510(k) Data Aggregation

    K Number
    K141284

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-23

    (68 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K081107, K101084, K033961, K043229, K122378

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and indicated for the following:

    · DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)

    • · Spondylolisthesis
    • Spinal stenosis
    • · Fracture/dislocation
    • · Revision of previous cervical spine surgery
    • · Tumors
    • · Atlantoaxial fracture with instability
    • · Occipitocervical dislocation

    The occipital bone screws are limited to occipital fixation only

    The rod and hook components are in the cervical/upper thoracic (CI-T3) spine. The pedicle screws are limited to placement in the TI-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.

    The CASPIAN OCT/MESA Mini/ DENALI Mini System can also be linked to the Range Spinal System using rod connectors or transitional rods.

    Device Description

    The CASPIAN OCT/MESA Mini/ DENALI Mini System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add 03.5 contoured rods to expand the rod offerings for the system.

    Materials: The subject devices are manufactured from Ti6Al4V Eli per ASTM and ISO standards.

    Function: The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine.

    AI/ML Overview

    The provided document is a 510(k) summary for a spinal fixation system, not a study report for a medical device that uses AI or reports on acceptance criteria and performance data in the typical sense of a diagnostic or screening device.

    Therefore, many of the requested categories for acceptance criteria and study design elements (such as sample size for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training set details, etc.) are not applicable to this type of submission.

    This 510(k) summary focuses on demonstrating substantial equivalence to previously marketed devices based on design, function, materials, and intended use, particularly for adding new rod sizes. The "acceptance criteria" here are implicitly related to mechanical testing and engineering rationale to ensure the new components maintain the same safety and effectiveness profile as the predicate devices.

    Here's a breakdown of the relevant information from the document in relation to your request:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied by 510(k))Reported Device Performance (from document)
    Mechanical PerformanceEquivalence to predicate devices in mechanical strength and stability under relevant biomechanical loads (e.g., as per ASTM F1717)."The worst case components of the CASPIAN OCT/MESA Mini/ DENALI Mini System were previously tested in accordance with ASTM F1717 and determined to be equivalent to predicate devices."
    New components (03.5 contoured rods) do not represent a new worst-case scenario."The subject rods were determined not to represent a new worst case via finite element analysis and engineering rationale."
    Material BiocompatibilityUse of biocompatible materials known to be safe for implantation."The subject devices are manufactured from Ti6Al4V Eli per ASTM and ISO standards."
    FunctionalityAbility to provide immobilization and stabilization of the posterior cervical and thoracic spine as an adjunct to fusion."The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine."
    Intended UseAlignment with established indications for use in spinal fixation.The Indications for Use statement in the document specifies conditions like DDD, Spondylolisthesis, Fracture/dislocation, etc. The device is deemed substantially equivalent for these uses.
    Design CharacteristicsNo significant differences from predicate devices that would adversely affect use."There are no significant differences between the CASPIAN OCT/MESA Mini/ DENALI Mini System and other systems currently being marketed which would adversely affect the use of the product."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable: This submission relies on mechanical testing standards (ASTM F1717) and engineering rationale/finite element analysis, not clinical 'test sets' of patient data in the context of AI or diagnostic devices. There is no mention of "patient data" test sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable: "Ground truth" in this context would relate to defining mechanical testing parameters and interpreting engineering analyses, handled by biomechanical engineers and regulatory specialists, rather than clinical experts establishing ground truth for diagnostic imaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: This relates to clinical data adjudication, which is not part of this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a submission for a spinal implant, not an AI-powered diagnostic or screening device, so MRMC studies are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is a medical device (spinal implant), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the mechanical performance, the "ground truth" is established by adherence to recognized biomechanical testing standards (e.g., ASTM F1717) and engineering principles/ finite element analysis (FEA), indicating that the device meets defined physical strength and durability criteria. The overall ground truth for regulatory clearance is substantial equivalence to existing, legally marketed devices.

    8. The sample size for the training set

    • Not Applicable: This is a spinal implant, not a machine learning model, so there is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable: No training set exists for this type of device.

