Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K103227

    Validate with FDA (Live)

    Date Cleared
    2012-04-11

    (527 days)

    Product Code
    Regulation Number
    862.3250
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K992918, K001197, K002375, K981341, K000399

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oratect Oral fluid Drug Screen Device is a one-step lateral flow immunoassay device for the qualitative detection of d-Methamphetamine (ME), Delta-9-Tetrahydrocannabinol (TH), Cocaine (CO), d-Amphetamine (AM), Morphine (OP) and Phencyclidine (PC) in human oral fluid. The Oratect tests detect these drugs at the cutoff concentration listed below.

    TestCutoff
    Oratect® Oral Fluid Drug Screen Device d-Amphetamine50 ng/mL
    Oratect® Oral Fluid Drug Screen Device d-Methamphetamine50 ng/mL
    Oratect® Oral Fluid Drug Screen Device Delta-9-Tetrahydrocannabinol40 ng/mL
    Oratect® Oral Fluid Drug Screen Device Cocaine20 ng/mL
    Oratect® Oral Fluid Drug Screen Device Morphine40 ng/mL
    Oratect® Oral Fluid Drug Screen Device Phencyclidine10 ng/mL

    These products are for in vitro diagnostic use and intended for prescription point of care use.

    The Orated® Oral Fluid Drug Screen Device provides only preliminary drug test results. A more specific alternative method must be used in order to obtain a confirmed analytical result. Liquid chromatography/mass spectrometry/mass spectrometry (LC/MS/MS) is the preferred confirmatory method. Samples for confirmatory testing should be collected with the Oratect® Oral Fluid Collection Tube (50 mL polypropylene tube) provided. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

    OratectCheck™ Oral Fluid Controls (Negative and Positive controls of the analytes) are available but not supplied with the Oratect Oral Fluid Drug Screen Devices. The OratectCheck™ Oral Fluid Controls are used as quality control materials with Oratect® Oral Fluid Drug Screen Devices.

    Device Description

    The Oratect® Oral Fluid Drug Screen Device is a one-step lateral flow immunoassay device for the qualitative detection of d-Methamphetamine (ME), Delta-9-Tetrahydrocannabinol (TH), Cocaine (CO), d-Amphetamine (AM), Morphine (OP) and Phencyclidine (PC) in human oral fluid. The test principle is a competitive lateral flow immunochromatographic assay. The presence of analyte will produce a negative signal.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Oratect® Oral Fluid Drug Screen Devices

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in the provided document. However, the document consistently uses "Accuracy results against reference method >90%" as the performance metric for the subject device and "Accuracy results against reference method >89%" for the predicate. For the purpose of this summary, we will infer the acceptance criterion to be >90% accuracy against a reference method.

    The reported device performance based on the "Accuracy" row in the comparison table is also >90% against reference method.

    CharacteristicAcceptance Criteria (Inferred)Reported Device Performance
    Accuracy (against reference method)>90%>90%
    PrecisionNot explicitly stated, but predicate's <10% imprecision is noted. For the subject device, "overall precision was above 90%""overall precision was above 90%"
    Cross-reactants and Interference compoundsNot explicitly stated, but "very specific antibodies" and lists in package insert are noted."very specific antibodies for their respective test. Only a limited number of structure related chemicals cross react with the tests."

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "analytical performance validation studies, precision site studies and in method comparison studies".

    • Sample Size: The document does not explicitly state the total sample size used for the test set in the "method comparison studies." It mentions "three (3) Point of Care (POC) sites" for the precision study, but not the number of samples at each site or for the overall accuracy study.
    • Data Provenance: The data provenance is not explicitly stated (e.g., country of origin, specific demographics). The studies are referred to as "analytical" and "clinical" performance tests, implying human samples were involved. It is likely retrospective data collected for validation, as no prospective clinical trial details are provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not provide details on the number of experts or their qualifications used to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided text. The study primarily focuses on comparing the subject device's performance against reference laboratory methods (LC/MS or GC/MS) and its predicate devices, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    The device itself, the "Oratect® Oral Fluid Drug Screen Device," is a standalone immunoassay device. The performance described (accuracy, precision, specificity) is inherently its standalone, algorithm-only (or device-only) performance, as it produces a visually interpreted result without human-in-the-loop assistance beyond the interpretation of the test lines.

