510(k) Data Aggregation

    K Number
    K150469

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-23

    (28 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K091623, K050082

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Integra Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) after laminoplasty procedures. The Integra Laminoplasty System is used to hold the allograft material in place in order to prevent the allograft material from expulsion, or impinging the spinal cord.

    Device Description

    The Integra Laminoplasty System consists of titanium alloy plates and screws that are attached to the lamina after a laminoplasty procedure. Implants are available in a range of sizes and shapes, as well as in various screw hole configurations to accommodate variations in patient anatomy. The system also includes instruments to assist with implantation, and a tray for organization and storage

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "Integra® Laminoplasty System." This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving performance against specific acceptance criteria through a study as would be common for AI/software devices.

    Therefore, many of the requested categories for AI/software studies are not applicable directly to this submission. However, I can extract the relevant information from the document regarding the device and its testing:

    Device Name: Integra® Laminoplasty System
    Regulation Name: Spinal interlaminal fixation orthosis (21 CFR 888.3050)
    Regulatory Class: Class II
    Product Code: NQW


    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: The primary acceptance criterion for a 510(k) submission is demonstrating "substantial equivalence" to a legally marketed predicate device. This means the new device is as safe and effective as the predicate device. For the Integra Laminoplasty System, this was demonstrated through non-clinical testing.
    • Reported Device Performance: The device "demonstrated equivalent performance to the predicate systems through static testing per ASTM F2193." This implies that the mechanical properties tested under ASTM F2193 met or exceeded the performance of the predicate device(s). Specific quantitative performance metrics (e.g., tensile strength values, compression resistance values) are not provided in this summary.
    Acceptance CriterionReported Device Performance
    Substantial Equivalence to Predicate DeviceDemonstrated equivalent performance to predicate systems through static testing per ASTM F2193.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document describes non-clinical testing (static testing per ASTM F2193), not a test set for an algorithm. Therefore, "sample size" here refers to the number of physical samples tested. The document does not specify the number of samples tested under ASTM F2193.
    • Data Provenance: Not applicable, as it's non-clinical lab testing of physical components, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This device is a mechanical implant; "ground truth" in the context of expert review for AI algorithms is not relevant here. The ground truth for mechanical performance is established by standardized testing methods (ASTM F2193) and engineering principles.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This concept pertains to resolving discrepancies among expert readers in clinical image review or similar tasks for AI evaluations. This document describes mechanical testing of a medical implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "No clinical testing was required to demonstrate equivalence." Therefore, no MRMC study, or any clinical study involving human readers with or without AI assistance, was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The "Integra® Laminoplasty System" is a physical medical device (implants and instruments) for surgery, not an algorithm or AI software. Therefore, there is no "standalone algorithm performance" to evaluate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance, in the context of regulatory submission, would be its mechanical properties meeting or exceeding those of the predicate devices as determined by standardized laboratory testing (ASTM F2193). This is a physical, objective "ground truth" based on engineering standards.

    8. The sample size for the training set

    • Not applicable. This is not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no training set for this type of device.

    Summary of what the document does provide regarding acceptance criteria and performance:

    The acceptance criterion for this 510(k) submission was substantial equivalence to predicate devices in areas including intended use/indications for use, technological characteristics, and performance. This equivalence was demonstrated through non-clinical static testing per ASTM F2193, showing "equivalent performance" to the predicate systems. Specific quantitative results from this testing are not included in the public 510(k) summary. No clinical testing or AI-related studies were performed or required for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122822

    Validate with FDA (Live)

    Date Cleared
    2013-02-07

    (146 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K032534, K050082

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectrum Spine Laminoplasty Plating System is intended for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Spectrum Spine Laminoplasty Plating System holds the allograft in place in order to prevent the allograft from expulsion or impinging on the spinal cord.

    Device Description

    The Spectrum Spine Laminoplasty Plating System is comprised of various sized, pre-bent plates that are designed to fit the anatomy of the vertebral arch (i.e., between the pedicle and spinous process). The plates have screw holes located at both ends of the plate to allow for attachment to the bone. The screws intended for use with the plates are available in a variety of lengths and diameter and are designed to match the anatomical requirements.

