510(k) Data Aggregation

    K Number
    K251316

    Validate with FDA (Live)

    Date Cleared
    2025-09-11

    (135 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-arms into a volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.

    The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.

    Device Description

    The Mazor X system combines robotic trajectory guidance with navigated surgical instruments (either guided or free hand navigation) to enable the surgeon to precisely position surgical instruments and/or implants according to predefined planning. With the imaging capabilities of the system, the user can also visualize the implants on the patients CT. Same as the predicate device the modified Mazor X consists of a workstation with dedicated software, the surgical system, navigation camera, accessories, instruments and disposable kits. The modified Mazor X, the subject of this 510(k) application, introduces software and hardware modifications to the Mazor X System cleared in 510(k) K230064.

    AI/ML Overview

    This 510(k) clearance letter describes a device, the Mazor X System / Mazor X Stealth Edition, and its substantial equivalence to a predicate device. It includes detailed information about new AI-enabled features: "2D Automatic Measurements" and "Plan Assist." The following analysis focuses on the acceptance criteria and study details for these AI components as presented in the provided text.


    Acceptance Criteria and Reported Device Performance

    Device: Mazor X System / Mazor X Stealth Edition (Software enhancements for 2D Automatic Measurements and Plan Assist)

    FeatureAcceptance CriteriaReported Device Performance
    2D Automatic MeasurementsICC (Intraclass Correlation Coefficient): At least 0.75 with a 95% confidence level for all spinopelvic measurements. F-test: Resulting p-value below 0.05, statistically demonstrating ICC is larger than 0.75. Bland-Altman analysis: 95% limits of agreement in range of a defined clinical range for all spinopelvic measurements.Achieved for all spinopelvic measurements. Achieved for all spinopelvic measurements. Achieved for all spinopelvic measurements.
    Plan Assist (Clinical Acceptability of Screw Placement)True Positive Rate (TPR) exceeding 80% with 95% confidence for AI-generated screw clinical acceptability.Exceeded 80% with 95% confidence.

    Detailed Study Information for AI-enabled Features

    1. 2D Automatic Measurements (AI-enabled Feature)

    • Sample Size for Test Set:
      • Clinical Evaluation: 146 AP images, 253 LAT images (for all spinopelvic parameters). Each specific spinopelvic parameter had its own sample size ranging from 24 to 126 images.
    • Data Provenance for Test Set: Not explicitly stated (e.g., country of origin). The note about "U.S. certified spine surgeons and radiologists" for reference standard implies some U.S. involvement, but doesn't specify data origin. It is explicitly stated that these datasets were independent from the development dataset, with data source site-level separation to ensure data independence.
    • Number of Experts for Ground Truth (Test Set): Three U.S. board-certified radiologists.
    • Qualifications of Experts (Test Set): U.S. board-certified radiologists. (No years of experience provided).
    • Adjudication Method (Test Set): Not explicitly stated, but implies individual evaluation by the three radiologists and comparison to AI measurements.
    • MRMC Comparative Effectiveness Study: No, a standalone performance of the AI algorithm was evaluated against expert annotations.
    • Standalone Performance: Yes, the accuracy of the endplate and femoral heads detection algorithms, and the accuracy of 13 spinopelvic parameters' measurements were evaluated.
    • Type of Ground Truth (Test Set): Expert annotations/measurements. Each spinopelvic parameter's ground truth was comprised of annotations (endplate lines, endplate endpoints, and/or femoral head circles) performed by the three U.S. board-certified radiologists using the Mazor X device.
    • Sample Size for Training Set:
      • Endplate Detection: 2327 AP images, 2651 LAT images.
      • Femoral Heads Detection: 2233 LAT images.
    • How Ground Truth for Training Set was Established: Endplate line and Circle femoral heads annotations were performed by trained labelers within a quality-controlled environment. In case of multiple annotations, they were aggregated. More than 65% of the data was reviewed by U.S. certified spine surgeons and radiologists.