    7. Type of Ground Truth Used

    The ground truth used for the comparison studies was established by "LC/MS or GC/MS reference test method." These are highly specific and quantitative laboratory methods considered the definitive gold standard for drug detection.

    8. Sample Size for the Training Set

    The document does not provide information regarding a "training set" or its sample size. Immunoassay devices like the Oratect® generally do not involve machine learning algorithms that require a distinct training set in the typical sense. Their performance is validated through analytical and clinical studies as described.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a "training set" in the context of machine learning, there is no information on how a ground truth for such a set was established. The performance validation relies on comparing the device's results to established reference methods (LC/MS or GC/MS) for individual samples.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011057

    Validate with FDA (Live)

    Date Cleared
    2001-06-06

    (61 days)

    Product Code
    Regulation Number
    862.1675
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intercept™ Oral Fluid Drug Test Oral Specimen Collection Device is intended for use in the collection, preservation, and transport of oral specimens. Oral specimens collected with the Intercept™ Oral Specimen Collection Device can be used to detect cocaine and cocaine metabolites, cannabinoids, phencyclidine, amphetamine, methamphetamine, opiates, barbiturates and methadone with the OraSure Technologies Intercept™ MICRO-PLATE EIAs.

    Device Description

    The Intercept™ Oral Fluid Drug Test Oral Specimen Collection Device consists of a treated absorbent cotton fiber pad affixed to a nylon stick (Collection Pad) and a preservative solution in a plastic container (Specimen Vial). The Collection Pad is impregnated with a mixture of common salts and gelatin, creating a hypertonic environment which produces an osmotic gradient across the buccal and gingival mucosae. The Pad is placed in contact with the gingival mucosa (between the lower cheek and gum) which enhances the flow of mucosal transudate onto the absorptive cotton fibers of the Pad. Following the collection period, the Collection Pad is removed from the mouth and placed into a Specimen Vial. The vial contains a preservative solution which serves to inhibit the growth of oral microorganisms recovered on the Collection Pad. The vial is sealed with a plastic cap and transported to a laboratory for processing and testing.

    AI/ML Overview

    This document pertains to the Intercept™ Oral Fluid Drug Test Oral Specimen Collection Device. It is a collection device, so the performance criteria focuses on the ability to collect and preserve samples for subsequent drug testing, rather than direct diagnostic accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document does not explicitly list quantitative acceptance criteria for the device's performance in a table format, nor does it present specific performance data from a study.

    However, the core of the submission (K011057) is a claim of substantial equivalence to a predicate device (OraSure® Oral Specimen Collection Device; K970357). The "performance" in this context is implicitly tied to demonstrating that the new device can effectively collect and preserve samples for the same range of drug detection as the predicate and extend this capability to additional drugs when paired with specific assays.

    Implicit Acceptance Criteria (derived from the text):

    Acceptance Criteria CategorySpecificsReported Device Performance
    I. Core Functionality (Substantial Equivalence)- Collection Capability: Ability to effectively collect an oral fluid specimen. - Preservation Capability: Ability to preserve the collected specimen. - Transport Capability: Ability to transport the specimen securely to a laboratory."Both devices share major components (collection apparatus and transport containing a preservative solution) and are intended for collecting an oral fluid specimen, and for containing and transporting that specimen." This implies the new device performs these core functions comparably to the predicate.
    II. Compatibility with Downstream Assays (Extended Use)- Detection of Cocaine & Metabolites: Compatibility with assays for these substances. - Detection of Cannabinoids: Compatibility with assays for these substances. - Detection of Phencyclidine: Compatibility with assays for this substance. - Detection of Amphetamine: Compatibility with assays for this substance. - Detection of Methamphetamine: Compatibility with assays for this substance. - Detection of Opiates: Compatibility with assays for these substances. - Detection of Barbiturates: Compatibility with assays for these substances. - Detection of Methadone: Compatibility with assays for these substances."The oral specimens collected with the Intercept™ device, however, can be used to detect cocaine and cocaine metabolites, cannabinoids, phencyclidine, amphetamine, methamphetamine, opiates, barbiturates and methadone with the OraSure Technologies Intercept™ MICRO-PLATE EIAs as demonstrated in the premarket notifications for the assays (K001197, K000399, K992918, K002375, K993208, K981341, K001976, K002010)." This indicates successful validation of compatibility with specific assays.

