Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181599

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-18

    (273 days)

    Product Code
    Regulation Number
    866.3215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unicel DxH 800 Coulter Cellular Analysis System with Early Sepsis Indicator Application is the quantitative measurement of Monocyte Distribution Width (MDW). The Early Sepsis Indicator is intended for use with adult patients presenting to the emergency department, on whom a white cell differential test has been ordered.

    MDW is measured from a (K2EDTA) whole-blood venous sample within 2 hours of collection. MDW values greater than 20.0 together with other laboratory findings and clinical information, aids in identifying patients with sepsis or at increased risk of developing sepsis within the first 12 hours of hospital admission.

    MDW values greater than 20.0 should be interpreted in association with other clinical information and diagnostic testing, as a proportion of patients without sepsis may have an elevated MDW value at baseline.

    MDW values less than or equal to 20.0 cannot rule out sepsis or the development of sepsis within 12 hours of hospital admission. The Early Sepsis Indicator should not be used as the sole basis to determine the absence of sepsis.

    The predictive value of the Early Sepsis Indicator for identifying sepsis in patients with hematological abnormalities has not been established.

    Device Description

    The Early Sepsis Indicator (ESI) requires the use of the UniCel DxH 800 Coulter Cellular Analysis System (DxH 800) and its reagents, controls and calibrators last cleared under 510(k) K140911.

    The UniCel DxH 800 Coulter Cellular Analysis System contains a quantitative, automated hematology analyzer (DxH 800) designed for in vitro diagnostic use in screening patient populations by clinical laboratories. The system provides a Complete Blood Count (CBC), Leukocyte 5 Part Differential (Diff), Reticulocyte (Retic), Nucleated Red Blood Cell (NRBC) on whole blood, as well as, Total Nucleated Count (TNC), and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial). This submission adds a new parameter, Monocyte Distribution Width (MDW) to those mentioned above. This parameter has been shown to aid in the early detection of Sepsis in emergency room patients.

    The system consists of two primary components, the workstation and the DxH 800 analyzer as shown in Figure 1. DxH 800 System Configuration. The primary function of the DxH 800 analyzer is to process samples and provide results to the workstation. The primary functions of the workstation are: user interface, system control, results processing and storage and external communications. The analyzer runs embedded code and the workstation runs Microsoft Windows 7 Operating System (OS).

    AI/ML Overview

    Here's the breakdown of the acceptance criteria and study detailed in the provided document:

    Acceptance Criteria and Device Performance for UniCel DxH 800 Cellular Analysis System with Early Sepsis Indicator Application

    The device is intended for the quantitative measurement of Monocyte Distribution Width (MDW) to aid in identifying adult patients with sepsis or at increased risk of developing sepsis within the first 12 hours of hospital admission. The key acceptance criterion for clinical accuracy revolves around sensitivity and specificity at a specified MDW cut-off.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria StudyAcceptance Criteria (Threshold)Reported Device Performance (Result)
    Clinical Accuracy (MDW Cut-off of 20.0)- Sensitivity within predefined lower limit- Validated the predefined cut-offs
    - Specificity within predefined lower limit- Demonstrated that the lower limit of sensitivity and specificity were within acceptance criteria for both cut-offs.

    Note: The document states that the study demonstrated that an MDW cut-off of 20.0 units provided an optimum diagnostic ability by balancing the ability to detect positive patients (sensitivity) and negative patients (specificity), and thus this cut-off was selected. Detailed numerical values for sensitivity and specificity at this cut-off are not explicitly provided in the table, but the text assures they met the acceptance criteria.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: Not explicitly stated for the "test set" in a distinct way, but the "Clinical Accuracy" study states it was a "multi-center prospective cohort study."
    • Data Provenance:
      • Country of Origin: Not specified in the provided text.
      • Retrospective or Prospective: Primarily prospective. The clinical accuracy study was a "multi-center prospective cohort study."

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    Note: The document mentions "clinical information and diagnostic testing" and "other laboratory findings" were used in conjunction with MDW values, implying expert judgment in establishing the ground truth for sepsis diagnosis.

    4. Adjudication Method

    • Adjudication Method: Not specified.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This device measures a quantitative biomarker (MDW) from a blood sample, not something interpreted by human readers from images or complex data.

    6. Standalone (Algorithm Only) Performance

    • Standalone Performance: Yes, the fundamental performance assessed is the standalone diagnostic ability of the MDW parameter as an aid in identifying sepsis. The clinical accuracy study directly evaluated the MDW parameter's performance in detecting sepsis.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for sepsis diagnosis was established based on "other laboratory findings and clinical information" and "diagnostic testing." This suggests a comprehensive clinical diagnosis of sepsis, likely involving a combination of clinical assessment, laboratory parameters (beyond MDW), and potentially culture results or other confirmatory tests, as determined by clinicians.

    8. Sample Size for Training Set

    • Sample Size for Training Set: Not explicitly stated as a separate "training set" in the context of machine learning model development. The document focuses on the validation of the MDW parameter and its cut-off.

    Note: The MDW parameter itself is a quantitative measurement derived from hematology analyzer data, not a machine learning model that would typically have a distinct training set in the conventional sense. The "training" or development of the algorithm to calculate MDW would have occurred prior to this validation study, but details are not provided here.

    9. How Ground Truth for Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not specified, as a distinct training set for a machine learning model is not explicitly mentioned or implied for the MDW parameter's development in this document. The MDW value is a quantitative measurement from the analyzer, not a diagnosis itself.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1