510(k) Data Aggregation

    K Number
    K230531

    Validate with FDA (Live)

    Device Name
    SOZO Pro
    Manufacturer
    Date Cleared
    2023-05-04

    (66 days)

    Product Code
    Regulation Number
    870.2770
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOZO Pro may be used as an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools to assess patients at risk of protein-calorie malnutrition (PCM).

    The SOZO Pro may be further used to estimate the following body composition parameters in humans to track clinically relevant body composition parameters over time:

    • · Fat mass
    • · Fat-free mass
    • Total body water
    • · Intracellular fluid
    • · Extracellular fluid
    • Skeletal muscle mass

    The following outputs are also presented:

    • · Body Mass Index (BMI)
    • · Basal metabolic rate (BMR; based on Mifflin St. Jeor's algorithm) displayed in calories per day

    • Protein and mineral (also known as 'dry lean mass') represents the content of a body that is not fat or fluid; calculated by subtracting total body water weight from fat-free mass weight.

    The SOZO Pro device measures current (1), voltage (V) and phase angle (Phi), and from these values calculates resistance (R).

    reactance (Xc), and impedance (Z), which are used to estimate the above body composition parameters. The device/ software will also display the Cole plot, subject height, weight, age and sex.

    Device Description

    The SOZO Pro system consists of a connected hand and footplate with built-in stainless-steel electrodes, paired with an Android tablet over Bluetooth connection. An app ("SOZOapp"), supplied with the tablet, controls the functionality of the hardware, and supplies the bioimpedance measurement data to a database ("MySOZO") managed on an external cloud- hosted database. Patient weight is measured with load cells located in the SOZO Pro foot unit or can be hand entered.

    Bioimpedance measurements require the patient's weight to be measured by a scale embedded in the base of the system or for the weight of the patient to be entered manually. Following the collection of the patient weight, the patient contacts the SOZO Pro with their bare hands and feet on stainless steel electrodes. The impedance measurement takes about 30 seconds, during which the SOZO Pro® system applies small levels of electrical energy (200µA RMS) to the body across 256 frequencies spaced from 3kHz to 1000kHz and measures the resulting voltage levels.

    AI/ML Overview

    The provided text is a 510(k) summary for the ImpediMed SOZO Pro device. It describes the device, its indications for use, technological characteristics, and performance data to demonstrate substantial equivalence to a predicate device. However, it does not contain detailed information about specific acceptance criteria or a study proving the device meets those criteria, especially in the context of AI-driven performance or human reader improvement.

    The "Performance Data" section primarily focuses on:

    • Electrical safety/EMC: Tested according to IEC 60601 standards.
    • Software V&V: Tested according to ISO 62304.
    • Biocompatibility: Tested according to ISO 10993.
    • Functional performance testing: "undertaken using fixed loads and comparing modified SOZO Pro to predicate SOZO measurements to demonstrate that outputs remained consistently accurate and precise." This is high-level and lacks specifics on metrics or thresholds.
    • Weight Scale Verification: Performed according to NIST Handbook 44 and EU Directive 2014/31/EU.

    Since the provided document does not include the specific details you've requested regarding acceptance criteria and performance study for, for example, an AI component's diagnostic accuracy, human-in-the-loop performance, or detailed ground truth establishment, I cannot populate all sections of your request.

    Based on the available information, here is what can be inferred and what is explicitly missing:

    1. Table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Generic/Inferred)Reported Device Performance (as stated or inferred from "no failures")
    Electrical Safety/EMCMet IEC 60601 standards; CB certificate granted.
    Software Verification & ValidationMet acceptance criteria and performed as intended (per ISO 62304).
    BiocompatibilityPassed ISO 10993 testing; no failures reported.
    Functional PerformanceOutputs remained consistently accurate and precise (compared to predicate, using fixed loads) - Specific metrics or thresholds not provided.
    Weight Scale AccuracyExperienced no failures (per NIST Handbook 44/EU Directive).

    2. Sample size used for the test set and the data provenance:

    • Sample Size (Test Set): Not specified. The document mentions "fixed loads" for functional testing and a comparison to the predicate device, but no human subject test set size for evaluating body composition parameters is detailed.
    • Data Provenance: Not specified. It's likely the testing occurred during device development/modification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not specified. The testing described focuses on engineering/functional validation (electrical, software, biocompatibility, scale accuracy, and comparison to predicate measurements using fixed loads), not on establishing diagnostic ground truth from human data for body composition parameters. The device provides "estimates" and tracks parameters over time, suggesting its performance is validated against its own internal consistency or engineering standards rather than external "expert" ground truth for individual body composition values in a clinical diagnostic sense based on this document.

    4. Adjudication method for the test set:

    • Not applicable/Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a Bioimpedance Spectroscopy (BIS) device for body composition analysis, not an AI-assisted diagnostic imaging device requiring human reader interpretation or MRMC studies. The document does not describe any AI component that would require human-in-the-loop evaluation for diagnostic improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The "Performance Data" section discusses "functional performance testing" using "fixed loads" and comparison to the predicate device, which would be a form of standalone performance evaluation for the device's measurement capabilities.
    • The software V&V confirms the software meets acceptance criteria and performs as intended.
    • However, specific standalone accuracy metrics (e.g., against a gold standard for body composition in human subjects) are not provided in this summary.

    7. The type of ground truth used:

    • For electrical, software, biocompatibility, and scale testing: Engineering standards, established regulatory requirements, and fixed measurement loads act as "ground truth."
    • For "functional performance testing" of body composition outputs: The "ground truth" used was a comparison to the predicate SOZO measurements and "fixed loads" to ensure "consistently accurate and precise" outputs. There is no mention of a clinical gold standard (e.g., DEXA, whole-body MRI, dilution methods) used as ground truth for the body composition parameters in human subjects for this specific K-summary documentation.

