Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K243449

    Validate with FDA (Live)

    Device Name
    INFINITT DPS
    Date Cleared
    2025-12-09

    (397 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K252762

    Validate with FDA (Live)

    Device Name
    HALO AP Dx
    Manufacturer
    Date Cleared
    2025-11-25

    (88 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below.

    Device Description

    HALO AP Dx, version 2.4 (abbreviated as v2.4), is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management, and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner or the Leica Biosystems Imaging, Inc. Aperio GT 450 DX.

    HALO AP Dx is typically operated as follows:

    1. Image acquisition is performed using the predicate device. The operator performs quality control of the digital slides per the instructions of the predicate device and lab specifications to determine if re-scans are necessary.

    2. Once image acquisition is complete, according to its Instructions for Use, the unaltered image is saved in an external image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.

    3. Depending upon a laboratory's workflow, the scanned images may first be reviewed by histotechnicians to confirm image quality and initiate any re-scans. After review, the histotechnician may modify the case status and make it available to the pathologist.

    4. The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.

    5. The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
      a. Zoom and pan the image.
      b. Measure distances and areas in the image.
      c. Annotate images.
      d. View multiple images side by side in a synchronized fashion.

    6. The above steps are repeated as necessary.

    After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System.

    HALO AP Dx operates and is validated for use with the FDA cleared components listed in the intended use statement table above.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243391

    Validate with FDA (Live)

    Device Name
    AISight Dx
    Manufacturer
    Date Cleared
    2025-06-26

    (238 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AISight Dx is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret, and manage digital images of these slides for primary diagnosis. AISight Dx is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. AISight DX is intended to be used with interoperable displays, scanners and file formats, and web browsers that have been 510(k) cleared for use with the AISight Dx or 510(k)-cleared displays, 510(k)-cleared scanners and file formats, and web browsers that have been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying interoperable devices.

    Device Description

    AISight Dx is a web-based, software-only device that is intended to aid pathology professionals in viewing, interpretation, and management of digital whole slide images (WSI) of scanned surgical pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue obtained from Hamamatsu NanoZoomer S360MD Slide scanner or Leica Aperio GT 450 DX scanner (Table 1). It aids the pathologist in the review, interpretation, and management of pathology slide digital images used to generate a primary diagnosis.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the AISight Dx device, based on the provided FDA 510(k) Clearance Letter:


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Pixel-wise ComparisonIdentical image reproduction (max pixelwise difference < 1 CIEDE2000)Maximum pixelwise difference was 0 CIEDE2000, indicating pixelwise identical output images. Meets criteria.
    Non-inferiority to Glass Slide Reads (Major Discordance Rate - Hamamatsu Scanner)Upper limit of 95% CI for difference in major discordance rate (MD vs. GT vs. MO vs. GT) less than 4%.Upper limit of 95% CI was 1.16%. Meets criteria.
    Non-inferiority to Glass Slide Reads (Major Discordance Rate - Leica Scanner)Upper limit of 95% CI for difference in major discordance rate (MD vs. GT vs. MO vs. GT) less than 4%.Upper limit of 95% CI was 2.52%. Meets criteria.
    Turnaround TimeAdequate for intended use (image processing, loading, panning, zooming).Test results showed these to be adequate for the intended use. Meets criteria.
    Measurement AccuracyAccurate distance and area measurements.Tests verified that distances and areas measured in AISight Dx accurately reflected those on a calibrated slide. Meets criteria.
    Human FactorsSafe and effective for intended users, uses, and use environments.AISight Dx has been found to be safe and effective for the intended users, uses, and use environments. Meets criteria.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document states that two separate clinical studies were conducted, one for each scanner (Hamamatsu NanoZoomer S360MD and Leica Aperio GT 450 DX).
      • The sample sizes for these clinical studies are not explicitly stated in the provided text.
      • Data Provenance: Not explicitly mentioned, but the study compares performance against "the original sign-out pathologic diagnosis using MO [ground truth, (GT)] rendered at the institution," suggesting the data is derived from clinical practice, likely retrospective or a mix, given the "original sign-out" aspect. The country of origin is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The study involved "3 reading pathologists" for assessing the differences in major discordance rates.
      • Qualifications of experts: Not explicitly stated, but they are referred to as "reading pathologists," indicating they are qualified to make primary diagnoses.
    3. Adjudication method for the test set:

      • The "reference (main) diagnosis" was the "original sign-out pathologic diagnosis using MO [ground truth, (GT)] rendered at the institution."
      • The document implies that this "original sign-out" acted as the ground truth. There's no explicit mention of an adjudication process (e.g., 2+1, 3+1 consensus) to establish this ground truth beyond the initial clinical diagnosis. The major discordance rate was calculated between MD (manual digital read), MO (manual optical read), and GT (original sign-out).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Yes, an MRMC-like study was done as it involved "3 reading pathologists" evaluating cases using both manual digital (MD) and manual optical (MO) methods.
      • Effect size of improvement with AI vs without AI assistance: This study did not measure the improvement of human readers with AI assistance. The AISight Dx is presented as a viewer and management software, not an AI-assisted diagnostic tool. The study aimed to demonstrate non-inferiority of digital viewing (MD) versus traditional optical viewing (MO) for primary diagnosis, where the software is simply the viewing platform, not an aid in interpretation itself. Therefore, no "effect size of human readers improving with AI vs without AI assistance" is reported because the device is not described as providing AI assistance for diagnostic tasks in this context.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. The AISight Dx is explicitly described as "an aid to the pathologist to review, interpret, and manage digital images." The clinical study evaluated "manual digital read (MD)" which is a human pathologist reading digital slides using the AISight Dx, compared to "manual optical (MO)" which is a human pathologist reading glass slides. The device is not an autonomous AI diagnostic algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth (GT) for the clinical study was the original sign-out pathologic diagnosis using manual optical microscopy (MO) rendered at the institution. This can be categorized as a form of expert (pathologist) ground truth based on clinical practice/standard of care.
    7. The sample size for the training set:

