510(k) Data Aggregation

    K Number
    K251293

    Validate with FDA (Live)

    Device Name
    CardioVision
    Manufacturer
    Date Cleared
    2025-11-21

    (210 days)

    Product Code
    Regulation Number
    870.2200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iCardio.ai CardioVision™ AI is an automated machine learning–based decision support system, indicated as a diagnostic aid for patients undergoing an echocardiographic exam consisting of a single PLAX view in an outpatient environment, such as a primary care setting.

    When utilized by an interpreting clinician, this device provides information that may be useful in detecting moderate or severe aortic stenosis. iCardio.ai CardioVision™ AI is indicated in adult populations over 21 years of age. Patient management decisions should not be made solely on the results of the iCardio.ai CardioVision™ AI analysis. iCardio.ai CardioVision™ AI analyzes a single cine-loop DICOM of the parasternal long axis (PLAX).

    Device Description

    The iCardio.ai CardioVision™ AI is a standalone image analysis software developed by iCardio.ai Corporation, designed to assist in the review of echocardiography images. It is intended for adjunctive use with other physical vital sign parameters and patient information, but it is not intended to independently direct therapy. The device facilitates determining whether an echocardiographic exam is consistent with aortic stenosis (AS), by providing classification results that support clinical decision-making.

    The iCardio.ai CardioVision™ AI takes as input a DICOM-compliant, partial or full echocardiogram study, which must include at least one parasternal long-axis (PLAX) view of the heart and at least one full cardiac cycle. The device uses a set of convolutional neural networks (CNNs) to analyze the image data and estimate the likelihood of moderate or severe aortic stenosis. The output consists of a binary classification of "none/mild" or "moderate/severe," indicating whether the echocardiogram is consistent with moderate or severe aortic stenosis. In cases where the image quality is insufficient, the device may output an "indeterminate" result.

    The CNNs and their thresholds are fixed prior to validation and do not continuously learn during standalone testing. These models are coupled with pre- and post-processing functionalities, allowing the device to integrate seamlessly with pre-existing medical imaging workflows, including PACS, DICOM viewers, and imaging worklists. The iCardio.ai CardioVision™ AI is intended to be used as an aid in diagnosing AS, with the final diagnosis always made by an interpreting clinician, who should consider the patient's presentation, medical history, and additional diagnostic tests.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) clearance letter for CardioVision™:

    Acceptance Criteria and Reported Device Performance

    MetricAcceptance CriteriaReported Device Performance (without indeterminate outputs)Reported Device Performance (including indeterminate outputs)
    AUROCExceeds predefined success criteria0.945Not explicitly stated but inferred to be similar to Sensitivity/Specificity
    SensitivityExceeds predefined success criteria and predicate device0.896 (95% Wilson score CI: [0.8427, 0.9321])0.876 (95% Wilson score CI: [0.8213, 0.9162])
    SpecificityExceeds predefined success criteria and predicate device0.872 (95% Wilson score CI: [0.8384, 0.8995])0.866 (95% Wilson score CI: [0.8324, 0.8943])
    PPVNot explicitly stated as acceptance criteria0.734 (95% Wilson score CI: [0.673, 0.787])Not explicitly stated
    NPVNot explicitly stated as acceptance criteria0.955 (95% Wilson score CI: [0.931, 0.971])Not explicitly stated
    Rejection RateNot explicitly stated as acceptance criteria1.077% (7 out of 650 studies)1.077%

    Note: The document explicitly states that the levels of sensitivity and specificity exceed the predefined success criteria and those of the predicate device, supporting the claim of substantial equivalence. While exact numerical thresholds for the acceptance criteria aren't provided in terms of specific values, the narrative confirms they were met.

