510(k) Data Aggregation

    K Number
    K201305

    Validate with FDA (Live)

    Date Cleared
    2021-09-03

    (476 days)

    Product Code
    Regulation Number
    868.5830
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K043140, K043582

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    · Evacuate air and/or fluid from the chest cavity or mediastinum.
    · Help re-establish lung expansion and restore breathing dynamics.
    Chest Drain Auto transfusion (ATS)
    To facilitate collection of autologous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in postoperative and trauma blood loss management.

    Device Description

    Atrium's Auto transfusion (ATS) Chest Drains are sterile, single use, disposable devices that mimic a traditional "3 Bottle Systems".

    A closed thoracic drainage system (chest drain and catheter together) are used to restore the chest to a more normalized condition after surgery, trauma, or spontaneous need for air and/or fluid to be removed. The Auto transfusion (ATS) features of the drains facilitate postoperative collection and reinfusion of autologous blood from the patient's pleural cavity or mediastinal area.

    AI/ML Overview

    This document describes a 510(k) premarket notification for Atrium Medical Corporation's Auto transfusion (ATS) Chest Drains, specifically the Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, and Express Dry Seal Chest Drain. The filing argues for substantial equivalence to previously cleared predicate devices (K043140 and K043582).

    Here's an analysis of the acceptance criteria and study information provided (or absence thereof):

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of acceptance criteria with specific quantitative performance metrics. This 510(k) submission is focused on demonstrating substantial equivalence to predicate devices, rather than meeting novel performance criteria through new clinical trials.

    The performance is implicitly described as being equivalent to the predicate devices and that the device "performs as well as the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable. No dedicated "test set" in the context of clinical performance evaluation is mentioned for the modified device. The document states, "There were no clinical studies of the modified device."
    • Data Provenance: Not applicable for a new clinical study. The submission relies on the established safety and performance of the predicate devices and extensive non-clinical testing of the modified devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. No ground truth establishment by experts for a new clinical test set is mentioned.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or adjudication method is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is an auto transfusion chest drain, which is a physical medical device, not an AI-assisted diagnostic or therapeutic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    Not applicable in the context of a new clinical performance study. The "ground truth" for this 510(k) submission is the established safety and effectiveness of the predicate devices (K043140 and K043582), as well as compliance with relevant voluntary standards and successful completion of non-clinical tests.

    8. The Sample Size for the Training Set

    Not applicable. The device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. The device is not an AI/ML algorithm.


    Summary of Acceptance Criteria and "Study" (Non-Clinical) Proving Equivalence:

    The primary "acceptance criterion" for this 510(k) submission is substantial equivalence to existing legally marketed predicate devices. The "study" that proves this involves a comprehensive non-clinical assessment, outlined as follows:

    Acceptance Criteria (Implied for Substantial Equivalence):

    • Same Intended Use: The modified devices must have the same indications for use as the predicate devices.
    • Similar Technological Characteristics: The modified devices must operate on the same fundamental scientific technology and have similar features. Any differences in technological characteristics must not raise new questions of safety or effectiveness.
    • Similar Materials: The major materials of construction must be the same or demonstrably equivalent in terms of safety and biocompatibility.
    • Performance: The modified devices must perform as well as the predicate devices.
    • Safety: The risks associated with the intended use must be acceptable and compatible with a high level of protection of health and safety.

    "Study" (Non-Clinical Testing) Proving Equivalence:

    The document states that Atrium Medical's development process required the completion of the following non-clinical activities:

    • Specification Review: Ensuring the designs meet predefined criteria.
    • Performance Testing: Verifying the functional aspects of the device.
    • Biocompatibility Testing: Evaluating the device's interaction with biological systems (e.g., ISO 10993 series).
    • Sterility Testing: Confirming the device is sterile and maintains sterility.
    • Stability Testing: Assessing the device's ability to maintain its properties over time (shelf life).
    • Design Validation: Ensuring the device meets user needs and intended uses.

    These non-clinical tests, along with compliance with voluntary standards (Section 9, though not provided in the excerpt), are the basis for demonstrating that the modified devices are safe and effective and perform as well as the predicate devices, thereby supporting the claim of substantial equivalence. The submission also relies on the low complaint rates and reportable events of the predicate devices over the last 5 years as evidence of acceptable clinical safety and performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130210

    Validate with FDA (Live)

    Device Name
    MEDELA THOPAZ
    Manufacturer
    Date Cleared
    2013-03-15

    (45 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K043140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medela® THOPAZ Suction Pump is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. The Medela® THOPAZ Suction Pump is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural effusion, pleural empyema or other related conditions. The Medela® THOPAZ Suction Pump is intended for use on patients in appropriate care settings.

    Device Description

    Powered Suction Pump

    AI/ML Overview

    The provided text is a 510(k) summary for the Medela® THOPAZ Suction Pump. It asserts substantial equivalence to predicate devices but does not contain information about specific acceptance criteria or a study proving that the device meets those criteria with performance metrics.

    Instead, it states:

    • "The Medela® THOPAZ Suction Pump is equipped with the identical technology like other marketed devices. These technological features do not affect safety and effectiveness of the device or the application (pleural and mediastinal drainage)."
    • "There are no differences in performance or technology which significantly affect the safety and effectiveness of the device or the application (pleural and mediastinal drainage)."
    • "The Medela® THOPAZ suction pump has the identical intended uses and, where applicable, the identical technological characteristics and performance data as the predicate devices."

    Therefore, I cannot populate the requested table or answer most of the questions because the document explicitly states the device is identical in technology and performance to predicate devices (K080212 and K043140) and does not present new performance data or studies to demonstrate meeting acceptance criteria.

    The submission is based on the premise that since it's identical to already-cleared devices, new performance studies are not required.

    Given the information provided, the following elements cannot be extracted:

    • A table of acceptance criteria and the reported device performance: Not provided; the device is stated to be identical to predicates.
    • Sample size used for the test set and the data provenance: Not applicable as no new performance study is detailed.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used: Not applicable.
    • The sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1