Search Results
Found 19 results
510(k) Data Aggregation
(29 days)
Ask a specific question about this device
(32 days)
The SKOUT system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT system is only intended to assist the gastroenterologist in identifying suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT is indicated for white light colonoscopy only.
The SKOUT® system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT® system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT® system by observing the software display, including the polyp detection box and device status indicator signal.
The provided FDA 510(k) clearance letter describes the SKOUT system, a software device designed to detect potential colorectal polyps in real-time during colonoscopy examinations. The document outlines acceptance criteria and summarizes performance testing.
1. A table of acceptance criteria and the reported device performance:
The document doesn't explicitly list "acceptance criteria" in a table format with numerical targets. Instead, it refers to performance objectives implicitly through comparisons to the predicate device and mentions passing results in applicable testing. Based on the "Performance Testing" section, the following can be inferred as the general performance areas that were assessed:
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Software Verification and Validation | SKOUT® system software meets design requirements for its intended use, in line with recommendations outlined in "General Principles of Software Validation, Guidance for Industry and FDA Staff." |
| Algorithm Performance (True Positives, False Positives, Polyp Detection Time) | Standalone and on-device algorithm performance testing was conducted, with passing results. Optimized performance with Endocuff, multiple polyps, and/or sessile serrated lesions, and reduced false positive rate. |
| Pixel Degradation | No pixel level degradation is introduced by SKOUT® to the Endoscopic System. |
| Video Delay (due to marker annotation) | Video delay due to marker annotation = 0ms (no standard error; all results were 0, minimum resolution 1.1ms). |
| Video Delay (due to device) | Video delay due to device = 0ms (no standard error; all results were 0, minimum resolution 1.1ms). |
| Endocuff Compatibility | Optimized performance in videos with Endocuff. |
| Multiple Polyps and Sessile Serrated Lesions | Optimized performance for multiple polyps and/or sessile serrated lesions. |
| Reduced False Positive Rate | Reduced false positive rate. |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document mentions "an expanded dataset for the added Endocuff compatibility" for standalone and on-device algorithm performance testing. However, it does not specify the exact numerical sample size (e.g., number of videos, number of polyps, or number of patients) of this expanded dataset.
- Data Provenance: The document does not specify the country of origin of the data used for testing, nor whether the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The document does not provide information on the number of experts used to establish ground truth or their specific qualifications.
4. Adjudication method for the test set:
The document does not specify the adjudication method used for the test set (e.g., 2+1, 3+1, none).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
The document explicitly states that the "clinical performance remains unchanged from the clinical performance submitted in K213686." This implies that a new MRMC comparative effectiveness study was not conducted for this submission (K253664). The existing clinical performance from K213686 was relied upon. Therefore, no new effect size for human reader improvement with AI vs. without AI assistance is reported here.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Yes, a standalone algorithm performance study was done. The "Performance Testing" section states: "Standalone and on-device algorithm performance testing was conducted for evaluation of true positives, false positives, and polyp detection time, with an expanded dataset for the added Endocuff compatibility."
7. The type of ground truth used:
The type of ground truth used is not explicitly specified in this document. Given it's a device for detecting polyps, it would typically be based on expert consensus (e.g., gastroenterologists reviewing videos and annotating polyps) or pathology (histopathological confirmation of removed lesions). However, the document does not confirm this.
8. The sample size for the training set:
The document does not specify the sample size used for the training set. It mentions "Retraining and refinement of inference algorithms... optimized performance," which indicates that a training process occurred, but details on the training set size are absent.
9. How the ground truth for the training set was established:
The document does not provide information on how the ground truth for the training set was established.
Ask a specific question about this device
(84 days)
ME-APDS (Magentiq Eye's Automatic Polyp Detection System) is intended to be used by endoscopists as an adjunct to the common video colonoscopy procedure (screening and surveillance), aiming to assist the endoscopist in identifying lesions during colonoscopy procedure by highlighting regions with visual characteristics consistent with different types of mucosal abnormalities that appear in the colonoscopy video during the procedure. Highlighted regions can be independently assessed by the endoscopist and appropriate action taken according to standard clinical practice.
ME-APDS is trained to process video images which may contain regions consistent with polyps.
ME-APDS is limited for use with standard white-light endoscopy imaging only.
ME-APDS is intended to be used as an adjunct to endoscopy procedures and is not intended to replace histopathological sampling as means of diagnosis.
ME-APDS™/MAGENTIQ-COLO is intended to be used as an adjunct to the common video colonoscopy procedure. The system application aims to assist the endoscopist in identifying lesions, such as polyps, during the colonoscopy procedures in real time. The device is not intended to be used for diagnosis or characterization of lesions, and does not replace clinical decision making.
The system acquires the digital video output signal from the local endoscopy camera and processes the video frames. It runs deep machine learning and additional supporting algorithms in real time on the video frames in order to detect and identify regions having characteristics consistent with different types of mucosal abnormalities such as polyps. The output video with the detected lesions is presented on a separate screen, highlighting the suspicious areas on the original video. The user can also take snapshots of the videos, with and without the highlighting of the suspicious areas, record videos and view in full screen mode. The ME-APDS™ includes an additional feature that removes the presentation of the bounding box once the physician has inserted an endoscopic tool (e.g. snare or forceps) into the endoscopic field of view.
