Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K202755

    Validate with FDA (Live)

    Date Cleared
    2022-11-05

    (775 days)

    Product Code
    Regulation Number
    866.3181
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaSorin Molecular Simplexa™ Congenital CMV Direct is a real-time PCR assay intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection of cytomegalovirus (CMV) from saliva swabs and urine from infants less than 21 days of age. Positive results from saliva are presumptive and should be confirmed with urine. The results of the Simplexa™ Congenital CMV Direct assay should be used in conjunction with the results of other clinical findings as an aid in the diagnosis of congenital CMV infection.

    This test has not been cleared for screening of blood products for the presence of CMV or for use with samples other than urine and saliva swabs.

    DiaSorin Molecular's Simplexa™ Congenital CMV Positive Control Pack is intended to be used as a control with the Simplexa Congenital CMV Direct kit for use on the LIAISON MDX instrument. This control is not intended for use with other assays or systems.

    Device Description

    The Simplexa™ Congenital CMV Direct assay is a real-time PCR system that enables the direct amplification and detection of CMV DNA from either saliva swab or urine specimens without nucleic acid extraction. The system consists of the Simplexa™ Congenital CMV Direct Reaction Mix, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc (DAD) and associated accessories.

    In the Simplexa™ Congenital CMV Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify CMV DNA. A well-conserved region of the CMV UL83 gene is targeted to identify CMV DNA. An internal control is used to detect PCR failure and/or inhibition.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in a dedicated table format. However, the reported performance metrics imply the criteria for acceptance. For the purpose of this response, I'll infer the implicit acceptance criteria based on the demonstrated performance, generally implying "high agreement" for positive and negative cases.

    Metric (Implicit Acceptance Criteria)Saliva Swab Performance (Retrospective)Urine Performance (Retrospective)Saliva Swab Performance (Prospective)Urine Performance (Prospective)
    Positive Percent Agreement (PPA)100.0% (95% CI: 93% - 100%)100.0% (95% CI: 93% - 100%)94.1% (95% CI: 73% - 99%)95.3% (95% CI: 85% - 99%)
    Negative Percent Agreement (NPA)100.0% (95% CI: 97% - 100%)98.4% (95% CI: 94% - 100%)99.9% (95% CI: 100% - 100%)100.0% (95% CI: 100% - 100%)
    Reproducibility (%CV)Between 0.5% and 1.6%Between 0.7% and 1.5%N/A (not directly from this study)N/A (not directly from this study)
    Analytical Sensitivity (LoD)500 Copies/mL (AD-169, Towne, Merlin) in UTM400 Copies/mL (AD-169), 800 Copies/mL (Towne), 6400 IU/mL (Merlin)N/AN/A
    Cross-Reactivity100% agreement (no cross-reactivity)100% agreement (no cross-reactivity)N/AN/A
    Interference100% agreement (no interference)100% agreement (no interference)N/AN/A
    Microbial Inhibition100% agreement (no inhibition)100% agreement (no inhibition)N/AN/A

    2. Sample Size and Data Provenance

    • Retrospective Study:

      • Saliva Swab: 173 total specimens (3 removed due to indeterminate CRM result, 170 analyzed)
      • Urine: 173 total specimens
      • Provenance: "collected during the clinical study", "stored at a central site", then distributed to three (3) laboratories. The document doesn't explicitly state the country of origin for these retrospective samples, though the testing sites were in the USA. These were pre-selected positive and negative samples based on routine laboratory results.
    • Prospective Study:

      • Saliva Swab: 1,859 initially collected, 6 deemed ineligible, resulting in 1,853 analyzed specimens.
      • Urine: 1,656 initially collected, 32 deemed ineligible, resulting in 1,624 analyzed specimens.
      • Provenance: Prospectively collected (frozen and/or fresh) from ten (10) collection sites across the USA and two (2) collection sites outside the USA. Testing was performed at six (6) testing sites located in the USA.

    3. Number of Experts and Qualifications

    The document states that a "Composite Reference Method (CRM)" was used, which involved "two (2) validated PCR followed by bi-directional sequencing assays." One (1) central laboratory performed these comparator assays.

    • The document does not specify the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience). It relies on the validation of the PCR and bi-directional sequencing assays as the basis for ground truth, implying that these are established and reliable laboratory methods.

