Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133155

    Validate with FDA (Live)

    Date Cleared
    2014-04-07

    (172 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K000011

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To facilitate the placement of devices during diagnostic and interventional procedures.

    Device Description

    Utilizing proprietary processes, these guides are constructed from a steerable, metallic core (Nitinol) with a radiopaque polymer (polyurethane) jacket and a hydrophilic coating is applied over the jacket. Guidewires are available up to 260cm length and in diameters from 0.018" to 0.038" depending on specific design requirements. Guidewires may have a straight or angled distal tip.

    AI/ML Overview

    This document summarizes the acceptance criteria and supporting studies for the Lake Region Medical Hydrophilic Coated Guidewires (K133155). The information primarily comes from the 510(k) Summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes qualification testing to demonstrate equivalence to a predicate device, rather than explicit acceptance criteria with numerical targets and reported performance values. The tests performed are listed, implying that the device was found to be acceptable against established criteria for each test. Due to the lack of specific numerical acceptance criteria and performance data in the provided text, the table below lists the tests performed and notes that they were conducted to demonstrate equivalence and found acceptable.

    Test CategoryTest NameAcceptance Criteria (Implicit)Reported Device Performance
    Bench TestingDimensionalMet established specific inspection criteria for visual/tactile, dimensional attributes.Tested and found acceptable to demonstrate equivalence.
    FDA Device CompatibilityMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Tensile StrengthMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Tip FlexibilityMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Coating Adherence/IntegrityMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Catheter CompatibilityMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Torque ControlMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    FDA Combined LoadMet established specific inspection criteria. Test methods developed using FDA guidance.Tested and found acceptable to demonstrate equivalence.
    Packaging StudyMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ParticulateMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Kink ResistanceMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Linear StiffnessMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Lateral StiffnessMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO VisualComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO FractureComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO FlexComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO CorrosionComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO Strength of UnionComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    ISO RadiopacityComplied with ISO 11070:1998 and established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Body StiffnessMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Tip IntegrityMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Jacket DurabilityMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Tip Puncture ResistanceMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    Coating Performance TestingMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    3 Point BendMet established specific inspection criteria.Tested and found acceptable to demonstrate equivalence.
    BiocompatibilityISO CytotoxicityPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Klingman Maximization TestPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Irritation / Intracutaneous ReactivityPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Systemic ToxicityPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Rabbit PyrogenPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ASTM HemolysisPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Complement Activation AssayPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO ThrombogenicityPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Lee and White CoagulationPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Prothrombin Time AssayPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.
    ISO Unactivated Partial Thromboplastin Time AssayPerformed per ISO 10993 series and found acceptable.Tested and found acceptable.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each individual bench or biocompatibility test. It states that "Test pieces were tested and inspected." This suggests a laboratory setting for testing physical and biocompatibility properties. The data provenance is presumed to be from internal lab testing conducted by Lake Region Medical ("Lake Region Medical performed testing"). The study is analogous to a pre-market, prospective validation of the device's physical and biological characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to the type of testing described (bench and biocompatibility). The "ground truth" for these tests would be defined by the specified test methods (e.g., ISO standards, FDA guidance documents) and the scientific principles behind them, rather than expert consensus on clinical cases.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the testing involves objective measurement and analysis against predefined standards for physical and chemical properties, not subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a guidewire, a physical medical instrument, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies comparing human reader performance with and without AI assistance are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. See point 5. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the bench testing is based on:

    • Established specific inspection criteria for visual/tactile and dimensional attributes.
    • FDA Coronary and Cerebrovascular Guidance for certain test method development.
    • ISO 11070:1998 for guidewire test methods.

    For biocompatibility testing, the ground truth is based on:

    • ISO 10993 series of standards.

    These standards and guidance documents define the acceptable limits and methodologies for evaluating the device's physical and biological properties.

    8. The Sample Size for the Training Set

    Not applicable. This product is a physical medical device, not an AI model, so there is no training set mentioned or implied.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1