Acceptance Criteria and Study for MAGENTIQ-COLO (ME-APDS)
Based on the provided FDA 510(k) clearance letter, here's a detailed breakdown of the acceptance criteria and the study conducted for MAGENTIQ-COLO (ME-APDS):
1. Table of Acceptance Criteria and Reported Device Performance
The document states that "All testing met pre-specified acceptance criteria," implying that the reported performance metrics achieved or exceeded predefined thresholds, though the specific numerical thresholds for acceptance are not explicitly stated within this document. The table below presents the reported performance metrics based on the standalone algorithm performance testing.
| Metric | Context | Reported Performance |
|---|---|---|
| Polyp Detection | ||
| Polyp-wise Recall (PRc1) | With Histology | 99.7% |
| Polyp-wise Recall (PRc3) | With Histology | 99.7% |
| Polyp-wise Recall (PRc5) | With Histology | 99.7% |
| Polyp-wise Recall (PRc7) | With Histology | 99.7% |
| Polyp Coverage (MPC) | With Histology | 81.7% (median) |
| Polyp-wise Recall (PRc1) | Without Histology | 100.0% |
| Polyp-wise Recall (PRc3) | Without Histology | 100.0% |
| Polyp-wise Recall (PRc5) | Without Histology | 100.0% |
| Polyp-wise Recall (PRc7) | Without Histology | 100.0% |
| Polyp Coverage (MPC) | Without Histology | 81.3% (median) |
| Frame-wise Recall (FRecall) | Full Testing Dataset | 73.0% |
| Polyp Coverage (MPC) | Full Testing Dataset | 72.0% (median) |
| Polyp-wise Recall (PRc1) | Full Testing Dataset | 98.1% |
| Polyp-wise Recall (PRc3) | Full Testing Dataset | 95.8% |
| Polyp-wise Recall (PRc5) | Full Testing Dataset | 93.8% |
| Polyp-wise Recall (PRc7) | Full Testing Dataset | 91.2% |
| False Positives Per Full Video (procedure) rate | Algorithm Performance | Target achieved (specific rate not stated) |
| False Positives Per Frame (TPP) | Algorithm Performance | 0.0334 (threshold achieved) |
| Tool Detection | ||
| Tool Detection Frame-wise Recall (FRecall) | Tool Detection | 85.9% |
| Median Tool Coverage (MTC) | Tool Detection | 96.5% |
| Tool Recall (TRc1) | Tool Detection | 99.7% |
| Tool Recall (TRc3) | Tool Detection | 99.7% |
| Tool Recall (TRc5) | Tool Detection | 99.7% |
| Tool Recall (TRc7) | Tool Detection | 99.7% |
| False Positives Per Frame (TPP) | Tool Detection | 0.0118 (calculated) |
| System Performance | ||
| Marker Annotation Delay | All cases (new system, with/without tool detection) | Adequate performance (specific metric not stated, but criteria met) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 252 unique full colonoscopy videos. These videos contained 806 polyps (660 small, 111 medium, 35 large). 24 videos contained no polyps.
- Data Provenance: The document does not explicitly state the country of origin. The test set appears to be retrospective, as it is referred to as "the algorithm was tested offline on... videos."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not explicitly state the number of experts or their specific qualifications (e.g., number of years of experience) used to establish the ground truth for the test set. It mentions "polyps verified by histology" and "polyps that were reported in the procedure report," implying a clinical and pathological review process, but details on the experts involved are missing.
4. Adjudication Method for the Test Set
The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). The ground truth for polyps is based on "histology findings" and "procedure report," suggesting a consensus or standard clinical reporting process but not a formal adjudication process by a panel for the study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study is not mentioned in the provided document. The study described is a standalone performance test of the algorithm.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, a standalone study was performed. The document explicitly states: "The algorithm was tested offline on 252 unique full colonoscopy videos..." and "Standalone algorithm performance testing was conducted..."
7. Type of Ground Truth Used
The ground truth used for the test set was primarily based on:
- Histology findings: This is considered a high-quality ground truth for polyp characteristics.
- Procedure report: This likely refers to the endoscopist's findings documented during the colonoscopy.
- The presence and size of polyps were noted, and 375 polyps had histology findings (302 adenoma polyps).
8. Sample Size for the Training Set
The document does not specify the sample size used for the training set. It only states that the device "is trained to process video images."
9. How the Ground Truth for the Training Set Was Established
The document does not specify how the ground truth for the training set was established. It only mentions that "ME-APDS is trained to process video images which may contain regions consistent with polyps."
Ask a specific question about this device
(28 days)
The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment.
CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only.
CADDIE is cloud based artificial intelligence medical device software. CADDIE interfaces with the video feed generated by an endoscopic video processor during a colonoscopy procedure
The software is intended to be used by trained and qualified healthcare professionals as an accompaniment to video endoscopy for the purpose of drawing attention to regions with visual characteristics consistent with colonic mucosal lesions (such as polyps and adenomas).
CADDIE analyses the data from the endoscopic video processor in real-time and provides information to aid the endoscopist in detecting suspected colorectal polyps, if they are in the field of view of the endoscope.
The areas highlighted by CADDIE are not to be interpreted as definite polyps or adenomas. The responsibility to make a decision as to whether or not a highlighted region contains a polyp or is an adenoma lies with the user. The endoscopist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp and its classification based on their own medical judgement.
Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter for CADDIE K252586:
1. Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria | Reported Device Performance (95% CI) |
|---|---|---|
| Overall Frame-level Accuracy | N/A (Not explicitly stated as an acceptance criterion, but represents overall performance) | 90.38% [90.34, 90.43] |
| Overall Frame-level TPR | N/A | 89.14% [89.06, 89.22] |
| Overall Frame-level FPR | N/A | 9.18% [9.13, 9.24] |
| AO Frame-level Accuracy | N/A | 93.93% [93.90, 93.97] |
| AO Frame-level TPR | N/A | 83.39% [83.27, 83.51] |
| AO Frame-level FPR | N/A | 4.19% [4.16, 4.22] |
| ICV Frame-level Accuracy | N/A | 94.32% [94.29, 94.36] |
| ICV Frame-level TPR | N/A | 83.78% [83.66, 83.91] |
| ICV Frame-level FPR | N/A | 4.57% [4.54, 4.61] |
| Overall AUC | N/A | 93.59 [93.55, 93.62] |
Note: The FDA letter explicitly states that for all changes (including the Cecum AI update), "design and development activities were performed in compliance with Odin's Quality Management System and FDA Design Control requirements (21 CFR 820.30), using the same methods and acceptance criteria described in the cleared submission (K240044)." However, the provided document for K252586 does not explicitly list the quantitative acceptance criteria (e.g., minimum TPR, maximum FPR) from the original K240044 submission for these specific metrics of the Cecum AI feature. It only reports the achieved performance. The implicit acceptance is that the device's performance is "satisfactory and do not raise any additional questions on the safety and effectiveness" as compared to the predicate.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The test set for the Cecum AI Convenience Feature Standalone Bench-testing Dataset comprised 5733 frames. This was broken down into:
- 838 Positive Frames (containing Cecal structures)
- 461 frames with Appendiceal Orifice (AO)
- 418 frames with Ileocecal Valve (ICV)
- 4016 Negative Frames (without Cecal structures)
- Data Provenance: The document states the dataset consists of "photo-documented frames from a standard colonoscopy procedure" and mentions "historical control (known cecal structure status per frame)." It does not explicitly state the country of origin or if it's retrospective or prospective. However, given it's "historical control" and "recorded colonoscopy frames," it is highly likely to be retrospective data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: A "team of trained clinical annotators" was used. The exact number is not specified.
- Qualifications: They are described as "trained clinical annotators." Specific professional qualifications (e.g., radiologist, gastroenterologist) or years of experience are not provided.
4. Adjudication Method for the Test Set
- The document states, "Annotation was performed on a per-frame basis, where a team of trained clinical annotators labelled cecal structures with a bounding box. These annotations were used as ground truth reference standards."
- It does not explicitly state an adjudication method (like 2+1 or 3+1). It implies that the "team of trained clinical annotators" collectively established the ground truth, but the process for resolving disagreements or reaching consensus is not detailed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done for the Cecum AI convenience feature.
- The primary submission (K252586) is a Special 510(k) addressing an update to a "convenience feature" (Cecum AI model algorithm). It explicitly states that "The baseline Clinical Performance Evaluation was conducted and reviewed in K240044 and is still applicable to the versions of the device that are the subject of this submission." This suggests that any MRMC studies related to the primary polyp detection functionality would have been part of the original K240044 clearance. The current submission focuses only on the AI model change for cecal landmark detection.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)
- Yes, a standalone performance study was done for the Cecum AI Convenience Feature.
- The document states, "Standalone performance testing was performed to assess the ability of the Cecum AI Convenience Feature to discriminate between normal mucosa and cecal landmarks... A set of recorded colonoscopy frames were analyzed by the Cecum AI Convenience Feature and the results were compared to the historical control (known cecal structure status per frame)." The reported metrics (accuracy, TPR, FPR, AUC) are all standalone performance metrics.
7. Type of Ground Truth Used
- The ground truth used for the Cecum AI convenience feature was expert consensus/annotation. Specifically, "Annotation was performed on a per-frame basis, where a team of trained clinical annotators labelled cecal structures with a bounding box. These annotations were used as ground truth reference standards."
8. Sample Size for the Training Set
- The document states that the "Non-clinical performance testing was performed on the standalone bench-testing dataset, which is separate to the development datasets."
- However, it does not provide the sample size for the training set used to develop the Cecum AI model.
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly describe how the ground truth for the training set was established. It only details the ground truth for the test set. It is reasonable to infer that a similar process of expert annotation would have been used for the training data, but this is not stated.
Ask a specific question about this device
(28 days)
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
The SKOUT® system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT® system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT® system by observing the software display, including the polyp detection box and device status indicator signal.
The provided FDA 510(k) clearance letter for the SKOUT® system (K251126) indicates that this submission is for a modified version of a previously cleared device (K241508) and asserts that the "inference algorithms have remained the same, therefore clinical performance remains unchanged from the clinical testing submitted in K213686." This means the detailed clinical performance data and ground truth establishment would likely reside in the K213686 submission, which is not provided in this document.
However, based on the information present in the K251126 clearance letter, we can describe the non-clinical performance testing that was conducted to demonstrate substantial equivalence to the predicate device. The letter explicitly states:
"Algorithm performance testing was performed for evaluation of true positives, false positives, and polyp detection time, with an expanded dataset for the added video processing tower."
This indicates that some form of performance evaluation was done for the algorithm, even if the core clinical performance from K213686 is simply being referenced.
Given the limitations of the provided document, here's a structured response based on the available information, with clear indications where information is not provided in K251126 and would require reviewing K213686:
Acceptance Criteria and Device Performance for SKOUT® System (K251126)
The SKOUT® system (K251126) is a modified version of a predicate device (K241508). The clearance letter explicitly states that "the inference algorithms have remained the same, therefore clinical performance remains unchanged from the clinical testing submitted in K213686." This implies that the core clinical performance metrics and their acceptance criteria were established and met in the K213686 submission.