    AI/ML Overview

    The provided text describes the Spectrum Spine Laminoplasty Plating System and its performance data for a 510(k) submission. However, it does not contain specific acceptance criteria, reported device performance in those terms (apart from passing certain ASTM standards), or details about a study that would address most of the requested information.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and mechanical testing according to established ASTM standards. It does not describe a clinical study or a study comparing the device's performance against specific, quantifiable acceptance criteria in a clinical context.

    Therefore, for many of your requested points, the information is not available in the provided text.

    Here is the information that can be extracted or deduced from the provided text, and where the information is not available, it is stated as such:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Performance:
    Static Axial Pullout (screws per ASTM F543-07)Tested and met the standard (implied, as conclusion states substantial equivalence)
    Static Four-Point Bend (system per ASTM F2193-02)Tested and met the standard (implied, as conclusion states substantial equivalence)
    Dynamic Four-Point Bend (system per ASTM F2193-02)Tested and met the standard (implied, as conclusion states substantial equivalence)
    Material Composition:
    Plates and screws manufactured from titanium alloy (ASTM F136)Manufactured from titanium alloy (ASTM F136)
    Indications for Use:
    Intended for use in lower cervical and upper thoracic spine (C3-T3) after laminoplasty to hold allograft and prevent expulsion/impingement on spinal cord.Indicated for use in lower cervical and upper thoracic spine (C3-T3) after laminoplasty to hold or buttress allograft and prevent expulsion/impingement on spinal cord.
    Technical Characteristics:
    No new technical characteristics compared to predicate devices.No new technical characteristics introduced.

    Note on "Reported Device Performance": The document states that the device "was tested" according to specific ASTM standards and that the data "lead to the conclusion that Spectrum Spine Laminoplasty Plating System is substantially equivalent to the predicate devices." This implicitly means the device met the requirements of these tests, but the specific numerical results or thresholds (acceptance criteria) for those tests are not detailed in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The testing described is mechanical (physical hardware testing), not clinical or involving human subjects/cases in the traditional sense of a "test set" for performance evaluation.
    • Data Provenance: The "performance data" refers to mechanical testing.
      • Country of origin: Not specified, but likely performed in a certified lab in the USA, given the FDA submission.
      • Retrospective or Prospective: Not applicable, as this refers to mechanical bench testing, not clinical data collection.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. The "ground truth" for this device's performance evaluation is based on established engineering standards (ASTM F543-07 and ASTM F2193-02) and material specifications (ASTM F136), not expert consensus from clinical cases. There's no mention of experts establishing a ground truth in the context of device performance as one would for an AI/diagnostic device.

    4. Adjudication Method for the Test Set

    • Not applicable. This concept typically applies to clinical studies where expert reviewers independently assess cases and then reconcile discrepancies. The reported testing is mechanical bench testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC study is not relevant for this type of mechanical implant device. This type of study is typically performed for diagnostic or AI-assisted interpretation devices.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This device is a mechanical surgical implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • Engineering Standards and Material Specifications: The ground truth for evaluating the device's safety and effectiveness for a 510(k) submission is based on adherence to established ASTM standards for mechanical properties (e.g., strength, durability) and material composition (titanium alloy as per ASTM F136), demonstrating substantial equivalence to previously marketed devices.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a mechanical surgical implant. There is no training set in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121732

    Validate with FDA (Live)

    Date Cleared
    2012-10-31

    (140 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080664, K050082, K091994, K032534

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CANOPY™ Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) in laminoplasty procedures. The CANOPY™ Laminoplasty Fixation System is used to hold bone allograft material in place in order to prevent the allograft from expulsion or impinging the spinal cord.

    Device Description

    The CANOPY™ Laminoplasty Fixation System consists of spinal fixation plates and screws for use in laminoplasty procedures. CANOPY™ implants are inserted through a posterior cervical or thoracic approach, and are available in various sizes and geometric options to fit individual patient anatomy. Fixation plates may be used with bone graft material. Hinge plates may be used to stabilize a weakened or displaced lamina. Screws are used to attach the plates to bone and are available in a variety of lengths and diameters to fit patient anatomy.

