510(k) Data Aggregation

    K Number
    K212563

    Validate with FDA (Live)

    Date Cleared
    2022-05-23

    (283 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K893794, K042769, K063237, K140247, K200175, K190510

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CalApex is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

    Calplus is a temporary or permanent root canal filling material for use following pulpectomy, or for apexogenesis or apexification.

    Cerafill RCS is a MTA (mineral trioxide aggregate) based root canal sealer that provides complete and permanent sealing of root canals. It can be used with or without root canal obturation materials.

    Endoseal is used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials.

    Nanoseal S is indicated in patients for permanent obturations of root canal after vital extirpation or after treatment of pulpal gangrene and temporary filling.

    Zical Ultra Paste is used for permanent obturation of the root canal space with the aid of obturating points.

    Zical Ultra Powder/Liquid is used for permanent obturation of the root canal space with the aid of obturating points.

    Device Description

    CalApex is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal filling material. It is a two part, base/catalyst – paste/paste system that is mixed in equal portions. The mixture is then carried to the root canal with endodontic obturation points or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe.

    Calplus can be used as a temporary or permanent root canal filling material.

    Cerafill RCS is an endodontic sealer based on MTA, providing a biocompatible and effective root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete filling of the entire root canal including accessory and lateral canals. The product is eugenol-free and will not impede adhesion inside the root canal.

    Endoseal is a Zinc Oxide/Eugenol root canal sealant. The product is made up of two separate components, a powder base and a liquid catalyst, to form the final device.

    Nanoseal S is a permanent root canal filling material which is silicone based (polydimethylsiloxane). It is cold flowable and is in a dual barrel cartridge.

    Zical Ultra Paste is a two-part, base/accelerator, paste/paste system that is mixed in equal portions. The mixture is carried to the root canal with endodontic obturation points. The product comes in a dual barrel cartridge.

    Zical Ultra Powder/Liquid is a two-part, powder/liquid system for permanent filling of root canals. The mixture is carried to the root canal with endodontic obturation points.

    AI/ML Overview

    The document describes the acceptance criteria and performance of several root canal sealers developed by Prevest Denpro Limited. The study involved a series of bench tests conducted according to ISO standards to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test ParameterAcceptance Criteria (ISO Standard & Specific Values)Reported Device Performance (Summary for all tested sealers)
    AppearancePass/Fail (based on acceptable appearance)All passed
    FlowBased on ISO 6876:2012, Section 5.2
    - CalApex, Cerafill RCS, Endoseal, Nanoseal S, Zical Ultra Paste, Zical Ultra Powder/LiquidNot less than 17mm> 17mm
    - Calplus25-28mm25-28mm
    Film ThicknessISO 6876:2012, Section 5.5, not more than 50µm<50µm
    Water SolubilityISO 6876:2012, Section 5.6, not more than 3% by massCalApex: 1.31%, Cerafill RCS: 1.61%, Endoseal: 1.09%, Nanoseal S: 1.06%, Zical Ultra Paste: 1.31%, Zical Ultra Powder/Liquid: 0.75% (All within <3% limit)
    Working TimeBased on ISO 6876:2012, Section 5.3
    - CalApex16-18min16-18min
    - Cerafill RCS, Endoseal, Zical Ultra Paste, Zical Ultra Powder/Liquid25-30min25-30min
    - Nanoseal S8-10min8-10min
    Setting TimeBased on ISO 6876:2012, Section 5.4
    - CalApex20-40min20-40min
    - Cerafill RCS24 hoursWithin 24 hours
    - Endoseal, Zical Ultra Paste, Zical Ultra Powder/Liquid45-60min45-60min
    - Nanoseal S10-15min10-15min
    DisintegrationISO 6876:2012, Section 4.3.5, no evidence of disintegrationNo evidence of disintegration (All passed)
    Radio-opacityISO 13116:2014, above or equal to 3 mm Al6 mm Al (All tested sealers, which is above 3mm Al)
    BiocompatibilityAssessment according to ISO 10993Yes (assessment provided for all subject devices)
    Shelf LifeSame as bench testing pass criteria (ISO 6876:2012)CalApex: 2 years, Others: 3 years (All met criteria)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify the exact sample sizes used for each bench test. It states that the tests were conducted "according to protocols based on ISO 6876:2012 and tested for radio-opacity according to a protocol based on ISO 13116:2014". The data provenance is implied to be from the manufacturer's testing, but no specific country of origin for the data or whether it was retrospective or prospective is mentioned. The manufacturer, Prevest Denpro Limited, is located in India.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This type of information is not applicable to this study. The tests performed are benchtop physical and chemical property evaluations rather than evaluations requiring expert human interpretation of medical images or patient outcomes. The "ground truth" for these tests is defined by the objective measurement protocols and acceptance criteria of the specified ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. The study involves objective bench testing, not subjective assessments requiring adjudication by multiple experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This study focuses on the physical and chemical properties of root canal sealers and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This study does not involve an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth for this study is defined by the objective measurement standards and acceptance criteria set forth by international ISO standards (ISO 6876:2012 for most physical properties and ISO 13116:2014 for radio-opacity). For biocompatibility, the ground truth is established by the assessment according to ISO 10993.