      • The document for AISight Dx does not mention a training set size. This is expected as AISight Dx is described as a viewing and management software, not an AI model that requires a training set for diagnostic capabilities. The performance data focuses on its function as a display and interpretation platform for human pathologists.
    8. How the ground truth for the training set was established:

      • As there's no mention of a training set for an AI model within the AISight Dx software (it's a viewer), there's no information on how a ground truth for such a set would be established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250414

    Validate with FDA (Live)

    Device Name
    CaloPix
    Manufacturer
    Date Cleared
    2025-05-14

    (90 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use Only

    CaloPix is a software only device for viewing and management of digital images of scanned surgical pathology slides prepared from Formalin-Fixed Paraffin Embedded (FFPE) tissue.

    CaloPix is intended for in vitro diagnostic use as an aid to the pathologist to review, interpret and manage these digital slide images for the purpose of primary diagnosis.

    CaloPix is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and the validity of the interpretation of images using CaloPix.

    CaloPix is intended to be used with the interoperable components specified in the below Table:

    Scanner HardwareScanner Output File FormatInteroperable Displays
    Leica Aperio GT 450 DX scannerSVSDell U3223QE
    Hamamatsu NanoZoomer S360MD Slide scannerNDPIJVC Kenwood JD-C240BN01A
    Device Description

    CaloPix, version 6.1.0 IVDUS, is a web-based software-only device that is intended to aid pathology professionals in viewing, interpreting and managing digital Whole Slide Images (WSI) of glass slides obtained from the Hamamatsu NanoZoomer S360MD slide scanner (NDPI file format) and viewed on the JVC Kenwood JD-C240BN01A display, as well as those obtained from the Leica Aperio GT 450 DX scanner (SVS file format) and viewed on the Dell U3223QE display.

    CaloPix does not include any automated Image Analysis Applications that would constitute computer aided detection or diagnosis.

    CaloPix is for viewing digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.

    As a whole, CaloPix is a pathology Image Management System (IMS) which brings case-centric digital pathology image management, collaboration, and image processing. CaloPix consists of:

    • Integration with Laboratory Information Systems (LIS): Allows to obtain automatically from the LIS patient data associated with the cases, scanned whole slide images and other related medical images to be analyzed. The data stored in the database is automatically updated according to the interface protocol with the LIS.

    • DataBase: After ingestion, scanned WSI can be organized in the CaloPix database consisting of folders (cases) containing patient identification data and examination results from a LIS.

    Ingestion of the slides is performed through an integrated module that allows their automatic indexation based on patient data retrieved from the LIS. After their ingestion, image files are stored in a CaloPix-specific file storage environment, that can be on premises or in the cloud.

    • The CaloPix viewer component to process scanned whole slide images, that includes functions for panning, zooming, screen capture, annotations, distance and surface measurement, and image registration. This viewer relies on image servers (IMGSRV) which extract image tiles from the whole slide image file and send these tiles to the CaloPix viewer for smooth and fast viewing.
    AI/ML Overview

    The FDA 510(k) clearance letter for CaloPix indicates that the device's performance was evaluated through a series of tests to demonstrate its safety and effectiveness. The primary study described in the provided document focuses on technical performance testing rather than a clinical multi-reader multi-case (MRMC) study.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Pixel-wise comparison (Image Reproduction Accuracy)The 95th percentile of the pixel-wise color differences (CIEDE2000, ΔE00) in any image pair between CaloPix and the predicate device's IRMS must be less than 3 (ΔE00 < 3).Test results showed that the 95th percentile of pixel-wise differences between CaloPix and the comparators (NZViewMD and Aperio WebViewer DX) were less than 3 CIEDE2000. The output images were considered pixel-wise identical and visually adequate.
    Measurements of Area and DistanceAccuracy of measurements made in CaloPix reviewer confirmed by comparing to predicate device viewers. (Specific numerical criteria not provided).Tests verified that the distance and area measurements made in the CaloPix viewer accurately reflected the measurements made in the NZViewMD viewer and the Aperio WebViewer DX viewer.
    Turnaround Time- Image loading: ≤ 10 seconds- Panning: ≤ 7 seconds- Zooming: ≤ 7 secondsTurnaround times for opening an image and panning or zooming have been determined and found to be adequate for the intended use of the subject device. (Specific measured times not provided, but stated to meet criteria).
    Human Factors Validation StudyCaloPix to be safe and effective for the intended users, uses, and use environments (per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices (2016)").CaloPix has been found to be safe and effective for the intended users, uses and use environments.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size for Pixel-wise Comparison: 30 whole slide images (WSIs). Specifically:
      • 25 H&E-stained (Formalin-Fixed Paraffin Embedded - FFPE) tissue glass slides.
      • 5 IHC-stained (Masson's trichrome stain, CD8, CD3, or CD20) FFPE tissue glass slides.
    • Data Provenance: Not explicitly stated regarding country of origin. The study appears to be a technical performance test using pre-scanned slides. It is unclear if these were from a retrospective or prospective collection for the purpose of this specific test. The mention of "FFPE tissue glass slides" implies clinical relevance.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: One pathologist.
    • Qualifications of Experts: The text states, "as verified by a pathologist," but does not specify the pathologist's experience level (e.g., years of experience, board certification). This was for identifying "relevant pathological features" for ROIs.

