Found 2 results

510(k) Data Aggregation

    K Number
    K182337

    Validate with FDA (Live)

    Device Name
    HYBRID Guidewire
    Manufacturer
    Date Cleared
    2018-10-04

    (37 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161803, K022357

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HYBRID Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in the coronary arteries.

    Device Description

    HYBRID Guidewire is a coiled wire that is designed to fit inside a percutaneous microcatheter for the purpose of directing the catheter through a blood vessel. It is used in association with catheters to enable insertion and easier navigation into the intracranial vascular branches for diagnostic or therapeutic use. Guidewires constitute an important element that conditions the success of the operation. They can be pushed and torqued to move forward through patient's tortuous anatomy.

    HYBRID Guidewire consists of a proximal coated stainless steel core wire and a distal coated nitinol core wire. The distal core wire is tapered at the distal tip and is contained within a 92/8 platinum/tungsten and stainless steel coil for maximum radiopacity. The distal tip comes in various tip designs; it can be straight or curved distal tip (forming a double angle).

    The coil section of the HYBRID Guidewire and the distal stainless steel section is coated with hydrophilic coating, while the proximal stainless steel section is coated with Polytetrafluoroethylene (PTFE). The purpose of the surface coatings is to provide lubricity when HYBRID Guidewire is passed through compatible microcatheters. A torquer and an introducer are also provided with the device.

    AI/ML Overview

    The HYBRID Guidewire is a Class II device intended for general intravascular use, including neuro and peripheral vasculature, to facilitate the selective placement of diagnostic or therapeutic catheters. It is not intended for use in coronary arteries.