    8. The sample size for the training set:

    Not applicable. This study involves bench testing of medical devices, not training of machine learning models.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212475

    Validate with FDA (Live)

    Date Cleared
    2022-05-13

    (280 days)

    Product Code
    Regulation Number
    872.3250
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K111668, K123265, K190091, K042769, K922721

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Apacal ART and is suitable for several indications including:

    • · Liner to be applied to the interior of a prepared cavity prior to the insertion of restorative materials.
    • · Indirect pulp capping or management of deep caries lesions, or
    • · Direct pulp capping

    CAL-LC is indicated for use as a cavity liner and pulp capping material. It is self-adhering and highly filled.

    Calcigel is a calcium hydroxide paste that has a creamy consistency and is suitable for several indications including;

    • · Indirect pulp capping or management of deep caries lesions, or
      · Direct pulp capping

    CalUltra is a hard-setting calcium hydroxide cavity liner and pulp capping agent to be used in conjunction with all permanent restorative techniques.

    Device Description

    Apacal ART is a light cured single-component paste reinforced with tricalcium phosphate and hydroxyapatite. It is sold in a syringe with accompanying applicator tips.

    CAL-LC is a highly filled cavity liner and pulp capping material. It is sold in a syringe with accompanying applicator tips.

    Calcigel is a creamy, ready -to -use calcium hydroxide paste with a pH above 11. It is syringe with accompanying applicator tips.

    CalUltra is a hard-setting calcium hydroxide cavity liner and pulp capping agent. CalUltra is a two-part base/catalyst paste/paste system. It is sold in a tubes with accompanying spatula and mixing pad.

    AI/ML Overview

    The provided text describes the substantial equivalence of Prevest Denpro Cavity Liners to predicate devices, focusing on physical parameters and material composition rather than AI performance. Therefore, many of the requested categories related to AI studies and expert review are not applicable to this document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All tests results met the criteria in standards." The acceptance criteria are implicit in the ISO and USP standards referenced for each test. The performance is reported as meeting these standards.

    Device NameTestStandard Reference (Implicit Acceptance Criteria)Reported Device Performance
    Apacal ART & Cal LCAppearanceISO 9917-2:2017, ISO 4049:2019, ISO 13116:2014Met standards
    Flexural StrengthISO 9917-2:2017 (e.g., for Apacal ART: 99.1 MPa; for CAL-LC: 96.3 MPa)Met standards
    Depth of CureISO 4049:2019Met standards
    Water SorptionISO 4049:2019 (e.g., for Apacal ART: 13.6 µg/mm³; for CAL-LC: Data not specified but implied to meet standards)Met standards
    Water SolubilityISO 4049:2019 (e.g., for Apacal ART: 4.5 µg/mm³; for CAL-LC: Data not specified but implied to meet standards)Met standards
    Radio-opacityISO 13116:2014Met standards
    CalcigelAppearanceISO 6876:2012Met standards
    FlowISO 6876:2012Met standards
    pHUSP 971 (e.g., 11-13)Met standards
    Radio-opacityISO 13116:2014Met standards
    CalUltraAppearanceISO 9917-1:2017Met standards
    Compressive StrengthISO 9917-1:2017Met standards
    Water SolubilityISO 6876:2012 (e.g., 3.8%)Met standards
    Setting TimeISO 6876:2012 (e.g., 1-3 minutes)Met standards
    Radio-opacityISO 13116:2014Met standards
    pHUSP 971 (e.g., 9.2)Met standards
    All DevicesShelf LifeRelevant protocols (not specified in detail, but implied to meet standards)3 years

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for the testing of these physical and chemical properties. The data provenance is presumed to be from the manufacturer's testing conducted in India (Prevest Denpro Limited, Jammu, India). The testing is described as being performed "according to protocols based on" relevant ISO and USP standards, indicating manufacturer-conducted in vitro/bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device clearance based on material properties and performance against established international standards, not an AI or diagnostic device requiring expert ground truth for image interpretation or similar tasks.