    2. Plan Assist (AI-enabled Feature)

    • Sample Size for Test Set: 326 screw plans from 25 spine images.
    • Data Provenance for Test Set: Not explicitly stated (e.g., country of origin). It is explicitly stated that these datasets were separated at the site level from the development dataset to ensure data independence.
    • Number of Experts for Ground Truth (Test Set): Three experts.
    • Qualifications of Experts (Test Set): Not explicitly stated, but are referred to as "experts" who evaluated clinical acceptability.
    • Adjudication Method (Test Set): Each AI-generated screw plan was evaluated by three experts for clinical acceptability.
    • MRMC Comparative Effectiveness Study: No, a standalone performance of the AI algorithm was evaluated against expert judgment of clinical acceptability.
    • Standalone Performance: Yes, the clinical acceptability of AI-generated screw positions was evaluated.
    • Type of Ground Truth (Test Set): Expert judgment of clinical acceptability.
    • Sample Size for Training Set: 160 studies (101 3D CT scans, 59 3D O-arm scans), 942 vertebrae, 5,795 unique screw placements, and over 23,000 screw plannings.
    • How Ground Truth for Training Set was Established: Annotations for pedicle screw planning (referred to as "reference Screw planning") were manually performed by 34 qualified Surgical support technicians using a released Mazor X software.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K050438, K201189, K182077, K203005, K111520, K123270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medtronic Surgical Instruments are intended to be used during preparation and placement of Medtronic implants during spinal surgery to assist in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Surgical Instruments are specifically designed for use with the Mazor X Stealth™ Edition. which is indicated for medical conditions in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Surgical Instruments can be navigated or non-navigated manual instruments that may or may not be guided through the Mazor X Stealth™ Edition Arm Guide. Medtronic surgical drills shall only be used through the Mazor X Stealth™ Edition arm guides, Medtronic cannulas, and Medtronic drill guides. Some of the Medtronic Surgical Instruments are also compatible with the IPC™ Powerease™ System or AO* style quick connect drilling motors. An instrument may incorporate a measuring function which has uses as described on the instrument. Medtronic cannulas may or may not be used with Midas Rex™ attachments and tools.

    Device Description

    The Catalyft™ PL Expandable Interbody System navigated instruments consist of an Inserter (Outer), Inserter Inner Sleeve, Trials, a NAV Verification Tool, and Navigated Rotating Shavers that were cleared in K210425. These instruments are navigated instruments manufactured from stainless steel and are currently cleared for use in procedures where the use of stereotaxic surgery may be appropriate with the StealthStation™ System (K050438. S.E. 06/02/2005; K201189. S.E. 05/29/2020). Per this 510(k) submission, Medtronic is seeking clearance of the subject devices to also be navigated with MAZOR X™ (K182077, S.E. 11/02/2018; K203005, S.E. 10/27/2020) with the devices remaining compatible with StealthStation™ System.

    Please note that the Navigated Rotating Shavers were previously cleared for use on MAZOR X™ (K203005, S.E. 10/27/2020), and per this 510(k) submission, the Rotating Shavers can be used for disc prep and to trial for Catalyft™ PL on MAZOR XTM.

    The Anteralign™ TL navigated instruments consist of an Inserter and Trials that were originally cleared in K212524. These instruments are navigated instruments manufactured from stainless steel and are currently cleared for use in procedures where the use of stereotaxic surgery may be appropriate with the StealthStation™ System (K050438, S.E. 06/02/2005; K201189, S.E. 05/29/2020). Per this 510(k) submission, Medtronic is seeking clearance of the subject devices to also be navigated with MAZOR X™ (K182077, S.E. 11/02/2018; K203005, S.E. 10/27/2020) with the devices remaining compatible with StealthStation™ System.

    The CD Horizon™ Solera™ Vovager™ Spinal System navigated instruments consists of new Drivers, which will be manufactured from stainless-steel. Medtronic is seeking clearance of the Drivers to be Navigated and Trajectory Guided via MAZOR X™ (K182077, S.E. 11/02/2018; K203005, S.E. 10/27/2020) and Powered via IPC™ POWEREASE™ System (K111520. S.E. 10/26/2011: K123270. S.E. 01/11/2013). To enable trajectory guidance compatibility, the Drivers have been designed to work with the MAZOR X™ Arm Guides. To enable power, the Drivers may also be attached to the POWEREASE™ Driver (K111520, S.E. 10/26/2011).

    AI/ML Overview

    The provided document is a 510(k) summary for Medtronic surgical instruments, outlining their substantial equivalence to predicate devices. It focuses on the compatibility of these instruments with the MAZOR X™ system. However, the document does not contain the detailed acceptance criteria and performance data typically found in a clinical study report for proving a device meets acceptance criteria.

