Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K031122

    Validate with FDA (Live)

    Date Cleared
    2003-05-19

    (41 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K012768, K981479

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor X Deficient Plasma is human plasma immunodepleted of Factor X and intended for the in vitro diagnostic quantitative determination of Factor X activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor X Deficient Plasma is human plasma immunodepleted of Factor X and intended for the in vitro diagnostic quantitative determination of Factor X activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Abnormalities of the extrinsic pathway factors are determined by performing a modified prothrombin time (PT) test. Patient plasma is diluted and added to a plasma deficient in Factor X. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the Factor X in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    Acceptance Criteria and Study for HemosIL Factor X Deficient Plasma

    This document describes the acceptance criteria and the study that demonstrates the HemosIL Factor X Deficient Plasma device meets these criteria, as derived from the provided 510(k) summary (K031122).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the performance metrics presented for substantial equivalence to predicate devices. These criteria are based on method comparison (slope and correlation coefficient) and precision (within-run and between-run coefficient of variation).

    Metric / Test TypePredicate Device 1 (Hemoliance Factor X Deficient Plasma on ELECTRA)Predicate Device 2 (IL Test Factor X Deficient Plasma on ACL Family)Acceptance Criteria (Implied by Predicate)Reported Device Performance (HemosIL Factor X Deficient Plasma)
    Method Comparisonvs. Predicate Hemoliance Factor X Deficient Plasma on ELECTRA:ELECTRA 1600C (n=63): Slope = 0.9461, r = 0.9948vs. Predicate IL Test Factor X Deficient Plasma on ACL Family:ACL 3000 (n=62): Slope = 1.0328, r = 0.9849ACL Futura (n=62): Slope = 1.0680, r = 0.9840
    SlopeA slope close to 1.0 (indicating proportional agreement)A slope close to 1.0 (indicating proportional agreement)Demonstrated to be close to 1.0 across different systems and predicates (e.g., 0.9461 to 1.0680).
    Correlation Coefficient (r)A high correlation coefficient (typically ≥ 0.975 as a common standard for method comparison in IVDs)A high correlation coefficient (typically ≥ 0.975)Demonstrated to be high across different systems and predicates (e.g., 0.9840 to 0.9948).
    Within Run Precision(Not explicitly stated for predicate in this summary, but typically a low CV% is expected for IVDs)(Not explicitly stated for predicate in this summary, but typically a low CV% is expected for IVDs)CV% for Normal Control: Low (e.g., < 5%)CV% for Low Abnormal Control: Low (e.g., < 7%)ACL 300: Normal Control: 3.5%Low Abnormal Control: 4.1%ACL 6000: Normal Control: 3.3%Low Abnormal Control: 3.0%ACL 9000: Normal Control: 1.4%Low Abnormal Control: 1.9%ACL Futura: Normal Control: 4.2%Low Abnormal Control: 5.0%ELECTRA 1400C: Normal Control: 0.9%Low Abnormal Control: 1.7%
    Between Run Precision(Not explicitly stated for predicate in this summary, but typically a low CV% is expected for IVDs)(Not explicitly stated for predicate in this summary, but typically a low CV% is expected for IVDs)CV% for Normal Control: Low (e.g., < 7%)CV% for Low Abnormal Control: Low (e.g., < 10%)ACL 300: Normal Control: 4.6%Low Abnormal Control: 6.0%ACL 6000: Normal Control: 4.3%Low Abnormal Control: 4.8%ACL 9000: Normal Control: 2.1%Low Abnormal Control: 4.4%ACL Futura: Normal Control: 4.5%Low Abnormal Control: 6.1%ELECTRA 1400C: Normal Control: 2.9%Low Abnormal Control: 2.4%