    CANOPY™ plates and screws are manufactured from titanium or titanium allov. as specified in ASTM F67, F136, F1295 and F1472. Optional graft chambers are manufactured from radiolucent polymer as specified in ASTM F2026 and contain tantalum or titanium alloy markers to permit radiographic visualization, per ASTM F67, F136, F560, F1295 or F1472.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the CANOPY™ Laminoplasty Fixation System. It describes the device, its intended use, and claims substantial equivalence to predicate devices based on performance data.

    However, the document does not contain information related to an AI/ML device study. Specifically, it lacks:

    • A table of acceptance criteria and reported device performance for an AI/ML model.
    • Sample sizes, data provenance, or details about test sets and training sets for an AI/ML model.
    • Information about experts, ground truth adjudication, or multi-reader multi-case studies related to an AI/ML model.
    • Any mention of standalone algorithm performance or human-in-the-loop performance with AI assistance.

    Instead, the performance data cited refers to "Mechanical testing (static and dynamic compression, static compression bending and expulsion) was conducted in accordance with ASTM F543 and F2193 and, the "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004." This indicates physical testing of the implant's mechanical properties, not an AI/ML diagnostic or prognostic study.

    Therefore, I cannot fulfill your request for information regarding acceptance criteria and studies related to an AI/ML device based on the provided text. The document describes a physical medical implant device, not an AI/ML software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113802

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-04-16

    (115 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K050082,K032534

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Escalate™ Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3-T3) in laminoplasty procedures. The system is intended to hold the lamina open following a laminoplasty procedure.

    Device Description

    The Escalate™ System consists of a comprehensive set of implants and instruments designed for a systematic approach to cervical laminoplasty procedures. The system features an expandable laminoplasty plate, a hinge (“Base”) plate, bone screws for fixation, and a set of instrumentation to assist in the implantation and removal of the implants. The plates have screw holes, which allow for attachment to the vertebral body. The screws to be used with the plates are available in a variety of lengths and diameters and are designed to match the anatomical requirements. The Escalate™ Laminoplasty plates and screws are manufactured from Titanium alloy and will be provided non-sterile.

    AI/ML Overview

    The Stryker Spine Escalate™ Laminoplasty System is a mechanical device intended to hold the lamina open following a laminoplasty procedure. The acceptance criteria and the study that proves the device meets those criteria are detailed below, as per the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    For mechanical devices like the Escalate™ Laminoplasty System, "acceptance criteria" generally refer to established engineering standards or performance thresholds that the device must meet to demonstrate safety and effectiveness for its intended use. "Reported device performance" refers to the results of tests designed to measure how the device performs against these standards.

    Acceptance Criteria (Standard/Test Type)Reported Device Performance (Outcome)
    Static Compression LoadingData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (Modified ASTM 1717-11a)
    Dynamic Compression LoadingData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (Modified ASTM 1717-11a)
    Screw Driving Insertion TorqueData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (ASTM F543-07)
    Screw Removal TorqueData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (ASTM F543-07)
    Screw Torque to FailureData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (ASTM F543-07)
    Axial Screw Pull-OutData demonstrates substantial equivalence to predicate devices; met all mechanical test requirements. (ASTM F543-07)
    Substantial EquivalenceThe Escalate™ Laminoplasty System was determined to be substantially equivalent to the predicate devices (Medtronic Centerpiece™ Plate Fixation System (K050082) and Synthes Arch™ Fixation System (K032534)). The subject system does not present any new issues of safety and effectiveness.