    Specifically, it mentions "Performance Testing" and "Discussion of the Performance Testing" but then lists generic test descriptions rather than specific quantitative acceptance criteria or reported device performance metrics. There is no information on sample sizes, ground truth establishment, expert qualifications, or MRMC studies.

    Therefore, based solely on the provided text, I cannot fill out the requested table or provide the detailed study information. The document serves as a high-level summary for regulatory submission, not a detailed performance study report.

    Here's a breakdown of what can and cannot be extracted from the provided text according to your request:

    Information NOT available in the provided text:

    • A table of acceptance criteria and the reported device performance: The document lists general "performance testing" but does not define quantitative acceptance criteria (e.g., "accuracy shall be within X mm") or specific numerical results.
    • Sample sizes used for the test set and the data provenance: No information on the number of cases, images, or subjects used in the "performance testing," nor any details on whether it was retrospective or prospective, or data origin.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as there's no mention of a ground truth established by experts for performance evaluation. The "ground truth" here is likely engineering validation against design specifications or a simulated environment, not clinical expert consensus.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable; no human adjudication described.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is surgical navigation instrumentation, not an AI / image analysis device that would assist human readers in diagnosis.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The "performance testing" described is likely functional verification of the instruments and their compatibility with the navigation system, not a standalone algorithm performance.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not explicitly stated, but inferred to be engineering specifications and simulated use assessments rather than clinical ground truth.
    • The sample size for the training set: Not applicable as this is not an AI/ML device that requires a training set.
    • How the ground truth for the training set was established: Not applicable.

    What can be inferred or extracted (though still lacking detail for your specific request):

    • Device Type: Surgical Instruments for spinal surgery, intended to assist in precisely locating anatomical structures using navigation systems (MAZOR X™ Stealth™ Edition and StealthStation™ System).
    • Nature of "Performance Testing": The document briefly lists categories of testing:
      • Navigation Accuracy Analysis
      • CAD Model Verification
      • Tools Package Functional Verification
      • NAV Simulated Use
      • Anatomical Simulated Use
    • Purpose of Testing: "To ensure the functionality and compatibility of the identified Medtronic products when used with the Mazor X™ system."
    • Conclusion: Based on the supporting information (which includes the performance testing), the subject instruments are deemed "substantially equivalent" to the predicate devices. This implies they met internal benchmarks/acceptance criteria, but the specific values are not provided.

    In summary, the provided document is a regulatory submission summary demonstrating substantial equivalence, not a detailed technical report of the performance studies with acceptance criteria and results for the device's navigation accuracy or other functionalities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181111

    Validate with FDA (Live)

    Date Cleared
    2018-05-17

    (20 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K163375, K111520, K123270, K050438, K170011, K124004

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC™ POWEREASETM System.

    Device Description

    The subject Navigated INFINITY™ drill bit is a sterile, single-use instrument that can be operated manually or under power. This instrument is intended to be used when implanting components of the INFINITY™ OCT System. The Navigated INFINITY™ Drill bit is also compatible with the StealthStation® and IPC® POWEREASE® Systems. Note: The subject devices are not intended to support occipital screw placement.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a device meets them. The document is an FDA 510(k) premarket notification for "Navigated INFINITY™ Instruments," detailing its indications for use, comparison to a predicate device, and sterilization information.

    Specifically, the document states:

    • No acceptance criteria or study details: The "Performance Data" section (VII) only mentions "Sterilization" and that "The addition of the sterile drill bit does not create a new 'worst case' and it was adopted in to the existing product family as detailed in the sterilization rationale provided in this submission since the design, material, manufacturing process, and manufacturing facilities of the sterile subject devices are equivalent to that of the referenced product family. The sterilization justification supports an SAL of 10-6. Based on the information provided in this submission, Medtronic believes that adequate evidence has been provided to show that we have fully validated the sterilization parameters." This indicates that the main "performance data" presented is a sterilization justification, not a study with acceptance criteria for device performance.
    • Focus on substantial equivalence: The document's goal is to demonstrate substantial equivalence to previously cleared devices, primarily based on identical indications, intended use, technology, materials, dimensions, and design features, with the only technological difference being that the subject device is sterile and single-use, while the predicate was non-sterile and single-use. This type of submission often relies on a comparison to a predicate device rather than extensive new performance studies if the changes are minor and do not introduce new questions of safety or effectiveness.