    In summary, the provided document is a regulatory submission for a spinal fixation system, cleared based on substantial equivalence demonstrated through mechanical testing and engineering rationale, not clinical study data in the context of diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133698

    Validate with FDA (Live)

    Device Name
    SYNAPSE SYSTEM
    Date Cleared
    2014-01-16

    (44 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K033961, K043229

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These Systems are intended for the following:

    Hooks, Plate/Rods, Plates, Rods and Screws

    When intended to provide stabilization as an adjunct to fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, plates, rod, hook and screw (3.2 mm cortex) components of the Synthes CerviFix, Axon; OC Fusion and Synapse Systems are indicated for skeletally mature patients using allograft and/or autograft for the following:

    • Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with . degeneration of the disc as confirmed by patient history and radiographic studies)
    • . Spondylolisthesis
    • Spinal stenosis .
    • Fracture/dislocation .
    • Atlantoaxial fracture with instability .
    • Occipitocervical dislocation .
    • Revision of previous cervical spine surgery .
    • Tumor .

    When used to treat these cervical and occipitocervical conditions, screws are limited to occipital fixation only.

    Hooks and Rods

    The rod and hook components are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

    Rods, Clamps, Screws, Nuts, Variable Axis Screws, Locking Screws and Transverse Bars

    The rods, clamps, screws, nuts, variable axis screws, locking screws and transverse bars are intended to promote fusion following reduction of fracture/dislocation or trauma in the upper thoracic spine (TI-T3).

    The use of these screws (3.5 mm. 4.0 mm and 4.5 mm cancellous and 3.5 mm and 4.2 mm cortex) is limited to placement in T1-T3 in treating thoracic conditions only. They are not intended to be placed in or treat conditions involving the cervical spine.

    The Synthes CerviFix, Axon, and Synapse Systems can also be linked to the Synthes Universal Spinal System using the 3.5 mm/6.0 mm and 4.0 mm/6.0 mm paralle1 connectors from that system and via the CerviFix tapered rods using lamina hooks, transverse process hooks, pedicle hooks; 4.2 mm screws, and the 5.0 mm/6.0 mm parallel connector.

    Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic (T4-T12), or lumbar spine.

    Device Description

    The Synapse System is a hook and pedicle screw spinal system which is designed to provide spinal fixation to allow for stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The Synapse System is composed of multiple components to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The Axon System and Cervifix System also included in the indications for use statement are additional hook and pedicle screw spinal systems. These systems consist of anchors including bone screws, pedicle screws and hooks: interconnection mechanisms including locking screws, set screws, parallel rod connectors, clamps, transverse bars, rod collars and nuts; longitudinal members including rods and rod/plates; and transverse connectors including transconnectors. These system components are implanted using class I surgical instruments.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Synapse System, specifically focusing on proposed curved rods. It is a submission intended to demonstrate substantial equivalence to previously cleared devices, rather than a study proving performance against acceptance criteria for a novel device.

    Therefore, many of the requested categories for a study proving device performance are not applicable (N/A) in this context, as this submission primarily relies on a "rationale" comparing the new components to existing ones, rather than new performance testing.

    Here's an analysis based on the provided text, addressing your questions:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Material Conformance: Curved rods manufactured from Ti-6A1-7Nb (TAN), conforming to ASTM F-1295-11 and ISO 5832-11.The proposed curved rods are manufactured from the same material, titanium aluminum niobium (Ti-6A1-7Nb (TAN)), as the previously cleared Synapse System curved rods. The material conforms to ASTM F-1295-11 and ISO 5832-11.
    Technological Characteristics: Indications for use, intended use, and technological characteristics (design, material) of proposed curved rods are the same as, or similar to, predicate devices.The indications for use, intended use and technological characteristics, such as design and material, of the proposed curved rods are the same as, or similar to, those of the predicate devices from the Synapse System, Sierra™ System and Altius™ OCT System.
    Performance Testing: (Implied: No new performance testing required if substantial equivalence is based on existing data/rationale).A rationale is provided in place of performance testing for the proposed curved rods. (This indicates that the "performance criteria" were met by demonstrating equivalence rather than actual new testing.)