    4. Adjudication Method

    The ground truth for the clinical agreement studies (both retrospective and prospective) was established via a "Composite Reference Method (CRM)".
    This CRM "utilized two (2) validated PCR followed by bi-directional sequencing assays. A sample had a final sequencing result of 'Detected' if one or both sequencing results were 'Detected'. Conversely a sample had a final sequencing result of 'Not Detected' if both results were 'Not Detected'." This implies a form of 2+0 or 1+1 adjudication model where if either reference method detects CMV, the sample is considered positive, and both must be negative for the sample to be considered negative.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This study focuses on the diagnostic performance of the device itself against a laboratory-based reference method, not on how human readers/clinicians improve with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance evaluation was done. The Simplexa™ Congenital CMV Direct assay is a real-time PCR assay and its performance was evaluated directly against the Composite Reference Method (CRM) without human-in-the-loop assistance. The reported PPA and NPA values represent the algorithm's standalone performance.

    7. Type of Ground Truth Used

    The ground truth used was expert consensus on laboratory results, specifically based on a "Composite Reference Method (CRM) utilized two (2) validated PCR followed by bi-directional sequencing assays." This is a highly robust and objective form of ground truth for nucleic acid detection devices, often considered a gold standard in molecular diagnostics.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. This is typical for submissions of this nature, where the focus is on the validation of the final device/algorithm using a separate, independent test set, rather than details of the developmental (training) phase.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as details about the training phase are not included in this summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN180040

    Validate with FDA (Live)

    Date Cleared
    2018-11-30

    (123 days)

    Product Code
    Regulation Number
    866.3181
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alethia CMV Assay Test System includes separately provided test kits for the Alethia CMV DNA Amplification Assay and the Alethia CMV External Control Reagents.

    The Alethia CMV DNA Amplification Assay, performed on the Alethia instrument, is a qualitative, in vitro diagnostic test system for the direct detection of Cytomegalovirus (CMV) DNA in saliva samples from neonates younger than 21 days of age. The test is used as an aid in the diagnosis of congenital CMV infection. The results of this test should be used in conjunction with the results of other clinical findings.

    Flocked swabs should be used to collect saliva from neonates. The swab can be collected dry, without viral transport media (VTM), or placed in no more than 1 mL VTM.

    The Alethia CMV External Control Reagents are used as part of a routine quality control program to aid the user in detection of unexpected conditions that may lead to test errors. The external controls are intended for use with the Alethia CMV DNA Amplification Assay; the controls are not intended for use with other assays or systems.

    Device Description

    The Alethia CMV Assay Test System, including the Alethia CMV DNA Amplification Assay and the Alethia CMV External Controls, is based on loopmediated amplification (LAMP) technology. The assay targets a region of the Cytomegalovirus genome that is conserved across multiple CMV strains. The Alethia CMV target is a 194 base pair (bp) sequence of the Human herpesvirus 5 genome.

    LAMP uses specially designed primers to provide for specific and continuous isothermal DNA amplification. A bv-product of amplification is magnesium pyrophosphate, which forms a white precipitate leading to a turbid solution. Reaction solution absorbance characteristics are monitored by the Alethia™ instrument.

    Changes in reaction solution turbidity created by precipitation of magnesium pyrophosphate indicate the presence of target DNA. The absence of target DNA results in no significant change in sample absorbance.

