Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K071572

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-23

    (45 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Capiox FX05 device is intended to be used during open heart surgical procedures to transfer oxygen and remove carbon dioxide from blood and to control the blood temperature during cardiopulmonary bypass for periods up to 6 hours. The Capiox FX05 is a Neonate/Infant oxygenator intended for use in procedures up to a maximum flow of 1.5 L/min. The patient weight and BSA should be considered upon use.

    The FX05 Hardshell Reservoir is also intended for use in vacuum assisted venous drainage procedures.

    The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

    Device Description

    The Capiox FX05 device is a modification of the Capiox RX05 device that has an arterial filter integrated into the design. The design of the oxygenator device is such that it utilizes an integrated oxygenator/heat exchanger module that provides for gas transfer (blood oxygenation and carbon dioxide removal) and for blood temperature control. The RX Oxygenator/Arterial Filter device also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. The arterial filter contained within the oxygenator module is comprised of 32 micron PET (polyethylene terephalate) mesh material that is wrapped around the outer circumference of the oxygenator fiber bundle.

    AI/ML Overview

    The provided document describes the Capiox® FX05 Hollow Fiber Oxygenator and Arterial Filter, a medical device intended for use during open heart surgical procedures. The submission is a 510(k) premarket notification, which means it aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than establishing completely new safety and effectiveness. As such, the "study" described is a series of in-vitro performance evaluations comparing the new device to its predicates.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific numerical targets. Instead, it states that the device "exhibited performance that is deemed to be substantially equivalent to the performance of the predicate devices." This substantial equivalence is based on a series of in-vitro tests where the new device's performance was compared to the established performance of the predicate devices. The implicit acceptance criterion is that the new device's performance on these tests is comparable to or better than the predicate devices, and that any differences do not raise new issues of safety or effectiveness.

    Test CategoryReported Device Performance
    Gas TransferPerformance "deemed to be substantially equivalent to predicate devices."
    Effects on Blood Components (Hemolysis)Performance "deemed to be substantially equivalent to predicate devices."
    Pressure DropPerformance "deemed to be substantially equivalent to predicate devices."
    Mechanical IntegrityPerformance "deemed to be substantially equivalent to predicate devices."
    Static Priming VolumePerformance "deemed to be substantially equivalent to predicate devices."
    Heat Exchanger PerformancePerformance "deemed to be substantially equivalent to predicate devices."
    Filtration EfficiencyPerformance "deemed to be substantially equivalent to predicate devices."
    Air HandlingPerformance "deemed to be substantially equivalent to predicate devices."
    Tubing Connection StrengthPerformance "deemed to be substantially equivalent to predicate devices."

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations."

    • Sample Size for Test Set: Not specified. The document repeatedly refers to "in-vitro performance evaluations" and "testing" without providing numerical sample sizes for each test (e.g., how many units were tested for pressure drop, or how many blood samples were used for hemolysis).
    • Data Provenance: In-vitro laboratory testing (likely conducted by Terumo Corporation or its designated testing facilities). The country of origin for the manufacturing and submission is Japan/USA, but the specific location of the in-vitro testing is not detailed. The data is prospective in the sense that these tests were conducted specifically for this submission to demonstrate equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable to this type of submission. The "ground truth" for mechanical and performance characteristics of medical devices is typically established through adherence to recognized international standards and scientifically validated test methods, rather than expert consensus on a test set. The predicate devices themselves represent the "ground truth" for comparison in a 510(k) process, as they are already legally marketed and presumed safe and effective.

    4. Adjudication method for the test set

    Not applicable. Device performance against technical specifications and predicate device performance is typically determined through direct measurement and comparison, not adjudication by human experts in the context of this 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical device (an oxygenator and filter), not an AI-powered diagnostic or decision support system. Therefore, MRMC studies and AI-assisted human reader performance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm.

    7. The type of ground truth used

    For the in-vitro performance evaluations, the "ground truth" is established by:

    • Predicate Device Performance: The primary "ground truth" for comparison is the documented and accepted performance characteristics of the predicate devices (Terumo's Capiox® RX05 Oxygenator/Reservoir K022115 and Terumo's Capiox® AF02X Arterial Filter K011804). The new device must perform comparably to these established devices.
    • Accepted Industry Standards: Implied adherence to relevant industry standards for medical device testing (e.g., for blood compatibility, mechanical integrity, gas exchange efficiency).
    • Biocompatibility Standards: Biocompatibility evaluation was done according to ISO 10993 guidelines, implying this standard serves as the ground truth for safety in this aspect.

    8. The sample size for the training set

    Not applicable. As this is a physical medical device and not an AI/ML algorithm, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. Since there is no training set, there is no ground truth for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071494

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-16

    (47 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Capiox FX Hollow Fiber Oxygenator and Arterial Filter is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.

    The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

    The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

    The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

    The Capiox FX15 is for use with patients when the required blood flow rate will not exceed 5.0 L/min. when used with a 4 Liter Reservoir; and when the required blood flow rate will not exceed 4.0 L/min. when used with a 3 Liter Reservoir.

