510(k) Data Aggregation

    K Number
    K183337

    Validate with FDA (Live)

    Date Cleared
    2019-05-29

    (177 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K101710, K013674, K924830, K073296

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LuxaCrown is a self-curing composite for the fabrication of semi-permanent crowns and bridges. The material is mixed automatically. Specifically: Crown restorations: For restoration of the anatomical form in order to provide durable protection for the remaining tooth; In order to restore chewing function; For esthetic corrections. Bridge restorations: In order to restore chewing function; For esthetic corrections.

    Device Description

    LuxaCrown is a bis-acrylic resin-based self-curing composite for the fabrication of semipermanent crowns and bridges. LuxaCrown is available in a range of shades to ensure an aesthetically-pleasing restoration. LuxaCrown is a non-sterile device intended for use by licensed dental professionals. Because LuxaCrown is intended for the fabrication of semi-permanent dental restorations, it may remain in the oral cavity for more than 30 days. LuxaCrown is supplied in dual-barreled syringes for use with the DMG Automix selfmixing dispenser and Automix Tips, which were previously cleared in K101710 (DMG LuxaTemp Ultra), K013674 (DMG LuxaTemp) and K924830 (DMG LuxaTemp Automix).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental material called LuxaCrown, a self-curing composite for fabricating semi-permanent crowns and bridges. The document focuses on demonstrating substantial equivalence to predicate devices based on indications for use, technical specifications, biocompatibility, and physical properties.

    However, the document does not contain any information about acceptance criteria or a study that specifically proves the device meets those criteria in the format requested. The "Discussion of Non-Clinical Tests" section mentions various tests performed (ISO 10447, internal DMG methods, biocompatibility testing) and states that "All samples met the requirements of internal product specifications and were within the limits specified in ISO 10477." This indicates that there are internal specifications and ISO limits, but the document does not list them as "acceptance criteria" in a table or provide detailed study results that meticulously "prove" the device meets them with specific performance metrics beyond a general statement of compliance.

    Therefore, many of the requested fields cannot be filled from the provided text.

    Here's a breakdown of what can and cannot be extracted:

    1. A table of acceptance criteria and the reported device performance

    The document states: "All samples met the requirements of internal product specifications and were within the limits specified in ISO 10477." However, it does not provide the specific numerical acceptance criteria nor detailed reported performance data in a tabular format. It only lists the types of tests conducted.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified for any of the tests mentioned (flexural strength, water sorption, water solubility, working time, setting time, compressive strength, biocompatibility, shelf life).
    • Data Provenance: The tests were performed by DMG Chemisch-Pharmazeutische Fabrik GmbH, which is a German company (Hamburg, Germany). It's implied these are laboratory tests, not clinical studies with human data, so "retrospective or prospective" is not fully applicable in the conventional sense for patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The tests are for physical and chemical properties of a material, not for diagnostic accuracy or human interpretation requiring expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there's no human interpretation or ground truth establishment in this context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a material science and biocompatibility assessment, not a study involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a material science assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the physical and chemical tests, the "ground truth" is established by the specified standards (ISO 10447, ISO 10993-3, -5, -10, ASTM F1980) and the manufacturer's internal product specifications. For biocompatibility, it's compliance with established biological safety requirements.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning or AI-based device.


    Summary Table (with available information):

    CriterionDetails from Document
    Acceptance Criteria & Reported Performance- Acceptance Criteria: "requirements of internal product specifications" and "limits specified in ISO 10477" for physical tests; "requirements of ISO 10993-1:2009 and ISO 7404:2008" for biocompatibility; compliance with ISO 10993-3 (mutagenicity), ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization/irritation); ASTM F1980 guidelines for shelf life.
    - Reported Performance: "All samples met the requirements of internal product specifications and were within the limits specified in ISO 10477." "LuxaCrown was determined to meet all biocompatibility requirements for dental materials." Shelf-life testing "demonstrated that LuxaCrown remains stable over the 2-year product shelf life." Specific numerical values for criteria or performance are not provided in the document.
    Sample Size (Test Set)Not specified.
    Data Provenance (Test Set)Laboratory testing conducted by DMG Chemisch-Pharmazeutische Fabrik GmbH (Germany). Not applicable as retrospective/prospective human data.
    Number of Experts for Ground Truth (Test Set)Not applicable (physical/chemical properties).
    Qualifications of ExpertsNot applicable.
    Adjudication Method (Test Set)Not applicable.
    MRMC Comparative Effectiveness StudyNo.
    Standalone Performance StudyYes, implied by the non-clinical tests. However, the performance is against material standards, not diagnostic accuracy.
    Type of Ground TruthEstablished standards (ISO 10447, ISO 10993, ASTM F1980) and manufacturer's internal product specifications.
    Sample Size (Training Set)Not applicable (not an AI/ML device).
    How Ground Truth for Training Set was EstablishedNot applicable.

