Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K073400

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-12-20

    (16 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K041517, K052738

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VariCath Peripheral Infusion Catheter is intended for the infusion of physician specified fluids, including thrombolytics, into the peripheral vasculature.

    Device Description

    The VeinRx VariCath Peripheral Infusion Catheter is comprised of a distal occlusion balloon, a variable infusion length catheter body with infusion holes, and a proximal trifurcated Luer connection hub. The trifurcated hub allows connection to three main systems of the device. The VariCath is packaged inside a protective tube mounted on a card and placed in a sealed protective pouch.

    AI/ML Overview

    This document is a 510(k) summary for the VariCath Peripheral Infusion Catheter. 510(k) summaries are primarily focused on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study results or performance metrics in the same way a PMA (Premarket Approval) would.

    Based on the provided text, here's an analysis against your requested information:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The submission is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to a legally marketed predicate device (InfusionCath Peripheral Infusion Catheter). It does not typically require the establishment of new acceptance criteria for clinical performance or present a detailed study comparing numerical performance metrics against such criteria.

    The document states:

    • "The VariCath Peripheral Infusion Catheter was tested in the same manner as the InfusionCath Peripheral Infusion Catheter."
    • "All components, subassemblies, and/or full devices met the required specifications for the completed tests."

    This indicates that internal performance testing was conducted and met specifications, but the specifics of those specifications (acceptance criteria) and the numerical results (reported device performance) are not included in this public summary.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission references "testing" but does not detail a clinical test set, its sample size, or its provenance. This level of detail is usually found in a full 510(k) submission, not the summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. The device is an infusion catheter, not an AI diagnostic tool that requires ground truth established by experts for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. The device is a medical catheter and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. The device is a medical catheter and does not involve an algorithm with standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable/not provided. The device is a medical catheter and does not involve diagnostic ground truth.

    8. The sample size for the training set

    This information is not applicable/not provided. The device is a medical catheter and does not involve an AI training set.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. The device is a medical catheter and does not involve an AI training set or its associated ground truth.

    In summary:

    This 510(k) summary indicates that the VariCath Peripheral Infusion Catheter was found substantially equivalent to its predicate, the InfusionCath Peripheral Infusion Catheter. The basis for this equivalence is stated to be similar indications for use, function, manufacturing methods, and materials. The testing referenced is primarily for device specifications (mechanical, material, etc.), not a clinical performance study with human subjects, expert adjudication, or AI performance metrics. Therefore, most of the detailed questions regarding acceptance criteria, study design, and ground truth for AI-related device evaluation are not relevant or answered in this type of submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052738

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-06

    (67 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K041517

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VeinRx Infusion Catheter is intended for the infusion of physician specified fluids, including thrombolytics, into the peripheral vasculature.

    Device Description

    The VeinRx InfusionCath Peripheral Infusion Catheter is comprised of a distal occlusion balloon, a fixed infusion length catheter body with infusion holes, and a proximal trifurcated Luer connection hub. The trifurcated hub allows connection to three main systems of the device. The distal black marker indicates the proximal edge of the infusion length. The proximal black marker is used to properly align the device with the recommended introducer. The InfusionCath is packaged inside a protective tube mounted on a card and placed in a sealed protective pouch.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    Important Note: The provided document is a 510(k) summary for a medical device (InfusionCath Peripheral Infusion Catheter) seeking substantial equivalence to a predicate device. As such, it does not contain the detailed study results, acceptance criteria for performance metrics, or information about AI/ML models typically found in submissions for novel AI/ML-powered devices. The primary goal of a 510(k) is to demonstrate that a new device is "as safe and effective as" a legally marketed predicate device, often by showing it meets similar performance specifications.


    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, specific acceptance criteria and reported device performance metrics in the context of quantitative measures (e.g., accuracy, sensitivity, specificity, F1-score) are NOT available. The document states:

    • "The modified InfusionCath Peripheral Infusion Catheter was tested in the same manner as the original InfusionCath Peripheral Infusion Catheter. All components, subassemblies, and/or full devices met the required specifications for the completed tests."

    This indicates that internal design and performance specifications were met, but the specific numerical values of these "specifications" are not detailed in this summary. The nature of this device (a catheter) suggests performance criteria would likely revolve around mechanical integrity, material compatibility, fluid flow rates, and safety, not diagnostic accuracy.

    Acceptance Criteria CategoryAcceptance Criteria (Specifics not provided in the document)Reported Device Performance (Specifics not provided in the document)
    Functional/MechanicalAll required specifications for components, subassemblies, and/or full devices.Met all required specifications. (Implies satisfactory performance against internal design specifications)
    Biocompatibility/MaterialMaterials suitable for intended use.Materials used in numerous previously cleared products. (Implies suitability and established safety profile)
    SafetyN/A (General safety is implied)N/A (General safety is implied)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document only mentions "Testing Summary" and that "All components, subassemblies, and/or full devices met the required specifications for the completed tests." It does not provide details on how many units or components were tested.
    • Data Provenance: Not applicable in the context of clinical data for an AI/ML device. For this physical medical device, the "data" would be results from engineering, bench, and potentially pre-clinical (animal) or limited human clinical compatibility testing. The country of origin of this testing is not specified, but it would have been conducted by or on behalf of VeinRx, Inc., a US-based company. The statement "The modified InfusionCath Peripheral Infusion Catheter was tested in the same manner as the original InfusionCath Peripheral Infusion Catheter" suggests a comparison to an existing product's performance profile rather than new, large-scale clinical trials.
    • Retrospective or Prospective: Not applicable in the context of clinical data for AI/ML. The testing would have been conducted for the purpose of the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. This device is a physical catheter, not a diagnostic device relying on expert interpretation of data to establish ground truth for a test set. Its performance is assessed through engineering testing, material characterization, and potentially pre-clinical studies, not clinical expert consensus for diagnostic accuracy.

    4. Adjudication Method for the Test Set

    • Not applicable. As above, this is not relevant for a physical medical device like a catheter.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No. This is a physical medical device (catheter) and does not incorporate AI. Therefore, an MRMC study and analysis of AI assistance is not relevant or applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No. This device is a physical medical device (catheter) and does not incorporate an AI algorithm. Therefore, standalone algorithm performance is not relevant or applicable.

    7. The Type of Ground Truth Used

    • Not applicable in the AI/ML sense. The "ground truth" for this device would be established by engineering specifications, material standards (e.g., ISO, ASTM, USP), and established biological compatibility requirements. For example, a "ground truth" for a material test would be whether a material passes a specific cytotoxicity test or particulate matter release limit.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not use AI/ML, and therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set (no AI/ML), this question is not relevant.

    In summary, the provided document is a 510(k) Pre-Market Notification for a conventional medical device (a catheter). It focuses on demonstrating substantial equivalence to a predicate device through manufacturing similarities and meeting established design specifications. It does not involve AI/ML technology, and therefore, many of the questions asked, which are relevant to AI/ML device submissions, are not applicable to this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1