    Therefore, I cannot populate the requested table and information as the necessary details are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173338

    Validate with FDA (Live)

    Date Cleared
    2018-01-18

    (87 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K163375, K111520, K123270, K050438, K170011, K124004

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC® POWEREASE® System.

    Device Description

    The Navigated INFINITY™ Instruments are non-sterile, reusable and single use instruments that can be operated manually or under power. These instruments are intended to be used when implanting components of the INFINITY™ OCT System. The Navigated INFINITY™ Instruments are also compatible with the StealthStation® and IPC® POWEREASE® Systems.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the "Navigated INFINITY™ Instruments":

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report for new device approval. As such, the information provided for some of your requested points is limited or not applicable in the context of a 510(k).


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Maintain Navigation AccuracyNavigation Accuracy Analysis Confirmed navigated instrument accuracy
    Instrument Functionality under Expected Use ConditionsAnatomical Simulated Use Confirmed instrument functionality under expected use conditions
    Navigation System Functionality under Expected Use ConditionsNavigation Simulated Use Confirmed navigation system functionality under expected use conditions
    Accurate Reflection of CAD Models in Application SoftwareCAD Model Evaluation Verified that the CAD models are accurately reflected in the application software
    Proper Mating of Instruments with Appropriate DevicesImplant/Instrument Mating Conditions Verified that the instruments can be assembled with the appropriate devices according to their intended use
    Spine Tools Package Meets Interface Needs of Spine Application SoftwareSpine Tools Package Functional Testing Verified that the Spine Tools package has met the required interface needs of the spine application software

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify the sample size (e.g., number of instruments tested, number of simulated cases) for any of the performance tests.
      • Data Provenance: Not specified. Given the nature of a 510(k) and the types of tests listed (simulated use, accuracy analysis, CAD evaluation, functional testing), it is highly probable that the data provenance is retrospective (using internal testing data) and likely in-house laboratory or engineering testing, rather than clinical data from a specific country of origin.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. For the types of engineering and functional tests described (e.g., navigation accuracy, instrument functionality, CAD evaluation), "ground truth" would likely be established through engineering specifications, validated measurement systems, and perhaps expert observation from R&D engineers, rather than clinical experts like radiologists.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided and is unlikely to be relevant for the non-clinical, performance-based tests described in the 510(k) summary. Adjudication methods like 2+1 are typically used for clinical endpoints, especially in image interpretation.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. The device (Navigated INFINITY™ Instruments) is a surgical navigation instrument, not an AI-powered diagnostic or assistive tool for "human readers." Therefore, this type of study and its associated effect size are not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The term "standalone" and "algorithm only" are typically used for AI/software devices. This device is a physical surgical instrument that works with existing navigation systems (StealthStation™ System, IPC® POWEREASE® System). The performance tests listed (e.g., Navigation Accuracy Analysis, Anatomical Simulated Use) assess the instrument's accuracy and functionality within the intended surgical navigation system, which inherently involves human interaction (the surgeon). Therefore, a "standalone algorithm only" performance is not applicable in the context of this device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the tests described:
        • Navigation Accuracy Analysis: Ground truth would likely be established through pre-defined engineering specifications and validated measurement systems (e.g., comparing navigated position to physically measured, known positions).
        • Functional/Simulated Use Tests: Ground truth would be defined by the device's design specifications and intended functionality (i.e., whether it performs as designed under various conditions).
        • CAD Model Evaluation: Ground truth would be the original CAD models themselves and the expectation of their accurate representation in the software.
        • No expert consensus, pathology, or outcomes data is explicitly mentioned or seems relevant for these specific engineering tests.
    7. The sample size for the training set:

      • This question is not applicable. The "Navigated INFINITY™ Instruments" are physical surgical instruments, not a machine learning model that requires a training set. The underlying navigation systems (StealthStation™) might have calibration data or algorithms, but the instruments themselves do not have a "training set" in the sense of AI.
    8. How the ground truth for the training set was established:

      • This question is not applicable as there is no "training set" for these physical instruments.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133665

    Validate with FDA (Live)

    Date Cleared
    2014-03-13

    (104 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K001284, K050438

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Malleable Suction™ Instrument is indicated for use in ENT procedures where surgical navigation or image-guided surgery may be necessary to locate and remove fluids, semi-fluid substances, tissue, and bone dust. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and Medtronic ENT software.