    Study Details

    FeatureDescription
    1. Sample size used for the test set and the data provenanceSample Size: 650 echocardiography studies from 608 subjects.Data Provenance: Retrospective, multi-center performance study from 12 independent clinical sites across the United States.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those expertsNumber of Experts: Not explicitly stated as a specific number, but referred to as "experienced Level III echocardiographers."Qualifications: "Experienced Level III echocardiographers."
    3. Adjudication method for the test setMethod: A "majority vote approach" was used in cases of disagreement among the experts.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistanceMRMC Study: No, an MRMC comparative effectiveness study is not detailed in this document. The study described is a standalone performance evaluation of the AI. (A "human factors validation study" was conducted to evaluate usability, where participants successfully completed the critical task of results interpretation without errors, but this is not an MRMC study comparing human performance with and without AI assistance on diagnostic accuracy).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was doneStandalone Performance Study: Yes, the document describes a "standalone study" with the primary objective to "evaluate the software's ability to detect aortic stenosis." The reported performance metrics (AUROC, Sensitivity, Specificity, etc.) are for the algorithm's performance alone.
    6. The type of ground truth usedGround Truth Type: Expert consensus based on "echocardiographic assessments performed by experienced Level III echocardiographers," with a majority vote for disagreements.
    7. The sample size for the training setTraining Set Size: Not specified in the provided document. The document states, "No data from these [test set] sites were used in the training or tuning of the algorithm."
    8. How the ground truth for the training set was establishedTraining Set Ground Truth: Not explicitly detailed in the provided document. It can be inferred that similar methods (expert echocardiographic assessments) would have been used for training data, but the specifics are not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240013

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-23

    (265 days)

    Product Code
    Regulation Number
    870.2200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EchoGo Heart Failure 2.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilised by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF).

    EchoGo Heart Failure 2.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 2.0 analysis.

    EchoGo Heart Failure 2.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%.

    Device Description

    EchoGo Heart Failure 2.0 takes as input a 2D echocardiogram of an apical four chamber tomographic view and reports as output a binary classification suggestive of the presence, or absence of heart failure with preserved ejection fraction (HFpEF). EchoGo Heart Failure 2.0 also provides users with an EchoGo Score ranging from 0 to 100% to support the binary classification. The EchoGo Score informs the binary classification when referenced against the pre-determined decision threshold (50%).

    To aid in the interpretation of the EchoGo Score, a comparative visual analysis is provided. A histogram format displays the reported EchoGo Score output against a population of patients with known disease status (Independent Testing Dataset). This allows the user to interpret the EchoGo Score relative to the decision threshold of 50%.

    EchoGo Heart Failure 2.0 should receive an input echocardiogram acquired without contrast and contain at least one full cardiac cycle.

    EchoGo Heart Failure 2.0 is fully automated and does not comprise a graphical user interface.

    EchoGo Heart Failure 2.0 is intended to be used by an interpreting clinician as an aid to diagnosis for HFpEF. The ultimate diagnostic decision remains the responsibility of the interpreting clinician using patient presentation, medical history, and the results of available diagnostic tests, one of which may be EchoGo Heart Failure 2.0.

    EchoGo Heart Failure 2.0 is a prescription only device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance LimitReported Device Performance
    I. Device Performance (Sensitivity & Specificity)Implicit within reporting of performance: The device must demonstrate sufficient sensitivity and specificity for detecting HFpEF as a diagnostic aid. The specific acceptance limits are not explicitly stated as numerical thresholds but are demonstrated by the reported performance being "substantively equivalent to the predicate device and met pre-specified levels of performance."Sensitivity: 90.3% (95% CI: 88.5, 92.4%) when removing "no classification" studies. 84.9% (95% CI: 83.0, 87.5%) when including "no classification" studies. Specificity: 86.1% (95% CI: 83.4, 88.3%) when removing "no classification" studies. 78.6% (95% CI: 75.3, 81.1%) when including "no classification" studies.
    II. Accuracy of EchoGo Score (AUROC & Goodness-of-Fit)Implicit within reporting of performance: The EchoGo Score must be accurate and align with known and expected proportions of HFpEF. Statistical significance (p-value > 0.05) for the Hosmer-Lemeshow Test and a sufficiently high AUROC are expected.Area Under the Receiver Operating Characteristic Curve (AUROC): 0.947 (95% CI: 0.934, 0.958) when removing "no classification" studies. 0.937 (95% CI: 0.924, 0.949) when considering all studies. Hosmer-Lemeshow Test for goodness-of-fit: p=0.304 (not significant, indicating acceptable fit).
    III. Proportion of Non-Diagnostic OutputsA priori acceptance limits: The proportion of "no classification" outputs must be within pre-specified limits (the exact numerical limit is not provided, but the text states it was "within a priori acceptance limits").7.4% (116 out of 1,578 studies) were categorized as "No Classification."
    IV. Precision (Repeatability and Reproducibility)Implicit within reporting of performance: The device must demonstrate high repeatability and acceptable reproducibility in its classification output.Repeatability: 100% in all measures. Reproducibility: 82.6% Positive Agreement and 82.4% Negative Agreement.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: 1,578 patients (785 controls and 793 cases).
      • Data Provenance: Retrospective case:control study. The data was collected from multiple independent clinical sites spanning five states in the US.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document states that the ground truth was established by "ground truth classifications of cases (HFpEF) or controls," but it does not specify the number or qualifications of experts who established this ground truth for the test set.
    3. Adjudication method for the test set:

      • The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1). It only refers to "ground truth classifications," implying these were already established.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study evaluating human readers with and without AI assistance was not done. The study focuses on the standalone performance of the device. The device is intended as a "diagnostic aid" for use "by an interpreting clinician," but its performance evaluation presented here is not an MRMC study.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done. The reported sensitivity, specificity, AUROC, and precision values are for the device (algorithm) itself without human intervention in the classification output for the test set. The device provides a "binary classification suggestive of the presence, or absence of heart failure with preserved ejection fraction (HFpEF)" and an "EchoGo Score."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth was based on "ground truth classifications of cases (HFpEF) or controls," and "known and expected proportions of HFpEF." While not explicitly stated as "expert consensus," this terminology strongly implies clinical diagnoses were used to establish the HFpEF status for each patient in the dataset. It does not mention pathology or outcomes data specifically for ground truth.
    7. The sample size for the training set:

      • The sample size for the training set is not explicitly stated in the provided text. It mentions that the "Subject device AI model was trained on more data and with additional preprocessing steps and data augmentations" compared to the predicate device, and the testing data cohort was a "22.9% increase in data beyond the testing data cohort utilized for the 510k submission of EchoGo Heart Failure 1.0." However, the exact size of the training set is not provided.
    8. How the ground truth for the training set was established:

      • The document does not explicitly describe how the ground truth for the training set was established. It only states that the AI model was "trained on more data" with "additional preprocessing steps and data augmentations." It is highly probable it was established similarly to the test set ground truth (i.e., using clinical diagnoses or expert classifications), given the nature of the diagnostic task.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223905

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-06

    (282 days)

    Product Code
    Regulation Number
    870.2200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vivio® System is indicated to non-invasively estimate left ventricular end-diastolic pressure (LVEDP) is above or below 18mmHg. This measurement can aid in the diagnosis of heart failure when used by qualified healthcare professionals as an adjunct alongside other clinically relevant information. For use in adults only.

    Device Description

    The Vivio System consists of an Arm Cuff system (BP cuff electronics, connection tubinq), EKG patch, and a software application that runs on an off-the-shelf computer tablet. Data is collected non-invasively at the brachial artery and via a single-lead EKG Patch. An off-the-shelf blood pressure arm cuff is attached to the patient's upper arm as if taking a blood pressure measurement. The Arm Cuff is attached to the blood pressure cuff with the quick connect end connected to the pneumatic pump within the Arm Cuff Electronics Enclosure. Two off-the-shelf electrodes are attached to the snap connections on the bottom of the EKG Patch and then connected on the left side of the patient's chest. Both the pneumatic pump attached to the Arm Cuff and EKG Patch are then powered on by pressing and releasing their respective power buttons. The Vivio App is opened on an off-the shelf computer tablet, and the on-screen prompts are followed to connect the Arm Cuff and the EKG Patch. The user is prompted to enter required patient and user data, initiating a patient recording session. Data from the recording session is processed by a proprietary algorithm and results reported on the computer tablet.

    AI/ML Overview

    The provided document focuses on the 510(k) summary for the Vivio® System, which is an "Adjunctive Cardiovascular Status Indicator" device. It outlines the regulatory classification, device description, indications for use, technological characteristics, and summaries of non-clinical and clinical tests.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Indications for Use: The Vivio® System is indicated to non-invasively estimate left ventricular end-diastolic pressure (LVEDP) as above or below 18mmHg. This measurement aids in the diagnosis of heart failure when used by qualified healthcare professionals as an adjunct alongside other clinically relevant information, for use in adults only.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states "The primary efficacy objectives for the study were met." The acceptance criteria for efficacy are directly linked to diagnostic performance metrics:

    Acceptance Criteria (Performance Metric)Reported Device Performance (95% Confidence Interval)
    Sensitivity80% (64-91)
    Specificity80% (73-86)

