Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K212761

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-17

    (259 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K033273, K182468, K151264, K062148, K201219, K193122

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Painful hip arthritis refractory to medical management resulting from post-traumatic arthritis, osteoarthritis, or rheumatoid arthritis.

    2. Painful femoral head. cup arthroplasty, or bi-polar or universal type femoral head replacement.

    3. Cases where more conventional arthroplasty techniques or arthrodesis are contraindicated because of a difficult clinical management problem, age, sex, occupation, or height of the patient

    The Biocore9 Acetabular Cup System components are intended for use in total hip arthroplasty in primary or revision surgery of skeletally mature patients. The Biocore9 Acetabular Cup System shells and liners are single use implants intended for cemented or cementless arthroplasty.

    Device Description

    The subject Biocore9 Acetabular Cup System includes acetabular shells, liners, and acetabular bone screws. Components are intended to replace the articular surface of the acetabular socket in the patient's hip joint. It is intended for the reconstruction of painful and/or severely disabled hip joints resulting from osteoarthritis or rheumatoid arthritis for patients who would be candidates for total hip procedure, whose acetabular socket has not been excessively damaged by disease or trauma and where damage is primarily associated with the articular surface damage. Acetabular components are available in nine sizes with available outside diameters from 50 mm to 66 mm in 2 mm increments. Bearing components are available in fourteen sizes from 39 mm to 52 mm ID in 1 mm increments. Screw components are available in 6.5 mm diameter in five sizes from 15 mm to 50 mm lengths.

    Biocore9 Cancellous Bone screws are manufactured from Ti6A14V alloy (ASTM F136) with low profile screw heads designed to fit the Acetabular shells with nominal diameter of 6.5 mm.

    The acetabular shell components are manufactured from Ti-6A1-4V alloy (ASTM F136) with a single radius spherical outer geometry coated with three layers of Commercially Pure (CP)-Ti (ASTM F67) spherical bead porous coating allowing for cemented or cementless fixation and an anatomical shaped rim to limit impingement with the femoral side of the joint or the psoas muscle. A crescent peripheral groove provides for assembly interlock and three rim tabs provide anti rotational interlock with the bearing liner. The acetabular shells are available in two configurations: one with no screw holes and a second with five screw holes for supplemental bone screw fixation. All surfaces are coated with Titanium Nitride (TiN) thin film ceramic coating. Porous structured acetabular shells are intended for cemented fixation.

    The acetabular cup bearing liner components are manufactured from UHMWPe (ASTM F648), GUR 1020 Highly cross linked UHMWPe (75KGy MRad /post irradiation annealed) which locks into the acetabular cup shell with ten flexible lip tabs and has three rotation resisting tabs. The liners have inner diameters (ID) intended for use with modular, unipolar, self-centering (bipolar), metallic or ceramic femoral heads within the 39-52 mm OD range and articulate with a femoral head of an appropriate corresponding diameter.

    The components of the subject device are compatible with the BioPro PSL Hip System femoral head and stems.

    The 510k numbers for the compatible BioPro devices are listed below:

    • BioPro PC Femoral Hip Component (K882146), O
    • O BioPro Hemi-Endo Modular Head (K895886),
    • BioPro Ziralloy Modular Femoral Head (K912641 & K925682), O
    • BioPro PSL Total Hip Replacement System (K922500), O
    • BioPro Hemi-Endo Modular Ceramic Head (K954768), O
    • BioPro BiPolar Head (K100761) O
    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Biocore9 Acetabular Cup System." As such, it does not contain the type of detailed study data and acceptance criteria information (such as AI model performance, expert ground truth establishment, or mMRMC studies) that would be found in a submission for an AI-powered device or diagnostic tool.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical performance testing of the physical components (e.g., material strength, wear simulation, screw testing). It does not describe an AI/ML component or a study to prove AI model performance.

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML device, as those details are not present in the provided text.