    8. The sample size for the training set:

    • Not applicable/Not specified. This document describes a medical device, not an AI model that undergoes "training" in the machine learning sense with a distinct training dataset. The device measures bioimpedance and calculates parameters using established physiological models and algorithms.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, this is not an AI model that undergoes training on a dataset with external ground truth. The device relies on biophysical principles and algorithms for its calculations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190529

    Validate with FDA (Live)

    Device Name
    SOZO
    Manufacturer
    Date Cleared
    2019-11-25

    (266 days)

    Product Code
    Regulation Number
    870.2770
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOZO system may be used as an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools to assess patients at risk of protein-calorie malnutrition (PCM).

    The SOZO system may be further used to estimate the following body composition parameters in humans to track clinically relevant body composition parameters over time:

    • Fat mass
    • Fat-free mass
    • Total body water
    • Intracellular fluid
    • Extracellular fluid
    • Skeletal muscle mass

    The following outputs are also presented:

    • Body Mass Index (BMI)
    • Basal metabolic rate (BMR; based on Mifflin - St. Jeor's algorithm) displayed in calories per day
    • Protein and mineral (also known as 'dry lean mass') represents the content of a body that is not fat or fluid; calculated by subtracting total body water weight from fat-free mass weight.

    The SOZO device measures current (I), voltage (V) and phase angle (Phi), and from these values calculates resistance (R), reactance (Xc), and impedance (Z), which are used to estimate the above body composition parameters. The device/software will also display the Cole plot, subject height, weight, age and sex.

    Device Description

    The SOZO system consists of a connected hand and footplate with built-in stainless steel electrodes, paired with an Android tablet over Bluetooth connection. An app ("SOZOapp"), supplied with the tablet, controls the functionality of the hardware and supplies the bioimpedance measurement data to a database ("SOZOhub") contained within the hospital/facility network.

    Measurements require the patient to make contact with bare hands and feet on stainless steel electrodes. The measurement takes about 30 seconds, during which the SOZO® system applies small levels of electrical current (200µA RMS) to the body across 256 frequencies spaced logarithmically from 3kHz to 1000kHz and measures the resulting voltage levels. Established algorithms are used to analyze data and calculate various body composition parameters and present the outputs for the clinician to review.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Electrical Safety/EMCMeets requirements set forth in IEC 60601 (subparts -1, -1-2, and -1-6). CB certificate granted.
    Software V&VMeets acceptance criteria and performs as intended, tested in accordance with ISO 62304.
    BiocompatibilityPassed testing according to ISO 10993 for a low-risk, limited-contact device; no failures reported.
    Functional PerformanceSystem expected to remain functional throughout its intended life; tested for design reliability by repeatedly placing weights on components.
    Comparative Performance vs. Predicate Device (SFB7)Very strong correlation (r > 0.99) with outputs from the SFB7 device when tested using a fixture with multiple fixed impedance loads representing different "humans."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state a specific "test set" sample size for a clinical study to directly prove the new indications for use. Instead, it states: "Clinical testing: clinical studies have identified the applicability of ImpediMed's BIS technology to act as an aid in body composition." This general statement implies that the underlying Bioimpedance Spectroscopy (BIS) technology has been validated through clinical studies, but it doesn't provide details about studies specific to the SOZO system's new indications or a dedicated test set with sample size.

    For the comparative performance with the predicate device, a "test fixture" was used with "multiple fixed impedance loads," which isn't a human sample size but rather an engineering test.

    Data Provenance: Not explicitly stated for specific clinical studies, only generally refers to "clinical studies."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The document refers to "clinical studies" for the applicability of BIS technology but does not detail how ground truth was established for these studies, nor the number or qualifications of experts involved.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Improvement with AI vs. Without AI Assistance

    No MRMC study is mentioned in the document. The SOZO system is a device for measuring body composition parameters, not an AI-assisted diagnostic tool for human readers in the traditional sense of an MRMC study. Its use is described as "an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools," but no study comparing human performance with and without SOZO assistance is described.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The device itself is a standalone measurement system that takes raw electrical measurements and processes them using "established algorithms" to calculate body composition parameters. The "comparative performance vs. predicate device" test (r > 0.99 correlation) can be considered a standalone performance assessment against a validated reference.

    7. The Type of Ground Truth Used

    • For Body Composition Parameters: The document states that "established algorithms are used to analyze data and calculate various body composition parameters." For the "comparative performance vs. predicate device," the predicate device (SFB7) is essentially the reference standard (ground truth surrogate). However, for the initial validation of the BIS technology to derive these parameters, it is generally understood that such systems are validated against gold standard body composition assessment methods (e.g., DEXA, isotope dilution, direct chemical analysis), though this specific detail is not provided for the SOZO system in this document. The document refers to "clinical studies have identified the applicability of ImpediMed's BIS technology to act as an aid in body composition," implying prior validation work.
    • For Protein-Calorie Malnutrition (PCM) Assessment: The device is used as an adjunct to Subjective Global Assessment (SGA) tools, implying that SGA is the primary assessment method clinicians use.

    8. The Sample Size for the Training Set

    No training set sample size is explicitly mentioned. The device uses "established algorithms," suggesting that the algorithms were developed and potentially trained on datasets prior to this submission. This document focuses on the performance and equivalence of the SOZO device itself, not the detailed development of its underlying algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document. As with the training set size, the document does not delve into the historical development and validation of the "established algorithms" that process the bioimpedance data.