    Note: The acceptance criteria for the predicate devices are not explicitly stated in this summary. However, for a 510(k) submission, the new device is typically considered to meet acceptance criteria if its performance is "substantially equivalent" to that of the predicate device. This usually implies that the reported performance metrics for the new device fall within acceptable ranges relative to the predicate, often demonstrated by meeting similar or better statistical measures (slopes close to 1, high correlation coefficients, and low coefficients of variation).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Sizes:
      • Method Comparison (vs. Hemoliance Factor X Deficient Plasma on ELECTRA): n = 63 citrated plasma samples (normal and abnormal)
      • Method Comparison (vs. IL Test Factor X Deficient Plasma on ACL Family): n = 62 citrated plasma samples (normal and abnormal) for both ACL 3000 and ACL Futura.
      • Precision Studies: n = 80 measurements (over multiple runs) for each control level (Normal Control and Low Abnormal Control) on each instrument (ACL 300, ACL 6000, ACL 9000, ACL Futura, ELECTRA 1400C).
    • Data Provenance: The studies were conducted as "field site studies." The country of origin of the data is not specified, but given the applicant's location (Lexington, MA, USA), it is likely to be primarily U.S. data. The data is retrospective in the sense that existing plasma samples were used, or it could be prospective if samples were collected for the purpose of the study. The summary does not specify this detail.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the given 510(k) summary. For in vitro diagnostic devices like HemosIL Factor X Deficient Plasma, "ground truth" is typically established by the reference method or comparative method (the predicate device) itself, or by accepted laboratory standards/controls, rather than expert consensus on diagnostic images or clinical outcomes. The performance is assessed by comparing the device's quantitative output to known values (controls) or to the results from the predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving subjective interpretation, such as image reading, where multiple experts interpret data independently and then resolve discrepancies. For a quantitative diagnostic like Factor X activity, the measurement itself is the primary output, and discrepancies would typically be handled by re-testing or applying established quality control rules, not expert adjudication in the same sense.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not applicable and not provided. MRMC studies are relevant for devices where human readers interpret data, often with or without AI assistance, and the impact of the AI on reader performance is evaluated. The HemosIL Factor X Deficient Plasma is a laboratory reagent used in an automated system, not a device that involves human reader interpretation in the context of an MRMC study.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

    The studies presented are effectively standalone performance studies for the device. The "algorithm" here is the chemical reaction and measurement process of the HemosIL Factor X Deficient Plasma test on the specified automated coagulation analyzers (IL Coagulation and ELECTRA Systems). The performance data (method comparison, precision) reflects the device's analytical performance independent of human interpretive input in the diagnostic process.

    7. Type of Ground Truth Used

    The "ground truth" for the test set was established through:

    • Comparative Analysis with Predicate Devices: For the method comparison studies, the results obtained using the HemosIL Factor X Deficient Plasma were compared against the results obtained using the legally marketed predicate devices (Hemoliance Factor X Deficient Plasma and IL Test Factor X Deficient Plasma). The predicate devices' results served as the reference or comparative "ground truth."
    • Control Materials: For the precision studies, Normal Control and Low Abnormal Control materials with known or expected Factor X activity levels were used. These control materials serve as an established reference for evaluating the accuracy and reproducibility of the device.

    8. Sample Size for the Training Set

    This information is not applicable and not provided in the given 510(k) summary. This device is a diagnostic reagent, not a machine learning or AI-based system that requires a "training set" in the typical sense of algorithm development. Its performance is based on chemical and electrochemical principles.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024082

    Validate with FDA (Live)

    Date Cleared
    2003-02-12

    (63 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K012768

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor VII Deficient Plasma is human plasma immunodepleted of Factor VII and intended for the in vitro diagnostic quantitative determination of Factor VII activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor VII Deficient Plasma is human plasma immunodepleted of Factor VII and intended for the in vitro diagnostic quantitative determination of Factor VII activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems. Abnormalities of the extrinsic pathway factors are determined by performing a modified prothrombin time (PT) test. Patient plasma is diluted and added to a plasma deficient in Factor VII. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the Factor VII in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HemosIL Factor VII Deficient Plasma device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device demonstrates substantial equivalence rather than meeting explicit acceptance criteria in the traditional sense of pre-defined thresholds. The performance is compared directly to predicate devices.