    4. Adjudication Method for Test Set

    • No "adjudication method" in the sense of multiple readers coming to a consensus for diagnostic ground truth is described. The ground truth for the pixel-wise comparison was based on pre-identified Regions of Interest (ROIs) verified by a single pathologist. For the measurement and turnaround time tests, the comparison was against predicate device software performance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done comparing human reader performance with and without AI assistance. The described tests are technical performance evaluations of the viewing and management software's rendering accuracy, measurement functions, and speed.

    6. Standalone (Algorithm Only) Performance

    • This device, CaloPix, is described as a "software only device for viewing and management of digital images." It "does not include any automated Image Analysis Applications that would constitute computer aided detection or diagnosis." Therefore, no standalone (algorithm only) performance, as typically understood for an AI/CADe/CADx device, was conducted or is applicable since the device's function is solely as a viewing platform, not an analytical algorithm.

    7. Type of Ground Truth Used

    • Technical Ground Truth: For the pixel-wise comparison, the "ground truth" was established by comparing CaloPix's image rendering output against the image rendering of the predicate device's Image Review Manipulation Software (IRMS) (i.e., NZViewMD and Aperio WebViewer DX), with the acceptance criteria based on CIEDE2000 values. The relevant pathological features within the slides were identified by a pathologist for ROI selection.
    • For measurements, the ground truth was the measurements obtained from the predicate device's IRMS.
    • For turnaround time, the ground truth was the pre-defined time limits.
    • For human factors, the ground truth was the satisfaction of usability and safety criteria.

    8. Sample Size for the Training Set

    • Not Applicable / Not Provided: As CaloPix is a viewing and management software without AI/CAD algorithms, there is no mention of a "training set" in the context of machine learning model development. The tests described are functional and technical performance evaluations, not AI model validation.

    9. How Ground Truth for Training Set was Established

    • Not Applicable / Not Provided: See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242244

    Validate with FDA (Live)

    Device Name
    Viewer+
    Manufacturer
    Date Cleared
    2025-03-14

    (226 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    Viewer+ is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. Viewer+ is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Viewer+ is intended for use with Hamamatsu NanoZoomer S360MD Slide scanner and BARCO MDPC-8127 display.

    Device Description

    Viewer+, version 1.0.1, is a web-based software device that facilitates the viewing and navigating of digitized pathology images of slides prepared from FFPE-tissue specimens acquired from Hamamatsu NanoZoomer S360MD Slide scanner and viewed on BARCO MDPC-8127 display. Viewer+ renders these digitized pathology images for review, management, and navigation for pathology primary diagnosis.

    Viewer+ is operated as follows:

      1. Image acquisition is performed using the NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
      1. Once image acquisition is complete and the image becomes available in the scanner's database file system, a separate medical image communications software (not part of the device) automatically uploads the image and its corresponding metadata to persistent cloud storage. Image and data integrity checks are performed during the upload to ensure data accuracy.
      1. The subject device enables the reading pathologist to open a patient case, view the images, and perform actions such as zooming, panning, measuring distances and areas, and annotating images as needed. After reviewing all images for a case, the pathologist will render a diagnosis.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Viewer+ device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Pixel-wise comparison (of images reproduced by Viewer+ and NZViewMD for the same file generated from NanoZoomer S360md Slide Scanner)The 95th percentile of pixel-wise differences between Viewer+ and NZViewMD was less than 3 CIEDE2000, indicating their output images are pixel-wise identical and visually adequate.
    Turnaround time (for opening, panning, and zooming an image)Found to be adequate for the intended use of the device.
    Measurement accuracy (using scanned images of biological slides)Viewer+ was found to perform accurate measurements with respect to its intended use.
    Usability testingDemonstrated that the subject device is safe and effective for the intended users, uses, and use environments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size of images or cases used for the "Test Set" in the performance studies. It mentions "scanned images of the biological slides" for measurement accuracy and "images reproduced by Viewer+ and NZViewMD for the same file" for pixel-wise comparison.

    The data provenance (country of origin, retrospective/prospective) is also not specified in the provided text.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts or their qualifications used to establish ground truth for the test set. It mentions that the device is "an aid to the pathologist" and that "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision." However, this relates to the intended use and not a specific part of the performance testing described.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) used for establishing ground truth or evaluating the test set results. The pixel-wise comparison relies on quantitative color differences, and usability is assessed according to FDA guidance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance is mentioned or implied in the provided text. The device is a "viewer" and not an AI-assisted diagnostic tool that would typically involve such a study.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

    The performance tests described (pixel-wise comparison, turnaround time, measurements) primarily relate to the technical functionality of the Viewer+ software itself, which is a viewing and management tool. These tests can be interpreted as standalone assessments of the software's performance in rendering images and providing basic functions like measurements. However, it's crucial to note that Viewer+ is an "aid to the pathologist" and not intended to provide automated diagnoses without human intervention. The "standalone" performance here refers to its core functionalities as a viewer, not as an autonomous diagnostic algorithm.