    The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT- or MR-based model or digitized landmarks of the anatomy.

    The system and its associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.

    The Malleable Suction™ Instrument is indicated for use in the following procedures:

    • . Pituitary tumor removal only ENT procedures
    • CSF leak repair only related or from an ENT procedure . ●
      • Skull base procedures for ENT access .
      • Transsphenoidal procedures .
    • · Intranasal procedures
      • Optic nerve decompression procedures .
      • . Orbital decompression procedures
      • . Polyposis procedures
      • Endoscopic dacryocystorhinostomy .
      • Encephalocele procedures .
      • Sinus procedures, such as maxillary antrostomies .
      • Ethmoidectomies .
      • Sphenoidotomies/sphenoid explorations .
      • Turbinate resections and frontal sinusotomies .
    Device Description

    The Malleable Suction instruments are sterile, single use instruments intended for use with the ENT software applications on a Medtronic computer assisted surgery systems, StealthStation® systems (S7, i7 and Fusion®). The Malleable Suction instruments are electromagnetically navigated devices that are designed to be bendable by hand into a shape by the user in order to customize the instrument to the needs of the surgeon for each individual patient's anatomy and disease.

    Each Malleable Suction instrument incorporates a tracking system within the instrument tip. The mobile emitter of the navigation system generates a low-energy magnetic field to locating the tracker mounted within the tip of the instrument. The navigation system software displays the location of the instrument's tip within multiple patient image planes and other anatomical renderings.

    The instrument attaches to, and functions with, standard surgical vacuum suction systems and comes with a cleaning stylet to assist in dislodging blockages.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Medtronic Malleable Suction™ Instruments based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    System Accuracy RequirementBench-top and simulated environment: 95% confidence / 99.5% reliability, as dictated by risk analysis, of ≤ 3.00 mm.
    Suction CapabilitiesCapable of removing fluids, semi-fluid substances, tissue, and bone dust.
    Shaft Flex/ConfigurationsBendable, configuration surgeon defined
    Tip ConfigurationsStandard (straight), Angle, Ball
    MaterialsAluminum shaft, Pebax Liner, LCP Tip, Polyester Shrink tubing
    Sterile/ReusableSterile, Single Use

    Note on "Reported Device Performance": The document primarily focuses on demonstrating substantial equivalence to predicate devices and detailing the types of tests performed. For some criteria, such as "Shaft Flex/Configurations," "Tip Configurations," "Materials," and "Sterile/Reusable," the "reported device performance" is essentially a description of the device's characteristics that meet the intended design and comparison with predicate devices. The accuracy criterion is the most quantitative.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for a "test set" in the context of an algorithm or AI performance evaluation. The "Accuracy" testing was conducted in "2D and 3D space" and "Accelerated Life Functionality" tests involved "accelerated life exposures and user conditions" in a "bench-top and simulated environment." The data provenance is described as "bench-top and simulated environment," implying laboratory testing rather than human patient data, and therefore, no specific country of origin or retrospective/prospective designation applies.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable and is not provided in the document. The device is a physical surgical instrument and not an AI or diagnostic algorithm that requires expert ground truth for its performance evaluation for this 510(k) submission.

    4. Adjudication Method

    This information is not applicable. The device is a surgical instrument, and its performance testing as described does not involve expert adjudication of results in the way an AI diagnostic tool would.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes a 510(k) submission for a surgical instrument, not an AI or software device that would typically undergo an MRMC study.

    6. Standalone (Algorithm Only) Performance Study

    No. This document describes a physical medical device, not an algorithm.

    7. Type of Ground Truth Used

    The ground truth for the device's performance, specifically its accuracy, would be established by metrology standards and measurements in a controlled laboratory environment. For functionality tests like suction capabilities and malleability, the ground truth is simply whether the device performs its intended mechanical function as designed and specified.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI or machine learning algorithm, and therefore there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122211

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-17

    (145 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K070106,K050438,K003111

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes Navigable Pedicle Preparation Instruments are intended to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or a vertebra can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. These procedures include but are not limited to spinal fusion.