    The Capiox FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

    The Capiox FX Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

    Device Description

    The Capiox FX device is a modification of the Capiox RX device that has an arterial filter integrated into the design. The design of the oxygenator device is such that it utilizes an integrated oxygenator/heat exchanger module that provides for gas transfer (blood oxygenation and carbon dioxide removal) and for blood temperature control. The RX Oxygenator/Arterial Filter device also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. The filter contained within the oxygenator module is comprised of 32 micron PET (polyethylene terephalate) mesh material that is wrapped around the outer circumference of the oxygenator fiber bundle. This design permits the oxygenation of blood and removal of carbon dioxide as blood passes through the fiber bundle and also facilitates blood filtration after the blood has been oxygenated. Air removal is accomplished by entrapment followed by permeation of the air into the hollow fibers of the oxygenator bundle - and subsequently is exhausted (along with carbon dioxide) via the gas outlet port. The materials used include polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene terephthalate, polyethylene and X-Coating™.

    AI/ML Overview

    Acceptance Criteria and Device Performance for TERUMO CAPIOX® FX Hollow Fiber Oxygenator

    The TERUMO CAPIOX® FX Hollow Fiber Oxygenator with Integrated Arterial Filter (K071494) was evaluated through a series of in-vitro performance tests to demonstrate substantial equivalence to predicate devices (Terumo's Capiox® RX15 Oxygenator/Reservoir K051997, Terumo's Capiox® RX25 Oxygenator/Reservoir K040210, and Terumo's Capiox® AF125X Arterial Filter K052205). Clinical studies were not deemed necessary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state numerical acceptance criteria for each test. Instead, it states that the device "exhibited performance that is deemed to be substantially equivalent to the performance of the predicate devices" based on the evaluations. The tests conducted and the general performance statements are summarized below:

    Performance Evaluation CategoryReported Device Performance
    Gas TransferDeemed substantially equivalent to predicate devices.
    Effects on Blood Components (Hemolysis)Deemed substantially equivalent to predicate devices.
    Pressure DropDeemed substantially equivalent to predicate devices.
    Mechanical IntegrityDeemed substantially equivalent to predicate devices.
    Static Priming VolumeDeemed substantially equivalent to predicate devices.
    Heat Exchanger PerformanceDeemed substantially equivalent to predicate devices.
    Filtration EfficiencyDeemed substantially equivalent to predicate devices.
    Air HandlingDeemed substantially equivalent to predicate devices.
    Tubing Connection StrengthDeemed substantially equivalent to predicate devices.
    Sterilization ConditionsValidated in accordance with AAMI guidelines to provide a SAL of 10-6. Ethylene oxide residues will not exceed maximum limits.
    BiocompatibilityBlood contacting materials found to be biocompatible as per ISO 10993.
    Polymer Coating (PMEA)Evaluated in an in-vivo animal study; no adverse conditions noted.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state specific sample sizes for the independent performance evaluation tests (test set). The studies were in-vitro evaluations.

    • Data Provenance: The studies were internal in-vitro performance evaluations conducted by Terumo Corporation (Ashitaka Factory), Fujinomiya City, Shizuoka Pref., Japan. The specific country of origin for the data generation is Japan. The studies are prospective in nature, as they were conducted to support the 510(k) submission for a new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable. The ground truth for the in-vitro physical and biological performance tests would be established by standard engineering and biocompatibility testing methodologies, not by expert consensus in the typical sense of diagnostic accuracy studies. The equivalence is drawn against established predicate devices, implying that their performance serves as a benchmark for "ground truth" or acceptable performance in these technical domains.

    4. Adjudication Method for the Test Set

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in diagnostic studies involving human interpretation. For in-vitro engineering and biological performance tests, the results are quantitative or qualitative assessments against defined standards or predicate device performance, not subject to subjective adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for evaluating human reader performance with and without AI assistance, which is not pertinent to the in-vitro performance evaluation of this medical device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance study was conducted. The performance evaluations listed (Gas Transfer, Hemolysis, Pressure Drop, etc.) are all conducted on the device itself, without human intervention for direct performance measurement. These are in-vitro tests designed to assess the device's inherent functional characteristics and safety.

    7. Type of Ground Truth Used

    The "ground truth" for the performance evaluations was based on:

    • Engineering Standards and Specifications: For tests like Pressure Drop, Mechanical Integrity, Static Priming Volume, and Tubing Connection Strength.
    • Biological Performance Benchmarks: For tests like Gas Transfer, Effects on Blood Components (Hemolysis), Filtration Efficiency, and Air Handling, the "ground truth" or acceptable performance was defined by demonstrating substantial equivalence to the performance of legally marketed predicate devices, and adherence to accepted medical device performance expectations for cardiopulmonary bypass equipment.
    • Regulatory Guidelines and Recognized Standards: For Sterilization (AAMI guidelines), Biocompatibility (ISO 10993), and the safety of the PMEA coating (demonstrated through prior FDA clearances and an in-vivo animal study).

    8. Sample Size for the Training Set

    This section is not applicable. There is no "training set" in the context of this device's performance evaluation. The device is not an AI/ML algorithm that learns from data. Its performance is inherent to its design and materials, and it is validated through direct testing.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1