    7. Type of Ground Truth Used

    • Pixel-wise comparison: The ground truth for this test was the image reproduced by the predicate device's software (NZViewMD) for the same scanned file. The comparison was quantitative (CIEDE2000).
    • Measurements: The ground truth would likely be established by known physical dimensions on the biological slides, verified by other means, or through precise calibration. The document states "Measurement accuracy has been verified using scanned images of the biological slides."
    • Usability testing: The ground truth here is the fulfillment of usability requirements and user satisfaction/safety criteria, as assessed against FDA guidance.

    8. Sample Size for the Training Set

    The document does not mention the existence of a "training set" in the context of the Viewer+ device. This is a software-only device for viewing and managing images, not an AI/ML algorithm that typically requires a training set for model development.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for this device, information on how its ground truth was established is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241273

    Validate with FDA (Live)

    Device Name
    FullFocus
    Manufacturer
    Date Cleared
    2025-01-09

    (248 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    FullFocus is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. FullFocus is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. FullFocus is intended to be used with the interoperable components specified in the below Table.

    Table: Interoperable components of FullFocus

    Scanner HardwareScanner Output file formatInteroperable Displays
    Leica Aperio GT 450 DX scannerDICOM, SVSDell UP3017Dell U3023E
    Hamamatsu NanoZoomer S360MD Slide ScannerNDPIDell U3223QEJVC-Kenwood JD-C240BN01A
    Device Description

    FullFocus, version 2.29, is a web-based software-only device that facilitates the viewing and navigating of digitized pathology images of slides prepared from FFPE-tissue specimens acquired from FDA cleared digital pathology scanners on FDA cleared displays. FullFocus renders these digitized pathology images for review, management and navigation for pathology primary diagnosis.

    Image acquisition is performed using the intended scanner (s), with the operator conducting quality control on the digital WSI images according to the scanner's instructions for use and lab specifications to determine if re-scans are needed. Please see the Intended Use section and below tables for specifics on scanners and respective displays for clinical use.

    Once a whole slide image is acquired using the intended scanner and becomes available in the scanner's database file system, a separate medical image communications software (not part of the device), automatically uploads the image and corresponding metadata to persistent cloud storage. Integrity checks are performed during the upload to ensure data accuracy.

    The subject device enables the reading pathologist to open a patient case, view the images, and perform actions such as zooming, panning, measuring distances and annotating images as needed. After reviewing all images for a case, the pathologist will render a diagnosis.

    FullFocus operates with and is validated for use with the FDA cleared components specified in the tables below:

    Scanner HardwareScanner Output file formatInteroperable Displays
    Leica Aperio GT 450 DX scannerDICOM, SVSDell UP3017Dell U3023E
    Hamamatsu NanoZoomer S360MD Slide ScannerNDPIDell U3223QEJVC-Kenwood JD-C240BN01A

    Table 1: Interoperable Components Intended for Use with FullFocus

    FullFocus version 2.29 was not validated for the use with images generated with Philips Ultra Fast Scanner.

    Table 2: Computer Environment/System Requirements for during the use of FullFocus

    EnvironmentComponentMinimum Requirements
    HardwareProcessor1 CPU, 2 cores, 1.6GHz
    Memory4 GB RAM
    NetworkBandwidth of 10Mbps
    SoftwareOperating System• Windows• macOS
    Browser• Google Chrome (129.0.6668.90 or higher)• Microsoft Edge (129.0.2792.79 or higher)
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Pixel-wise comparison: The 95th percentile of pixel-wise color differences in any image pair across all required screenshots must be less than 3.0 ΔE00 when compared to comparator (predicate device's Image Review Manipulation Software - IRMS) for identical image reproduction. This indicates visual adequacy for human readers.The 95th percentile of pixel-wise differences between FullFocus and the comparators were less than 3 CIEDE2000, indicating that their output images can be considered to be pixel-wise identical. FullFocus has been found to visually adequately reproduce digital pathology images to human readers with respect to its intended use.
    Turnaround time (Case selection): It should not take longer than 10 seconds until the image is fully loaded when selecting a case.System requirements fulfilled: Not longer than 10 seconds until the image is fully loaded.
    Turnaround time (Panning/Zooming): It shall not take longer than 7 seconds until the image is fully loaded when panning and zooming the image.System requirements fulfilled: Not longer than 7 seconds until the image is fully loaded.
    Measurement Accuracy (Straight Line): The 1mm measured line should match the reference value exactly 1mm ± 0mm.All straight-line measurements compared to the reference were exactly 1mm, with no error.
    Measurement Accuracy (Area): The measured area must match the reference area exactly 0.2 x 0.2 mm for a total of 0.04 mm² ± 0 mm².All area measurements compared to the reference value were exactly 0.04mm², with no error.
    Measurement Accuracy (Scalebar): 2mm scalebar is accurate.All Tests Passed.
    Human Factors Testing: (Implied from previous clearance) Safe and effective use by representative users for critical user tasks and use scenarios.Human factors study designed around critical user tasks and use scenarios performed by representative users were conducted for previously cleared FullFocus, version 1.2.1, in K201005, per FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices (2016)". Human factors validation testing is not necessary as the user interface hasn't changed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Pixel-wise Comparison: 30 formalin-fixed paraffin-embedded (FFPE) tissue glass slides, representing a range of human anatomical sites.
    • Sample Size for Turnaround Time & Measurements: Not explicitly stated as a number of distinct cases or images beyond the 30 slides used for pixel-wise comparison. For measurements, a "1 Calibration Slide" was used per test.
    • Data Provenance: The text does not explicitly state the country of origin. The slides are described as "representing a range of human anatomical sites," implying a diverse set of real-world pathology samples. It is a retrospective study as it states "30 formalin-fixed paraffin-embedded (FFPE) tissue glass slides... were scanned".

