Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K230694

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-13

    (31 days)

    Product Code
    Regulation Number
    876.1310
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN190037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NaviCam Stomach Capsule is intended for visualization of the stomach of adults (>/=22 years) with BMI < 38. The system can be used in clinics and hospitals, including ER settings.

    The NaviCam Tether is an accessory of the NaviCam Stomach Capsule. It is intended to aid the Capsule for visualizing the esophagus (not magnetically maneuvered) prior to the Capsule's release into the stomach capsule endoscopy (MCCE) procedure.

    Device Description

    The NaviCam Xpress Stomach System is a novel endoscopic capsule imaging system intended to obtain images of the stomach. It differs from passive capsule endoscopy systems in that it uses magnetic fields to allow the position of the capsule within the stomach to be controlled by an operator.

    The NaviCam Xpress Stomach System includes the following key components:

      1. Ingestible capsule (CP-US-7005) for obtaining images.
      1. Data recorder (MC-US-1006) for logging image data.
      1. Locator (SB-US-2006) for turning on the capsule and for determining if the capsule is still in the body.
      1. Controller, NaviEC-2000 (MC-US-1002) with the NaviCtrl software that allows the navigation of the capsule within the stomach.
      1. ESView software for review of the images obtained by the capsule and generating reports.
    • Optional Accessory: NaviCam Tether (ES-US-7005) for aiding the capsule in visualizing 6. the esophagus.
    AI/ML Overview

    This document is a 510(k) summary for a software addition, not a comprehensive clinical study report. Therefore, much of the requested information about clinical performance, ground truth, and expert involvement is not present in this document.

    Here's a breakdown of what can be extracted based on the provided text:

    Device: NaviCam Xpress Stomach System (addition of GastroScan feature to NaviCtrl software)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) for the device's diagnostic capabilities. Instead, it focuses on demonstrating that the software addition does not negatively impact the existing cleared device's performance or introduce new risks. The reported "performance" is that the device "performs as intended based on the non-clinical bench testing."

    Acceptance Criteria (Implied by the submission)Reported Device Performance
    Software updates function as intendedVerified through Software Verification Testing
    Device remains safe and effective (no new risks introduced by software update)Demonstrated by Human Factors Engineering/Usability Testing/Evaluation, performance bench testing, software risk analysis, and cybersecurity FMEA risk analysis.
    No changes to existing indications for useThe device retains the same Indications for Use as the predicate device.
    No degradation in performance compared to the predicate deviceNot explicitly quantified, but implied by successful non-clinical testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified for any performance testing related to diagnostic accuracy. The non-clinical performance and bench testing would have used specific test cases, but the number of such cases is not provided.
    • Data Provenance: The new software features were evaluated through non-clinical performance and bench testing. This is typically internal testing. No information about country of origin for this testing or whether it was retrospective or prospective is provided. The document states that "No additional Animal and clinical performance testing were conducted."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this submission, as no new clinical performance testing was conducted. Ground truth for previous clinical evidence (mentioned in point 9) would have been established, but details are not provided here.

    4. Adjudication Method for the Test Set

    Not applicable for this submission, as no new clinical performance testing was conducted.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a MRMC comparative effectiveness study was not done for this submission. The document explicitly states "No additional clinical literature has been provided in this submission." It refers to previous clinical evidence from predicate devices but does not provide details of those studies in this document.

    6. Standalone Performance Study

    A standalone study of the algorithm's core diagnostic performance (e.g., sensitivity/specificity for detecting stomach abnormalities) was not performed as part of this specific submission. The submission focuses on the safety and efficacy of a software addition to an already cleared device. The "performance bench testing" mentioned would assess the functionality of the new software feature (GastroScan) rather than clinical diagnostic accuracy.

    7. Type of Ground Truth Used

    For this specific submission, the "ground truth" for the non-clinical performance and bench testing would be defined by the expected outcomes of the software functions and hardware interactions as specified in design documents and testing protocols. No clinical ground truth (e.g., pathology, expert consensus on patient cases) was used for this particular submission's evaluation.

    8. Sample Size for the Training Set

    No training set information is provided, as this submission is for a software feature addition to a previously cleared device. It's unlikely that machine learning model training was a primary component of this specific update.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for this submission (see point 8).

