such as balloon angioplasty. With ChromoFlo® (not available with Revolution catheter), a two-dimensional color map of relative blood velocities is overlaid on the grayscale image, providing additional information for vessel analysis. The In-Line Digital feature displays a two-dimensional, longitudinal view of the vessel. The angle of the longitudinal cut can be varied around the full 360 degrees.

    When operating in pressure mode, the system acquires intraluminal data from the pressure guidewire (SmartWire II/ComboWire) while simultaneously taking aortic pressure data from the established ECG/EKG catheterization lab equipment. In conjunction with the procedure, the system measures pressure and calculates pressure differences between the aoritic pressure and the SmartWire/ComboWire pressure transducer typically located distal to the vascular lesion and calculates the fractional flow reserve (FFR).

    AI/ML Overview

    The provided text is a 510(k) Summary for the Volcano s5/s5i Series Intravascular Imaging and Pressure System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting new clinical study data to establish performance against acceptance criteria.

    Therefore, the document does not contain the detailed information requested regarding specific acceptance criteria, study designs, sample sizes for test and training sets, expert qualifications, or adjudication methods. The "Performance Data" section explicitly states that "A risk analysis was conducted... Applicable testing was performed as required by the Quality System to evaluate the modifications... The test results were found to be acceptable as required by the respective test plans and protocols." This indicates that verification and validation testing was performed internally according to established quality system procedures, but the specifics of those tests and their results against quantitative acceptance criteria are not included in this summary document.

    Here's a breakdown of what can be extracted and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Missing. The document mentions that "test results were found to be acceptable as required by the respective test plans and protocols," but it does not provide the specific acceptance criteria or the reported device performance measurements for any of its features (IVUS, ChromaFlo, VH IVUS, Pressure, Rotational 45MHz).

    2. Sample size used for the test set and the data provenance:

    • Missing. The document does not describe any specific clinical or bench test sets with sample sizes or data provenance (e.g., country of origin, retrospective/prospective). The performance claims are based on substantial equivalence to predicate devices and internal verification/validation testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Missing. No information on experts or ground truth establishment for a test set is provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Missing. No information on adjudication methods is provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Missing. This device is an imaging and pressure system, not an AI-assisted diagnostic tool in the sense of a software algorithm interpreting images. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Missing. This device is a hardware system with integrated software; the concept of a "standalone algorithm" performance as typically understood for AI/CAD devices doesn't apply directly here. The system's functionalities (IVUS, pressure measurement, etc.) rely on the device hardware and software working together to produce data for human interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Missing. As no specific performance study against a ground truth is described, this information is not available.

    8. The sample size for the training set:

    • Missing. No information about a training set is provided. The device likely uses algorithms that are well-established for intravascular ultrasound and pressure measurements, rather than machine learning models that require a distinct "training set" in the modern sense.

    9. How the ground truth for the training set was established:

    • Missing. As no training set is mentioned, the method for establishing its ground truth is also not provided.

    What the document does state regarding performance evaluation:

    • Performance Data: "A risk analysis was conducted according to 803475-001 Risk Management which was written to comply with ISO 14971 and IEC 60601-1-4 as specific risk management standards. Also taken into consideration in this procedure are 21 CFR 820.30 and the Medical Device Directive of the European Union (93/46/EEC). Applicable testing was performed as required by the Quality System to evaluate the modifications to the Volcano s5/s5i Intravascular Imaging and Pressure System. The test results were found to be acceptable as required by the respective test plans and protocols."
    • Conclusion: "The discussion and data presented in this 510(k) and conformance with Design Controls and the Quality System Regulations establishes the Volcano s5/s5i Series Intravascular Imaging and Pressure System to be substantially equivalent for its intended use to the predicate devices listed in this submission."

    In summary, this 510(k) focuses on demonstrating substantial equivalence based on the device's technological characteristics and intended use being similar to predicate devices, and internal validation testing compliant with quality system regulations. It does not provide the specific detailed performance study information typically found in submissions for novel diagnostic algorithms or AI-driven devices.