Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K051890

    Validate with FDA (Live)

    Date Cleared
    2005-08-04

    (23 days)

    Product Code
    Regulation Number
    866.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K000155

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elecsys Anti-TPO immunoassay is for the in vitro quantitative determination of antibodies to thyroid peroxidase in human serum and plasma. The anti-TPO determination is used as an aid in the diagnosis of autoimmune thyroid diseases.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 1010/2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.

    Device Description

    The COBAS Elecsys® Anti-TPO Test System is based on a competitive immunoassay with streptavidin microparticles and electrochemiluminescence detection. First and second incubations are nine minutes in duration. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode, unbound substances are removed. Voltage is applied to the electrode inducing chemiluminescent emission which is measured by a photomultiplier. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve provided with the reagent bar code.

    AI/ML Overview

    This K051890 submission for the COBAS Elecsys® anti-TPO device is a modification of an existing device (K000155). The acceptance criteria and supporting evidence are presented for the modified device by showing its equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly met by demonstrating that the modified device performs "Same" as the predicate device (K000155) across several key technical specifications and performance characteristics. The primary change highlighted is related to Biotin Interference, where the predicate allowed up to 60 ng/ml and the modified device has an acceptance criteria of up to 10 ng/ml. This suggests an improvement or a more stringent and defined limit for the modified device.

    Acceptance Criteria CategoryPredicate Device (K000155) PerformanceCOBAS Elecsys® Anti-TPO (Modified Device) PerformanceMet Acceptance Criteria?
    Intended Usefor the quantitative determination of antibodies to thyroid peroxidase in human serum and plasma.SameYes
    Indication for Useas an aid in the diagnosis of autoimmune thyroid diseasesSameYes
    Sample Typehuman serum and plasmaSameYes
    Dilution PerformanceAutoantibodies are heterogeneous and this may lead to non-linear dilution phenomena for certain individual samples.SameYes
    Traceability / StandardizationWHO 66/387SameYes
    Test PrincipleCompetitive chemiluminescenceSameYes
    Measuring Range5 - 600 IU/mlSameYes
    Reagent Stability (Unopened)at 2-8°C up to the expiration date.SameYes
    Reagent Stability (Opened, 2-8°C)up to six weeks.SameYes
    Reagent Stability (Opened, on E170 / 2010)two weeksSameYes
    Reagent Stability (Opened, on 1010)one week (store alternately in refrigerator and on analyzer - at ambient temperature 20-25 °C; up to 20 hours in total).SameYes
    Interference - Bilirubinup to 66 mg/dlSameYes
    Interference - Hemolysisup to 1.5 g/dlSameYes
    Interference - Lipemiaup to 2100 mg/dlSameYes
    Interference - Biotinup to 60 ng/mlup to 10 ng/mlYes (More stringent/defined)
    Interference - Rheumatoid factorsup to 1500 U/mlSameYes

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The submission relies on demonstrating equivalence to a predicate device, assuming the predicate's performance claims, which would have been supported by its own testing data during its original clearance. The listed "Limitations" for interference are typically derived from dedicated interference studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. For in vitro diagnostic devices like this immunoassay, "ground truth" often refers to the clinical diagnosis or a well-established reference method. The document implicitly uses the predicate device as the comparison for "ground truth" in terms of performance characteristics.

    4. Adjudication method for the test set

    This information is not provided. Given the nature of an immunoassay, the "adjudication method" would likely involve comparing results to a reference method or clinical diagnoses, rather than human expert agreement on interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is an immunoassay device, not an AI-assisted diagnostic imaging or interpretation device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone immunoassay system (algorithm/device only). Its performance is evaluated intrinsically through various analytical studies (e.g., precision, measuring range, interference) to ensure it meets performance specifications, and in this case, demonstrates equivalence to a predicate device. The document implies these standalone performances were assessed to confirm the "Same" claims.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the purpose of this 510(k) submission, the "ground truth" is largely implied to be the established performance characteristics and validation of the predicate device (Elecsys Anti-TPO, K000155). The submission is based on demonstrating substantial equivalence to this legally marketed device across all critical performance parameters. For inherent analytical validation, the "ground truth" for parameters like accuracy and precision would typically rely on reference methods, spiked samples, and internal controls. For clinical utility, the "aid in the diagnosis of autoimmune thyroid diseases" would be linked to clinical diagnoses validated by other means (e.g., patient history, other diagnostic tests).

    8. The sample size for the training set

    This information is not provided. Immunoassay development typically involves extensive characterization and optimization, but the concept of a "training set" in the machine learning sense is not directly applicable to a traditional immunoassay. The development and optimization data would serve a similar purpose.

    9. How the ground truth for the training set was established

    As with the previous point, the concept of a "training set" with a defined "ground truth" in the AI/ML context doesn't directly apply here. Instead, during the development of an immunoassay, the "ground truth" for various analytical studies (e.g., sensitivity, specificity, linearity, interference) would be established using:

    • Reference materials/standards: For quantitative measurements and calibration.
    • Known positive and negative samples: For initial analytical performance characterization.
    • Clinical samples with established diagnoses: To assess clinical performance and aid in setting cut-off values.

    The submission focuses on demonstrating that the modified device performs identically to the predicate device across critical parameters, with a specified change in biotin interference tolerance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1