    Summary of this Submission's Focus:

    This 510(k) submission (K230694) is for a software update (GastroScan feature) to an already cleared device (NaviCam Xpress Stomach System, K203192). The primary goal of this submission is to demonstrate that this software addition does not change the indications for use, does not introduce new risks, and does not negatively impact the performance of the existing device. Therefore, the evidence focuses on non-clinical software verification, human factors, and risk analysis, rather than new clinical trials or diagnostic performance metrics. The clinical evidence for the overall system's diagnostic capabilities would have been established in previous submissions (DEN190037 and K203192).

    Ask a Question

    Ask a specific question about this device

    K Number
    K221590

    Validate with FDA (Live)

    Date Cleared
    2022-12-02

    (183 days)

    Product Code
    Regulation Number
    876.1300
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    DEN190037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

    Device Description

    The NaviCam Small Bowel Capsule Endoscopy System is an endoscopic capsule imaging system intended to obtain images of the small bowel. It is comprised of the following components:

    1. Capsule (AKES-11SW, AKES-11SI): The disposable, ingestible NaviCam Small Bowel Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via an RF communication channel to the NaviCam Data Recorder located outside the body.
    2. Data recorder (AKR-1, AKRI-1): The Data Recorder is an external receiving and recording unit that receives and stores the acquired images from the capsule.
    3. ESView Software: The ESView is a software application for processing, analyzing, storing, and viewing the acquired images collected from the NaviCam Data Recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports and a capsule endoscopy atlas.
    4. Locator: The Locator is a handheld device that is used to turn the NaviCam Capsule on. It is also used for determining if the capsule is still in the body when the patient is not sure whether he/she expelled it.
    AI/ML Overview

    The NaviCam Small Bowel Capsule Endoscopy System was assessed for its performance primarily through a comparative clinical study against a predicate device, the PillCam SB3 Capsule Endoscopy System, and also through various bench/in-vitro tests.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Stated or Implied)Reported Device Performance
    Bench/In-Vitro TestsSuccessfully passed all listed tests.
    Biting TestAbility to withstand applied forces similar to accidental biting.Pass
    Angular Resolution TestMeasurement of MTF using ISO 12233 slanted edge methodology and new angular resolution method using LEDs. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Temperature Safety TestTemperature change during operation within safe limits. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    pH TestIntegrity of the capsule during exposure to simulated extreme pH levels. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Image Intensity UniformityUniformity of image intensity. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Image Frame Rate TestHigher frame rate provides good transmission property. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Geometric Distortion TestDetermination of geometric distortion and local magnification. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Field of View (FOV) TestDetermination of FOV value. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Battery Life TestBattery life of at least 8 hours and capturing over 57,500 images.Pass (demonstrated to last at least 8 hours and capture over 57,500 images).
    Image Resolution TestTesting of image resolution. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    Magnetic Field TestMeasurement of magnetic flux density on capsule surface and non-optical bottom, and determination of safety distance. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
    DOV TestMeasurement of MTF in air and underwater at different distances within claimed DOV range using ISO 12233 slanted edge methodology and angular resolution method. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
    Color and Gray Scale TestEvaluation of optical performance. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
    Data Integrity TestData transmission between capsule, data recorder, and ESView software. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
    Clinical Study
    Diagnostic Overall Percent Agreement with predicate deviceNot explicitly stated as a strict threshold, but achieving substantial equivalence to the predicate device (PillCam SB3) in diagnostic performance. The document implies that the observed agreement rate demonstrates similarity.89.66% (81.50%, 94.46%) overall percent agreement with the PillCam SB3. Overall percent agreement Kappa of 0.6652 (0.4653, 0.8652). This was deemed to demonstrate similar performance and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions that the clinical study was a prospective study (NCT05086471).
    However, the specific sample size (number of patients or cases) used for the test set is not provided in the given text.
    The data provenance (e.g., country of origin) for the clinical study is also not stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the given text.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text. The text only states that the NaviCam system was compared to the PillCam SB3 in terms of diagnostic performance, but it doesn't detail how discrepancies or consensus was handled for establishing ground truth or comparing diagnostic findings.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement with AI vs Without AI Assistance