    Here's a breakdown of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual and Dimensional InspectionExternal surface free from process and surface defects; dimensions (overall length, radiopaque length, outer diameter, distal shape) per product specification.All test samples passed testing.
    Catheter Compatibility & Simulated UseIn vitro performance under simulated use conditions using a worst-case tortuous ICA vessel model.All test samples passed testing.
    Coating Integrity and AdherenceCoating quality (after dyeing and simulated use) inspected under a microscope.All test samples passed testing.
    Tensile StrengthMinimum force to break the guidewire meets ISO 11070. Tensile strength comparable to the predicate device.All test samples passed testing. Tensile strength was comparable to predicate devices.
    Corrosion ResistanceNo visual evidence of corrosion after immersion in sodium chloride solution and boiling distilled/deionized water.All test samples passed testing.
    Torque StrengthNumber of turns-to-failures comparable to the predicate device.All test samples passed testing. The torque strength of the subject device is equivalent to the predicate device.
    TorqueabilityRotational input to output ratio comparable to the predicate device.All test samples passed testing. The torqueability value of the subject device is equivalent to the predicate device.
    Fracture Resistance and Flexing TestResistance to damage by flexing and resistance to fracture per ISO 11070.All test samples passed testing.
    Tip FlexibilityForce required to deflect the distal tip (45° and 90°) comparable to the predicate device.All test samples passed testing. The tip flexibility of the subject device is equivalent to the predicate device.
    Particulate Matter CharacterizationParticulate matter in injections quantified after simulated use comparable to the predicate device.All test samples passed testing. Particulate matter characterization was comparable to predicate device.
    Design Verification and Packaging ValidationDevice design and packaging design meet product specification requirements at t=1 year time-point after exposure to 1x EtO sterilization.All test samples passed testing.
    Packaging Integrity (Visual Inspection)Packaging free of damage; all seals free of incomplete seals, voids, channels, wrinkles, foreign material, and/or over-heated sections. Labels legible, intact, no damage or delamination.All test samples met the acceptance criteria.
    Packaging Integrity (Bubble Leak Test)No leaks per ASTM F2096-11.All test samples met the acceptance criteria.
    Packaging Integrity (Seal Strength)Meets seal strength per ASTM F88/F88M-15.All test samples met the acceptance criteria.
    Animal Testing (GLP)In vivo performance in an acute porcine model: Trackability, handling, radiopacity, and microcatheter compatibility comparable to predicate devices.All test samples passed testing. Trackability and handling of the guidewire, radiopacity, and compatibility with microcatheter were comparable to predicate device.
    Biocompatibility (Cytotoxicity)Non-cytotoxic (percent viability of test article not less than 70% compared to cells exposed to control).The percent viability of the test article was 110.6%. Non-cytotoxic.
    Biocompatibility (Sensitization)Did not induce delayed sensitization.Did not elicit sensitization response.
    Biocompatibility (Irritation)Non-irritant.Non-irritant.
    Biocompatibility (Acute Systemic Toxicity)No evidence of significant systemic toxicity or mortality.Non-toxic.
    Biocompatibility (Pyrogenicity)Non-pyrogenic (rabbit temperature rise not exceeding 0.5°C).Non-pyrogenic.
    Biocompatibility (Hemolysis)Non-hemolytic (mean hemolytic index for direct contact and extract not exceeding 2%).Non-Hemolytic. (Mean hemolytic index for direct contact was 0.71%; for extract was 0.61%).
    Biocompatibility (Thrombogenicity)Minimal thrombus formation; similar thromboresistance characteristics as control devices.Passed. Minimal thrombus formation associated with test devices; neither animal's clotting abilities were compromised. Similar thromboresistance characteristics to control devices.
    Biocompatibility (Complement Activation)Statistically similar to negative control data.Passed. Statistically similar to the reference negative control data (p > 0.050).
    Biocompatibility (Partial Thromboplastin Time)Within the same thrombogenicity category (minimal activator); responses considered similar to predicate.Passed. Test article and predicate percent plasma control values fell within the same thrombogenicity category (minimal activator). Responses considered similar.
    Biocompatibility (Platelet and Leukocyte Counts)Test article comparable to reference control and comparison article (not statistically significant, p > 0.05).Passed. Test article results for leukocyte and platelet counts showed 100.6-105.6% of negative control. Not statistically significant (p > 0.05) when compared to reference material or comparison article. Comparable to both.
    Shelf Life (Accelerated Aging)All acceptance criteria met at T=1 year accelerated aging, no new questions of safety or effectiveness raised. Performs as intended to Design Specification.All acceptance criteria were met. Concluded that the HYBRID Guidewire will perform as intended to the Design Specification. Labeled for 1-year shelf life.
    Sterilization ValidationAchieves a minimum Sterility Assurance Level (SAL) of 10-6 in accordance with ISO 11135:2007.The validation study demonstrated that the sterilization process and equipment are capable of reliably and consistently sterilizing the devices to a minimum SAL of 10-6. Fractional cycles, half cycles, and full cycles met the acceptance criteria.
    EO and ECH ResidualsResidual traces of Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) below limits specified in ISO 10993-7:2008.The residual traces of EO and ECH for the worst-case device were well below the limits specified in ISO 10993-7.
    Bacterial Endotoxin LevelsBelow 2.15 EU/device, in accordance with FDA guidance, USP <85>, and European Pharmacopeia BET 2.6.14.< 2.15 EU/device.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the exact sample sizes for each bench test. However, it indicates "All test samples" in the results for most bench tests, implying an adequate number of samples were tested to demonstrate compliance.

    For the Animal Testing (GLP), it refers to an "acute porcine model" but does not specify the number of animals used. The provenance of the data is from these controlled laboratory (preclinical) studies, as implied by "Performance Bench Testing and Animal Testing." These are prospective studies conducted specifically for regulatory submission.

    For Biocompatibility testing, various in vitro and in vivo tests were performed:

    • Cytotoxicity: L-929 mouse fibroblast cells.
    • Sensitization: Guinea pig (not specified number but implies multiple for statistical significance).
    • Irritation: Rabbits (not specified number but implies multiple for statistical significance).
    • Acute Systemic Toxicity: Mice (not specified number but implies multiple for statistical significance).
    • Pyrogenicity: Rabbits (not specified number but implies multiple for statistical significance).
    • Hemocompatibility (Hemolysis, Thrombogenicity, Complement Activation, PTT, Platelet and Leukocyte Counts): Human blood or canine models. For thrombogenicity, it mentions "neither animal's clotting abilities," suggesting at least two animals were used, likely canine.