    Note: The document does not explicitly state these as acceptance criteria but rather as "primary efficacy objectives for the study." In the context of a 510(k) submission, meeting these pre-specified performance thresholds demonstrates the device's ability to achieve its intended use. The LVEDP threshold for diagnosis is 18mmHg.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set):

      • Cath Lab Cohort: 195 subjects scheduled for non-emergent left heart catheterization.
      • Non-Aged Healthy Cohort: 101 subjects.
      • Aged Healthy Cohort: 40 subjects.
      • Total Enrolled: 195 + 101 + 40 = 336 subjects.
      • However, the statistical analysis was performed on a "subset of validation data that had passed the completeness and quality requirements outlined in the data management plan and procedure." The exact size of this final validation subset, specifically used for the reported sensitivity and specificity, is not explicitly stated, but it would be derived from the 195 subjects in the "Cath Lab Cohort" used for comparison to invasive LVEDP. The sensitivity and specificity numbers are specifically tied to the LVEDP estimate, which would rely on the cohort with ground truth LVEDP measurements (Cath Lab Cohort).
    • Data Provenance:

      • Geographic Origin: The Cath Lab Cohort was enrolled "across 9 sites" (locations not specified, assuming within the US for FDA clearance), and the two Healthy Cohorts were enrolled "at a single site" (location not specified).
      • Nature of Data: Prospective validation study.

    3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications

    • The document states that the ground truth for the Cath Lab Cohort was "gold-standard invasive LVEDP obtained via the Millar Mikro-Cath." This is a direct physiological measurement, not established by human experts in a subjective reading context.
    • For the "reference-standard non-invasive echocardiographic indices" used for the Aged Healthy Cohort, the document does not specify the number or qualifications of experts involved in establishing these indices. However, echocardiographic indices are objective measurements and calculations, typically performed by trained sonographers and interpreted by cardiologists/echocardiographers.

    4. Adjudication Method for the Test Set

    • Since the primary ground truth for LVEDP was an invasive, objective measurement (Millar Mikro-Cath), no expert adjudication method (like 2+1 or 3+1) would have been necessary for this part of the ground truth establishment.
    • For the echocardiographic indices, the document does not provide details on adjudication, but as noted above, these are typically objective measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document does not indicate that an MRMC comparative effectiveness study was done to show how human readers improve with AI vs. without AI assistance.
    • The Vivio System is described as a device that "estimates LVEDP" and "aids in the diagnosis of heart failure when used by qualified healthcare professionals as an adjunct." It provides a calculated clinical parameter, not necessarily an AI-driven image interpretation system that assists human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, the reported "sensitivity of 80% (64-91), and specificity of 80% (73-86)" appear to be a measure of the Vivio System's (algorithm's) standalone performance in classifying LVEDP as above or below 18mmHg against the invasive gold standard. The device generates the estimate, and these metrics directly assess its accuracy.

    7. Type of Ground Truth Used

    • Primary Ground Truth: "Gold-standard invasive LVEDP obtained via the Millar Mikro-Cath" for the Cath Lab Cohort. This is direct physiological outcomes data.
    • Reference Standard: "Reference-standard non-invasive echocardiographic indices" for the Aged Healthy Cohort. This relies on established clinical measurements, not subjective expert consensus.

    8. Sample Size for the Training Set

    • The document does not specify the sample size used for the training set. It mentions a "proprietary algorithm" and a "fixed parameter model" (as opposed to AI/machine learning), but no details on its development data are provided.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not specify how the ground truth for the training set was established. Given the device uses a "fixed parameter model" rather than a machine learning model, it implies that the algorithm's parameters were likely determined based on physiological principles and pre-existing clinical data/models, rather than a large-scale data-driven training process with established ground truth labels in the same way as an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222463

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-11-23

    (100 days)

    Product Code
    Regulation Number
    870.2200
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EchoGo Heart Failure 1.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilised by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF).

    EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis.

    EchoGo Heart Failure 1.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%.

    Device Description

    EchoGo Heart Failure 1.0 is a software-only medical device manufactured by Ultromics Limited and granted breakthrough status by the FDA under Q212613.