    A comparative clinical study was performed, but it was to compare the device's performance (NaviCam Small Bowel Capsule Endoscopy System) to a predicate device (PillCam SB3 Capsule Endoscopy System), not specifically an MRMC study assessing human reader improvement with or without AI assistance. The study evaluates the diagnostic agreement between the two capsule endoscopy systems. Therefore, the effect size of human readers improving with AI vs. without AI assistance is not applicable/not reported in this context.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The NaviCam Small Bowel Capsule Endoscopy System includes "ESView Software...for processing, analyzing, storing, and viewing the acquired images... The software also includes a reporting function". While the software performs analysis, the clinical study appears to evaluate its overall diagnostic performance in detecting abnormalities, which would typically involve human review of the generated images/reports. The text does not explicitly state if a standalone algorithm-only performance study was conducted separate from human interpretation. The reported "diagnostic Overall Percent Agreement" is for the system, which implies the combined interpretation of the images.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study is based on the detection of abnormalities of the small bowel by both the NaviCam system and the predicate PillCam SB3 system. The phrase "diagnostic Overall Percent Agreement" implies that the agreement was measured against the findings of another diagnostic tool (the predicate device), which often serves as a form of "ground truth" in equivalence studies when a gold standard (like pathology) is not universally available for every finding. The document does not explicitly state that pathology or outcomes data were used as the definitive ground truth for every finding. It strongly suggests the predicate device's findings were used as the reference point for comparison.

    8. The Sample Size for the Training Set

    This information is not provided in the given text. The document focuses on the performance study data, not the training dataset for any underlying AI/software components.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the given text, as details about a training set are absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203192

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-05-04

    (188 days)

    Product Code
    Regulation Number
    876.1310
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    DEN190037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NaviCam Xpress Stomach System is intended for visualization of the stomach of adults (≥22 years) with BMI <38. The system can be used in clinics and hospitals, including ER settings.

    Device Description

    The NaviCam Xpress Stomach System is a endoscopic capsule imaging system intended to obtain images of the stomach. It differs from passive capsule endoscopy systems in that it uses magnetic fields to allow the position of the capsule within the stomach to be controlled by an operator. The NaviCam Xpress Stomach System includes the following key components: Ingestible capsule (AKEM-11SW) for obtaining images, Data recorder (AKR-1) for logging image data, Locator (AKS-1) for turning on the capsule and for determining if the capsule is still in the body, Controller (NaviEC-2000) with the NaviCtrl software that allows the navigation of the capsule within the stomach, ESView software for review of the images obtained by the capsule and generating reports.

    AI/ML Overview

    The provided text is a 510(k) summary for the NaviCam Xpress Stomach Capsule Endoscope System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the acceptance criteria or a study proving the device meets those criteria, especially in terms of diagnostic performance or clinical accuracy. The "Performance Data" section solely discusses bench/in-vitro testing related to the controller's functionality and compliance with electrical safety standards, not the diagnostic performance of the capsule endoscope system itself.

    Therefore, most of the requested information regarding acceptance criteria, performance, sample sizes, expert involvement, and ground truth for a diagnostic study cannot be extracted from this document.

    Here's what can be inferred or explicitly stated based on the provided text, alongside what is missing:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Diagnostic Performance: (e.g., Sensitivity, Specificity, Accuracy for detecting gastric abnormalities)NOT PROVIDED IN THIS DOCUMENT. The document focuses on the controller's performance and system compliance, not the diagnostic capabilities of the capsule endoscope.
    NaviCam Xpress Controller meets performance requirements in specifications and implements software functions.Passed: Verified that the NaviCam Xpress Controller meets its specifications and implements software functions as stated in its User Manual.
    Magnetic field intensity range meets specifications with various examination beds.Passed: Verified that the magnetic field intensity range meets specifications when used with examination beds/tables provided by users.
    Mechanical motion range meets specifications.Passed: Verified that the mechanical motion range meets specifications.
    Titanium clips do not affect data communication, and clips do not move during magnetic head rotation.Passed: Data communication not affected, and clip position does not change.
    Compliance with IEC 60601-1, ANSI/AAMI ES60601-1, CAN/CSAC22.2 No. 60601-1:14, EN 60601-1:2006, AAMIIEC 60601-1-2:2014, EN 60601-1-2:2015.Passed: The system was found to comply with all applicable requirements of these standards.

    The following information cannot be extracted from the provided text:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    • 8. The sample size for the training set
    • 9. How the ground truth for the training set was established

    Reasoning for Missing Information:

    This 510(k) summary focuses on demonstrating substantial equivalence for a modification to an already cleared device (the NaviCam Stomach System). The "Performance Data" section details bench/in-vitro testing and compliance with electrical safety and EMC standards. It explicitly states, "The NaviCam Xpress Stomach System successfully passed the following test," and then lists functional and safety-related tests for the controller. It does not discuss any clinical diagnostic performance studies comparing the NaviCam Xpress Stomach System's ability to detect stomach abnormalities against a ground truth, which would involve patient data, expert review, and metrics like sensitivity or specificity.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1