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Pixel-wise Comparison: "For each WSI, three regions of interest (ROIs) were identified to highlight relevant pathological features, as verified by a pathologist."
      • Number of Experts: At least one pathologist.
      • Qualifications: "A pathologist" (specific qualifications like years of experience are not provided).
    • Measurements: No expert was explicitly mentioned for establishing ground truth for measurements; it relies on a "test image containing objects with known sizes" (calibration slide) and "reference value."

    4. Adjudication Method for the Test Set

    • The text does not mention an explicit adjudication method (like 2+1 or 3+1 consensus) for the pixel-wise comparison or measurement accuracy. For the pixel-wise comparison, ROIs were "verified by a pathologist," suggesting a single-expert verification rather than a consensus process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done in this context. The study focused on demonstrating identical image reproduction (pixel-wise comparison) and technical performance (turnaround time, measurement accuracy) of the FullFocus viewer against predicate devices' viewing components. It did not directly assess the improvement in human reader performance (e.g., diagnostic accuracy or efficiency) with or without AI assistance. The device is a "viewer and management software," not an AI diagnostic aid in the sense of providing specific findings or interpretations.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone "algorithm only" performance was effectively done for the technical aspects. The pixel-wise comparison directly compares the image rendering of FullFocus with the predicate viewer's rendering without human intervention in the comparison process itself (though a pathologist verified ROIs). Similarly, turnaround times and measurement accuracy are intrinsic technical performances of the software.

    7. The Type of Ground Truth Used

    • Pixel-wise Comparison: The ground truth for this test was the digital image data as rendered by the predicate device's IRMS. The goal was to show that FullFocus reproduces the same image data. The "relevant pathological features" within ROIs were "verified by a pathologist" which served as a reference for what areas to test, not necessarily a diagnostic ground truth for the device's output.
    • Measurements: The ground truth was based on known physical dimensions within a calibration slide and corresponding "reference values."

    8. The Sample Size for the Training Set

    • The provided text does not mention a training set. This is expected because FullFocus is a viewer and management software for digital pathology images, not an AI or machine learning algorithm that is "trained" on data to make predictions or assist in diagnosis directly. Its core function is to display existing image data accurately and efficiently.

    9. How the Ground Truth for the Training Set Was Established

    • As no training set is mentioned (since it's a viewer software), this question is not applicable based on the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240303

    Validate with FDA (Live)

    Date Cleared
    2024-10-28

    (269 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    MetaLite DX Digital Pathology Software is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides.

    MetaLite DX Digital Pathology Software is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. MetaLite DX Digital Pathology Software is intended for use with Philips Ultra Fast Scanner and the Barco MDPC-8127 display.

    Device Description

    MetaLite DX Digital Pathology Software, Model MLDXUS, version 1.2.1 is software designed for viewing digital pathology images of glass slides from the Philips IntelliSite Pathology Solution Ultra-Fast Scanner (PIPS-UFS), version 1.8.4 on Barco MDPC-8127 display.

    MetaLite DX Digital Pathology Software is operated as follows:

    Before scanning the slide on the PIPS-UFS, the technician performs quality control on the tissue of interest. The images captured by the PIPS-UFS are compressed using Philips' proprietary iSyntax format and are transmitted to the Philips Image Management System (IMS).

    (1) After the Whole Slide Images (WSIs) are successfully, acquired by using PIPS-UFS, the WSIs are stored in the Local file system. A qualified pathologist will upload compatible iSyntax format digital pathology images, and the software will load them to the "Main Viewer" area of the graphical interface for the pathologist to view.

    (2) Once properly loaded, the pathologist will use the inherent features of the device (including tools that allow for adjusting the position and viewing angle of the image, measuring lengths between two coordinates, and adding annotations to specific regional areas).

    (3) After viewing all images for a patient (case), the pathologist will make a diagnosis. The diagnosis will be documented in another system, e.g., a Laboratory Information System (LIS).

    The software has various features such as zoom-in and zoom-out functions, scale display, thumbnail view, measurement function, annotation function, and panning function to help pathologists interpret, diagnose and manage digital whole slide images. The MetaLite DX Digital Pathology Software is validated for use with the components specified the tables below.

    AI/ML Overview

    Let's break down the acceptance criteria and the study proving the device meets them based on the provided text.

    Based on the provided text, the "MetaLite DX Digital Pathology Software" (MLDXUS) is a software-only device intended for viewing and managing digital images of scanned surgical pathology slides for primary diagnosis. The performance testing section describes the studies conducted to demonstrate its safety and effectiveness.