For the K251126 submission, non-clinical performance testing was conducted to demonstrate substantial equivalence. This included "Algorithm performance testing... for evaluation of true positives, false positives, and polyp detection time, with an expanded dataset for the added video processing tower." However, the specific quantitative acceptance criteria and the reported performance metrics (true positives, false positives, detection time) are NOT provided in this document.
1. Table of Acceptance Criteria and Reported Device Performance
Note: The specific quantitative acceptance criteria and reported device performance (True Positives, False Positives, Polyp Detection Time) for the algorithm from the K251126 submission are NOT explicitly provided in this document. The document refers to the unchanged clinical performance from K213686. The table below represents the types of metrics that would have been evaluated, as stated in the document, but the actual numerical values are missing.
| Performance Metric (Non-Clinical, for K251126) | Acceptance Criteria (Not Explicitly Stated in Doc) | Reported Device Performance (Not Explicitly Stated in Doc) |
|---|---|---|
| True Positives (Algorithm) | Conformance with predicate performance from K213686 | "Passing results" (qualitative confirmation) |
| False Positives (Algorithm) | Conformance with predicate performance from K213686 | "Passing results" (qualitative confirmation) |
| Polyp Detection Time (Algorithm) | Conformance with predicate performance from K213686 | "Passing results" (qualitative confirmation) |
| Video Delay (Marker Annotation) | 0 ms | 0 ms (no standard error, minimum resolution 1.1ms) |
| Video Delay (Device) | 0 ms | 0 ms (no standard error, minimum resolution 1.1ms) |
| Pixel Level Degradation | No degradation introduced to Endoscopic System | No pixel level degradation |
2. Sample Size and Data Provenance for Test Set
- Sample Size Used for Test Set: The document mentions "an expanded dataset for the added video processing tower" for algorithm performance testing. However, the specific sample size (number of cases, images, or polyps) is NOT provided in this document for either the K251126 testing or the referenced K213686 clinical testing.
- Data Provenance: Not explicitly stated in this document. For K251126, it implies additional video data for specific hardware validation. For the core clinical performance (K213686), this information would be detailed in that submission. It is generally assumed, for FDA clearance, that data would represent diverse patient populations.
- Retrospective or Prospective: Not explicitly stated in this document for the algorithm performance testing in K251126. Clinical studies underpinning K213686 might have been retrospective or prospective, but this is not discussed in the provided text.
3. Number and Qualifications of Experts for Ground Truth
- Number of Experts: Not provided in this document for either the K251126 testing or the referenced K213686 clinical testing.
- Qualifications of Experts: Not provided in this document. For ground truth in medical imaging, experts are typically board-certified specialists (e.g., gastroenterologists or pathologists) with significant experience.
4. Adjudication Method for Test Set
- Adjudication Method: Not provided in this document. For K251126, the performance testing was focused on validating the algorithm with expanded data, not necessarily re-establishing clinical ground truth for a human-in-the-loop study. For K213686, the adjudication method for ground truth establishment for polyp detection would be detailed in that submission. Common methods include consensus reading (e.g., 2+1, where two readers agree, or a third adjudicates disagreement) or pathology correlation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Conducted: The document for K251126 does not describe an MRMC comparative effectiveness study directly for this submission. It states that "clinical performance remains unchanged from the clinical testing submitted in K213686." An MRMC study would be common for a device that assists human readers (CADe), and such a study would likely have been part of the original K213686 submission to demonstrate clinical benefit.
- Effect Size of Human Reader Improvement: Since an MRMC study is not detailed for K251126, the effect size is not provided here. If such a study was performed for K213686, the effect size (e.g., improvement in adenoma detection rate, per-lesion sensitivity) would be reported in that submission.
6. Standalone (Algorithm Only) Performance
- Standalone Performance Done: Yes, "Algorithm performance testing was performed for evaluation of true positives, false positives, and polyp detection time." This indicates a standalone evaluation of the algorithm's detection capabilities. However, the specific quantitative metrics (e.g., sensitivity, specificity, FROC curves) are NOT provided in this document. It only states that the testing demonstrated "passing results." The document also notes that "SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients," reinforcing its role as a CADe tool for human assistance.
7. Type of Ground Truth Used
- Type of Ground Truth: The document implies that ground truth for "potential colorectal polyps" was established for the algorithm performance testing. For a device detecting polyps, the most robust ground truth would generally be histopathology (pathology results) from biopsied/resected lesions. Clinical consensus from expert endoscopists is also commonly used, especially for cases where biopsy might not be performed. The document does not explicitly state the type of ground truth used, but for polyp detection, pathology is gold standard.
8. Sample Size for Training Set
- Sample Size for Training Set: Not provided in this document. The document primarily focuses on the substantial equivalence and validation of changes rather than a full re-description of the training process for the core AI algorithm, as the algorithm itself is stated to be unchanged from K213686.
9. How Ground Truth for Training Set Was Established
- Ground Truth Establishment for Training Set: Not provided in this document. Similar to the test set, the methods for establishing ground truth for the training data (e.g., expert annotations, pathology reports, adjudicated consensus) would have been detailed in the original K213686 submission.
In summary, the provided FDA clearance letter for K251126 confirms that algorithm performance testing (for true positives, false positives, and detection time) was conducted for this submission, particularly with an "expanded dataset for the added video processing tower." However, it relies on the clinical performance established in the predicate's prior submission (K213686) because the core inference algorithms are unchanged. Therefore, many details regarding the clinical acceptance criteria, sample sizes, expert qualifications, and ground truth establishment for the clinical performance aspect are referred to the previous submission and are not present in this document.