    Performance MetricAcceptance Criteria (Implicit - Substantial Equivalence to Predicate)Reported Device PerformanceComments
    Method Comparison
    Slope (vs. Hemoliance on ELECTRA E1400C)Close to 1.0 (indicating good agreement)0.9683This value is very close to 1.0, suggesting strong agreement with the predicate on the ELECTRA system.
    Correlation (r) (vs. Hemoliance on ELECTRA E1400C)Close to 1.0 (indicating strong linear correlation)0.9967This high correlation coefficient (r) indicates a very strong linear relationship between the new device and the predicate.
    Slope (vs. IL Test on ACL 300)Close to 1.01.0045Excellent agreement.
    Correlation (r) (vs. IL Test on ACL 300)Close to 1.00.9994Very strong correlation.
    Slope (vs. IL Test on ACL 6000)Close to 1.00.9646Strong agreement.
    Correlation (r) (vs. IL Test on ACL 6000)Close to 1.00.9989Very strong correlation.
    Slope (vs. IL Test on ACL 9000)Close to 1.00.9778Strong agreement.
    Correlation (r) (vs. IL Test on ACL 9000)Close to 1.00.9996Very strong correlation.
    Slope (vs. IL Test on ACL Futura)Close to 1.00.9678Strong agreement.
    Correlation (r) (vs. IL Test on ACL Futura)Close to 1.00.9943Very strong correlation.
    Within Run PrecisionNot explicitly stated as acceptance criteria, but generally low CV% values are desirable.The CV% values demonstrate good precision across various instruments and control levels. The ACL Advance shows slightly higher CV%s for both within-run and total precision compared to other instruments.
    Normal Control Mean % Factor VIINot explicitly stated, but expected to be around the nominal value.79.4 - 100.5%Values are within expected ranges for normal controls.
    Normal Control Within Run CV%Not explicitly stated, but low CV% is desirable.0.8 - 4.7%Generally low, with ACL Advance showing the highest (4.7%).
    Normal Control Total CV%Not explicitly stated, but low CV% is desirable.1.9 - 5.7%Generally low, with ACL Advance showing the highest (5.7%).
    Low Abnormal Control Mean % Factor VIINot explicitly stated.37.8 - 50.9%Values are within expected ranges for low abnormal controls.
    Low Abnormal Control Within Run CV%Not explicitly stated, but low CV% is desirable.1.2 - 4.6%Generally low, with ACL Advance showing the highest (4.6%).
    Low Abnormal Control Total CV%Not explicitly stated, but low CV% is desirable.2.5 - 5.9%Generally low, with ACL Advance showing the highest (5.9%).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Method Comparison: 60 citrated plasma samples (30 normal / 30 abnormal)
    • Sample Size for Precision: The text states "multiple runs (n=80)" using two levels of control. This means 80 measurements were taken for each control level on each instrument for the precision study.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective measurements of samples (either previously collected or prepared for the study), as opposed to prospective collection specifically for this study. The phrasing "evaluating 60 citrated plasma samples" and "using two levels of control" suggests existing or prepared samples within a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this device and study. The device is for quantitative determination (measuring Factor VII activity), not for diagnostic interpretation by human experts. The "ground truth" for method comparison and precision studies typically relies on the established accuracy and precision of the predicate device and the known concentrations of control materials.

    4. Adjudication Method for the Test Set

    This information is not applicable to this type of device and study, as there is no human interpretation or subjective assessment that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is an in vitro diagnostic reagent, not an AI-powered diagnostic imaging or interpretive tool. There are no human readers involved in the direct determination of Factor VII activity using this reagent.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable in the same way it would be for an AI algorithm. However, the performance data presented (method comparison and precision) is essentially "standalone" in the sense that it measures the performance of the device (reagent and instrument system) independently, without human interpretive input affecting the quantitative result. The device itself performs the measurement.

    7. The Type of Ground Truth Used

    • For Method Comparison: The measurements from the predicate devices (Hemoliance Factor VII Deficient Plasma and IL Test Factor VII Deficient Plasma) served as the comparative "ground truth" to establish substantial equivalence.
    • For Precision: The known, established concentrations or expected values of the control materials (Normal Control and Low Abnormal Control) served as the ground truth against which precision (variability) was assessed.

    8. The Sample Size for the Training Set

    This information is not applicable. This is an in vitro diagnostic reagent, not a machine learning or AI algorithm that requires a "training set" in the traditional sense. The development of such reagents involves chemical and biological formulation and optimization, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023839

    Validate with FDA (Live)

    Date Cleared
    2003-01-14

    (57 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K012768

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor V Deficient Plasma is human plasma immunodepleted of factor V and intended for the in vitro diagnostic quantitative determination of factor V activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor V Deficient Plasma is human plasma immunodepleted of factor V and intended for the in vitro diagnostic quantitative determination of factor V activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Abnormalities of the extrinsic pathway factors are determined by performing a modified prothrombin time (PT) test. Patient plasma is diluted and added to a plasma deficient in factor V. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the factor V in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the HemosIL Factor V Deficient Plasma device:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in terms of pre-defined thresholds for slope, correlation, or CV%. Instead, the implied acceptance criteria for substantial equivalence are based on demonstrating comparable performance to the predicate devices. The study data presented serves as the evidence that these implicit criteria (i.e., performance similar to predicate devices) are met.