    4. Adjudication method for the test set

    Not applicable. This type of adjudication is typically used in clinical or image-based studies. The "adjudication" here is the assessment against specified ISO/USP standard requirements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a dental material, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a dental material, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is based on the objective measurements against established international and national standards (ISO and USP) for material properties (e.g., flexural strength, depth of cure, water sorption/solubility, pH, setting time, radio-opacity).

    8. The sample size for the training set

    Not applicable. There is no training set mentioned, as this is not an AI or machine learning device.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190510

    Validate with FDA (Live)

    Device Name
    GuttaSil
    Date Cleared
    2019-05-29

    (89 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K042769, K171449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GuttaSil is a material for permanent obturation of root canals after treatment of pulpal gangrene and temporary filling of the canal.

    Device Description

    GuttaSil, a product in the form of a paste which combines gutta-percha with a sealer. The gutta-percha powder is mixed in a matrix of polyvinylsiloxane. It is convenient since sealing and obturation are simultaneously possible without heating and the root canal can be filled in a quick manner. This device contains a syringe, auto-mix& Endo tips, spatula, and mixing pads.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the device meets those criteria for the GuttaSil, a root canal filling material.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Standard/Test)Device Performance
    Mechanical/PhysicalAppearance (ISO 6876:2012)Conformed to standard
    Weight (ISO 6876:2012)Conformed to standard
    Packaging (ISO 6876:2012)Conformed to standard
    Flow (ISO 6876:2012)Conformed to standard
    Working Time (ISO 6876:2012)Conformed to standard
    Setting Time (ISO 6876:2012)Conformed to standard
    Film Thickness (ISO 6876:2012)Conformed to standard
    Solubility (ISO 6876:2012)Conformed to standard
    Radio-opacity (ISO 6876:2012)Conformed to standard
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Biocompatible
    Sensitization (ISO 10993-10:2010)Biocompatible
    Irritation (ISO 10993-10:2010)Biocompatible
    Systemic Toxicity (ISO 10993-11:2006)Biocompatible
    Overview (ISO 10993-1:2009)Biocompatible
    Shelf LifeISO 6876 tests (Appearance, Package, Flow, Setting Time, Solubility) after agingConformed to standard for 2 years

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for each specific test within the non-clinical testing. It only generally refers to "the following testing was conducted." The provenance of the data is not specified (e.g., country of origin, retrospective or prospective), but it is implied to be laboratory testing conducted by or for Meta Biomed Co., Ltd.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The tests conducted are non-clinical, primarily physical, chemical, and biocompatibility tests based on ISO standards. These types of tests typically do not involve human expert adjudication of results in the way clinical performance studies do.

    4. Adjudication Method for the Test Set

    This information is not applicable as the tests are non-clinical performance and biocompatibility tests against quantitative ISO standards, not evaluations requiring human rater adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance and biocompatibility testing. The device is a material for root canal filling, not an AI-assisted diagnostic or treatment planning tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not Applicable. This device is a material (gutta-percha with a sealer) used in a medical procedure, not an algorithm or software. Therefore, the concept of "standalone algorithm performance" is not relevant.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests, the "ground truth" is defined by the specifications and acceptable ranges outlined in the ISO 6876:2012 standard for root canal sealing materials.

    For the biocompatibility tests, the "ground truth" is adherence to the biocompatibility criteria specified in the relevant ISO 10993 series standards.

    8. The Sample Size for the Training Set

    Not Applicable. This device is a physical material, not an AI algorithm or model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132123

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-03-20

    (253 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K032603, K042769, K960548

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TRC-Pex: The TRC-Pex is a temporary root canal sealer for use in the treatment of root canals, following a pulpectomy or for apexegenesis or apexification.
    Radic-Sealer: The Radic-Sealer is a root canal sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). The Radic-Sealer is intended for use by qualified healthcare personnel trained in its use.