    The provenance for biocompatibility testing is also from controlled laboratory (preclinical) studies, conducted prospectively for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information (number and qualifications of experts) is typically relevant for studies involving subjective human interpretation, such as clinical trial imaging reads or pathology reviews. For a medical device like a guidewire, the "ground truth" is established through objective, quantifiable engineering and biological tests against established standards (e.g., ISO, ASTM) and comparison to predicate devices, rather than expert consensus on individual "cases." Therefore, this section is not directly applicable to the provided data. The "experts" are implicitly the engineers and scientists conducting and analyzing these performance and biocompatibility tests according to their respective standard operating procedures and regulatory guidelines.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies when multiple human readers interpret data, and there's a need to resolve disagreements to establish a definitive "ground truth." Since the provided data details bench and animal testing, which rely on objective measurements and established protocols, an adjudication method in this sense is not applicable. The "adjudication" is inherent in the test methodology (e.g., meeting a specified numerical threshold, visual inspection against a clear defect standard).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This section is not applicable as the HYBRID Guidewire is a physical medical device, not an AI/Software as a Medical Device (SaMD) that involves human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This section is not applicable as the HYBRID Guidewire is a physical medical device, not an AI/SaMD. The performance described is inherent to the device itself.

    7. The Type of Ground Truth Used

    The ground truth for the HYBRID Guidewire's performance is established by:

    • Established Performance Requirements: Meeting engineering specifications that are commonly accepted for guidewires (e.g., tensile strength, torque strength, flexibility).
    • International Standards: Compliance with relevant ISO (e.g., ISO 11070, ISO 10993) and ASTM standards.
    • Comparison to Predicate Devices: Demonstrating that the subject device's performance is "comparable" or "equivalent" to legally marketed predicate devices, which have a history of safe and effective use.
    • Biological Endpoints: For biocompatibility, the ground truth is the absence of adverse biological reactions (e.g., cytotoxicity, sensitization, irritation, pyrogenicity, significant thrombus formation, etc.) as defined by the applicable ISO 10993 standards.

    8. The Sample Size for the Training Set

    This section is not applicable as the HYBRID Guidewire is a physical device, not a machine learning model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable as the HYBRID Guidewire is a physical device, not a machine learning model requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052526

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-15

    (62 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Novus 3000 Laser System is intended for use in the treatment of ocular pathology. The Novus 3000 Laser System is indicated for use in photocoagulation of both anterior and posterior segments including:

    • Retinal photocoagulation, panretinal photocoagulation and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroid including:
    • proliferative and nonproliferative diabetic retinopathy;
    • choroidal neovascularization;
    • branch retinal vein occlusion;
    • age-related macular degeneration;
    • retinal tears, detachments;
    • retinopathy of prematurity.
    • Iridotomy, iridectomy and trabeculoplasty in angle closure glaucoma and open angle glaucoma.

    Laser Indirect Ophthalmoscope
    The Laser Indirect Ophthalmoscope is indicated for the delivery of laser energy in eyes with retinal pathology. The Laser Indirect Ophthalmoscope is indicated for use in the following ophthalmic treatments and conditions: diabetic retinopathy (panretinal photocoagulation); peripheral neovascularization, retinal breaks, detachments, and tears, lattice degeneration, pneumatic retinopexy reattachment procedures, segmental peripheral photocoagulation; segmental photocoagulation; cloudy vitreous cavities; pediatric retinal repairs (under general anesthesia), delivery of laser energy through small pupils or to eyes with focal lens opacities.

    Endophotocoagulation
    When used with a Novus 3000 Laser System, the Acculite EndoOcular Probes (Acculite EndoOcular Probe, Acculite Aspirating EndoOcular Probe, and Acculite Illuminating EndoOcular Probe) are intended for use in the following ophthalmic applications: photocoagulation of the anterior and posterior segment, including: anterior segment treatment in the surgical management of glaucoma; endophotocoagulation in vitreoretinal surgery, including panretinal photocoagulation, retinopexy, and treatment of neovascularization.

    The laser delivery function of the Acculite Acculite EndoOcular Probes (Acculite EndoOcular Probe, Acculite Aspirating EndoOcular Probe and Acculite Illuminating EndoOcular Probe) is indicated for use in ocular surgery to deliver laser energy to the treatment area selected by the surgeon.

    The aspiration function of the Acculite Aspirating EndoOcular Probe is indicated for use when unwanted fluid is present in the eye during ocular surgery, causing refraction or scattering of the laser beam from the intended treatment site.

    The illumination function of the Acculite Illuminating EndoOcular Probe is indicated for use during ocular surgery to illuminate the interior of the eye.