    2. Sample size used for the test set and the data provenance

    • Sample Size: N/A. No new test set involving patient data or specimens was described. The submission relies on a "rationale" for equivalence to predicate devices.
    • Data Provenance: N/A. No new data from a test set (e.g., country of origin, retrospective/prospective) is presented. The "performance data" section explicitly states "A rationale is provided in place of performance testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. As no new test set data requiring ground truth establishment was presented, this is not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. No test set requiring expert adjudication was described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This device is a spinal fixation orthosis (physical implant system), not an AI-assisted diagnostic or treatment planning tool. Therefore, MRMC studies and AI effect sizes are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is a physical implant, not an algorithm or software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. No new ground truth establishment was described, as the submission focused on comparing the new components to established predicate devices based on design, materials, and existing performance data (implied from the predicates).

    8. The sample size for the training set

    • N/A. No training set for an algorithm was involved; this is a physical medical device.

    9. How the ground truth for the training set was established

    • N/A. No training set or associated ground truth was established for this submission.

    Summary of the Submission's Approach:

    The K133698 submission for the Synapse System (specifically, proposed curved rods) is a substantial equivalence (SE) submission. This means the applicant demonstrated that the new device components (curved rods) are as safe and effective as a legally marketed predicate device.

    Instead of conducting new performance studies, the manufacturer provided a "rationale" arguing that:

    • The materials (Ti-6A1-7Nb (TAN)) are identical to those used in previously cleared components.
    • The technological characteristics (design, intended use, indications for use) of the proposed curved rods are the same or similar to predicate devices (Synapse System, Sierra™ System, Altius™ OCT System).

    This approach is common in 510(k) submissions where modifications to existing devices are minor and do not raise new questions of safety or effectiveness. The FDA's clearance (as indicated in the letter dated January 16, 2014) confirms their agreement that the device is substantially equivalent to predicates.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113593

    Validate with FDA (Live)

    Date Cleared
    2012-04-17

    (134 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When intended for stabilization as an adjunct to fusion of the cervical spine and occipito-cervicothoracic junction (occiput -T3), the Nextgen Altius OCT System is intended for use with allograft or autograft and indicated for: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; deformities or curvatures (i.e., scoliosis, kyphosis and/or lordosis); pseudoarthrosis; trauma, (i.e., fracture or dislocation); atlanto/axial fracture with instability; occipitocervical dislocation; revision of previous cervical spine surgery; and tumors.

    The occipital bone screws are limited to occipital fixation only.

    The use of pedicle screws, is limited to placement in T1-T3 in treating thoracic conditions only. They are not intended to be placed in or treat conditions involving the cervical spine.

    The Nextgen Altius OCT System can also be linked to the Biomet Polaris Systems via transitional rods or using Altius Rod Connectors or Polaris Dominoes. Please refer to the individual system's package insert for a list of indications for use for each system.

    Device Description

    The Nextgen Altius OCT System is an occipito-cervico-thoracic spinal fixation system. This submission is a line extension to Altius OCT System to add CoCr rods and alternate styles of rods, and screws, as well as other system components to the system and to update the indications for use for the system as a whole.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided document:

    This 510(k) summary is for a spinal fixation system, which falls under the category of a medical device and not an AI/ML algorithm. Therefore, many of the requested points related to AI/ML studies (like MRMC, standalone algorithm performance, AI improvement over human readers, training set details) are not applicable to this document. The study described is a series of mechanical tests to demonstrate substantial equivalence to existing predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from FDA Guidance)Reported Device Performance
    Mechanical Test Requirements:
    Static compression bendingMet all mechanical test requirements based on worst-case construct testing
    Static torsion fatigueMet all mechanical test requirements based on worst-case construct testing
    Compression bending fatigueMet all mechanical test requirements based on worst-case construct testing
    Substantial Equivalence to Predicate Devices (design, indications, operational principle, safety, effectiveness)The subject components are substantially equivalent to predicate devices (Altius OCT System, Depuy Mountaineer OCT Spinal System, Vertex Max® & Select™ Reconstruction Systems); did not present any new issues of safety or effectiveness.
    Compliance with ASTM F2706Conducted in accordance with ASTM F2706
    Compliance with ASTM F1717Conducted in accordance with ASTM F1717
    Compliance with ASTM F1798Conducted in accordance with ASTM F1798