Ask a specific question about this device
(28 days)
ME-APDS (Magentig Eye's Automatic Polyp Detection System) is intended to be used by endoscopists as an adjunct to the common video colonoscopy procedure (screening and surveillance), aiming to assist the endoscopist in identifying lesions during colonoscopy procedure by highlighting reqions with visual characteristics consistent with different types of mucosal abnormalities that appear in the colonoscopy video during the procedure. Highlighted regions can be independently assessed by the endoscopist and appropriate action taken according to standard clinical practice.
ME-APDS is trained to process video images which may contain regions consistent with polyps.
ME-APDS is limited for use with standard white-light endoscopy imaging only.
ME-APDS is intended to be used as an adjunct to endoscopy procedures and is not intended to replace histopathological sampling as means of diagnosis.
ME-APDS™MAGENTIQ-COLO is intended to be used as an adjunct to the common video colonoscopy procedure. The system application aims to assist the endoscopist in identifying lesions, such as polyps, during the colonoscopy procedures in real time. The device is not intended to be used for diagnosis or characterization of lesions, and does not replace clinical decision making.
The system acquires the digital video output signal from the local endoscopy camera and processes the video frames. It runs deep machine learning and additional supporting algorithms in real time on the video frames in order to detect and identify regions having characteristics consistent with different types of mucosal abnormalities such as polyps. The output video with the detected lesions is presented on a separate screen, highlighting the suspicious areas on the original video. The user can also take snapshots of the videos, with and without the highlighting of the suspicious areas, record videos and view in full screen mode.
Here's an analysis of the acceptance criteria and study details for the MAGENTIQ-COLO device, based on the provided document:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the reported performance metrics, particularly "Polyp-wise Recall" and "False Positives Per Frame (FPPF)". The study aims to demonstrate that the device performs comparably to or better than the predicate device.
| Acceptance Criteria / Metric | Reported Device Performance (Full Testing Dataset) |
|---|---|
| Polyp-wise Recall (PRecall1) | 97.9% [96.63%, 98.94%] |
| Polyp-wise Recall (PRecall3) | 95.3% [93.39%, 96.96%] |
| Polyp-wise Recall (PRecall5) | 93.2% [91.01%, 95.15%] |
| Polyp-wise Recall (PRecall7) | 90.6% [88.19%, 92.91%] |
| False Positives Per Frame (FPPF) | 0.0333 (threshold achieved) |
| Polyps with Histology: PRecall1 | 99.7% [99.12%, 100.0%] |
| Polyps with Histology: PRecall7 | 99.7% [99.11%, 100.0%] |
| Median Coverage of Polyps (with histology) | 81.7% |
| Marker Annotation Latency (Median) | 133 msec for FHD, 157 msec for 4K |
Note: The document states that "The testing results were observed to be as expected and support that the device has similar performance to the predicate device," implying that these reported values met the implicit acceptance criteria for substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): 212 unique full colonoscopy videos, containing 702 polyps (16 videos contained no polyps).
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, it references polyps "verified by histology" and "reported in the procedure report," implying clinical expert input.
4. Adjudication Method for the Test Set
The document does not describe a specific adjudication method like 2+1 or 3+1. The ground truth seems to be derived from documented polyps in the "procedure report" and "histology findings," suggesting a standard clinical reporting process rather than a specific consensus method for this study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported in this document. The study described is a standalone performance test of the algorithm. The document mentions that the clinical validation used to support the device's polyp detection functions was conducted in a previous submission (K223473). This K223473 submission might contain an MRMC study, but it's not detailed here.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, a standalone performance study was done. The "Standalone Performance Testing" section describes how "The algorithm was tested offline" on an independent dataset to evaluate its recall, false positive performance, and false positives per full video rate without direct human interaction during the test.
7. Type of Ground Truth Used
The ground truth used for the test set was a combination of:
- Histopathology findings: For polyps with histology reports.
- Procedure reports: For polyps identified and documented during the colonoscopy procedure.
8. Sample Size for the Training Set
The document does not provide the sample size for the training set. It only states that "ME-APDS is trained to process video images which may contain regions consistent with polyps."
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. It only broadly states that the system "runs deep machine learning" and is "trained to process video images."
Ask a specific question about this device
(27 days)
Ask a specific question about this device
(201 days)
The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment.
CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of action for patients. The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only.
CADDIE is cloud based artificial intelligence medical device software. CADDIE interfaces with the video feed generated by an endoscopic video processor during a colonoscopy procedure
The software is intended to be used by trained and qualified healthcare professionals as an accompaniment to video endoscopy for the purpose of drawing attention to regions with visual characteristics consistent with colonic mucosal lesions (such as polyps and adenomas).
CADDIE analyses the data from the endoscopic video processor in real-time and provides information to aid the endoscopist in detecting suspected colorectal polyps, if they are in the field of view of the endoscope.
The areas highlighted by CADDIE are not to be interpreted as definite polyps or adenomas. The responsibility to make a decision as to whether or not a highlighted region contains a polyp or is an adenoma lies with the user. The endoscopist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp and its classification based on their own medical judgement.