    AI/ML Overview

    The Alethia CMV DNA Amplification Assay is a qualitative in vitro diagnostic test for the direct detection of Cytomegalovirus (CMV) DNA in saliva samples from neonates younger than 21 days of age. It aids in the diagnosis of congenital CMV infection.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implicit from study results and regulatory requirements)Reported Device Performance (from "Clinical studies" section)
    Clinical Performance:
    Positive Percent Agreement (PPA) with Composite Reference Method100% (39/39) [95% CI: 91.0%; 100%]
    Negative Percent Agreement (NPA) with Composite Reference Method99.8% (1,472/1,475) [95% CI: 99.4%; 99.9%]
    Invalid Rate (Initial)1.7% (27/1,548)
    Invalid Rate (Final, after re-testing)0.06% (1/1,548) [95% CI: 0.01%; 0.37%]
    Analytical Performance:
    Limit of Detection (LoD) - Dry Swab1,025 copies/mL
    Limit of Detection (LoD) - Swab in VTM15,686 copies/mL
    Inclusivity (tested for 3 additional CMV strains)100% positive for AD-169, Toledo, and Towne strains at 2-3X LoD.
    Cross-Reactivity (panel of 40 microorganisms and human genomic DNA)No cross-reactivity observed.
    Microbial Interference (panel of 40 microorganisms and human genomic DNA)No microbial interference observed (all CMV positive samples remained positive).
    Chemical/Biological Interfering SubstancesNo interference observed with specified substances at tested concentrations (except for mucin at 50 mg/mL, mitigated by labeling).
    Sample Stability (room temp, refrigerated, frozen, freeze-thaw)Supports storage for up to 48 hours at 19-30°C, 7 days at 2-8°C, or frozen at <-20°C. Supports up to 2 freeze-thaw cycles.
    Carryover ContaminationNo carryover observed in alternating high positive and negative samples.
    Precision/ReproducibilityDemonstrated acceptable repeatability and reproducibility across different operators, days, sites, and kit lots.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Prospective Test Set:
      • Sample Size: 1,514 specimens.
      • Data Provenance: Prospectively collected from August 2017 to March 2018 at seven clinical study sites representing geographically distinct regions throughout the United States, Canada, Europe, and Australia.
    • Archived Preselected Positive Samples:
      • Sample Size: 34 specimens.
      • Data Provenance: De-identified samples previously evaluated from prospective clinical studies and found to have CMV infection. Collected from infants less than 21 days of age and stored at -80 ℃. The provenance is retrospective, as they were existing samples used to supplement positive cases.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologist with X years of experience) used to establish the ground truth.

    Instead, the ground truth (Composite Reference Method - CRM) was established by a technical algorithm:

    • Two manufacturer-developed and validated PCR assays were used.
    • Samples positive by either PCR assay were further tested by bidirectional sequencing (BDS).
    • Ground Truth: Samples were considered positive when BDS confirmed the presence of CMV DNA, and negative when neither PCR assay produced amplicon after 40 cycles or BDS was negative.

    4. Adjudication Method for the Test Set:

    The adjudication method for the test set relied on a defined algorithm using two PCR assays and bidirectional sequencing.

    • If both PCRs were positive and BDS was positive, it was CRM positive.
    • If both PCRs were positive and BDS was negative, it was CRM negative.
    • If one PCR was positive and BDS was positive, it was CRM positive.
    • If one PCR was positive and BDS was negative, it was CRM negative.
    • If both PCRs were negative, it was CRM negative (BDS not applicable).

    This is a rule-based algorithmic adjudication, not human consensus-based.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This device is an automated in vitro diagnostic test (qualitative LAMP technology) for direct detection of CMV DNA, not an AI-powered image analysis or diagnostic support tool intended to improve human reader performance. Therefore, the concept of human readers improving with AI assistance is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies presented are for the standalone performance of the Alethia CMV Assay Test System. The device is a qualitative, automated system that detects CMV DNA, providing a POSITIVE, NEGATIVE, or INVALID result based on its internal algorithms and measured absorbance characteristics. Human intervention is limited to sample preparation and operating the instrument, not interpretation of raw signals to determine positivity or negativity. The clinical performance (PPA, NPA) directly reflects the algorithm's accuracy against a molecular ground truth.

    7. The Type of Ground Truth Used:

    The ground truth used was a Composite Reference Method (CRM) based on molecular testing results (two PCR assays and bidirectional sequencing). This is a type of technical/laboratory ground truth, not expert consensus, pathology, or direct outcomes data.

    8. The Sample Size for the Training Set:

    The document does not explicitly state the sample size for a training set. This type of in vitro diagnostic assay, based on LAMP technology, would typically be developed and optimized using analytical studies (determining LoD, inclusivity, specificity, precision, etc.) and then validated in clinical studies, rather than relying on a separate "training set" in the machine learning sense. The performance characteristics represent the validation of the assay's design.

    9. How the Ground Truth for the Training Set Was Established:

    As no explicit training set is detailed for an algorithm in the machine learning sense, the method for establishing its ground truth is not described. The analytical performance evaluations (LoD, inclusivity, specificity) are based on contrived samples with known CMV status and concentration, which serve as the "ground truth" for evaluating those specific analytical characteristics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1