    Here's the information organized as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied)Reported Device Performance
    Pixel-wise comparisonThe output images of the MetaLite DX Digital Pathology Software should be visually identical to those produced by the predicate device (PIPS IMS) for the same file.The 95th percentile of pixel-wise differences between MetaLite DX Digital Pathology Software and PIPS IMS was less than 3 CIEDE2000, indicating that their output images are pixel-wise identical and visually adequate.
    Turnaround timeOpening, panning, and zooming an image should be within an adequate timeframe for intended use (implicitly, within 5 seconds based on the outcome).The turnaround time for opening, panning, and zooming an image is within 5 seconds. This was determined and found to be adequate for the intended use.
    MeasurementsThe software should perform accurate measurements.Measurement accuracy was verified using a scanned image of a calibration scale slide. MetaLite DX Digital Pathology Software was found to perform accurate measurements with respect to its intended use. (Note: Predicate device also measures area, but this device only explicitly states distance measurement in the comparison table, although the performance statement is general for "measurements").
    Usability testingThe device should be safe and effective for its intended users, uses, and use environments.Conducted per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices (2016)". The test result demonstrated that the subject device has been found to be safe and effective for the intended users, uses, and use environments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for the test sets used in the pixel-wise comparison, turnaround time, measurement accuracy, or usability testing.

    The document does not specify the provenance of the data (e.g., country of origin, retrospective or prospective). It only states that the images used were "iSyntax file generated from UFS 1.8.4," which refers to the Philips Ultra Fast Scanner, a compatible component.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used or their specific qualifications for establishing ground truth for any of the performance tests.

    • For the pixel-wise comparison, the ground truth seems to be the output of the predicate device (PIPS IMS) for the same initial image file, rather than expert judgment on clinical images.
    • For turnaround time and measurements, the ground truth would be objectively measurable (time, known distances on a calibration slide).
    • For usability testing, ground truth typically involves observing user interactions and identifying errors or difficulties, rather than a clinical ground truth established by experts.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly conducted for the MetaLite DX Digital Pathology Software as described in this document. The studies performed focus on technical performance (pixel comparison, speed, measurement accuracy) and usability of the software as a viewing and management tool, not on its impact on human reader diagnostic accuracy or efficiency with and without AI assistance. The device is purely a viewer/manager and does not incorporate AI for diagnosis.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The studies described are primarily standalone in the sense that they evaluate the software's technical performance attributes (pixel reproduction, speed, measurement accuracy) independent of a pathologist's diagnostic performance. The usability test involved human interaction but assessed the usability of the software interface, not the diagnostic accuracy of the human using it. The device itself is described as "software only" and an "aid to the pathologist," rather than an AI diagnostic algorithm.

    7. Type of Ground Truth Used

    • Pixel-wise comparison: The ground truth appears to be the output of a reference system (PIPS IMS) for the same iSyntax file. This is a technical ground truth based on image fidelity.
    • Turnaround time: The ground truth is objective measurement of time.
    • Measurements: The ground truth is objective, known distances on a calibration scale slide.
    • Usability testing: The ground truth is based on observed user interactions, identification of use errors, and compliance with human factors principles, as guided by FDA guidelines.

    There is no mention of clinical ground truth (e.g., expert consensus on pathology diagnoses, or outcomes data) being used for these particular performance tests, as the device is not a diagnostic AI algorithm.

    8. Sample Size for the Training Set

    The document does not provide any information regarding a training set size. This is consistent with the device being a viewer and manager of digital images, not an AI algorithm that requires a training set for model development.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described for this type of device, this information is not applicable and not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233126

    Validate with FDA (Live)

    Device Name
    aetherSlide
    Manufacturer
    Date Cleared
    2024-05-30

    (246 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    aetherSlide is a software-only device intended for viewing and managing digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. aetherSlide is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. aetherSlide is intended for use with the Philips Ultra Fast Scanner (UFS) and the Philips PS27QHDCR monitor.

    Device Description

    aetherSlide, version 101692 is a web-based, software only device that is intended to aid pathology professionals in viewing, interpretation and management of digital whole slide images (WSI) of scanned surgical pathology slides prepared from formalin-fixed paraffin-embedded (FFPE) tissue obtained from Philips Ultra Fast Scanner (UFS). It aids the pathologist in the review, interpretation, and management of pathology slide digital images used to generate a primary diagnosis.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the aetherSlide device, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Pixel-wise comparison (Image Fidelity): Images reproduced by aetherSlide should be visually adequate.The 95th percentile of pixel-wise differences between aetherSlide and PIPS IMS was less than 3 CIEDE2000, indicating pixel-wise identical output images, and color images were visually adequate.
    Turnaround time - Case Selection: Not longer than 10 seconds until the image is fully loaded after selecting a case.Turnaround times for opening an image were determined and found to be adequate for the intended use.
    Turnaround time - Panning: Not longer than 7 seconds until the image is fully loaded when panning one-quarter of the monitor.Turnaround times for panning were determined and found to be adequate for the intended use.
    Measurements (Accuracy): Perform accurate measurements.Measurement accuracy was verified using a scanned image of a grid micrometer and found to be accurate for the intended use.
    Usability: Safe and effective for intended users, uses, and use environments.The usability test demonstrated the subject device is safe and effective for the intended users, uses, and use environments.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated. The pixel-wise comparison mentions "the same iSyntax file" but does not quantify the number of such files or cases used. Similarly, for turnaround time, measurements, and usability, specific sample sizes are not provided.
      • Data Provenance: Not explicitly stated. The document mentions the device is intended for use with "scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue" obtained from the Philips Ultra Fast Scanner (UFS). This implies the data would be clinical pathology slides, but their origin (country, retrospective/prospective collection) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The studies described are primarily technical performance assessments (pixel comparison, turnaround time, measurement accuracy, usability) rather than diagnostic accuracy studies requiring expert consensus as ground truth. The "ground truth" for these tests would be the reference values or expected outcomes based on the technical specifications of the images or system.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided as the studies are focused on technical performance rather than diagnostic outcomes.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. The aetherSlide is described as a "software-only device intended for viewing and managing digital images... It is an aid to the pathologist to review, interpret and manage digital images...". It is a WSI viewer and manager, not an AI-assisted diagnostic tool designed to improve human reader performance (e.g., by detecting abnormalities). Therefore, this type of study would not be applicable for this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, standalone technical performance assessments were done. The pixel-wise comparison, turnaround time tests, and measurement accuracy tests are evaluations of the algorithm's direct performance in rendering images, speed, and accuracy, independent of a pathologist's diagnostic performance. The usability testing, while involving human users, focuses on the system's interface and interaction, not diagnostic accuracy.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Technical Reference Standards:
        • Pixel-wise comparison: The ground truth was the image reproduced by the predicate device (Philips IntelliSite Pathology Solution - PIPS IMS) for the same iSyntax file.
        • Turnaround time: The ground truth was the pre-defined target times (10 seconds for case selection, 7 seconds for panning).
        • Measurements: The ground truth was the known dimensions on a "scanned image of the grid micrometer."
        • Usability: The ground truth would be the safety and effectiveness criteria outlined in the usability engineering guidance (e.g., task completion rates, error rates, user feedback conforming to safety and effectiveness).
    7. The sample size for the training set:

