Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K083056

    Validate with FDA (Live)

    Device Name
    NEO-STAT CO2 <KG
    Manufacturer
    Date Cleared
    2009-02-23

    (132 days)

    Product Code
    Regulation Number
    868.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K031411,K944400,K024300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neo-StatCO2 <Kg is indicated to provide a semi-quantitative visualization of the CO2 in the patient airway. It is an adjunct in patient assessment to be used in conjunction with other methods to determine clinical signs and symptoms by or on the order of a physician. Intended for use with neonate and infant patients from 250g to 6 kg.

    Device Description

    Mercury Medical, Inc. Neo-StatCO2 <Kg is a colorimetric breath indicator for visualization of exhaled CO2. It is designed to connect between an endotracheal tube and a breathing device to help verify proper intubation. Exhaled gas passes through the indicator to detect approximate ranges of end-tidal CO2 by color comparison. Color change is rapid and the detector has the ability to show color changes with 1 ml tidal volume and up to a respiratory rate of 100 breaths per minute. The detector may be used during patient transport or in locations when intubations are performed. While the indicator functions at relative humidity (RH) of up to 100%, it is not recommended for use below 10% RH. It may be used for up to 24 hours.

    AI/ML Overview

    This document describes the Neo-StatCO2 device, a colorimetric breath indicator for CO2 that helps verify proper intubation. The submission aims to demonstrate substantial equivalence to predicate devices, particularly the Mercury Medical Mini-StatCO2 (K031411).

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document implicitly defines acceptance criteria by comparing the new device (Neo-StatCO2) to predicate devices in terms of its physical, performance, and operating characteristics. The primary new "performance" aspect is the expanded patient population, which required specific testing.

    Feature / Criteria (Derived from Predicate Comparison)Acceptance Criteria (Predicate)Reported Device Performance (Neo-StatCO2)
    Indications for Use"semi-quantitative visualization of the CO2 in the patient airway. Adjunct in patient assessment... in conjunction with other methods... by or on the order of a physician."Identical to predicate (Mercury Mini-StatCO2 K031411)
    Environment of UseHospital, sub-acute, pre-hospital & transportIdentical to predicate (Mercury Mini-StatCO2 K031411)
    Patient PopulationNeonate and Pediatric (Mini-StatCO2), Adult, Pediatric, Infant/Neonatal (Oridion MicroCap)Neonate and infant (250g to 6 kg) Expanded from predicate, similar to Oridion MicroCap.
    Weight5 g3 g
    Internal Volume3 cc1 cc
    Resistance to Flow1.23 cm H2O @ 10 Lpm; 8.8 cm H2O @ 30 Lpm1.86 cm H2O @ 10 Lpm; 15.6 cm H2O @ 30 Lpm
    Duration of CO2 DetectionUp to 24 hoursUp to 24 hours
    Operating Conditions-10°C to +50°C-10°C to +50°C
    Storage ConditionsColor change function after storage at 0°C and 60°CColor change function after storage at 0°C and 60°C
    Contraindications & WarningsDo not use for detection of hypercapnia, main-stem bronchial intubation, mouth-to-tube ventilation, oropharyngeal tube placement. Standard clinical assessment must be used.Same as predicate
    Effectiveness of color change for intended populationDemonstrated ability of color change for predicate's population.Performance testing to demonstrate ability of indicator to change color for the intended body weight, tidal volume and breath rate under conditions of use. (This is the specific study mentioned).

    Study Proving Device Meets Acceptance Criteria:

    The document explicitly states: "The only difference between the Mercury Mini-Stat CO2 is the expansion of the patient population to include patients with a body weight of 250 g to 6 Kg which we have demonstrated as being effective via comparative bench testing."

    This implies that the "comparative bench testing" was the study conducted to ensure the device's effectiveness for the expanded patient population (neonate and infant - 250g to 6 kg) under the specified tidal volumes and breath rates.

    Additional Information (Based on the Provided Text):

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified. The document mentions "comparative bench testing" but does not give specific numbers for the test set used in this bench testing.
    • Data provenance: The testing was described as "bench testing," suggesting a laboratory environment. There is no information about the country of origin or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the testing was "bench testing." Ground truth would likely be established by objective measurements of CO2, tidal volume, and breath rate, rather than expert interpretation of the device's color change.

    4. Adjudication method for the test set:

    • Not applicable for bench testing. The evaluation would be based on direct measurement outcomes rather than human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is a colorimetric indicator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The "comparative bench testing" would be a standalone performance evaluation of the device's ability to change color under specific simulated conditions, independent of human interpretation for the physical manifestation of the color change itself. The human-in-the-loop aspect refers to a clinician interpreting the color output. The bench testing would confirm the device produces the correct color change.

    7. The type of ground truth used:

    • For the "comparative bench testing," the ground truth would be precise, known values of CO2 concentration, tidal volume, and respiratory rate, simulated in a controlled environment. The device's color change would then be observed and compared against these known conditions to verify its accuracy.

    8. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI/machine learning algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established:

    • Not applicable as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1