Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143085

    Validate with FDA (Live)

    Date Cleared
    2015-02-09

    (105 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker Orthopaedic Acetabular Shells for use with the Total Hip Instruments are intended for use in total hip arthroplasty and are intended for either primary or revision total hip arthroplasty. The Stryker Orthopaedic Hip Implant Stems for use with the Total Hip Instruments are intended for use in total or hemi-arthroplasty of the hip and are intended for either primary or revision procedures. The Direct Superior Approach does not change the intended use of the subject devices.

    Indications for Use (Trident PSL Shell, Tritanium Shell, Trident Hemispherical Shells):

    • Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
    • Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
    • Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
    • Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
      The Trident shells are intended for cementless fixation within the prepared acetabulum.

    Indications for Use (Tritanium Acetabular Shell System):

    • Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
    • Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
    • Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
    • Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
      The Tritanium Acetabular Shell System is intended for cementless use only.

    Indications for Use (Accolade II Femoral Stem):
    The indications for use for total hip arthroplasty with stems include:

    • Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis:
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Revision procedures where other treatments or devices have failed; and,
    • Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
      Additional indication specific to use of the Accolade II Femoral Stems with compatible Howmedica Osteonics Constrained Liners:
    • When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
      Accolade II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.

    Indications for Use (Secur-Fit Advanced Hip Stem):
    The indications for use for total hip arthroplasty with stems include:

    • Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis:
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Revision procedures where other treatments or devices have failed; and,
    • Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
      Additional indication specific to use of the Secur-Fit Advanced Hip Stems with compatible Howmedica Osteonics Constrained Liners:
    • When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
      Secur-Fit Advanced Hip Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.

    Indications for Use (Anato Hip Stem):
    The indications for use for total hip arthroplasty with stems include:

    • Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Revision procedures where other treatments or devices have failed; and,
    • Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
      Additional indication specific to use of the Anato Hip Stem with compatible Howmedica Osteonics Constrained Liners:
    • When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
      The Anato Hip Stem is intended for cementless use only and is intended for total and hemiarthroplasty procedures.
    Device Description

    The devices covered by this submission include Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments. The Instruments consist of Retractors, Cup Impactors, U-Joint Bolt Driver, Alignment Guide, Angled Reamer Handle, Instrument Trays and a Case. All instruments are hand-held surgical instruments used during orthopaedic surgery and are nonsterile, reusable devices. All Class II devices in this submission have been previously determined substantially equivalent in prior 510(k) submissions and are commercially available. The acetabular shells and femoral stems are manufactured from the following materials: titanium (Ti-6Al-4V) alloy, commercially pure (CP) titanium plasma spray, commercially pure titanium and PureFix® hydroxylapatite (HA).

    AI/ML Overview

    This document is a 510(k) premarket notification from Stryker Orthopaedics to the FDA for their Acetabular Shells, Femoral Stems, and Total Hip Instruments for use with the Direct Superior Approach.

    It is important to note that this submission is for a labeling modification to introduce a new surgical approach (Direct Superior Approach) for existing devices, not for a new device requiring extensive performance studies. Therefore, the details requested for acceptance criteria and studies proving device performance (like sample sizes, expert qualifications, etc., which are typically associated with evaluating a new device's diagnostic or predictive performance) are largely not applicable in the context of this specific regulatory submission.

    Here's an attempt to address your request based on the provided document, highlighting what is (and isn't) present:

    1. Table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria or report on device performance based on a study, because the submission is for a labeling modification of already cleared and marketed devices. The core of the submission is to demonstrate that the new surgical approach (Direct Superior Approach) does not fundamentally change the intended use, safety, or effectiveness of the previously cleared devices.

    Instead of performance metrics, the document focuses on demonstrating substantial equivalence to predicate devices. This means that the new labeling (for the Direct Superior Approach) does not raise new questions of safety or effectiveness and that the devices, when used with this new approach, are as safe and effective as existing legally marketed predicate devices.

    The "performance" in this context is the general safety and efficacy of the existing devices, which was established in their original 510(k) clearances.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. No specific test set to evaluate device performance for a new device was used. The submission focuses on the surgical approach for existing devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No new test set requiring expert ground truth establishment for a diagnostic or predictive device was involved. The safety and effectiveness of the devices themselves were established in prior 510(k)s.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no test set requiring ground truth adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/diagnostic device submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI/diagnostic device submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the sense of a new performance evaluation for a new device. The "ground truth" for this submission is the established safety and efficacy of the predicate devices.

    8. The sample size for the training set

    Not applicable. No new device or algorithm requiring a training set was involved.

    9. How the ground truth for the training set was established

    Not applicable. No new device or algorithm requiring a training set was involved.


    Summary of the Study (as per the document):

    The document explicitly states:

    • Non-Clinical Testing: "Performance testing was not required in support of this submission because this submission covers a labeling modification to introduce an additional surgical protocol, Direct Superior Approach."
    • Clinical Testing: "Clinical testing was not required as a basis for substantial equivalence."

    Conclusion of the Study:

    The "study" in this case is the review of the modified labeling and comparison to predicate devices, rather than experimental testing of the device itself.