Here's a breakdown of the acceptance criteria and study details for the CADDIE device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Polyp Detection (Standalone Bench-testing Data):
| Name | Description | Acceptance Criteria (Success Criteria) | Reported Device Performance [95% CI] |
|---|---|---|---|
| Object-level True Positive Rate (TPR)* | Proportion of polyps detected by the device (>0.5 seconds & IoU >20%) and confirmed by pathology. | > 80% (to show a lower miss rate than in clinical practice of 25%) | 98.27% [97.33, 99.20] |
| Frame-Level False Positive Rate (FPR)* | Proportion of frames (%) in which CADDIE detects a box (>0.5 seconds) that is not a histopathologically confirmed polyp. | < 5 % (to provide a good user experience, avoid distractions, and reduce false predictions) | 1.54% [1.31, 1.76] |
| Object-level False Positive Rate (FPR)** | Average of suspected regions per minute that CADDIE finds per procedure (> 0.5 seconds) that are not histopathology-confirmed polyps. | Not applicable | 0.89 [0.79, 1.00] |
| Frame-Level TPR** | Proportion of frames (%) with confirmed polyps in which CADDIE detects the polyp (>0.5 seconds & IoU >20%). | Not applicable | 54.92% [53.02, 56.81] |
*Primary Endpoints; **Secondary Endpoints
Cecum AI (Standalone Bench-testing Data):
| Name | Description | Acceptance Criteria (Success Criteria) | Reported Device Performance [95% CI] |
|---|---|---|---|
| Frame-Level true positive rate (TPR) | The proportion (%) of all the frames annotated with cecum which the Cecum AI identifies correctly. | Frame-level TPR > 80% | 94.05% [91.58, 96.52] |
| Frame-level false positive rate (FPR)* | The proportion (%) of all the frames annotated without cecal landmarks which the Cecum AI incorrectly identifies the cecum. | Frame-level FPR < 20% | 12.08% [10.14, 14.02] |
Clinical Study (Multi-Reader, Multi-Case Comparative Effectiveness Study):
| Endpoint | Study Arm | N patients / N polyps | Acceptance Criteria | Reported Performance | P-value |
|---|---|---|---|---|---|
| Primary Endpoints: | |||||
| Adenomas Per Colonoscopy (APC) | CADe vs SoC | CADe: 417, SoC: 424 | Significantly higher APC in CADe group (implied by superiority over SoC) | CADe: 0.82 ± 1.40, SoC: 0.62 ± 1.19 (Ratio 1.33 [1.06, 1.67]) | 0.01 |
| Positive Percent Agreement (PPA) | CADe vs SoC | CADe: 700, SoC: 517 | Non-inferiority to SoC (lower bound of 95% CI for difference > -15%) | CADe: 53.9%, SoC: 53.4% (Difference 0.5% [-5.0%, ∞]) | - |
Conclusion: The device met all stated acceptance criteria in the standalone testing and demonstrated superiority in APC with non-inferiority in PPA in the clinical study.
2. Sample Size Used for the Test Set and Data Provenance
Polyp Detection Standalone Bench-testing Dataset:
- Sample Size (Subjects): 389 subjects.
- Data Provenance: Not explicitly stated, but the demographics include African American, American Indian, Asian, Caucasian, and Hispanic races/ethnicities, suggesting a diverse multi-region dataset, potentially US-based given the specific racial categories listed. This was a retrospective analysis as it used recorded colonoscopy videos and compared results to historical control (known polyp status per frame).
Cecum AI Standalone Bench-testing Dataset:
- Sample Size (Frames): 5092 total frames (2833 positive frames, 2259 negative frames).
- Data Provenance: Not explicitly stated, but it uses recorded colonoscopy frames, implying a retrospective analysis.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Polyp Detection Standalone Bench-testing Dataset:
- Experts: A team of trained clinical annotators initially labeled polyp structures, followed by an additional layer of review by a separate team of experts.
- Qualifications: The "separate team of experts" had "over 2000 endoscopic procedures experience."
Cecum AI Standalone Bench-testing Dataset:
- Experts: A team of trained clinical annotators labeled cecal structures.
- Qualifications: Not explicitly stated beyond "trained clinical annotators."
4. Adjudication Method for the Test Set
Polyp Detection Standalone Bench-testing Dataset:
- Annotation was performed on a per-frame basis. A "team of trained clinical annotators" labeled polyp structures, followed by an "additional layer of review by a separate team of experts." This indicates a multi-reader review process, likely with a consensus or hierarchical adjudication, though the exact method (e.g., 2+1, 3+1) is not specified.
Cecum AI Standalone Bench-testing Dataset:
- Annotation was performed on a per-frame basis by a "team of trained clinical annotators." No additional layer of review or specific adjudication method (like 2+1) is mentioned.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance
- Yes, a prospective, multi-center, MRMC, randomized controlled, parallel group trial was done.
- Effect Size of Human Improvement with AI vs. Without AI Assistance:
- Adenomas Per Colonoscopy (APC): CADDIE (with AI) resulted in an APC of 0.82 ± 1.40, while Standard of Care (without AI) had an APC of 0.62 ± 1.19. The ratio of CADe to SoC was 1.33 (95% CI: 1.06, 1.67), meaning 33% more adenomas per colonoscopy were detected with AI assistance.
- Adenoma Detection Rate (ADR): CADe group had an ADR of 42.9%, SoC had 35.9%. The difference was 7.1% (95% CI: 0.5%, 13.7%), meaning AI assistance led to a 7.1% absolute increase in the proportion of examinations with at least one adenoma detected.
- AI assistance also led to significant increases in detection of diminutive (≤5 mm) adenomas/adenocarcinomas (29% more) and large (≥10 mm) adenomas/adenocarcinomas (93% more).
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, standalone performance testing was performed for both the Polyp Detection component and the Cecum AI component.
- For Polyp Detection, a set of recorded colonoscopy videos was analyzed by CADDIE, and the results were compared to historical controls.
