Found 3 results

510(k) Data Aggregation

    K Number
    K240651

    Validate with FDA (Live)

    Device Name
    MRI Universal
    Date Cleared
    2024-07-12

    (127 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Screw Washers - K980364, Stryker Universal SMARTLock Hybrid MMF System - K122313, Stryker MMF Screw K050535

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Universal CMF System Intended use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. MP Lefort I Plates Intended Use The Stryker Leibinger Universal CMF and Universal 2.0 Mini Plating Systems are Craniomaxillofacial (CMF) plate and screw systems intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Indications for Use The MP LeFort I Plates as part of the predicate subgroup Universal 2.0 Mini Plating System are implants indicated for osteotomy, stabilization, and rigid fixation of LeFort I fractures of the maxillofacial skeleton. Stryker Upper-Face AXS Screws And Mid-Face AXS Screws Indications for Use The Stryker Universal CMF System is a Cranio-maxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Stryker MP, Mandible, HMMF And MMF AXS Screws -Universal CMF System: Intended Use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for Use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. -Stryker Universal SMARTLock Hybrid MMF System: Intended Use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for Use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. -Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Stryker Universal Orbital Floor System Intended Use The Stryker Universal Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit. Indications for Use The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older. Stryker Universal Smartlock Hybrid MMF System Intended use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion.

    Device Description

    This Bundled. Traditional 510(k) is for a single change the MR status that affects multiple devices and can be assessed during one review. There are no changes to the predicate devices themselves, just the addition to allow MR Conditional use of the implants in scope after MRI compatibility testing was completed. Associated labeling for each Predicate Device is also updated to reflect the MR conditional use of the implants.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K231599

    Validate with FDA (Live)

    Date Cleared
    2023-08-24

    (84 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K221855, K050535, K172572

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    -Stryker Universal SMARTLock Hybrid MMF System: The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted.
    -Universal CMF System: Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction.
    -Stryker MMF screw: The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion.

    Device Description

    The Subject Device screws are screw versions modified with the Stryker AXS Screw technology, where the abbreviation "AXS" stands for axial stability. This submission is adding the AXS Screw technology to the Subject Device. The Subject Device screws have modified geometry and features which will be implemented as AXS Screws in each of the existing Predicate/Reference Device systems. This includes screws with a length range of 4-20 mm and a diameter range from 2.0-2.7 mm. The materials (titanium and stainless steel) are the same as for the corresponding Predicate/Reference Devices. There are no other changes to the Predicate/Reference Device systems, only the implementation of the AXS Screw technology to each existing product line. The Subject Device screws will have the same clinical application, treat the same craniomaxillofacial anatomical locations, and are for the same patient population as the corresponding Predicate/Reference Devices.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Stryker MP, Mandible, HMMF and MMF AXS Screws, which are intraosseous fixation screws. This submission is for a medical device (screws), not an AI/software as a medical device (SaMD). Therefore, the questions related to AI device performance, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance are not applicable to this document.

    The clearance is based on demonstrating substantial equivalence to predicate devices, primarily through engineering performance bench testing.

    Here's the relevant information that can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document states that "The Subject Device met all pre-defined acceptance criteria, and the results of the V&V tests support the substantial equivalence of the Subject Device to the Predicate/Reference Devices." However, the specific numerical acceptance criteria for each test are not detailed in this summary document. The document lists the types of performance bench tests conducted:

    Test NameReported Device Performance (as stated in document summary)
    Pull-out Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Full Insertion and Removal Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Shaft Shear Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Head Shear TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Locking Cantilever Bending TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Locking Torque to Failure TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Self-Retention TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the summary document. The document refers to "Verification and Validation (V&V) testing" and "test reports" being included within the full submission, but the summary itself does not detail specific sample sizes for these bench tests or the provenance of any data beyond the general statement that testing was performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not Applicable. This is a medical device (screws) and does not involve AI or software requiring expert review for ground truth establishment. The "acceptance criteria" are engineering performance metrics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. This is a medical device (screws) and does not involve subjective human interpretation or adjudication for its performance evaluation in the way an AI diagnostic tool would. Performance is based on objective engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is a medical device (screws) and does not involve AI or human "readers."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This is a medical device (screws) and does not involve an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For mechanical performance tests, the "ground truth" or reference standard would be the physical properties and behavior of the devices as measured by standardized testing methods (e.g., ASTM F543) and compared against pre-defined engineering specifications and the performance of the predicate device.

    8. The sample size for the training set

    Not Applicable. This is a medical device (screws) and does not involve a training set as would be used for AI/machine learning.

    9. How the ground truth for the training set was established

    Not Applicable. This is a medical device (screws) and does not involve a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230733

    Validate with FDA (Live)

    Date Cleared
    2023-08-05

    (142 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K050535

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental
    The self-tapping screws and low profile emergency screws are intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction, and can be used in pediatric patients older than 29 days and up to two (2) years of age (infants). The self-tapping screws and low profile emergency screws are designed to be compatible with the components (plates, meshes) of the Delta Resorbable Fixation System K213777.

    Device Description

    The subject devices, the Delta System self-tapping screw (STS) and low profile emergency screw (LPES), are intended for use in the fixation of bones of the maxillofacial and midfacial skeleton, affected by trauma or for reconstruction. The subject device can be used in pediatric patients older than 29 days and up to two (2) years of age (infants), but is not intended for use in the mandible and/or full load bearing procedures.

    The scope of this submission covers the addition of the subject device screws to the previously cleared Delta System. Through this submission there is no change to the existing articles within the Delta System. The subject device screws are designed to be compatible with the existing components of the Delta System (plates, meshes), which have been previously cleared through K213777. Compared to the original Delta System, the subject STS and LPES have a limited patient population of pediatric patients older than 29 days and up to two (2) years of age (infants).

    AI/ML Overview

    The document provided is a 510(k) summary for a medical device called the "Stryker Resorbable Fixation System". This summary outlines the device, its intended use, and comparative information to a predicate device to demonstrate substantial equivalence to a legally marketed device.

    However, the provided text does not contain information about an AI/ML device or a study to prove that a device meets acceptance criteria related to AI/ML performance. The document focuses on regulatory approval for an orthopaedic fixation system, specifically resorbable screws, based on traditional engineering performance testing (e.g., mechanical tests, material properties) and comparison to a predicate device.

    Therefore, I cannot provide the requested information regarding AI/ML acceptance criteria, study details, sample sizes, expert involvement, or MRMC studies, as these concepts are not applicable to the content of the provided document.

    The document discusses "acceptance criteria" in the context of:

    • Regulatory Acceptance: The FDA's determination of substantial equivalence.
    • Performance Testing: Bench test "acceptance criteria" are implied by the "Passed" results for tests like Insertion, Shear, and Pull-out, and "Passed" for Inherent Viscosity. These are mechanical and material property tests, not AI/ML performance metrics.
    • User Validation Testing: A validation study was done to confirm usability, comprehension of Instructions for Use, and that user needs are met, concluding a "passing result." This relates to human factors and usability, not AI/ML performance.

    In summary, the provided text does not contain the information needed to answer the prompt's specific questions about AI/ML device acceptance criteria and proving device performance related to AI/ML.