    Device Description

    The Navigable Pedicle Preparation Instruments are manual surgical instruments which are designed to interface with already-cleared surgical navigation systems. Instruments in this system may be pre-calibrated to already-cleared surgical navigation systems, or may be manually calibrated to already-cleared surgical navigation systems using manufacturers' instructions. These instruments are intended to be used in spine applications to perform general manual functions within the orthopaedic surgical environment.

    AI/ML Overview

    This is a 510(k) summary for a set of surgical instruments, not an AI/ML device. Therefore, the requested information regarding acceptance criteria, study data, expert involvement, and ground truth for an AI device is not applicable and cannot be extracted from the provided text.

    The document describes the Synthes Navigable Pedicle Preparation Instruments, which are manual surgical instruments designed to interface with existing surgical navigation systems.

    Here's a breakdown of the relevant information provided:

    1. Acceptance Criteria and Reported Device Performance (Non-AI/ML):

    Acceptance Criteria (Implied)Reported Device Performance
    Meet performance requirements for intended use"The Navigable Pedicle Preparation Instruments met the performance requirements, providing assurance of device performance for their intended use."
    No safety or effectiveness issues"No safety or effectiveness issues were raised by the performance testing."

    2. Sample size used for the test set and data provenance (Not applicable for AI/ML):

    The document mentions "usability testing" as a validation activity in the "Performance Data" section. However, it does not specify sample sizes for test sets, data provenance (e.g., country of origin), or whether the data was retrospective or prospective, as these details are typically associated with clinical studies or AI/ML performance evaluations, not characterization of manual surgical instruments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (Not applicable for AI/ML):

    This information is not provided because the validation involved usability testing of manual instruments, not the establishment of ground truth for an AI/ML algorithm.

    4. Adjudication method (Not applicable for AI/ML):

    This concept is not relevant to the validation described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance (Not applicable for AI/ML):

    No MRMC study was conducted as this device is not an AI/ML system or a reading assistance tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done (Not applicable for AI/ML):

    This is not applicable as the device is a set of manual surgical instruments.

    7. The type of ground truth used (Not applicable for AI/ML):

    Ground truth as it relates to AI/ML performance is not relevant here. The validation focused on the functional performance and usability of the physical instruments.

    8. The sample size for the training set (Not applicable for AI/ML):

    There is no "training set" in the context of manual surgical instrument validation, as this is a concept associated with AI/ML model development.

    9. How the ground truth for the training set was established (Not applicable for AI/ML):

    This is not applicable for the reasons stated above.

    In summary: The provided 510(k) summary focuses on demonstrating the substantial equivalence of manual surgical instruments through non-clinical performance testing (usability testing). It does not contain the detailed information related to AI/ML device validation, such as specific acceptance criteria for algorithm performance, sample sizes for test/training sets, expert involvement in ground truth establishment, or comparative effectiveness studies of AI assistance. The document explicitly states: "Clinical data was not needed for the Navigable Pedicle Preparation Instruments."

    Ask a Question

    Ask a specific question about this device

    K Number
    K073467

    Validate with FDA (Live)

    Date Cleared
    2008-05-23

    (165 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K050438

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SpineAssist™ System is indicated for precise positioning of surgical instruments or implants during general spinal surgery. The SpineAssist™ System may be used in either open or percutaneous procedures.

    Device Description

    The SpineAssist™ system is a computer controlled miniature medical image-guided surgery (IGS) system, which serves as a technological platform for solutions that provide unprecedented levels of accuracy, precision and accessibility in performing orthopedic procedures. The SpineAssist™ is designed to assist surgeons in precisely guiding handheld surgical tools in line with a computerized, image-based pre-operative plan along given trajectories. The system's software processes fluoroscopic and CT images via proprietary algorithms and automatically exports the desired coordinates to the SpineAssist device, which positions its articulating arm and tool guide. Using a special bone attachment component (i.e., a clamp and bridge, the Hover-T / Bi-lateral Hover-T bridge, the Bed Mount Hover-T or the Cervical Kit) the SpineAssist device attaches to the bone in the area where the procedure is being performed and assists surgeons in precisely guiding handheld surgical tools according to the computerized, image-based, pre-operative plan.

    The main components of the SpineAssist™ system include:
    A. SpineAssist™ device
    B. Workstation
    C. Accessories including the Clamp Kit for less invasive procedures, the MIS platforms: Hover-T, Bi-lateral Hover-T, Bed Mount Hover-T and the Cervical Kit, the last one intended for less invasive procedures.