    In conclusion, while the document confirms that testing was done and met specified requirements, it does not provide the granular detail needed to fill out all aspects of the requested table, particularly the specific numerical acceptance criteria and the detailed reported performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171729

    Validate with FDA (Live)

    Device Name
    Fixtemp C&B
    Date Cleared
    2017-12-13

    (184 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122039, K033022, K042820, K013674

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fixtemp® C&B is a resin based material used to fabricate temporary crowns and bridges.

    Device Description

    Fixtemp® C&B is a two-component automatically mixable composite on the basis of multifunctional (meth)acrylates provided in standard double cartridges. The material is dispensed and mixed by cartridge/mixing tip combination and it is used for the manufacturing of temporary crowns and bridges for use until the permanent restoration is fabricated. It can be also used for inlays, onlays and veneers and is available in tooth-colors A1, A2, A3, A3.5, B1, Bleach X and D2. This is a prescription only material. The labeling and working instructions are designed for health care professionals.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Fixtemp® C&B

    The Fixtemp® C&B device is a resin-based material used to fabricate temporary crowns and bridges. Its performance was evaluated against a set of non-clinical criteria to establish substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (or Reference Standard)Reported Device Performance
    Working / Processing time≥ 45 sec≥ 45 sec
    Curing time in mouth (flexible phase)≥ 2 - 3 min≥ 2 - 3 min
    Setting time≥ 6 min≥ 6 min
    Hardness (1h)≥ 75 Shore D≥ 75 Shore D
    Hardness (24h)≥ 80 Shore D≥ 80 Shore D
    Compressive strength≥ 200 MPa (according to DIN EN ISO 4049:2010)≥ 200 MPa (according to DIN EN ISO 4049:2010)
    Flexural strength≥ 60 MPa (according to DIN EN ISO 4049:2010)≥ 60 MPa (according to DIN EN ISO 4049:2010)
    E-modulus> 1500 (according to DIN EN ISO 4049:2010)> 1500 (according to DIN EN ISO 4049:2010)
    Water absorptionPass (according to DIN EN ISO 4049:2010)Pass (according to DIN EN ISO 4049:2010)
    Radiopacity≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)
    CytotoxicityNo biological reactivity (Grade 0) based on ISO 10993-5No biological reactivity (grade 0) of the cells exposed to the test article. (Test report: "Toxikon Europe Final GLP Report: 17-01389-G1")
    Carcinogenic potentialNo evidence for causing cancer within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for causing cancer within the time period of use in the patients mouth for up to 29 days.
    Mutagenic potentialNo evidence for mutagenic reactions within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for mutagenic reactions within the time period of use in the patients mouth for up to 29 days.
    Irritating potentialNo evidence for irritation within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for irritation within the time period of use in the patients mouth for up to 29 days.
    Sensitizing potentialNo evidence for sensitization within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for sensitization within the time period of use in the patients mouth for up to 29 days.
    Shelf life2 yearsThe results of stability tests at 23 °C/37 °C justify to assume a shelf life of 2 years for Fixtemp® C&B.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of specimens for compressive strength). It generally refers to "a series of in-house tests" and adherence to specific ISO standards (e.g., DIN EN ISO 4049:2010).