      • The document does not mention a training set size. The aetherSlide is a viewer and management software, not a machine learning model that typically requires a dedicated training set. Its functionality is based on rendering and interacting with existing digital slide images.
    8. How the ground truth for the training set was established:

      • As there is no mention of a machine learning component or a training set for diagnostic purposes, the concept of establishing ground truth for a training set is not applicable to the information provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212361

    Validate with FDA (Live)

    Device Name
    Novo
    Manufacturer
    Date Cleared
    2022-08-11

    (377 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Novo is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret, and manage digital images of these slides for primary diagnosis. Novo is not intended for use with frozen sections, cytology, or non- FFPE hematopathology specimens.

    It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Novo is intended for use with the Philips Ultra Fast Scanner and the Barco PP27QHD or Philips PS27QHDCR display.

    Device Description

    The PathAI Novo device is a web-based software-only device that is intended to aid pathology professionals in the viewing, interpretation, and management of digital whole slide images (WSIs) of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue using the Philips IntelliSite Pathology Solution (PIPS) Ultra Fast Scanner (UFS).

    The proposed device is typically operated as follows:

      1. A user prepares and scans slides and reviews the slide quality in accordance with the PIPS UFS IFU and standard lab procedures. The Novo device workflow is initiated when a user uploads WSIs from the local file system to the cloud storage using Novo.
      1. After uploading WSIs to cloud storage using Novo, a user builds a patient accession using the patient's medical record number (MRN), date of birth (DOB) and accession ID to support linkage of one or more slides from a single procedure using patient identifiers in Novo.
      1. A pathologist uses the slide viewer to perform their primary diagnosis workflow including zooming and panning images.

    After viewing all images belonging to a particular accession, the pathologist will make a diagnosis.

    AI/ML Overview

    The provided text describes the regulatory clearance for the "Novo" device, a software-only whole slide imaging system, and references a clinical study conducted to establish its substantial equivalence to a predicate device. However, the document primarily focuses on regulatory approval and does not contain the detailed acceptance criteria table or comprehensive study breakdown as requested in the prompt.

    Therefore, the following response will extract what is available and highlight where information is missing based on your request.


    Acceptance Criteria and Device Performance for Novo (as described by available information)

    Based on the provided FDA 510(k) summary, details regarding specific quantifiable acceptance criteria and performance beyond a non-inferiority finding are limited. The document focuses on demonstrating substantial equivalence to a predicate device (Philips IntelliSite Pathology Solution - PIPS).

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric (Inferred/Stated)Acceptance Threshold (Inferred/Stated)Reported Device Performance
    Clinical EquivalenceMajor Discordance RateUpper limit of 95% CI for difference in major discordance rates < 4%.-0.1% (95% CI, -2.05, 1.78) for all organs (MD vs. MO compared to GT). Met.
    Image Loading SpeedLoad time for selected images< 7 seconds"Images load in less than 7 seconds when selected for viewing." Met.
    Image Panning/Zooming SpeedLoad time during panning/zooming< 10 seconds"Images load in less than 10 seconds when panning or zooming." Met.
    Image Reproduction (Pixel-wise)Color differences ($\Delta$E00)Implicitly, not significantly different from PIPS/IMS with JPEG compression."Color differences ($\Delta$E00) between Novo and PIPS/IMS are not zero." "Novo-generated images are similar to PIPS/IMS-generated images that had been JPEG-compressed at quality 95."
    Human FactorsSafety and EffectivenessImplicitly, found safe and effective for intended users, uses, and environments."Novo has been found to be safe and effective for the intended users, uses, and use environments." Met.

    Missing Information (Not Available in the Provided Text):

    • Explicit, pre-specified quantitative acceptance criteria beyond the non-inferiority margin for major discordance.
    • Detailed quantitative performance for features like color differences beyond a statement of similarity to JPEG-compressed images.