- For Cecum AI, a set of recorded colonoscopy frames were analyzed by Cecum AI, and the results were compared to historical controls.
7. The Type of Ground Truth Used
Polyp Detection Standalone Bench-testing:
- Histology: Each polyp was "histologically confirmed." The ground truth for polyp annotations was based on these confirmed histology reports.
Cecum AI Standalone Bench-testing:
- Expert Annotation: Ground truth reference standards were "annotations performed on a per-frame basis, where a team of trained clinical annotators labelled cecal structures with a bounding box."
Clinical Study (MRMC):
- Histology/Pathology: The primary and secondary endpoints (APC, PPA, ADR, etc.) were based on "histologically confirmed" findings of polyps, adenomas, adenocarcinomas, and sessile serrated lesions.
8. The Sample Size for the Training Set
Polyp Detection Development Datasets:
- Number of Polyps: 1711 polyps.
- Number of Patients: 906 patients.
- Number of Frames: 318,603 frames (162,207 polyp frames and 156,396 non-polyp frames).
- This dataset was used for training, tuning, and testing (development data, separate from bench-testing data).
Cecum AI Development Datasets:
- Number of Patients: 1467 patients.
- Number of Images: 17,844 images.
- This dataset was used for training, tuning, and testing (development data, separate from bench-testing data).
9. How the Ground Truth for the Training Set Was Established
Polyp Detection Development Datasets:
- Ground truth was based on a combination of histology (for 1296 polyps from 714 patients) and optical confirmation by additional endoscopists (for 415 polyps confirmed through resection or photo-documentation, but not histopathology).
Cecum AI Development Datasets:
- The ground truth for the Cecum AI development dataset was established by using "informative static photo-documentation images, as well as images extracted from videos of cecal landmarks including appendiceal orifice (AO), ileocecal valve (ICV)." While not explicitly stated as "expert annotation" for the training set, this description implies that the landmarks were identified and labeled. The standalone test set confirmed ground truth by "a team of trained clinical annotators," suggesting a similar method for development.
Ask a specific question about this device
(36 days)
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
The SKOUT system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT system by observing the software display, including the polyp detection box and device status indicator signal.
The provided text describes an FDA 510(k) clearance for the SKOUT® system, a software device designed to detect potential colorectal polyps during colonoscopy. However, it focuses on demonstrating substantial equivalence to a predicate device (K240781), which itself was a predicate to an earlier device (K213686). The current submission (K241508) mainly highlights minor software refinements and states that the "clinical performance remains unchanged from the clinical performance submitted in K213686." Therefore, the details requested about acceptance criteria and the study proving the device meets them would primarily refer to the data supporting K213686, which is not fully detailed in this document.
Based on the provided K241508 document, here's the information that can be extracted, and where the information is missing:
1. A table of acceptance criteria and the reported device performance
The document states, "The inference algorithms the same architecture and meet the same performance requirements as the predicate device, therefore clinical performance remains unchanged from the clinical performance submitted in K213686." This implies that the acceptance criteria and reported performance for K241508 are identical to those established for K213686. However, the specific acceptance criteria and numerical performance metrics are not provided in this document.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated for K241508. The document mentions "new data representing 61% of the cumulative data" from 27 new clinical sites compared to the predicate, used for retraining and refinement. However, the size of the test set that explicitly demonstrated performance against acceptance criteria for this specific submission is not detailed. The phrase "clinical performance remains unchanged from the clinical performance submitted in K213686" suggests that the original clinical performance evaluation from K213686 is referenced, but its test set details are not here.
- Data Provenance: The document states "Utilization of data from 30+ unique clinical sites, of which 27 were new compared to the predicate device." It does not specify the countries of origin or if the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not explicitly stated in this document. This information would likely be found in the original K213686 submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not explicitly stated in this document. This information would likely be found in the original K213686 submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document mentions "The inference algorithms the same architecture and meet the same performance requirements as the predicate device, therefore clinical performance remains unchanged from the clinical performance submitted in K213686." This suggests that if such a study was performed, it was for K213686. However, the details of whether an MRMC study was done, its effect size, or human reader improvement are not provided in this document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document states that the system "is only intended to assist the gastroenterologist" and "is not intended to replace a full patient evaluation." This indicates its role as a human-in-the-loop tool. While standalone performance data might have been collected as part of the technical evaluation, the document does not explicitly describe a standalone performance study as the primary means of demonstrating effectiveness. It alludes to "algorithm performance" being assessed as part of "additional bench software testing" to meet special controls.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not explicitly stated in this document. This information would likely be found in the original K213686 submission. For polyp detection, pathology is a common ground truth, but expert consensus is also frequently used for live video analysis without immediate pathology.
8. The sample size for the training set
The document mentions "Utilization of data from 30+ unique clinical sites, of which 27 were new compared to the predicate device, with new data representing 61% of the cumulative data." This composite data was used for "Refinement/retraining of polyp detection algorithm." However, the total numerical sample size (e.g., number of colonoscopies, video frames, or polyps) for the training set is not explicitly stated.
9. How the ground truth for the training set was established
Not explicitly stated in this document. This information would likely be found in the original K213686 submission.
Summary of Missing Information and Recommendation:
The provided document (K241508) is a 510(k) summary for a modified device. It heavily relies on the performance demonstrated by an earlier predicate device (K213686) by asserting "clinical performance remains unchanged from the clinical performance submitted in K213686." To answer most of your detailed questions regarding acceptance criteria, study design, ground truth establishment, expert qualifications, and specific performance metrics, you would need to access the information contained in the K213686 FDA submission. The current document primarily confirms the substantial equivalence of the modified SKOUT® system (K241508) to its immediate predicate (K240781), which itself points back to K213686 for clinical performance.