    AI/ML Overview

    Here's an analysis of the provided text regarding the SpineAssist™ System's acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test / RequirementReported Device Performance / Outcome
    StabilityBed Mount Hover-T Stability TestPerformed, intent to validate stability
    Relative Movement of C2 TestPerformed, intent to validate stability
    Cervical Bridge Rigidity TestPerformed, intent to validate stability
    Bi-lateral Hover-T Rigidity TestPerformed, intent to validate stability
    AccuracyCervical Accuracy C1-C6 TestPerformed, intent to validate accuracy
    Bed Mount Hover-T Accuracy T6-T12Performed, intent to validate accuracy
    Software ValidationSoftware Validation (IEC 60601-1-4 & FDA Guidelines)Meets design and performance specifications
    BiocompatibilityBiocompatibility Testing (ISO 10993)Complies with standard
    Overall ConclusionSafety and EffectivenessSystem may be safely and effectively used; substantially equivalent to previously cleared devices.

    Note: The document states that the performance tests "demonstrate that SpineAssist system meets its design and performance specifications." However, it does not provide specific quantitative acceptance criteria or detailed results for each test (e.g., "accuracy within X mm"). It only indicates that these tests were performed to validate stability and accuracy for the extended intended use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a numerical sample size. The document mentions tests were performed "on cadavers to simulate real clinical procedures." The number of cadavers used is not specified.
    • Data Provenance: The cadaver tests were described as simulating clinical procedures, but the origin of the cadavers (e.g., country) is not mentioned. These were non-clinical tests. The study is prospective as it involved performing tests to gather data for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not provided in the given text. The tests were "stability and accuracy tests... performed on cadavers," implying a technical assessment rather than a human expert consensus for "ground truth" related to device performance metrics.

    4. Adjudication Method for the Test Set

    • This information is not applicable and not provided. The performance tests described (stability, accuracy) are typically measured objectively by technical instrumentation rather than requiring a human adjudication process like in a diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This submission is for a surgical navigation system, not an AI-assisted diagnostic tool. There is no mention of human readers or AI assistance in a diagnostic context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, standalone (algorithm/device only) performance tests were done. The "Non-Clinical Performance Data" section describes various stability and accuracy tests performed on the SpineAssist™ system itself (e.g., "Bed Mount Hover-T Stability Test," "Cervical Accuracy C1-C6 Test"). These tests evaluate the device's technical specifications without necessarily involving human surgical performance in a comparative study. The software validation is also a standalone assessment of the algorithmic components.

    7. The Type of Ground Truth Used

    • For the accuracy and stability tests, the ground truth would likely be established through pre-defined anatomical landmarks or precise measurements taken by a high-precision measurement system (e.g., optical tracking, CMM). For example, in an "accuracy" test, the deviation of the navigated tool from a known target trajectory or point would be measured against a gold standard reference. The text doesn't explicitly state the methodology for establishing this ground truth but implies it's based on physical measurements and engineering specifications.

    8. The Sample Size for the Training Set

    • This information is not applicable and not provided. The SpineAssist™ system is described as a "computer controlled miniature medical image-guided surgery (IGS) system" that processes images and exports coordinates. While it uses "proprietary algorithms," it is not described as a machine learning or AI system that requires a distinct "training set" to develop its core functionality in the sense that a diagnostic AI model would. Its software validation focuses on compliance with standards rather than a machine learning training paradigm.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable and not provided for the reasons stated in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063607

    Validate with FDA (Live)

    Date Cleared
    2007-08-23

    (262 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K033413, K051676, K954276, K050438

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SpineAssist™ System is indicated for precise positioning of surgical instruments or implants during thoracic and lumbar spinal surgery. The SpineAssist™ system may be used in either open or percutaneous procedures.

    Device Description

    The SpineAssist™ system is a computer controlled miniature medical image-guided surgery (IGS) system which serves as a technological platform for solutions that provide unprecedented levels of accuracy, precision and accessibility in performing orthopedic procedures. The SpineAssist™ is designed to assist surgeons in precisely guiding handheld surgical tools or implants in line with a computerized, image-based pre-operative plan along given trajectories. The system's software processes fluoroscopic and CT images via proprietary algorithms and automatically exports the desired coordinates to the SpineAssist device, which positions its articulating arm and tool guide. Using a special bone attachment component (i.e., a clamp and bridge or the Hover-T bridge) the SpineAssist device attaches to the bone in the area where the procedure is being performed and assists surgeons in precisely guiding handheld surgical tools or implants according to the computerized, image-based, preoperative plan.