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: This information is not explicitly stated in the summary. For mechanical tests, the "sample size" would refer to the number of constructs tested for each configuration.
    • Data Provenance: The data is generated from mechanical laboratory testing of the device components, not from patient data. Therefore, concepts like "country of origin" or "retrospective/prospective" studies are not applicable.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not Applicable. As this is a mechanical testing study for a physical device, expert-established ground truth (in the clinical sense, e.g., for image interpretation) is not relevant. The "ground truth" here is adherence to established mechanical testing standards (ASTM standards) and the comparison of performance metrics against predicate devices.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements in expert clinical assessments or diagnostic interpretations. This study involves objective mechanical performance measurements, not subjective expert reviews.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No. An MRMC study is relevant for evaluating the impact of an AI/ML system on human reader performance. This document describes mechanical testing of a spinal fixation device, which does not involve human readers for diagnostic interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This is not an AI/ML algorithm. It is a physical medical device.

    7. The Type of Ground Truth Used

    • The "ground truth" in this context is based on established mechanical testing standards (ASTM F2706, F1717, F1798) and performance characteristics of legally marketed predicate devices. The goal is to demonstrate that the new device's components meet these engineering and regulatory benchmarks, indicating that they are at least as safe and effective as existing devices.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" in the context of mechanical testing for a spinal fixation system. This concept applies to the development of AI/ML algorithms.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As there is no training set, there is no ground truth established for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072729

    Validate with FDA (Live)

    Device Name
    SIERRA
    Manufacturer
    Date Cleared
    2008-04-17

    (204 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062934, K022048, K033961, K043229

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Sierra spinal system is to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3). The indications for use are as follows:

    • degenerative disc disease (DDD) as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies,
    • spondylolisthesis. .
    • trauma (i.e., fracture or dislocation), .
    • . spinal stenosis.
    • atlantoaxial fracture with instability, .
    • occipitocervical dislocation, .
    • revision of previous cervical spine surgery, and/or
    • . spinal tumor.

    The occipital bone screws are limited to occipital fixation only.

    The use of the polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.

    Device Description

    Rod connectors, crossbars and set screws are being added to the Sierra spinal system. All products are fabricated from titanium alloy. All products are supplied "NON-STERILE" and must be sterilized prior to use.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Sierra™ spinal system, specifically referring to the addition of rod connectors, crossbars, and set screws. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting a full clinical study to prove novel efficacy or safety. Therefore, the provided information does not include details on acceptance criteria, a study proving performance against those criteria, or specific metrics like sensitivity, specificity, or reader improvement with AI.

    Here's an analysis based on the provided text, highlighting the absence of the requested information due to the nature of a 510(k) submission:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as performance metrics in the provided document. For 510(k) submissions, the “acceptance criteria” typically revolve around demonstrating substantial equivalence in terms of functionality, technological characteristics, and safety to a predicate device. This is often achieved through comparative testing (e.g., mechanical, material) and analysis rather than clinical performance metrics.
    • Reported Device Performance: Not reported in terms of clinical outcomes or specific performance metrics (e.g., sensitivity, specificity, accuracy). The submission asserts that the new components have "substantially equivalent technological characteristics to the predicate devices."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable/not provided. This document does not describe a clinical study in the traditional sense involving a "test set" of patients or data for performance evaluation. The "testing" likely involved mechanical and material characterization of the new components against established standards or predicate device specifications.
    • Data Provenance: Not applicable/not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/not provided. Ground truth establishment by medical experts is typically part of clinical performance studies, which are not detailed in this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/not provided. Adjudication methods are relevant for clinical studies that involve expert review of cases, which is not the focus of this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This document concerns a conventional medical device (spinal fixation system components), not an AI-powered diagnostic or assistive tool.
    • Effect size of human reader improvement with AI: Not applicable, as no AI component is described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable, as no algorithm or AI component is described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. The "ground truth" for a 510(k) submission like this would involve demonstrating that the new components meet established engineering and material standards, and perform comparably to predicate devices in mechanical and functional tests. It is not about clinical ground truth derived from patient data.

    8. The sample size for the training set

    • Not applicable. There is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable. There is no mention of an algorithm or AI requiring a training set or ground truth establishment in this context.