    2. Sample Size and Data Provenance for Test Set:

    The provided 510(k) summary does not detail a specific "test set" sample size or data provenance for a direct clinical performance study of the collection device itself.

    Instead, the submission relies on:

    • Substantial Equivalence: By claiming substantial equivalence to the OraSure® Oral Specimen Collection Device (K970357), it implicitly asserts that the new device performs similarly for its core functions.
    • Referenced PMA/510(k)s for Assays: For the extended detection capabilities, the document refers to multiple separate premarket notifications (K001197, K000399, K992918, K002375, K993208, K981341, K001976, K002010) for the OraSure Technologies Intercept™ MICRO-PLATE EIAs. The performance data for detecting the specific drugs with oral specimens collected by this device would be detailed within those referenced assay submissions.

    Therefore, without access to those referenced documents, we cannot provide details on the test set sample size or data provenance for the drug detection capabilities.

    3. Number of Experts and Qualifications for Ground Truth of Test Set:

    The document does not mention the use of experts to establish ground truth specifically for the collection device's performance in this 510(k) summary. This is consistent with a substantial equivalence claim for a collection device, where the focus is on physical and functional similarity to an already cleared device, and the analytical performance (e.g., drug detection accuracy) is tied to the separate assays.

    4. Adjudication Method for Test Set:

    Not applicable, as no direct "test set" and ground truth establishment by experts for the collection device itself is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. The Intercept™ Oral Fluid Drug Test Oral Specimen Collection Device is a collection device, not an AI-assisted diagnostic tool or a device that requires human interpretation in the way an imaging AI might. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant.

    6. Standalone Performance Study (Algorithm Only):

    Not applicable. This device is a manual specimen collection device and does not involve an algorithm for standalone performance. Its "performance" is its ability to successfully collect and preserve an oral fluid specimen for laboratory analysis.

    7. Type of Ground Truth Used:

    For the collection device's core functionality, the "ground truth" is implied by its functional equivalence to the predicate device and its ability to deliver a specimen suitable for subsequent laboratory analysis.

    For the drug detection claims, the ground truth would be established during the development and validation of the specific EIAs (Enzyme Immunoassays) as detailed in the referenced K numbers. This type of ground truth typically involves:

    • Analytical Standards: Known concentrations of drugs and their metabolites.
    • Spiked Samples: Oral fluid samples to which known amounts of drugs have been added.
    • Clinical Samples with Confirmed Results: Oral fluid samples from actual subjects, where drug presence/absence and concentration have been independently confirmed by more definitive methods (e.g., GC/MS or LC/MS-MS, considered the gold standard for drug testing).

    8. Sample Size for the Training Set:

    The document does not specify a training set sample size. As a collection device, it does not typically involve traditional machine learning "training sets" in the same way an AI diagnostic algorithm would. Its design and functional validation would be based on engineering principles, materials science, and compatibility testing with chemical assays.

    9. How Ground Truth for Training Set Was Established:

    Not applicable, as a traditional "training set" with ground truth in the context of an algorithm or machine learning is not described or relevant for this type of device. The "ground truth" for the device's design and functionality would be based on:

    • Established methods for oral fluid collection.
    • Chemical stability validation for the preservative solution ensuring drug integrity.
    • Compatibility testing with the OraSure Technologies Intercept™ MICRO-PLATE EIAs (which themselves would have their own established ground truth).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1