    The main components of the SpineAssist™ system include:

    • A. SpineAssist™ device
    • B. Workstation
    • A. Accessories including clamp, bridge, Hover-T bridge, targets, prism, wedge, etc.
    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a comprehensive study report with the specific information requested in the prompt (e.g., specific quantitative benchmarks for acceptance, detailed performance metrics with numerical results, sample sizes for test and training sets, expert qualifications, or adjudication methods).

    However, based on the summary of non-clinical performance data and conclusions, I can infer the general acceptance criteria and report the mentioned performance.

    Here's an analysis of what can be extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in a table format. However, it implicitly suggests that the system's performance for "accuracy" and "safety and effectiveness" during "general spinal surgical procedures" were the key criteria.

    Acceptance Criteria (Inferred)Reported Device Performance
    Safe and Effective for general spinal surgical proceduresPerformance tests demonstrate that SpineAssist system may be safely and effectively used in general spinal surgical procedures.
    Meets design and performance specificationsSoftware validation and accuracy performance tests demonstrate that the SpineAssist system meets its design and performance specifications.
    Substantial Equivalence to predicate devices (original SpineAssist™ and StealthStation System)The system is found to be substantially equivalent to the predicate devices.
    Compliance with relevant standardsComplies with IEC 60601-1-4 (Software Validation) and ISO 10993 (Biocompatibility Testing).
    Demonstrates "precision" in guiding instruments/implantsIndicated for "precise positioning of surgical instruments or implants."
    Accuracy in new imaging scenarios"New Imaging and Lateral to 30 degree Accuracy Test" completed.
    Accuracy with Hover-T and Prisms"Hover-T Accuracy Test Results Report" and "Use of Prisms in Translaminar Facet Cases Study" completed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "non-clinical performance data" and lists several "case studies" and "accuracy tests." It does not specify the sample size (e.g., number of cases, number of measurements) for any of these tests.

    The data provenance is not explicitly stated. However, given the nature of "non-clinical performance data" for a surgical device involving "accuracy tests" and "case studies," it is highly likely that these were conducted in a controlled laboratory or cadaveric setting, rather than on human patients (since "Clinical Performance Data" is marked as "Not Applicable"). It is likely prospective for the specific tests conducted for this submission. The company is based in Israel, so the studies were likely conducted there or in a partnered facility.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide any information regarding the number of experts, their qualifications, or how ground truth was established for the "accuracy tests" or "case studies."

    4. Adjudication Method for the Test Set

    The document does not provide any information regarding adjudication methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done or reported in this summary. The device is a surgical navigation system, not an AI diagnostic tool that assists human readers in interpreting medical images. "Clinical Performance Data" is explicitly stated as "Not Applicable," further reinforcing that human reader studies are outside the scope of this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the "Non-Clinical Performance Data" listed are all standalone algorithm/system performance tests. The device itself is designed to guide surgical tools based on algorithmic processing of imaging data, and the tests like "General Spinal Accuracy Test," "Hover-T Accuracy Test," and "New Imaging and Lateral to 30 degree Accuracy Test" directly assess the system's performance in this standalone capacity. The lack of "Clinical Performance Data" and MRMC studies further supports that the reported performance reflects the algorithm/system without human-in-the-loop performance for this specific submission.

    7. The Type of Ground Truth Used

    While not explicitly stated, for "accuracy tests" in a surgical navigation system, the ground truth would typically be established by:

    • Precise measurements using highly accurate calibration tools or coordinate measuring machines in a controlled environment (e.g., phantom, cadaveric models).
    • Known physical dimensions or fiducial markers on phantoms or cadavers which the system is attempting to target or guide to.
    • For "case studies" (e.g., Osteoid Osteoma, Thoracic Hover-T), the "ground truth" might refer to the actual anatomical position or intended trajectory as confirmed by intra-operative imaging or post-operative assessment against the surgical plan.

    8. The Sample Size for the Training Set

    The document does not provide any information regarding the sample size for the training set.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information regarding how the ground truth for the training set was established.

    Ask a Question