    The conclusion is that the Stryker Orthopaedics Acetabular Shells, Femoral Stems, and Total Hip Instruments for use with the Direct Superior Approach are substantially equivalent to the predicate devices identified in the premarket notification. This substantial equivalence is based on:

    • The intended use and indications for the implants with the Direct Superior Approach being identical to those of the cleared predicate devices.
    • The materials, design features, and functionality for the components being substantially equivalent to previously cleared predicate devices.
    • The fundamental scientific technology of the modified device (i.e., the surgical instruments for the new approach) not having changed relative to the predicate devices.
    • The material and design of the Total Hip Surgery Instruments for use with the Direct Superior approach being identical to those of the predicate devices cleared for use with other approaches (Posterolateral Approach and Direct Anterior Approach), and their operational principles being similar.

    In essence, the "study" for this submission was a comparative analysis of design, materials, intended use, indications for use, and operational principles between the proposed devices (with the new approach) and their already cleared predicates, confirming that no new safety or effectiveness concerns were raised by adding the Direct Superior Approach to the labeling.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142606

    Validate with FDA (Live)

    Date Cleared
    2015-01-26

    (133 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K112802, K081171, K983382, K013676

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trident® Tritanium® PST® Acetabular Shell System is in skeletally mature individuals undergoing surgery for total hip replacement due to:

    · A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis, or congenital hip dysplasia;

    · Acute traumatic fracture of the femoral head or neck;

    · Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.

    The Trident® Tritanium® PST® Acetabular Shell is intended for cemented fixation. The porous structured surface provides biological fixation when used in a cementless application.

    Device Description

    The subject Trident® Tritanium® PST® Acetabular Shell is the predicate PST® Acetabular Shell, which has been modified to incorporate the acetabular bearing mating features of the predicate Trident® Tritanium®, Trident® PSL®, and Trident® Hemispherical Shells, to allow compatibility with existing Stryker Orthopaedics acetabular bearings including the Trident polyethylene and Modular Dual Mobility (MDM) bearing families.

    The subject Trident® Tritanium® PST® Acetabular Shells are manufactured from titanium alloy and feature a porous structured surface. The subject shells feature a dome hole, and are available either in a solid shell (no screw holes) configuration, or in a cluster screw hole configuration for optional supplemental bone screw fixation. The subject Tritanium® PST® Acetabular Shells are compatible with the optional predicate PST Acetabular Shell system bone screws (titanium alloy), with the optional predicate Stryker Orthopaedics Torx head acetabular dome hole occluder (CP titanium), and with an optional new hex head acetabular dome hole occluder (titanium alloy).

    AI/ML Overview

    The provided text is a 510(k) Summary for the Trident® Tritanium® PST® Acetabular Shells. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, this document does not contain information about acceptance criteria or a study proving the device meets specific performance criteria in the way an AI/software device would.

    This document describes a medical device (an acetabular shell for hip replacement), not a software or AI product. The "Performance Testing" section refers to comparisons of materials, manufacturing methods, design features, and disassembly testing per ASTM F1820 (including push-out, and torsional testing) to demonstrate substantial equivalence to predicate devices.

    Therefore, most of the requested information (related to AI performance, sample sizes for test/training sets, experts for ground truth, MRMC studies, standalone performance, etc.) is not applicable to this type of regulatory submission for a physical medical device.

    I can only extract the following relevant information:


    1. A table of acceptance criteria and the reported device performance

    Based on the document, specific acceptance criteria and reported device performance in terms of quantifiable metrics like sensitivity, specificity, accuracy, etc., are not applicable as this is a physical medical device, not an AI/software device with such performance metrics.

    The performance testing mentioned is mechanical in nature, focused on demonstrating substantial equivalence to predicate devices.

    Acceptance Criteria CategoryReported Device Performance / Evaluation Method
    Mechanical Performance- Comparison of materials: Same materials as predicate PST® Acetabular Shells.- Comparison of manufacturing methods: Same manufacturing methods as predicate PST® Acetabular Shells.- Comparison of design features: Same outer profile and porous structured surface as predicate PST® Acetabular Shells.- Disassembly Testing: Per ASTM F1820, including push-out and torsional testing (for worst-case style and size of compatible Stryker Orthopaedics acetabular bearings).
    Compatibility- Incorporates acetabular bearing mating features of predicate Trident® Tritanium®, Trident® PSL®, and Trident® Hemispherical Shells.- Compatible with existing Stryker Orthopaedics acetabular bearings, including Trident polyethylene and Modular Dual Mobility (MDM) bearing families.
    SterilizationSame sterilization methods as predicate PST® Acetabular Shells.
    PackagingSimilar packaging to predicate PST® Acetabular Shells.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable to this device. The testing described is mechanical performance (e.g., disassembly testing per ASTM F1820) of physical components, not data analysis or clinical studies in the typical sense of AI/software device evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth, in the context of expert review for AI/software, is not relevant for the mechanical testing of a physical implant.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically for resolving discrepancies in expert interpretations for ground truth establishment, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. MRMC studies are for evaluating human performance with and without AI assistance for tasks like image interpretation, which is not relevant for an orthopedic implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This describes a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable. The "ground truth" for a physical device is its adherence to specified material properties, mechanical strength characteristics, and design specifications as evaluated through engineering tests, not clinical or pathological outcomes in the context of expert consensus as described for AI/software.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" as this is not an AI/software device.

    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1