    For the cytotoxicity testing, the test was conducted by the independent laboratory Toxikon Europe. The provenance of the data is indicated as "Toxikon Europe Final GLP Report: 17-01389-G1". The country of origin for the manufacturing company (Dreve Dentamid GmbH) is Germany. It is implied that these are prospective tests performed on the Fixtemp® C&B material.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The performance evaluation relies on standardized tests (e.g., ISO standards) and laboratory findings rather than expert consensus on a test set in the typical sense of diagnostic imaging or clinical studies. The cytotoxicity test was conducted by an "independent laboratory," but the qualifications of the individuals establishing the ground truth (i.e., interpreting the cytotoxicity results) are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable. The reported tests are objective measurements based on standardized procedures and laboratory analysis, not subjective interpretations requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. Fixtemp® C&B is a material for fabricating dental prosthetics, and its performance is evaluated through material properties and biocompatibility, not reader interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. Fixtemp® C&B is a dental material, not an algorithm or software. Its performance is inherent to its physical and chemical properties.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests consists of:

    • Standardized measurement directly from testing: This applies to properties like working time, curing time, setting time, hardness, compressive strength, flexural strength, E-modulus, water absorption, and radiopacity, where the "ground truth" is a measured physical or chemical property that meets or exceeds a predefined numerical threshold from an established standard (e.g., DIN EN ISO 4049:2010).
    • Laboratory findings: For cytotoxicity, the ground truth is established by the results of the ISO 10993-5 test conducted by an independent laboratory, indicating the absence of biological reactivity.
    • Evaluations and assessments: For carcinogenic, mutagenic, and irritating/sensitizing potentials, the ground truth is based on "different evaluations and assessments" indicating no evidence of these effects within the specified time frame. While the specifics of these evaluations are not detailed, they likely rely on established toxicological principles and potentially other in-vitro or in-vivo studies not explicitly listed.

    8. The Sample Size for the Training Set

    Not applicable. Fixtemp® C&B is a physical material, not an AI model or software that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073296

    Validate with FDA (Live)

    Device Name
    TEMPXN28
    Date Cleared
    2008-01-09

    (47 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K033022, K013674, K992645

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fabrication of temporary crowns, bridges, inlays, onlays and veneers.
    Fabrication of long-lasting temporary restorations. Lining material for prefabricated temporary crowns made of composite and metal.

    Device Description

    TempXN28 is a temporary crown and bridge resin intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated. Temporary crown and bridge resin is designated at 21 C.F.R § 872.3770 as a Class II device.
    Like Protemp™ 3 Garant™, TempXN28 is available in the proven Garant™ mixing and dispensing system.

    AI/ML Overview

    The provided text is a 510(k) Summary for a dental material (TempXN28) and not a study describing the performance of an AI/ML powered device. As such, the requested information pertaining to AI/ML device acceptance criteria, study design, and performance metrics (e.g., sample size for test set, ground truth establishment, MRMC studies, standalone performance) is not available in the given document.

    The document discusses the safety and effectiveness of the dental material TempXN28 by demonstrating its substantial equivalence to predicate devices (Protemp™ 3 Garant™, Luxatemp Fluorescence, and Sinfony).

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: Not explicitly stated as quantifiable metrics for a software/AI device. For this dental material, acceptance criteria would typically involve biocompatibility, mechanical properties (e.g., flexural strength, wear resistance), polishability, color stability, and ease of use, all within acceptable ranges or comparable to predicate devices.
    • Reported Device Performance: The document generally states: "To provide evidence for safety biocompatibility testing was carried out. The results show that TempXN28 is a safe device. The comparison for chemistry, performance data and indications for use shows that TempXN28 is substantially equivalent to the predicate devices."
      • No specific performance values (e.g., strength in MPa, fill rate, etc.) are provided in this summary.
    Acceptance Criteria (Implied for Dental Material Equivalence)Reported Device Performance (Implicit from Summary)
    Biocompatibility (Safety)Biocompatibility testing carried out, results show TempXN28 is a safe device.
    Performance Data (e.g., mechanical, chemical properties)Comparison for chemistry and performance data shows substantial equivalence to predicates.
    Indications for UseSubstantially equivalent to predicate devices for stated indications.

    The following questions cannot be answered from the provided text as they relate to AI/ML device evaluation, which is not the subject of this 510(k) summary:

    1. Sample size used for the test set and the data provenance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
    3. Adjudication method for the test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, and the effect size.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    7. The sample size for the training set.
    8. How the ground truth for the training set was established.