    Study Proving Acceptance Criteria are Met:

    The study proving the device meets the acceptance criteria was a series of performance tests conducted to demonstrate substantial equivalence to existing legally marketed predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The provided 510(k) summary does not specify the exact sample sizes (e.g., number of implants tested, number of screws tested) for each biomechanical test. It only states that "performance testing was conducted."
    • Data Provenance: This information is not explicitly stated in terms of country of origin. Given that Stryker Spine is in New Jersey, USA, and is submitting to the US FDA, it is highly probable that the testing was conducted in the USA or by a testing facility compliant with US regulatory standards. The data is retrospective in the sense that it describes tests performed on physical device prototypes, not data gathered from a clinical trial in patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable to this type of device and study. The "ground truth" for mechanical performance is established by the specifications of the ASTM standards and engineering principles, not by expert consensus in the way a medical image diagnosis would be. The "experts" involved would be biomechanical engineers and technicians conducting and analyzing the tests, but their specific number and detailed qualifications are not typically reported in this summary format.

    4. Adjudication Method for the Test Set

    This section is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation or clinical outcomes where there might be disagreement, such as in reading medical images. For biomechanical testing, the results are quantitative measurements against established standards, not subject to subjective adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. MRMC studies are relevant for diagnostic devices where human readers interpret medical data, often with or without AI assistance, to assess diagnostic performance. The Escalate™ Laminoplasty System is a surgical implant, and its evaluation focuses on mechanical integrity and equivalence, not diagnostic accuracy or human interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable to this device. A "standalone" performance evaluation typically refers to the performance of an AI algorithm independent of human input. The Escalate™ Laminoplasty System is a physical implant; it does not involve an algorithm.

    7. The Type of Ground Truth Used

    For this device, the "ground truth" is defined by:

    • Established engineering standards and specifications: Specifically, the ASTM standards (ASTM 1717-11a for compression loading and ASTM F543-07 for screw-related tests) provide the benchmarks for acceptable performance.
    • Performance of predicate devices: The "ground truth" for substantial equivalence is that the subject device performs similarly to or better than predicate devices already on the market and deemed safe and effective.

    8. The Sample Size for the Training Set

    This question is not applicable. The Escalate™ Laminoplasty System is a mechanical implant, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The design and manufacturing processes are informed by established engineering principles and prior device experience, but not by a machine learning training dataset.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as #8. There is no training set for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113218

    Validate with FDA (Live)

    Date Cleared
    2012-01-26

    (86 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K091994,K050082

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pioneer Release Laminoplasty Plating System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Pioneer Release Laminoplasty Plating System is used to hold the graft material in place in order to prevent the graft material from expulsion, or impinging the spinal cord.

    Device Description

    The Pioneer Release Laminoplasty Plating System consists of implantable plates and screws that will act as a buttress to maintain decompression after a laminoplasty procedure. The system also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.

    AI/ML Overview

    This document describes the acceptance criteria and the study that proves the Pioneer Release Laminoplasty Plating System meets those criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Standard)Reported Device Performance
    Static Four-Point Bend per ASTM F2193Functioned as intended
    Fatigue Four-Point Bend per ASTM F2193Functioned as intended
    Axial Screw Pull-Out per ASTM F543Functioned as intended
    Substantial Equivalence to Predicate SystemAchieved

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: Not explicitly stated in the provided document. The document mentions "standard testing" per ASTM F2193 and ASTM F543, implying that the number of samples would conform to the requirements of these standards.
    • Data Provenance: The study appears to be a laboratory-based performance test conducted by the manufacturer, Pioneer Surgical Technology. There is no information regarding the country of origin of the data or whether it was retrospective or prospective in a clinical setting. The tests were performed on the device components themselves.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    Not applicable. This device is a medical implant, and its performance was evaluated through mechanical testing, not by expert review of imaging or clinical data to establish ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the evaluation was based on mechanical test results against established ASTM standards, not on expert adjudication of clinical or imaging data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not performed. This study focuses on the mechanical performance of a spinal implant system. MRMC studies are typically used for diagnostic or screening devices where human readers interpret data, often with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is a physical medical implant, not an algorithm or AI system.

    7. Type of Ground Truth Used

    The "ground truth" for this study was established by recognized ASTM standards (F2193 and F543) for the mechanical performance of spinal implants. The device's ability to maintain its structural integrity and mechanical properties under specified loads, as defined by these standards, constituted the success criteria.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical implant, not a machine learning model that requires a training set.

    9. How Ground Truth for Training Set was Established

    Not applicable, as no training set was used.