    The information provided is relevant to the safety and effectiveness of the physical hip implant components, not to the performance of an algorithm or AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180973

    Validate with FDA (Live)

    Date Cleared
    2018-09-28

    (168 days)

    Product Code
    Regulation Number
    888.3350
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K000306, K033273

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components.

    Total hip replacement is indicated for the following conditions:

    • Non-inflammatory degenerative joint diseases including osteoarthritis and avascular necrosis.
    • Rheumatoid arthritis.
    • Congenital hip dysplasia.
    • Acute traumatic fracture of the femoral head or neck.
    • Certain cases of Ankylosis.
    • Dislocation of the hip.
    • Correction of functional deformity.
    • Revision of failed joint reconstruction or treatment.
    • Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur.

    NOTE:

    • The Modular Shell and Uncemented Stem are intended for press-fit, uncemented use only.
    • The Cemented stem is intended for cemented use only.
    Device Description

    The Libertas™ Hip Replacement System consists of the following components:

    • Acetabular Cup System
      • Modular Shell
      • Modular Liner HXLPE
      • Cobalt Chromium alloy Modular Femoral Head
      • Biolox® delta Modular Femoral Head
    • Femoral Stem
      • Uncemented Femoral Stem
      • Cemented Femoral Stem
    • Accessories (Sub components)
      • Bone Screw
      • Apical Hole Occluder
      • Centralizer
      • Cement Restrictor
      • Screw Hole Occluder
    AI/ML Overview

    This document describes the Libertas™ Hip Replacement System, a medical device for total hip arthroplasty, and its substantial equivalence to predicate devices. It does not contain information about a study proving the device meets acceptance criteria or detailed performance data in the format requested.

    Therefore, I cannot provide the requested table of acceptance criteria and device performance, nor the details regarding sample sizes, ground truth establishment, expert qualifications, adjudication methods, multi-reader multi-case studies, or standalone algorithm performance.

    The document primarily focuses on establishing substantial equivalence through non-clinical performance testing and similarity to predicate devices, rather than a clinical study with detailed acceptance criteria and performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132959

    Validate with FDA (Live)

    Date Cleared
    2014-03-10

    (171 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K033273, K033338

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Pinnacle AltrX Acetabular Cup Liners are indicated for use in total hip replacement procedures.

    Total hip replacement is indicated in the following conditions:

    1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
    2. Avascular necrosis of the femoral head.
    3. Acute traumatic fracture of the femoral head or neck.
    4. Failed previous hip surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
    5. Certain cases of ankylosis.

    The Pinnacle AltrX Acetabular Cup Liners are indicated for use with Pinnacle Acetabular Cups in cementless applications.

    Device Description

    The DePuy Pinnacle ALTRX Acetabular Liners are part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The liner is manufactured from ultra high molecular weight polyethylene (UHMWPE), which locks into a porous coated, hemispherical outer shell component manufactured from titanium alloy (Ti-6Al-4V). The liner component articulates with a metal or ceramic femoral head of an appropriate diameter. The subject devices represent additional sizes and style combinations of the predicate acetabular liners.

    AI/ML Overview

    The DePuy Pinnacle ALTRX Acetabular Liners are a medical device used in total hip replacement procedures. The 510(k) summary provides information on the device's acceptance criteria and the study that proves it meets these criteria.

    Acceptance Criteria and Device Performance

    The acceptance criteria for the DePuy Pinnacle ALTRX Acetabular Liners are implicitly demonstrated by showing substantial equivalence to legally marketed predicate devices. The study conducted was a non-clinical test that focused on specific performance requirements.