    Device Description

    TRC-Pex: TRC-Pex is an immediately available pre-mixed root canal filling material based on Calcium Hydroxide and lodoform. This pre-mixed temporary filling material maintains a constant flow, which makes it easy to inject into the canal. Depending on apicoectomy or root canal therapies, it can be used as a temporary root canal filling material. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, a rubber, a protective cap, an indo stop and a holder for direction control of the tip.
    Radic-Sealer: Radic-Sealer is a root canal sealing & filling material based on epoxy-amine resin. Radic-Sealer is contained in a dual syringe which makes it easy to mix and inject. It makes permanent obturation of root canals using adhesion properties between root canal walls and the root canal filling materials. The base and catalyst are contained separately in the chambers of two separate plastic syringes. The product consists of a mixing tip with a mixing joint, a protective cap, a mixing pad, and a plunger.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the TRC-Pex and Radic-Sealer devices:

    This document is a 510(k) summary for medical devices, which focuses on demonstrating substantial equivalence to predicate devices rather than proving a new device's safety and effectiveness from scratch through extensive clinical trials. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed within the context of showing equivalence to existing, legally marketed devices.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the standards and properties of the predicate devices. The study aims to show that the new devices meet the established performance parameters defined by these standards.

    Acceptance Criteria (Based on ISO Standards)Reported Device Performance (TRC-Pex & Radic-Sealer)
    ISO 6876: RadiopacityConfirmed to protocols and recommended values
    ISO 6876: LiquidityConfirmed to protocols and recommended values
    ISO 6876: Setting timeConfirmed to protocols and recommended values
    ISO 6876: Solubility testConfirmed to protocols and recommended values
    ISO 6876: Film ThicknessConfirmed to protocols and recommended values
    ISO 6876: FlowabilityConfirmed to protocols and recommended values
    ISO 10993-5: CytotoxicitySimilar biocompatibility properties (comparative test with predicate)
    ISO 10993-10: Oral Mucous Irritation & SensitizationBiocompatibility data provided
    ISO 10993-11: Short-term systemic toxicity (Oral)Biocompatibility data provided
    Other Bench Testing: Shelf lifeConfirmed
    Other Bench Testing: AppearanceConfirmed
    Other Bench Testing: Volume/weight specConfirmed
    Other Bench Testing: Packaging testsConfirmed

    Study Proving Acceptance Criteria:

    The study conducted was a series of non-clinical performance tests and biocompatibility tests. The document states:

    • "Performance testing that confirmed to the protocols and recommended values described in ISO 6876 was performed."
    • "Biocompatibility data were provided to demonstrate substantial equivalence to the predicate devices."
    • "Side-by-Side cytotoxicity testing was performed that compared the TRC-Pex and a predicate device, and the test results showed that they have similar biocompatibility properties."
    • "The bench and biocompatibility tests provided evidence that the chemical and physical properties of TRC-Pex and Radic-Sealer are substantially equivalent to the predicate devices."

    Additional Information Not Present in the Document:

    The provided text describes a 510(k) submission for dental materials (root canal filling and sealing materials). For this type of device, the performance evaluation primarily relies on bench testing and biocompatibility studies as per ISO standards, which is what the document reports. It does not involve studies typically associated with AI/software devices or diagnostic image analysis. Therefore, many of the requested points below are not applicable or cannot be answered from the provided text.


    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for each test, but standard protocols for material testing (e.g., ISO 6876, ISO 10993 series) involve a specified number of samples for each property being evaluated (e.g., usually 3-5 samples per test condition for material science, but this is not detailed here).
    • Data Provenance: The tests were conducted by KM Corporation, likely in a laboratory setting. No country of origin for "data" in the sense of patient data is applicable. These are material properties tests.
    • Retrospective or Prospective: Not applicable. These are laboratory-based material characterization tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For material properties and biocompatibility, "ground truth" is established by adherence to standardized testing methods (e.g., ISO standards) and instrument measurements, not expert consensus in the human interpretation sense.

    4. Adjudication method for the test set:

    • Not applicable. Material property tests are objective measurements, not subject to human adjudication or interpretation disputes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an AI/software device.

    7. The type of ground truth used:

    • Bench Test Ground Truth: Measured physical and chemical properties according to ISO standards.
    • Biocompatibility Ground Truth: Cellular responses (e.g., cytotoxicity) and tissue reactions according to ISO 10993 standards and comparison to established predicate device performance.

    8. The sample size for the training set:

    • Not applicable. These materials are not "trained" in the machine learning sense. The "training" might refer to the development and refinement process of the material formulations, which is not detailed here.

    9. How the ground truth for the training set was established:

    • Not applicable for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1