    Study Details (Extracted from the provided text):

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated. The text mentions "WSIs of H&E stained FFPE tissue slides."
    • Data Provenance: Not specified for the test set. (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. The reference diagnosis (ground truth) was the "original sign-out pathologic diagnosis using MO [manual optical] rendered at the institution," implying involvement of qualified pathologists for routine diagnosis, but specific qualifications for ground truth establishment are not given.

    4. Adjudication method for the test set:

    • Adjudication Method: Not specified. The study compared "major discordance rates between MD [manual digital read, using Novo] and MO [manual optical] when compared to the reference (main) diagnosis, which was the original sign-out pathologic diagnosis using MO [ground truth, (GT)] rendered at the institution." This suggests a comparison against an existing diagnosis, not necessarily a separate adjudication process for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: Yes, a clinical study was conducted. It involved human pathologists making diagnoses using both the Novo device (MD) and conventional microscopy (MO).
    • Effect Size of Improvement: The study focused on non-inferiority of the manual digital read (MD) using Novo compared to manual optical (MO) read. It did not measure "how much human readers improve with AI vs without AI assistance" because Novo is described as a "viewing and management" tool, not an AI-assisted diagnostic tool. Its purpose is to present the image for the pathologist's primary diagnosis, making it a replacement for conventional microscopy, not an enhancement tool for human readers in the context of an AI-assisted workflow. The primary outcome was maintaining diagnostic accuracy (non-inferiority) when switching from MO to MD.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: No, this device is a "viewer and management" software for human pathologists. It does not perform diagnostic algorithms independently. The performance data presented (non-inferiority) is explicitly human-in-the-loop (MD: "manual digital read").

    7. The type of ground truth used:

    • Ground Truth Type: "The original sign-out pathologic diagnosis using MO [manual optical] rendered at the institution." This can be categorized as expert consensus/clinical outcomes data based on standard practice.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable/Not specified. Novo is a viewing and management system, not described as an AI/machine learning algorithm that requires a "training set" in the conventional sense for a diagnostic prediction model. The "pixel-wise comparison" tests were likely technical assessments, not dependent on a "training set."

    9. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: Not applicable, as there is no mention of an algorithm requiring a "training set" with ground truth in the provided information.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210811

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-03-01

    (349 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamyx Digital Pathology Software is intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis.

    Dynamyx Digital Pathology Software is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Dynamyx Digital Pathology Software.

    The Dynamyx Digital Pathology Software consists of the Installed Pathologist Client and the Pathologist Workstation Web Client. The Installed Pathologist Client is intended for use with Leica's Aperio AT2 DX scanner and Dell MR2416 monitor as well as Philips' Ultra Fast Scanner and Philips PP27QHD monitor. The Pathologist Workstation Web Client is intended for use with Philips' Ultra Fast Scanner and Philips PP27QHD monitor.

    Device Description

    Dynamyx Digital Pathology Software is a client-server software device used for importing, displaying, navigating, and annotating whole slide images obtained from the Leica Aperio AT2 DX scanner or the Philips Ultra Fast Scanner.

    Whole slide images are created by scanning glass microscope slides using a digital slide scanner which are then imported into the Dynamyx Digital Archive server. Dynamyx uses the image decoding libraries licensed by Leica and Philips for the native images. Dynamyx then uses lossless compression to send the images to the Dynamyx viewer.

    Note that Dynamyx has two different applications for two different inputs as specified below.

    1. The Dynamyx Web Application running in the Chrome browser can only display WSI from the Philips Ultra Fast Scanner.

    2. The Dynamyx Installed Client Application can display WSI from both the Leica AT2 DX Scanner and the Philips Ultra Fast Scanner.

    Whole slide image files are viewed in the Dynamyx image viewer window by histologists and by pathologists who can also navigate (pan and zoom) and annotate the images.

    Dynamyx incorporates typical histology/pathology workflow and is operated as follows:

      1. Dynamyx receives whole slide images from the scanner as specified above and extracts a copy of the images' metadata. The unaltered images are then sent to the external image storage (Digital Archive). A copy of the image metadata (e.g. the pixel size) is stored in the subject device's database to increase the operational performance (e.g. response times) of Dynamyx.
      1. Depending upon a laboratory's workflow, whole slide images may be reviewed first by histologists to confirm image quality and initiate any slide rescans as necessary prior to being viewed by pathologists. The digital slide review OC status determined by the histologist indicates which slides have been reviewed and approved. The OC status is available to the reading pathologist.
      1. The reading pathologist selects a patient case from a selected worklist within Dynamyx whereby the case images are retrieved from the digital archive.
      1. The reading pathologist uses Dynamyx to view, navigate, annotate, and interpret the digital images. The pathologist can perform the following actions to displayed image:
      • a. Zoom and pan the image at will;
      • b. Adjust the apparent image observed magnification level;
      • c. Measure distances and areas;
      • d. Annotate images and cases;
      1. The above steps are repeated as required.

    After viewing all images, the pathologist will make a diagnosis which is documented in a laboratory information system.

    AI/ML Overview

    There is no information regarding acceptance criteria and a study proving the device meets it for an AI/ML clinical decision support function in the provided text. The document refers to the Dynamyx Digital Pathology Software, which is a viewing and management software for digital images of pathology slides, not an AI/ML device.

    The document discusses non-clinical performance testing for image reproduction, turnaround time, measurement accuracy, and usability, demonstrating substantial equivalence to predicate devices, but this is for the core functionality of a digital pathology viewer, not an AI feature.

    Therefore, I cannot provide the requested information for acceptance criteria and a study proving an AI device meets them.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2