Ask a specific question about this device
(29 days)
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
The SKOUT® system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT system by observing the software display, including the polyp detection box and device status indicator signal.
The provided document, an FDA 510(k) summary for the SKOUT® system (K240781), primarily focuses on demonstrating substantial equivalence to a predicate device (K230658) and does not contain the detailed acceptance criteria or the specific study results from a primary clinical performance study.
The document indicates that "the inference algorithms have remained the same, therefore clinical performance remains unchanged from the clinical performance submitted in K213686." This suggests that the clinical performance evaluation was conducted for a previous version or submission (K213686), and the current submission relies on that prior assessment.
Therefore, I cannot provide all the requested information using only the text you provided. The document explicitly states: "Performance data demonstrates that the SKOUT system is as safe and effective as the predicate device." However, it does not explicitly show the full performance data, acceptance criteria, sample sizes, or ground truth establishment details for that primary performance study (K213686).
Based on the provided text, here is what can be extracted and what information is missing:
Information Extracted from the Provided Text:
- Device Performance Reported: The document states that "SKOUT system demonstrated passing results in all applicable testing." and "Performance data demonstrates that the SKOUT system is as safe and effective as the predicate device."
- Adjudication Method: "None" is inferred for the listed "Performance Testing" which are non-clinical tests (Software verification and validation, bench software testing). For the clinical performance from K213686, the adjudication method is not described in this document.
- Standalone Performance: The non-clinical testing described seems to be for algorithm-only performance ("bench software testing was performed to confirm the device meets the special controls in 21 CFR 876.1520 for true and false positives, pixel degradation and video delays."). However, the specific metrics (e.g., sensitivity, specificity for polyp detection) are not reported here.
- Ground Truth Type: For the non-clinical testing, the "ground truth" seems to be defined by the design requirements and special controls for software (e.g., "true and false positives"). For the clinical performance (K213686), the type of ground truth is not specified.
Missing Information (Not Present in the Provided Text):
- A table of acceptance criteria and the reported device performance: While it states "passing results," the specific numerical acceptance criteria and the corresponding numerical performance values are not provided.
- Sample sized used for the test set and the data provenance: Not described for the underlying clinical performance study (K213686).
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not described for the underlying clinical performance study (K213686).
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not described for the underlying clinical performance study (K213686). The device is a CADe system, which suggests human-in-the-loop, but specific MRMC study results are not here.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not described for the underlying clinical performance study (K213686).
- The sample size for the training set: Not described.
- How the ground truth for the training set was established: Not described.
Based on the provided document, here's a structured response (with noted limitations):
The provided document (FDA 510(k) Summary for SKOUT® system, K240781) primarily asserts substantial equivalence to a predicate device (K230658) and refers to prior performance data from K213686. It does not contain the detailed acceptance criteria or the specific study results from the primary clinical performance evaluation.
The document states: "Performance data demonstrates that the SKOUT system is as safe and effective as the predicate device. Thus, the SKOUT system is substantially equivalent." and "the inference algorithms have remained the same, therefore clinical performance remains unchanged from the clinical performance submitted in K213686." This implies that the definitive study proving device performance against acceptance criteria was conducted for the K213686 submission, not detailed within this current document.
Here's what can be gathered:
1. Table of Acceptance Criteria and Reported Device Performance:
The document mentions "Additional bench software testing was performed to confirm the device meets the special controls in 21 CFR 876.1520 for true and false positives, pixel degradation and video delays." and "SKOUT system demonstrated passing results in all applicable testing." However, the specific numerical acceptance criteria (e.g., minimum sensitivity, maximum false positives per minute) and the quantified reported device performance values against these criteria are not provided in this document.
2. Sample Size and Data Provenance (for the test set):
Not explicitly stated for the underlying clinical performance study (K213686). The "Performance Testing" section describes non-clinical software verification and validation, which usually involves test cases rather than patient sample sizes.
3. Number of Experts and Qualifications for Ground Truth:
Not explicitly stated for the underlying clinical performance study (K213686).
4. Adjudication Method for the Test Set:
Not explicitly stated for the underlying clinical performance study (K213686). For the "Performance Testing" described in this document (non-clinical bench software testing), an adjudication method is not applicable in the human-reader sense.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
The document does not provide information about an MRMC comparative effectiveness study or the effect size of human readers improving with AI vs. without AI assistance. While the device is a Computer-Aided Detection (CADe) tool, which inherently assists human readers, the details of such a study are absent here.
6. Standalone (Algorithm Only) Performance:
The document states "Additional bench software testing was performed to confirm the device meets the special controls in 21 CFR 876.1520 for true and false positives, pixel degradation and video delays." This indicates that the algorithm's performance in detecting polyps and managing system lags was tested independently. However, the specific metrics (e.g., standalone sensitivity, specificity, or FPs/min rate) from this testing are not numerically reported in this document.
7. Type of Ground Truth Used:
For the clinical performance (referred to as K213686), the type of ground truth (e.g., expert consensus, pathology, follow-up outcomes) is not specified in this document. For the non-clinical performance testing, the ground truth is defined by the design requirements and regulatory standards for "true and false positives, pixel degradation and video delays."
8. Sample Size for the Training Set:
Not provided in this document.
9. How the Ground Truth for the Training Set Was Established:
Not provided in this document.
Ask a specific question about this device
Page 1 of 2