    Here's a table summarizing the characteristics of the subject device in comparison to its predicate and reference devices, which effectively serve as the acceptance criteria based on equivalence:

    CharacteristicAcceptance Criteria (Predicate/Reference)Reported Device Performance (Subject Device)
    Intended UseTotal Hip ArthroplastyTotal Hip Arthroplasty (Same)
    MaterialUHMWPE, GUR 1020 (Predicate)/GUR 1050 (Reference)UHMWPE, GUR 1020
    FixationUncementedUncemented (Same)
    Compatible Acetabular Shell Diameters44-76 mm (Predicate), 38-76 mm (Reference)52-76 mm (New sizes added within acceptable range)
    Compatible Femoral Head Diameters28, 32, 36, 40, 44, and 48 mm (Predicate/Reference)28, 32, 36, and 40 mm (Subset of compatible sizes)
    Minimum Cross-sectional Thickness3.14 mm (Predicate), 3.25 mm (Reference)3.05 mm (Slightly thinner but justified by testing)
    Sterile MethodGas PlasmaGas Plasma (Same)
    PackagingDouble PETG blister with Tyvek peel lidDouble PETG blister with Tyvek peel lid (Same)
    Shelf Life5 years5 years (Same)
    Impingement PerformanceMet applicable performance requirements (implied by predicate clearance)Demonstrated to meet applicable performance requirements
    High Angle Fatigue PerformanceMet applicable performance requirements (implied by predicate clearance)Demonstrated to meet applicable performance requirements
    Push-out PerformanceMet applicable performance requirements (implied by predicate clearance)Demonstrated to meet applicable performance requirements
    Torque-out PerformanceMet applicable performance requirements (implied by predicate clearance)Demonstrated to meet applicable performance requirements

    The device performance, as reported, indicates that the subject devices "met the applicable performance requirements and are as safe and effective as a legally marketed device."

    Study Details

    Given that this is a 510(k) submission for a line extension, the primary "study" is a demonstration of substantial equivalence through non-clinical testing.

    1. Sample size used for the test set and the data provenance:

      • No specific sample sizes for test sets are provided in the document. The testing mentioned (Impingement, High Angle Fatigue, Push-out, and Torque-out) would involve specific numbers of physical samples of the liners and associated components, but these numbers are not detailed here.
      • Data Provenance: The data is from non-clinical tests conducted by the manufacturer, DePuy Orthopaedics. The country of origin of the data is not explicitly stated but is implied to be within the company's testing facilities, likely in the US (Warsaw, IN). The tests are prospective in the sense that they were conducted for the purpose of this submission to demonstrate performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For non-clinical performance testing of medical devices like acetabular liners, the "ground truth" is typically established by engineering standards and validated test methodologies. There isn't a concept of medical "experts" establishing ground truth in the same way as, for example, reviewing medical images. The acceptance criteria for these tests (e.g., maximum force, cycles to failure) are derived from industry standards, regulatory requirements, and historical data from similar, cleared devices. The engineers and technicians who conduct and interpret these tests are qualified in biomechanical testing and materials science, but their specific number or qualifications are not provided here.
    3. Adjudication method for the test set:

      • Not applicable in the traditional sense. Non-clinical mechanical tests are generally quantitative, and their outcomes are determined by meeting predefined engineering specifications and statistical analysis, rather than through expert human adjudication.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done:

      • No, a MRMC comparative effectiveness study was explicitly stated as not conducted. The document states, "No clinical tests were conducted to demonstrate substantial equivalence." This type of study is typically used for diagnostic or screening devices involving human interpretation of data (e.g., radiology images).
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical implant (acetabular liner) for hip replacement, not a software algorithm.
    6. The type of ground truth used:

      • The ground truth for the non-clinical tests (Impingement, High Angle Fatigue, Push-out, and Torque-out) is based on established engineering performance requirements and industry standards for such orthopedic implants. The reported performance of the subject device was compared against these predefined acceptable limits (derived from the predicate devices and general orthopedic device standards) to ensure it performs "as safe and effective as a legally marketed device."
    7. The sample size for the training set:

      • Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070756

    Validate with FDA (Live)

    Date Cleared
    2007-06-06

    (78 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K061253, K033273

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The REFLECTION 3 Acetabular System is indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, osteonecrosis, avascular necrosis, post traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

    Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseased and anomalies, and congenital dysplasia; old remote osteomyelitis with and an extended drainage-free period, in which case, the patient should be warned of an above normal danger of infection postoperatively; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity.