    Performance MetricImplied Acceptance Criteria (Demonstrate equivalence to predicate)Reported Device Performance (HemosIL Factor V Deficient Plasma)
    Method Comparison vs. Hemoliance F-V Deficient Plasma (ELECTRA E1400C)Slope and 'r' comparable to predicateSlope: 1.0161, r: 0.9877
    Method Comparison vs. IL Test F-V Deficient Plasma (ACL 300)Slope and 'r' comparable to predicateSlope: 1.0005, r: 0.9932
    Method Comparison vs. IL Test F-V Deficient Plasma (ACL 6000)Slope and 'r' comparable to predicateSlope: 0.9771, r: 0.9940
    Method Comparison vs. IL Test F-V Deficient Plasma (ACL 9000)Slope and 'r' comparable to predicateSlope: 0.9746, r: 0.9945
    Method Comparison vs. IL Test F-V Deficient Plasma (ACL Futura)Slope and 'r' comparable to predicateSlope: 1.1331, r: 0.9814
    Within Run Precision (ACL 300 - Normal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 110.6, Within run CV%: 1.0, Total CV%: 3.1
    Within Run Precision (ACL 300 - Low Abnormal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 30.1, Within run CV%: 1.8, Total CV%: 3.5
    Within Run Precision (ACL 6000 - Normal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 115.1, Within run CV%: 1.5, Total CV%: 3.0
    Within Run Precision (ACL 6000 - Low Abnormal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 31.5, Within run CV%: 1.9, Total CV%: 2.8
    Within Run Precision (ACL 9000 - Normal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 113.7, Within run CV%: 1.2, Total CV%: 1.7
    Within Run Precision (ACL 9000 - Low Abnormal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 29.0, Within run CV%: 4.0, Total CV%: 5.6
    Within Run Precision (ACL Advance - Normal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 132.4, Within run CV%: 6.0, Total CV%: 7.6
    Within Run Precision (ACL Advance - Low Abnormal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 27.1, Within run CV%: 4.1, Total CV%: 5.7
    Within Run Precision (ELECTRA 1400C - Normal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 101.0, Within run CV%: 2.1, Total CV%: 3.5
    Within Run Precision (ELECTRA 1400C - Low Abnormal Control)CV% comparable to expected diagnostic precision for this assay typeMean % Factor V: 24.0, Within run CV%: 1.5, Total CV%: 2.6

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Method Comparison: 60 citrated plasma samples (30 normal / 30 abnormal).
    • Sample Size for Precision: Controls run over multiple runs (n=80) for each instrument.
    • Data Provenance: The document does not specify the country of origin or whether the samples were retrospective or prospective. It only states "citrated plasma samples."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device is an in vitro diagnostic reagent. The "ground truth" for Factor V activity would typically be established by the reference method (the predicate device, in this case) or a validated laboratory method using established controls.
    • The concept of "experts" in the sense of clinical reviewers (e.g., radiologists) and their qualifications is not applicable to this type of device and study. The ground truth is the measured laboratory value.

    4. Adjudication method for the test set:

    • Not applicable. This is a comparison of quantitative laboratory measurements, not a qualitative assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-assisted diagnostic device, nor does it involve human readers interpreting images or data. It's an in vitro diagnostic reagent.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is a standalone diagnostic reagent. Its performance is evaluated intrinsically through comparison to predicate devices and precision studies, without human interpretation as part of the primary diagnostic result.

    7. The type of ground truth used:

    • The ground truth for the method comparison study is implicitly the results obtained from the predicate devices (Hemoliance Factor V Deficient Plasma and IL Test Factor V Deficient Plasma) on their respective analyzer systems, using a common PT reagent (HemosIL RecombiPlasTin).
    • For the precision studies, the ground truth for the "normal" and "low abnormal" controls would be their assigned target values (or range) based on their manufacturing and calibration processes.

    8. The sample size for the training set:

    • This device is a reagent and not a machine learning or AI model, so there is no "training set" in that context. The development of the reagent itself would involve formulation, characterization, and quality control, but not in the sense of a data-driven training set for an algorithm.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no "training set" in the context of an algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1