    EchoGo Heart Failure 1.0 takes as input a DICOM file containing an echocardiogram and reports a classification decision suggestive of the presence or absence of heart failure with preserved ejection fraction (HFpEF). The output of this device is based on an artificial intelligence (Al) model developed using a convolutional neural network that produces the classification result. The model takes as input a 2D echocardiogram in which an apical 4-chamber view of the heart has been captured and assessed to have an ejection fraction ≥50% (this would normally be computed using a medical device for the assessment of cardiac function of the left ventricle, for example K213275). The echocardiogram should be acquired without contrast and contain at least one full cardiac cycle.

    Independent training, validation and test datasets were used for training and performance assessment of the device. EchoGo Heart Failure 1.0 is fully automated and does not comprise a user interface.

    EchoGo Heart Failure 1.0 produces a report containing the result of the classification, and this report is intended to be used by an interpreting clinician as an aid to diagnosis for HFpEF. The results are intended as an additional input to standard diagnostic pathways and should only be used by an interpreting clinician. The device is a diagnostic aid and thus according to common medical sense and the principles of differential diagnosis any diagnostic finding derived from usage of this product must be confirmed by additional diagnostic investigations, if in doubt. The ultimate diagnostic decision remains the responsibility of the interpreting clinician using patient presentation, medical history, and the results of available diagnostic tests, one of which may be EchoGo Heart Failure 1.0.

    EchoGo Heart Failure 1.0 is a prescription only device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text only explicitly states the reported device performance and the p-values for one-sided binomial exact tests against a priori acceptance criteria. The specific numerical acceptance criteria (e.g., minimum sensitivity and specificity thresholds) are not directly stated in the document. However, since the p-values are reported as p<0.001, it implies that the reported performance met or exceeded the pre-specified thresholds.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Sensitivity(Not explicitly stated, but met with p<0.001)87.8% (95% CI: 85.0, 90.3%)
    Specificity(Not explicitly stated, but met with p<0.001)82.0% (95% CI: 78.6, 85.0%)
    Repeatability(Not explicitly stated)100%
    Reproducibility (Positive Agreement)(Not explicitly stated)86.7%
    Reproducibility (Negative Agreement)(Not explicitly stated)76.9%
    Reproducibility (No Classification Agreement)(Not explicitly stated)45.5%
    Non-diagnostic outputs ("No Classification")Within a priori acceptance limits7.3% (94 out of 1,285 studies)

    Note on Acceptance Criteria: The document mentions "a priori acceptance criteria" for sensitivity and specificity but does not specify their numerical values. The p-values of <0.001 indicate that the observed sensitivity and specificity are statistically significantly higher than these unstated thresholds.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for the Test Set: 1,285 patients (comprising 639 controls and 646 cases).
    • Data Provenance: Retrospective case-control study. The data was collected from multiple independent clinical sites spanning five states in the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts used to establish the ground truth for the test set or their specific qualifications (e.g., "radiologist with 10 years of experience"). It mentions that the "device output was compared to the ground truth classifications of cases (HFpEF) or controls."
    We can infer that the ground truth for human-in-the-loop performance would be established in conjunction with "accredited cardiac physiologists (N=2) and cardiologists (N=5)" or similar, as cited in Section 7 for Usability for the formative and summative evaluations. However, this is for usability assessment and not explicitly tied to the ground truth creation for the performance study.

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. Therefore, no effect size of human readers improving with AI vs. without AI assistance is reported. The study described focuses on the standalone performance of the AI device against a ground truth.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The device's output (classification) was directly compared to the ground truth classifications without human intervention in the device's output decision. The text explicitly states, "EchoGo Heart Failure 1.0 is fully automated and does not comprise a user interface" (Page 4), which further supports this.

    7. The Type of Ground Truth Used

    The ground truth used was "ground truth classifications of cases (HFpEF) or controls," implying a clinical diagnosis or consensus based on established diagnostic criteria for Heart Failure with Preserved Ejection Fraction. The document does not specify if it was solely expert consensus, pathology, or outcomes data, but it refers to "ground truth classifications," which typically involves expert review of comprehensive patient data.

    8. The Sample Size for the Training Set

    The document states, "Independent training, validation and test datasets were used for training and performance assessment of the device" (Page 4) but does not specify the sample size for the training set.

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It only mentions that the AI model was "developed using a convolutional neural network" and used "independent training, validation and test datasets." It can be inferred that a similar method to the test set (clinical diagnosis/consensus) would have been used for the training data ground truth.