    The R3 Acetabular System is for single use only. The R3 Acetabular System is intended for cementless use.

    Device Description

    The REFLECTION 3 Acetabular System consists of Acetabular shells and liners. The R3 shells are manufactured from titanium alloy. The design of the shells and liners are similar to the REFLECTION 3 Acetabular Shell, the 10MRad REFLECTION 3 Acetabular Liner cleared under K061253 and the DePuy Pinnacle Acetabular Shell cleared under K033273. The REFLECTION 3 liners are manufactured from cross-linked polyethylene. The proposed modifications will include:

    • · The addition of Asymmetric Porous Coating to the outside of the R3 Acetabular Shell
    • · The addition of the R3 MAX shell with Asymmetric Porous Coating to the outside of the shell
    • · The addition of the R3 MAX shell with Titanium Bead Porous Coating to the outside of the shell

    The Acetabular shells feature two locking grooves that correspond to locking beads on the XLPE liners, as cleared in K061253. There are also 12 indentions in the face of the shell that mate with twelve tabs of the XLPE liner to prevent rotation. The R3 Acetabular Shells and the R3 MAX shells will be offered in sizes 40mm-68mm.

    AI/ML Overview

    This document is a 510(k) summary for the Smith & Nephew REFLECTION 3 Acetabular System. It states that the device is substantially equivalent to previously cleared devices.

    Here's an analysis based on the provided text:

    Key Takeaway: This 510(k) submission primarily relies on substantial equivalence to previously cleared predicate devices, rather than presenting a de novo study with specific performance acceptance criteria and a detailed study proving those criteria were met. Therefore, many of the requested sections will state "Not applicable" or reflect that a comparative effectiveness study was not the basis of this particular submission.

    Here's a breakdown of the requested information:


    1. A table of acceptance criteria and the reported device performance

      Acceptance CriteriaReported Device Performance
      Design Features: Similar to predicate devicesSimilar to Smith & Nephew REFLECTION Acetabular System (K920430, K932755, K990666), REFLECTION Interfit Shell (K9640094, K990666), 10MRad REFLECTION Acetabular Liner (K002747), REFLECTION 3 Acetabular System (K061253), and DePuy Orthopaedics Inc. Pinnacle Acetabular System (K033273).
      Overall Indications: Similar to predicate devicesSimilar to predicate devices (as listed in Section A. Intended Use and Indications for Use).
      Materials: Similar to predicate devicesR3 shells: titanium alloy. Liners: cross-linked polyethylene. Similar to predicate devices.
      Safety and Effectiveness: Demonstrated through design control activities and substantial equivalence."Design Control Activities have been completed and the results indicated that the subject device is safe and effective." The FDA's substantial equivalence determination implies acceptance of this claim based on comparison to predicates.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Sample Size: Not applicable. This submission relies on substantial equivalence and design control activities, not a clinical test set with a specific sample size.
      • Data Provenance: Not applicable for a dedicated test set. The data presented here is descriptive of the device modifications and comparisons to existing devices.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable. There was no specific test set requiring expert ground truth establishment for performance evaluation in this 510(k) summary.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. No test set was described that required an adjudication method.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This is not an AI-based device, and no MRMC comparative effectiveness study was performed or described in this document.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable. This is a physical medical device (hip prosthesis components), not an algorithm or AI system.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not applicable in the context of a performance study for this submission. The "ground truth" for this 510(k) is the established safety and effectiveness of the predicate devices. The modifications were assessed through "Design Control Activities."
    8. The sample size for the training set

      • Not applicable. This is not a machine learning or AI device that requires a training set.
    9. How the ground truth for the training set was established

      • Not applicable. As above, no training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1