    Summary of what the document does provide:

    • Predicate Devices: SeaSpine, Inc.'s Sierra Spinal System (K062934) and Interpore Cross International's Altius OCT System (K022048, K033961 and K043229).
    • Basis for Equivalence: "The overall designs of the rod connectors, crossbars and set screws in this submission are substantially equivalent to the predicate devices." and "The devices in this submission have substantially equivalent technological characteristics to the predicate devices."
    • Materials: All products are fabricated from titanium alloy.
    • Intended Use: Promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) for various conditions (degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, atlantoaxial fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, and spinal tumor).
    • Limitations: Occipital bone screws limited to occipital fixation; polyaxial screws limited to T1-T3 for thoracic conditions only, not for cervical spine.

    In conclusion, this document is a 510(k) summary for a conventional medical device modification, which focuses on demonstrating substantial equivalence to existing devices based on technological characteristics and intended use, rather than presenting clinical study results with performance metrics or AI algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080526

    Validate with FDA (Live)

    Device Name
    SIERRA SYSTEM
    Manufacturer
    Date Cleared
    2008-04-16

    (50 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062934, K022048, K033961, K043229

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Sierra spinal system is to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3). The indications for use are as follows:

    • degenerative disc disease (DDD) as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies,
    • spondylolisthesis. .
    • trauma (i.e., fracture or dislocation), .
    • spinal stenosis, .
    • atlantoaxial fracture with instability, .
    • occipitocervical dislocation, .
    • revision of previous cervical spine surgery, and/or .
    • . spinal tumor.
      The occipital bone screws are limited to occipital fixation only.
      The use of the polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.
    Device Description

    A new occipital plate design is being added to the Sierra spinal system. All products are fabricated from titanium alloy and cobalt alloy. All products are supplied "NON-STERILE" and must be sterilized prior to use.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Sierra™ OCT Spinal System, specifically for the addition of a new occipital plate design. It is a submission to the FDA (Food and Drug Administration) for regulatory clearance. Since this is a submission for a Special 510(k) and the information provided points to substantial equivalence with existing predicate devices, the focus is on design and material equivalency, rather than clinical performance studies involving a test set, ground truth, or human-in-the-loop assessments as would be done for AI/ML-driven devices.

    Therefore, many of the requested categories for acceptance criteria and study details are not applicable to this type of submission. The "acceptance criteria" here refer to satisfying FDA's requirements for substantial equivalence based on design, materials, and intended use.

    Here's the breakdown of the information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance/Characteristics
    Overall Design Equivalence to Predicate Devices (Sierra Spinal System K062934 and Interpore Cross International's Altius OCT System K022048, K033961, K043229)"The overall design of the occipital plates in this submission is substantially equivalent to the predicate devices."
    Materials Equivalence"All products are fabricated from titanium alloy and cobalt alloy." (Implied to be equivalent to predicate device materials, as no material change is highlighted as a differentiator)
    Intended Use Equivalence"The statements of intended use are identical." The intended use is to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) for various conditions (DDD, spondylolisthesis, trauma, spinal stenosis, atlantoaxial fracture with instability, occipitocervical dislocation, revision, spinal tumor). Specific limitations for occipital bone screws (occipital fixation only) and polyaxial screws (T1-T3 only, not cervical spine) are also identical.
    Technological Characteristics Equivalence"The addition of the new devices has not altered the fundamental scientific technology of the previously cleared system. The devices in this submission have substantially equivalent technological characteristics to the predicate devices."
    Sterilization Status Equivalence"All products are supplied "NON-STERILE" and must be sterilized prior to use." (Implied to be equivalent to predicate devices, as this is a common practice for such implants.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable (N/A). This submission is for a medical device (occipital plate) and relies on substantial equivalence to predicate devices, not clinical performance data from a test set of patients in the way an AI/ML device would. No patient data or test set in the traditional sense was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. Ground truth establishment is not relevant for this type of submission, which focuses on engineering and design equivalency.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. No test set or human adjudication process as described is relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is not an AI/ML device, so no MRMC study or AI assistance evaluation was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. The "ground truth" equivalent in this context is the regulatory acceptance of the predicate devices based on their established safety and effectiveness. The new device is deemed safe and effective because its characteristics are substantially equivalent to those already cleared devices.

    8. The sample size for the training set

    • N/A. No training set is relevant for this device.

    9. How the ground truth for the training set was established

    • N/A. No ground truth for a training set was established.