    Device Description

    Lumenis, Inc.'s Novus® 3000 Laser System with the Delivery Devices is an air cooled, diode-pumped, solid state, Nd: YAG Laser System, which produces a wavelength of 532 um of laser light with a treatment beam output ranging from 50 mw to 3.0 W. The main parts of the Novus® 3000 system include the laser console, a footswitch, a remote control and assorted laser accessories.

    AI/ML Overview

    This 510(k) summary for the Lumenis Novus® 3000 Laser System and Delivery Devices does not include any performance data or a study to demonstrate that the device meets specific acceptance criteria.

    The document explicitly states:

    "No performance data is required for this Class II device nor requested by the Food and Drug Administration (Office of Device Evaluation). No data was submitted for section 807.92 6[(b)(1)(2)(3c)]."

    Therefore, I cannot provide the requested information. The submission relies on demonstrating substantial equivalence to predicate devices based on similarities in intended use, indications for use, and technological characteristics, rather than on new performance data for this specific device.

    If this were a submission that included a performance study, the requested information would typically be provided as follows:


    Hypothetical Example if performance data were submitted:

    Here's how the information would be presented if a study was conducted and acceptance criteria were defined, based on a typical medical device submission structure:

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical Example)

    Performance MetricAcceptance CriteriaReported Device Performance
    Laser Output Power AccuracyWithin ±10% of set value±5% (e.g., for 50 mW to 3.0 W range)
    Beam Spot Size AccuracyWithin ±15% of nominal size±10% (e.g., for various delivery devices)
    Pulse Duration AccuracyWithin ±5% of set value±2% (e.g., for 10 ms to 1000 ms range)
    Aiming Beam Co-axialityWithin 0.5 mm relative to treatment beam<0.2 mm
    Thermal Performance (e.g., temperature rise of handpiece)Below X°C for Y continuous operationZ°C (well within limits)

    2. Sample Size Used for the Test Set and Data Provenance (Hypothetical Example)

    • Sample Size for Test Set: E.g., 5 units of the Novus® 3000 Laser System and a representative sample of each delivery device (e.g., 10 Laser Indirect Ophthalmoscopes, 15 Endophotocoagulation Probes).
    • Data Provenance: Prospective, collected at the manufacturer's facility in Santa Clara, CA, USA, under controlled laboratory conditions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications (Hypothetical Example)

    This type of assessment would typically involve engineering and quality control experts rather than clinical experts for ground truth establishment.

    • Number of Experts: 3-5 Engineers and Quality Control personnel.
    • Qualifications: E.g., Lead Optical Engineer (15 years experience in laser systems), Senior Electrical Engineer (10 years experience in medical device electronics), Quality Assurance Manager (12 years experience in medical device testing and regulatory compliance).

    4. Adjudication Method for the Test Set (Hypothetical Example)

    • Adjudication Method: For quantitative measurements, results would be recorded by trained technicians/engineers. Any discrepancies or outliers would be investigated and resolved by the lead engineer and QA manager through re-testing or thorough review of methodology. For subjective assessments (e.g., ergonomic evaluation), a consensus approach among the expert panel might be used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical Example)

    • Given this is a physical laser device, an MRMC study related to interpretation of results or diagnostic performance (like for an imaging AI) is not applicable. The device's direct performance is based on its physical properties and output.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop) (Hypothetical Example)

    • Again, this term is more relevant for AI/software devices. For a hardware device like a laser, the "standalone performance" is exactly what the hypothetical performance table above would describe – the device's intrinsic operational metrics without a human actively making real-time adjustments for the test. The clinician is always "in the loop" during actual treatment.

    7. Type of Ground Truth Used (Hypothetical Example)

    • Ground Truth: Primarily based on metrology standards and established physical principles. This would involve using calibrated measurement equipment (e.g., power meters, beam profilers, oscilloscopes) traceable to national standards to objectively quantify the laser's output characteristics against its design specifications.

    8. Sample Size for the Training Set (Hypothetical Example)

    • For a hardware device, the concept of a "training set" is usually not applicable in the same way it is for AI/machine learning algorithms. Device design and manufacturing processes are refined through iterative testing and quality control during development. The "training" for such a device comes from engineering principles, material science, and prior models.

    9. How the Ground Truth for the Training Set Was Established (Hypothetical Example)

    • As above, not directly applicable in the AI sense. Ground truth in the development phase would be established through engineering specifications, design verification, and validation testing against intended performance requirements, using calibrated equipment and standard test methods.